TESTO 6321 User manual

testo 6321 · differential pressure transmitter
P2A software · Parameterizing, adjusting
and analyzing software
Instruction manual

2

1 Safety and the environment
3
Pos: 1 /TD /Über schri ften/M UF/Si cherh eit und Um welt @ 3\mod_1234793958627_79.doc @ 26223@ 1
1 Safety and the environment
Pos: 2/ TD /Si ch erh eit und U mw el t/S ic her heit gew ährl ei st en/MU F 63xx /El ektr is ch e G ef ahr en v er mei de n @ 3\mod_1234794609299_79.doc@ 26280 @ 5
Avoiding electrical hazards
> Never use the instrument and connected probes to measure on
or near live parts!
> Damaged mains cables must only be replaced by authorized
personnel.
> Only have the transmitter wired and connected by authorized
personnel with the voltage disconnected.
> You must always comply with the regulations applicable in your
country for opening and repairing electrical equipment.
Pos: 3/ TD /Si ch erh eit und U mw el t/S ic her heit gew ährl ei st en/MU F 63xx /P erson en- u nd Sac hschäd en verm eiden @ 3 \mod_1234794744768_79.doc@ 26299@ 5
Avoiding personal injury and damage to equipment
> Installation, setting and calibration work must only be carried
out by qualified and authorized personnel!
> Only open the instrument when this is expressly described in
the instruction manual for installation, maintenance or repair
purposes.
> Observe the permissible storage, transport and operating
temperature.
Pos: 4 /TD /Sich erhei t undU mwel t/Sic herhei t gewähr leis ten/Ni cht mit Lö sungsm ittel n lager n @0 \mod_1175692375179_79.doc@ 583 @
> Do not store the product together with solvents. Do not use any
desiccants.
Pos: 5/ TD /Si c h erh eit und U mw el t/ Sic her hei t g ew ähr l eis ten/ MU F 63xx /B ei W ar tu ng M UF nic ht z ur Reg el ung verw en de n @ 3\mod _1234794852377_79.doc@ 26318 @
> Do not use the instrument for control purposes at the same time
as operating or servicing the transmitter.
Pos: 6/ TD /Si ch erh eit und U mw el t/S ic her heit gew ährl ei st en/ Pro duk t b esti mm ung sg em äß verw en den @ 0\mod_1173781261848_79.doc @ 38 6 @
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
Pos: 7/ TD / Sic h er h eit und U m w el t/ Sic her hei t gew ährl eis t en/ Nur bes chr i eb e ne W ar t ung s arb ei t en d urc hfü hr en @ 0\m od_1175692705195_79.doc @ 601@
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
Pos: 8/ TD /Si c h erh eit un d U mw el t/Si c her heit gew ährl ei st en/MU F 63xx /F ach per so nal @ 3\mod_1234794940409_79.doc @ 26337@
Any additional work must only be carried out by authorized
personnel. Otherwise testo will not accept any responsibility for the
proper functioning of the instrument after repair and for the validity
of certifications.

2 About this document
4
Pos: 9/ TD /Ü ber sc hr ift en /M UF /Um w elt s c hütz en @ 3\mod_1234858757571_79.doc@ 26363@ 5
Protecting the environment
Pos: 10 /T D/ Si c her heit un d Um w el t/Um w elt s c hüt z en/P r oduk t e nt sor g en @ 0\m od _1173780307072_79.doc@ 357 @
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
Pos: 11 /T D/Ü b er schr if te n/M UF /Z u di es em Dok um ent @ 3\mod_1234793991331_79.doc@ 26242@ 1
2 About this document
Pos: 12 /T D/ Si c her heit un d Um w el t/Z u di es em D ok um e nt /Ver we nd u ng ( St and ard) @ 0\mod_1173775068554_79.doc@ 337 @ 5
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Pos: 13 /T D/ Si c her heit un d Um w el t/Z u di es em D ok um e nt /Sym bol e und S c hrei bk onv e nti on en /War nhi nw eis WARN UN G @ 2\mod_1207646966234_79.doc @ 14398@
WARNING Indicates potential serious injuries
Pos: 14 /T D/ Si c her heit un d Um w el t/Z u di es em D ok um e nt /Sym bol e und S c hrei bk onv e nti on en /War nhi nw eis VOR SI CH T @ 2\mod_1207651416515_79.doc@ 14416 @
CAUTION indicates potential minor injuries
Pos: 15 /T D/ Si c her heit un d Um w el t/Z u di es em D ok um e nt /Sym bol e und Sc hr ei bk onv . Softwar e [Sta nd ard ] @ 0\mod_1190203332543_79.doc @ 4883@ 5
Symbols and writing standards
Representa-
tion
Explanation
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu
Elements of the program interface.
[OK]
Buttons of the program interface.
... | ... Functions/paths within a menu.
“...” Example entries
Pos: 16 /TD /--- Seit e nwec hs el ---@ 0\mod_1173774430601_0.doc@ 283 @

