
Page 19
12. Troubleshooting
Review the information in the table below to troubleshoot operating problems.
PROBLEM CAUSE SOLUTION
I. Optical Section:
Bulb operates, but eld of view re-
mains dark. Power supply is unplugged. Connect
Brightness is too low Set brightness to a proper level
Field of view is obscured or not evenly
illuminated Revolving nosepiece is not correctly
engaged. Make sure that the revolving nosepiece
clicks properly into place.
Filter holder, polarizer or analyzer are
partially inserted Insert or remove the lter slider
Dirt or dust is visible in the field of view. Dirt/dust on the specimen Clean the specimen
Dirt/dust on the eyepieces Clean the eyepieces
Image looks double Aperture iris diaphragm is stopped
down too far. Open aperture iris diaphragm.
The field diaphragm is not well cen-
tered Set the diaphragm according to Koe-
hler settings.
Visibility is poor.
• Image is not poor.
• Contrast is poor.
• Details are indistinct.
Revolving nosepiece is in an incorrect
position Move the nosepiece to a click stop
Aperture iris diaphragm is too closed
or too open. Adjust aperture iris diaphragm.
Dust or dirt on lenses (objectives, eye-
pieces) Clean thoroughly.
Focus is not even Specimen holder is not at. Move the
specimen to a at position.
One side of the image is unfocused Revolving nosepiece is in an incorrect
position Move the nosepiece to a click stop
Specimen is not in a at position Place the specimen in a at position
on the stage
II. Mechanical Section:
Coarse focus knob is hard to turn Tension adjustment ring is too tight Loosen tension adjustment ring
Focus is unstable Tension adjustment ring is too loose Tighten tension adjustment ring
III. Electrical Section:
Bulb doesn’t turn on. Power supply not connected Check for proper connection
Brightness is not enough Brightness setting is too low Adjust brightness
Light blinks Power supply not well connected Check for proper connection
IV. Observation tube:
Field of view of one eye does not match
that of the other. Interpupillary distance is incorrect. Adjust interpupillary distance.
Incorrect diopter adjustment. Adjust diopter.
Your view is not accustomed to micro-
scope observation. Upon looking into eyepieces, try look-
ing at overall field before concentrating
on specimen range. You may also find
helpful to look up and into distance for
a moment before looking back into mi-
croscope.
V. Microphotography:
Image edge is unfocused To a certain extent it is due to achro-
matic objectives features To minimize the problem, set the aper-
ture diaphragm in a proper position
Bright spots appear on the image Stray light entering in the microscope
through eyepieces or camera viewn-
der
Cover eyepieces and viewnder with a
dark cloth