Pari Vios User manual

REMEMBER
REPLACE YOUR REUSABLE NEBULIZER CUP EVERY SIX MONTHS
INSTRUCTIONS FOR USE
For single patient use (one person) only
INCLUDES INSTRUCTIONS FORTHE PARI LC®SPRINT NEBULIZER
IMPORTANT: READ INSTRUCTIONS AND SAFETY PRECAUTIONS BEFORE USE
READ ALL WARNINGS AND INSTRUCTIONS BEFORE USE. USE THE PRODUCT ONLY FOR ITS INTENDED USE
AS DESCRIBED IN THESE INSTRUCTIONS FOR USE. IMPROPER USE CAN RESULT IN SERIOUS OR FATAL ILLNESS/
INJURY, IMPROPER TREATMENT OR PROPERTY DAMAGE.
Vios IFU- Sprint EN 0029.indd 1Vios IFU- Sprint EN 0029.indd 1 6/16/10 11:46:59 PM6/16/10 11:46:59 PM

A. SAFETY PRECAUTIONS - LC®SPRINT
CAUTION
US Federal Law restricts this device to sale by or on the
order of a physician. Use only when prescribed by a
physician and with the prescribed medication.
The LC®Sprint is a handheld nebulizer, designed to
aerosolize medication approved for nebulization and
prescribed by a physician. The LC®Sprint is intended for
adult and pediatric patients consistent with the indications
for the aerosol medication. Improper use can result in
serious or fatal illness or injury. Nebulizer is not sterile.
WARNING
Use only when prescribed by a physician and with the
prescribed medication. In order to achieve the specified
droplet characteristics the compressor or central air supply
must provide 3.5 - 8 lpm at a minimum of 20psi.
WARNING
1. To reduce the risk of increased bacterial growth,
infection, illness, or injury from contamination:
thoroughly clean and dry all parts of the nebulizer and
thoroughly dry any moisture or condensation in the
tubing at the end of every treatment following the
enclosed instructions.
2. Cleaning the nebulizer after each treatment is not
enough---you must regularly disinfect or sterilize the
nebulizer between treatments to prevent serious or
fatal illness caused by contamination of the nebulizer.
3. Never use a nebulizer with a clogged nozzle. If the
nozzle is clogged, the aerosol mist will be reduced,
altered or stopped, which will diminish or eliminate
the effectiveness of the treatment. If clogging occurs,
replace with the same model PARI Reusable Nebulizer.
4. Close supervision is necessary when this product is
used by or near children and physically or mentally
impaired. Never use when sleeping or drowsy.
5. This product contains small parts that may present
a choking hazard to small children. The tubing also
presents a strangulation hazard. ALWAYS USE CLOSE
ADULT SUPERVISION WHEN ADMINISTERING A
TREATMENT TO A CHILD. DO NOT LEAVE CHILD
UNATTENDED DURING TREATMENT.
6. Patients under the age of 4 or any patients who have
special needs should use a facemask instead of the
valved mouthpiece.
7. Before use, check PARI Reusable Nebulizer for proper
assembly. All parts should be seated firmly in place.
Use of an improperly assembled nebulizer could
diminish or prevent ad equate delivery of medication
and the effectiveness of the treatment.
Welcome to the Vios experience! The Vios™
Aerosol Delivery System is designed from
PARI’s long legacy of clinically proven products
and consists of two important components:
a Vios™compressor and a PARI nebulizer.
Vios™was inspired by the Latin and Greek
words for “Life” and our goal is to improve
your quality of “Life” through the efficient
delivery of aerosol.
Your physician has prescribed the Vios™to
treat your respiratory symptoms because
of PARI’s reputation for consistent particle
size and fast treatments resulting in targeted
delivery to the lungs. PARI products have
been used in more than 25 clinical trials for
new nebulized medications. Designed with the
patient in mind, these instructions will help you
achieve the best aerosol therapy experience.
We hope the Vios experience allows for a
“Life Inspired”.
SAVE THESE INSTRUCTIONS
FOR FUTURE USE
INDICATIONS FOR USE: LC®SPRINT
The LC®Sprint is a handheld nebulizer, designed to
aerosolize medication approved for nebulization and
prescribed by a physician. The LC®Sprint is intended for
adult and pediatric patients consistent with the indications
for the aerosol medication.
EN-1
Vios IFU- Sprint EN 0029.indd 2Vios IFU- Sprint EN 0029.indd 2 6/16/10 11:47:07 PM6/16/10 11:47:07 PM