3 Contents
5
Pos: 17 /T D/ Ü b ers chr if t en/ MU F /I nh al t @ 3\mod_1234794019831_79.doc @ 26261 @ 1
3 Contents
1Safety and the environment ........................................................ 3
2About this document.................................................................... 4
3Contents ........................................................................................ 5
4Transmitter.................................................................................... 7
4.1. Specifications ..................................................................................7
4.1.1. Functions and use............................................................................7
4.1.2. Scope of delivery .............................................................................7
4.1.3. Dimensions......................................................................................7
4.1.4. Technical data .................................................................................8
4.2. Product description........................................................................11
4.2.1. At a glance.....................................................................................11
4.2.2. Scaling...........................................................................................12
4.3. Commissioning..............................................................................13
4.3.1. Assembling the instrument.............................................................13
4.3.1.1. Wall mounting....................................................................13
4.3.2. Wiring the instrument .....................................................................14
4.3.2.1. 4-wire system ....................................................................16
4.3.2.2. 3-wire system ....................................................................16
4.4. Maintenance and cleaning.............................................................17
4.4.1. Cleaning housing ...........................................................................17
4.4.2. Namur fault conditions ...................................................................17
5Parameterizing, adjusting and analyzing software (P2A
software)...................................................................................... 18
5.1. Specifications ................................................................................18
5.1.1. Functions and use..........................................................................18
5.1.2. System requirements .....................................................................19

3 Contents
6
5.2. First steps ..................................................................................... 20
5.2.1. Installing the software/driver ..........................................................20
5.2.1.1. Installing P2A software......................................................20
5.2.1.2. Installing USB driver..........................................................20
5.2.1.3. P2A software upgrade....................................................... 20
5.2.2. Starting the software......................................................................21
5.2.2.1. Starting the program.......................................................... 21
5.2.2.2. Establishing a connection with the instrument ...................21
5.2.2.3. Activating the connection with the instrument ....................21
5.3. Using the software ........................................................................ 22
5.3.1. User interface ................................................................................22
5.3.2. Editing instrument/parameter file....................................................24
5.3.2.1. Changing instrument/parameter file................................... 24
5.3.2.2. Creating a new instrument file ...........................................26
5.3.2.3. Saving parameters ............................................................26
5.3.2.4. Opening the parameter file ................................................ 27
5.3.2.5. Copying and pasting parameters....................................... 27
5.3.2.6. Deleting instrument/parameter file .....................................27
5.3.3. Analyzing/testing the transmitter ....................................................28
5.3.3.1. Analyzing/testing the instrument ........................................ 28
5.3.3.2. Carrying out factory reset ..................................................28
5.3.3.3. Testing analog output........................................................29
5.3.3.4. Displaying min./max. values ..............................................30
5.3.4. Adjusting the transmitter ................................................................32
5.3.4.1. n-point adjustment............................................................. 32
5.3.4.2. Adjusting the analog output ............................................... 33
5.3.5. Transmitter history .........................................................................35
6Tips and assistance ................................................................... 37
6.1. Questions and answers ................................................................ 37
6.2. Accessories and spare parts......................................................... 37
6.2.1. Ordering options for testo 6321 transmitter (0555 6321) ................ 38
Pos: 18 /TD /--- Sei te nw ec hs el --- @ 0\m od_1173774430601_0.doc@283 @

