
3 1377486-IN2-A
INSTALLATION LAYOUT
START UP
OPERATION & FILTER CHANGEOUT
TROUBLESHOOTING
• Clean the filter either with compressed air or flush the housing with pure water to avoid debris or dirt to float in the system.
• Position the filter in such a way that the fluid enters the inlet connection.
• Close the outlet valve and open the vent on top of the filter housing.
• Partially open the inlet valve at pipe before the filter housing and allow fluid to fill completely into housing along with venting
of air from top of the housing.
• After complete venting of air, close the vent valve. Fully open the inlet valve and open the outlet valve simultaneously.
1. Changing the Filter Cartridge
• Follow regulations in force for the disposal of used cartridges.
• Before opening the filter housing turn-off, the water main.
• Release the pressure from the product by un-screwing the vent-valve of the head. Then screw and tighten the vent-valve.
• Open the housing unscrewing the housing using a proper wrench.
• Remove the used cartridge.
• Wipe clean the filter housing with cold water and a soft sponge.
• Open the new cartridge wrap and discard wrap.
• Insert the new cartridge and place it into the housing.
• Tighten the housing bowl to housing’s head using a proper wrench. Do not over-tighten.
• Turn-on the water main.
• Slowly turn-on a water supply (tap) downstream the filter.
• Wait for air purge from the vent-valve, then screw and tighten the vent-valve.
Remark: Let the water flow for at least 5 minutes before the use.
1. Troubleshooting for Leaks
• Check the O-ring for any damage and wear & tear, dirt or swelling etc. If required, replace 0-ring with new one.
• Properly close the housing. Hand tightening is good enough to seal.
• Carefully check all threaded connections. If connections not proper, unscrew and connect again as and when needed.
• Check to see if filter cartridge is dirty and choked up. If this is the case, replace the cartridges as per instructions.
2. Poor Filtration Results
• Check to make sure the filter cartridges is properly seated in the support basket.
• Check all O-rings and sealing surfaces are cleaned before installing.
• Frequent Cartridges Clogging - Selection of proper filter cartridge with proper micron rating.
• Lower filtration flow or High Differential Pressure- Change the Cartridgeas per recommended Differential Pressure.
If Differential Pressure is not achieved -Open the system and check the cartridge condition for damages (if any).
NPT
FITTINGS
BSP
IN
OUT
IN
OUTIN IN OUT
a.
b.
General information.
IN-OUT directions presentation in different lter housings models.