INDICATIONS FOR USE: VIOSTM AEROSOL DELIVERY SYSTEM
The Vios™is a tabletop, AC-powered air compressor intended to provide a source of compressed air for use with
jet nebulizers. The Vios™Compressor is intended for adult and pediatric patients for use in hospital, clinic, or home
environments.
WARNING
Use only with nebulizers and tubing authorized by PARI
for this device. Other nebulizers and tubing may result in
improper treatment.
When using electrical and medical products, especially
when children are present, basic safety precautions must
be followed.
DANGER
To reduce the risk of serious or fatal injury from
electrocution:
1. Do not place or store product where it can fall or be
pulled into a bathtub, sink, water, or other liquid. Do
not place in or drop into water or other liquid. Do not
use while bathing.
2. Do not reach for a product that has fallen into water or
other liquid. Unplug immediately.
WARNING
To reduce the risk of serious or fatal injury from
electrocution, fire or burns and to reduce the risk of
damage and malfunction to the product:
1. Before use, check the serial number label on the
bottom of the PARI Vios™Compressor to ensure that
the voltage and current indicated on the compressor
label correspond to the voltage and current available.
2. Never operate this product if it has a damaged cord
or plug, if it is not working properly, if it has been
dropped or damaged, or if it has been exposed to any
liquids inside the case. Return the product to PARI for
examination and repair.
3. Keep all electrical cords away from heated surfaces. Do
not overload wall outlets or use extension cords.
4. This appliance has a polarized plug (one blade is wider
than the other). As a safety feature, this plug will fit in
a polarized outlet only one way. If the plug does not fit
fully in the outlet, reverse the plug.
5. Never block the air vents of this compressor or place it
on a soft surface, such as a bed or couch, where the air
vents may be blocked. Keep the air vents free of lint,
hair, and similar substances. Blocked vents could cause
excessive heat buildup. This product should never be
left unattended when plugged in.
WARNING
6. Do not operate outdoors.
7. Not suitable for use in the presence of flammable
anesthetic mixture with air or oxygen or nitrous oxide.
8. Do not operate where oxygen is being administered in
a closed environment such as an oxygen tent.
9. Never drop or insert any object into any openings.
10. Always unplug this product immediately after using.
11. Be sure the PARI Vios™Compressor is unplugged prior
to cleaning.
12. Do not open or disassemble the PARI Vios™
Compressor. Refer all servicing to PARI.
WARNING
To reduce risk that the compressor will fail to operate
properly as a result of decreased electromagnetic
interference, resulting in improper treatment:
1. This device needs special precautions regarding
electromagnetic compatibility (“EMC”) and must be
installed and put into service according to the EMC
information provided in these Instructions For Use.
2. Portable and mobile radio frequency (“RF”)
communication devices can disrupt electrical medical
equipment. Technical EMC information is available on
pages 9-10.
3. Use only with accessories and cables authorized
by PARI for this device. Use of this device with
unauthorized accessories and cables may result in
increased electromagnetic emissions or decreased
electromagnetic immunity of the device.
4. This device should not be used adjacent to, or stacked
with other equipment. If adjacent or stacked use is
necessary, this device should be observed to verify
normal operation.
B. SAFETY PRECAUTIONS -VIOSTM AEROSOL DELIVERY SYSTEM
Questions? 1.800.FAST.NEB (327.8632) |PARI.com/VIOS EN-2
Vios IFU- Sprint EN 0029.indd 3Vios IFU- Sprint EN 0029.indd 3 6/16/10 11:47:09 PM6/16/10 11:47:09 PM

C. GETTING STARTED
Check your Vios™package to ensure you have the
following items.
It is important to have a second PARI Reusable Nebulizer
on hand at all times.
Wing Tip™Tubing
PARI LC®Sprint Reusable Nebulizer (2 included)
PARI Reusable Nebulizer Holder & Clip
Air Outlet
Air Filter
Power Switch
Timestrip®
(Timestrip®Location)
Inspiratory Valve Cap
Nebulizer Outlet
Mouthpiece with Expiratory Valve
Nebulizer Top
Nebulizer Insert
Nebulizer Cup
Air Intake
Questions? 1.800.FAST.NEB (327.8632) | PARI.com/VIOS
EN-3
Air Vents
Vios IFU- Sprint EN 0029.indd 4Vios IFU- Sprint EN 0029.indd 4 6/16/10 11:47:11 PM6/16/10 11:47:11 PM