4 Transmitter
7
Pos: 19 /T D/ Ü b ers chr if t en/ MU F /1 Mess um f orm er @ 3\mod_1234258401060_79.doc@ 23894@ 1
4 Transmitter
Pos: 20 /T D/ Ü b ers chr if t en/ MU F /1 .1/ 2. 1/ 3. 1 Leist ung s bes chr eib ung @ 3\mod_1234258595211_79.doc@ 23951@ 2
4.1. Specifications
Pos: 21 /T D/ L ei st ung s bes c hr ei b ung/ V er we nd u ng /M UF 63x x/M UF 63 21 @ 3\m od_1241608910962_79.doc@ 32625@ 3
4.1.1. Functions and use
The testo 6321 transmitter is suitable for the following applications,
amongst others:
• Air conditioning and ventilation technology
◦Monitoring ventilation and filtration systems
◦Monitoring fans
◦Monitoring exhaust air volumetric flow rates
Pos: 22 /T D/ L ei st ung s bes c hr ei b ung/ Li e fer um f ang /M UF 63xx /M UF 678 1 @ 3\mod_1238162497483_79.doc @ 30275@ 3
4.1.2. Scope of delivery
The scope of delivery of the testo 6321 transmitter includes the
following:
• Assembly accessories
• Operating instructions
• Calibration report
Pos: 23 /T D/ L eist ung s bes chr eib ung /T ec h nis ch e D at e n/MU F 63xx/M UF 678 1 A bm ess u nge n @ 3\mod_1239806330094_79.doc @ 31183@ 3
4.1.3. Dimensions
Pos: 24 /T D/ L eist ung s bes chr eib ung /T ec h nis ch e D at e n/MU F 63xx/M UF 632 1 @ 3\m od_ 1241601145321_79.doc@ 32604@ 35555555555555555555

4 Transmitter
8
4.1.4. Technical data
Parameter
• Differential pressure
Accuracy1
The specifications are only valid if the positive pressure is
applied at the positive pressure connection.
• 1.2 % of measuring range, additional ±0.3 Pa intrinsic error 2
• TK slope drift = 0.05 % of measuring range per degree Kelvin of
deviation from nominal temperature 22 °C
• TK zeroing drift= 0 % (by means of automatic zeroing)
Measuring range, resolution and overload
Measuring range
depending on
version ordered
Resolution Overload
0 to 100 Pa 0.1 Pa 20,000 Pa
0 to 10 hPa 0.01 hPa 200 hPa
0 to 20 hPa 0.01 hPa 200 hPa
0 to 50 hPa 0.01 hPa 750 hPa
0 to 100 hPa 0.1 hPa 750 hPa
0 to 500 hPa 0.1 hPa 2500 hPa
0 to 1000 hPa 1 hPa 2500 hPa
0 to 2000 hPa 1 hPa 2500 hPa
1At 25 °C. A stabilization period of approx. 30 min must be taken into account upon commissioning the instrument.
2Measuring uncertainty in accordance with GUM: ±1.2 % of measuring range final value ±0.3 Pa.
GUM (Guide to the Expression of Uncertainty in Measurement): ISO guideline for determining the
measuring uncertainty in order to render global measurement results comparable.
The following uncertainties are used during the inquiry:
• Hysteresis
• Linearity
• Reproducibility
• Long-term stability
• Adjustment area/factory calibration
• Test location

4 Transmitter
9
Measuring range
depending on
version ordered
Resolution Overload
-100 to 100 Pa 0.1 Pa 20,000 Pa
-10 to 10 hPa 0.01 hPa 200 hPa
-20 to 20 hPa 0.01 hPa 200 hPa
-50 to 50 hPa 0.01 hPa 750 hPa
-100 to 100 hPa 0.1 hPa 750 hPa
-500 to 500 hPa
0.1 hPa
2500 hPa
-1000 to 1000 hPa 1 hPa 2500 hPa
-2000 to 2000 hPa 1 hPa 2500 hPa
Upon delivery and following a factory reset the readings are
shown in the display in the unit that was ordered via the
KMAT option Fxx, see Ordering options for testo 6321
transmitter (0555 6321) page 38.
Meas. cycle
• 1/sec
Zeroing cycle
• Set to 1 min at the factory
Interface
• Mini-DIN for P2A software (adjustment and parameterization
software)
Voltage supply
• 3 or 4-wire (separate signal and supply lines): 20 to 30 V
AC/DC, 300 mA power consumption

