K3TJ K3TJ
5
Accessories (Order Separately)
Name Appearance Model Applicable Unit
Water-resistive Soft Front Cover K32-L49SC K3TE
K3TF
K3TL
K3TJ
Terminal Cover K32-L49TC K3TF
K3TL
K3TJ
Note: 1. Use the Mounting Adapters provided with the K3TJ to mount the above accessories.
2. The Terminal Cover is used to protect against electric shock but is not waterproof.
Specifications
Ratings
Supply voltage 100 to 240 VAC (50/60 Hz); 24 VDC (internally insulated)
Operating voltage range --15% to 10% of supply voltage
Power consumption Approx. 7.7 VA (max. AC load); approx. 2.8 W (max. DC load)
Insulation resistance 10 M!min. (at 500 VDC) between external terminals and case
Dielectric strength 2.000 VAC for 1 minute between input terminals and power supply
2,000 VAC for 1 minute between external terminals and case
Noise immunity AC Model: "1,500 V on power supply terminals in normal or common mode
DC Model: "480 V on power supply terminals in normal mode and "1,500 V on power supply
terminals in common mode
Vibration resistance Malfunction: 10 to 55 Hz, 0.5-mm single amplitude for 10 minutes each in X, Y, and Z
directions
Destruction: 10 to 55 Hz, 0.75-mm single amplitude for 2 hours each in X, Y, and Z directions
Shock resistance Malfunction: 98 m/s2(10G) for 3 times each in 6 directions.
Destruction: 294 m/s2(30G) for 3 times each in 6 directions
Ambient temperature Operating: --10#Cto55
#C (with no icing)
Storage: --20#Cto65
#C (with no icing)
Ambient humidity Operating: 35% to 85% (with no condensation)
Ambient atmosphere Must be free of corrosive gas
Weight 200 g
Case material Heat-resistive ABS/PC
EMC Emission Enclosure: EN55011 Group 1 class A
Emission AC Mains: EN55011 Group 1 class A
EN61000-4-2: 4 kV contact discharge (level 2)
8 kV air dischanrge (level 3)
Immunity RF-interference: ENV50140: 10 V/m (amplitude modulated, 80 MHz to
1GHz)(level3)
10 V/m (pulse modulated, 900 MHz)
Immunity Conducted Disturbance: ENV50141: 10 V (0.15 to 80 MHz) (level 3)
Immunity Burst: EN61000-4-4: 2 kV power-line (level 3)
2 kV I/O signal-line (level4)
Approved standards UL (File No. E4151), CSA (File No. LR67027)
Conforms to EN50081-2 and EN50082-2
Note: An initial current of approximately 0.25 A will flow for approximately 35 ms the moment the K3TJ is turned on.