D. VIOS™ SET-UP
You are 10 Steps away from starting your treatment with
Vios™. Please follow step by step.
1. Clean/Disinfect Nebulizer Cup: Refer to Section F to
properly clean and/or disinfect your nebulizer prior to
first use. Re-assemble nebulizer.
2. Set-up Compressor: Place the compressor on a
flat, stable surface free of dust. Avoid cushioned,
upholstered or carpeted surfaces. Make sure power
switch is turned to , the “OFF” position. Plug power
cord into wall outlet (Fig A). Activate and attach the
Timestrip®label to the front on the Vios™Compressor
as explained in Section E.
3. ConnectTubing To Compressor: Insert one end of
Wing Tip™tubing to the air outlet on the compressor.
It should fit snugly (Fig. B).
4. Load Medication: Open the inspiratory valve cap of
nebulizer by lifting the cap (Fig. C). Pour medication into
nebulizer cup. Never fill past the max. fill line of 8 ml.
5. Attach Mouthpiece or Mask: If using the mouthpiece,
attach the mouthpiece to the nebulizer outlet with the
expiratory valve facing up. Then close the inspiratory
valve cap by pressing firmly on the nebulizer top. If
using a mask, refer to the mask instructions for use.
6. Attach TubingTo Nebulizer Cup: Firmly press the other
end of the Wing Tip™tubing to bottom of PARI
reusable nebulizer (Fig. D).
7. Starting Treatment: Sit in a relaxed, upright position and
turn the compressor “ON” .
8. Nebulizer Position: Place the mouthpiece on top of
tongue and close lips around it. Take slow, deep
breaths. As you inhale, the inhalation valve opens.
9. Breathing: For mouthpiece, inhale and exhale slowly
through your mouth. As you exhale, the valve on
the mouthpiece opens to allow the exhaled mist to
escape. For mask, place mask over nose and mouth
and inhale and exhale slowly.
10. CompleteTreatment: Continue until the medication is
gone or you hear a slight “sputtering” sound. Turn
compressor “OFF” .
A NOTE ON TREATMENT TIME
Please note that many factors affect treatment time,
such as a patient’s age, condition, breathing pattern, and
medication. The Vios™was designed for intermittent use
to deliver respiratory medications, not continuous use
for humidification or prolonged treatments (for example,
over 35 minutes at one time). It is possible with prolonged
treatments that the tubing may get warm and disconnect
from the Vios™air outlet. If your physician prescribed
consecutive (back to back) treatments, resulting in the
use of your Vios™for longer than 35 minutes at a time,
please contact PARI Patient Connection at 1.800.FAST.NEB
(327.8632).
B
C D
Questions? 1.800.FAST.NEB (327.8632) |PARI.com/VIOS EN-4
A
Vios IFU- Sprint EN 0029.indd 5Vios IFU- Sprint EN 0029.indd 5 6/16/10 11:47:12 PM6/16/10 11:47:12 PM
Table of contents
Languages:
Other Pari Respiratory Product manuals

Pari
Pari ProNEB Ultra User manual

Pari
Pari 041G0749 User manual

Pari
Pari 044F7251 User manual

Pari
Pari LC SPRINT STAR User manual

Pari
Pari KINDERMASKE SOFT User manual

Pari
Pari Dura-Neb 3000 User manual

Pari
Pari SMARTMASK Kids User manual

Pari
Pari Lamira User manual

Pari
Pari Vios PRO User manual

Pari
Pari Compact User manual

Pari
Pari Trek S User manual

Pari
Pari PEP S Reference guide

Pari
Pari LC SPRINT SPA User manual

Pari
Pari LC Sprint User manual

Pari
Pari VORTEX 051B0100 User manual

Pari
Pari LC series User manual

Pari
Pari Tolero 678G8208 User manual

Pari
Pari LC Sprint User manual

Pari
Pari SINUS User manual

Pari
Pari SMARTMASK 041G0730 User manual