4 Transmitter
10
Maximum load
• 4-wire: 500 Ω(power output)
Maximal load
• 10 kΩ(voltage output)
Analog output
• 0 to 1 V ± 2.5 mV (4-wire) or
• 0 to 5 V ± 12.5 mV (4-wire) or
• 0 to 10 V ± 25 mV (4-wire) or
• 4 to 20 mA ± 0.05 mA (4-wire)
• TK= 0.05 %K of measuring range per degree Kelvin of
deviation from nominal temperature 22 °C
Resolution of analog output
• 12 bit
Display
• 2-line LCD (optional)
Operating temperature
• -5 to 50 °C

4 Transmitter
11
Storage temperature
• -20 to 60 °C
Application humidity
• 0 to 90 % RH
Housing, weight
• Plastic, approx. 160 g
Protection class
• IP 65 only if the transmitter is wired and/or sealing plugs are
inserted
Directives, standards and tests
• EC Directive: 2014/30/EC
Pos: 25 /T D/ Ü b ers chr if t en/ MU F /1 .2/ 2. 2 Prod ukt be s chr ei b ung @ 3\m od_1234258723551_ 79. d oc @ 24008 @ 2
4.2. Product description
Pos: 26 /T D/ P ro d ukt bes c hr ei b ung/ Ü b ers ic ht /MU F 63xx / Auf ein en Bl ic k MUF 632 1 @ 3\mod_1241599865646_79.doc @ 32583@ 3
4.2.1. At a glance
1 Display for showing reading
(optional, cannot be retrofitted)
2 Service interface (mini-DIN port)
3 Wall bracket (on rear)
4 Negative pressure connection
5 Positive pressure connection
6 M 16 x 1.5 screw connection, e.g.
analog output
Pos: 27 /T D/ P ro d ukt bes c hr ei b ung/ Ü b ers ic ht /MU F 63xx / Sk ali er ung 63 21 @ 4\mod_1252938938948_79.doc@ 50283@ 3

4 Transmitter
12
4.2.2. Scaling
There are three types of min./max. values:
1 The measuring range: The maximum sensor performance is in
this range. Measuring range, see table (below).
2 Standard scaling: The output signals are assigned to this
measuring range as standard:
◦during delivery if no entries are made in the order code
◦after exchanging the unit, the measuring range recorded in
the instrument is applied as standard.
The transmitter even retains its scaling with the voltage
disconnected.
Measuring range, see table (below).
3 The maximum settings for the manual scaling
◦The maximum limits can be calculated as follows:
X = difference between MIN. and MAX. value of the
standard scaling
(Max. value of standard) + (50 % of X)
(Min. value of standard) - (50 % of X)
◦It is thus possible to scale beyond the measuring range, e.g.
for the adjustment of the scaling limits to standard values of
a PLC.
With the alarm definition, however, the physical measuring
range limits are decisive.
Pos: 28 /T D/ Pro duk tbes c hr eib ung/Ü b ers ich t/M UF 63xx/ T abel l e Sk ali er ung MUF 6321 @ 3\mod_1241617696391_79.doc @ 32646@
Measuring range/
standard scaling
Maximum scaling
0 to 100 Pa -50 to 150
0 to 10 hPa -5 to 15
0 to 20 hPa
-10 to 30
0 to 50 hPa -25 to 75
0 to 100 hPa -50 to 150
0 to 500 hPa -250 to 750
0 to 1000 hPa -500 to 1500
0 to 2000 hPa -1000 to 3000
-100 to 100 Pa -200 to 200
-10 to 10 hPa -20 to 20

4 Transmitter
13
Measuring range/
standard scaling
Maximum scaling
-20 to 20 hPa -30 to 30
-50 to 50 hPa -100 to 100
-100 to 100 hPa -200 to 200
-500 to 500 hPa
-1000 to 1000
-1000 to 1000 hPa -2000 to 2000
-2000 to 2000 hPa -4000 to 4000
Pos: 29 /T D/ Ü b ers chr if t en/ MU F /1 .3/ 2. 3 Inb etri e bn ahm e @ 3\m od_1234258805768_79.doc@ 24027@ 2
4.3. Commissioning
Pos: 30 /T D/ E rs t e Sc hri t te /M UF 63xx /W an dm o nt ag e 6 35 x_ 63 2 1 @ 3\mod_1236065568006_79.doc@ 27063 @ 3455
4.3.1. Assembling the instrument
4.3.1.1. Wall mounting
Attaching rear panel bracket
1. Remove locking screw (see item (4) of drawing below) and
detach rear panel bracket from plastic bracket (see item (2) of
drawing below).
2. Hold rear panel bracket in assembly position and mark the three
drill holes.
3. Drill three holes (Ø 5 mm) and insert dowels where necessary.
4. Screw on rear panel bracket.
Remember that the clamping brackets (1) must face the wall.

4 Transmitter
14
Fastening instrument to rear panel bracket
1. Slide plastic bracket (2) on the back of instrument onto rear
panel bracket until it engages (see arrows).
2. Insert screw (4) through hole (3) and screw into rear panel
bracket.
Pos: 31 /T D/ E rs t e Sc hri t te /M UF 63xx /G er ät v erdr ah te n 6321 @ 3\mod_1241618087115_79.doc@ 32667@ 344
4.3.2. Wiring the instrument
WARNING
Electrical voltage!
> When routing cables, ensure that there is a space between the
signal line and the interfering external lines.
> If electromagnetic interference is likely, use a shielded and/or
twisted cable. Connect the shield to the earth on the side
facing away from the transmitter.
> If overvoltages are likely, install overvoltage protection devices.
WARNING
Electrical voltage
Danger of injury!
> De-energize the mains connection before connecting the
transmitter.

4 Transmitter
15
CAUTION
Damage to electronic components!
> The terminal strip can be removed from the circuit board in
order to screw on the cable ends. After wiring, be sure to
completely attach the terminal strip on the contact pins as pre-
assembled.
Only have the transmitter wired and connected by
authorized personnel with the voltage disconnected.
1. Loosen and remove housing screws (1).
2. Remove upper part of housing (2) from lower part of housing (3)
and place on a clean surface.
3. Wire instrument (see 3-wire system page 16, 4-wire system
page 16).
4 Place upper part of housing (2) on lower part of housing (3) and
tighten with housing screws (1).
5. On instruments with display: Remove protective film (4) from
the display cutout of the housing cover.

4 Transmitter
16
4.3.2.1. 4-wire system
Voltage output (4-wire, 0 to 1 V/0 to 5 V/0 to 10 V)/voltage output
(4-wire, 4 to 20 mA):
U = 20 to 30 V DC/AC
4.3.2.2. 3-wire system
All ground connections are connected to one another (= one
collective ground connection).
CAUTION
Destruction of the instrument in the event of incorrect
polarity!
> When connecting the voltage supply, place collective ground
connection on PIN 2!
Pos: 32 /T D/ Ü b ers chr if t en/ MU F /1 .6 W art ung u nd R ei ni g ung @ 3\mod_1234443039129_79.doc@ 24982@ 2

4 Transmitter
17
4.4. Maintenance and cleaning
Pos: 33 /T D/ P ro d ukt i nst a nd halt en /MU F 63xx /G e häus e r ei ni g e n 63 2 1 @ 3\mod_1241092795617_79.doc@ 32303@ 3
4.4.1. Cleaning housing
• Only clean the housing carefully with a moist cloth.
• Do not use aggressive cleaning agents.
• Do not use any solvents.
Pos: 34 /T D/Pro dukt ver wende n/M UF 63x x /S ta tus -W ar nm eld ung en/N am ur F ehl er bedi ng u nge n 63 2 1 @ 4 \mod_1253024846257_79.doc @ 50363@ 3
4.4.2. Namur fault conditions
If the faults named in the following table occur, the analog outputs
output special values that enable a general fault warning in the
higher-level control system. The values correspond to the "Namur"
industry standard.
Status
message in
the display
Class
Display
value in the
display
Analog output
4 to 20 mA
1 V
5 V
10 V
Watchdog
error
Error Previous
value stops
3.8 mA 1.1 V 5.5 V 11 V
Value below
min. scale
Underrange Reading 3.8 mA 0 V 0 V 0 V
Value above
max. scale
Overrange Reading 20.5 mA 1.1 V 5.5 V 11 V
Pressure too
high
Overrange ooooo · 20.5 mA 1.1 V 5.5 V 11 V
Pos: 35 /T D/Ü berschr ifte n/MUF /3 Param etri er-, A bg lei ch und A nalys esoftw are (P2 A-Softwar e) @ 3\m od_1234258523713_79.doc@ 23932@ 1

5 Parameterizing, adjusting and analyzing software (P2A software)
18
5 Parameterizing, adjusting and
analyzing software (P2A
software)
Pos: 36 /T D/ Ü b ers chr if t en/ MU F /1 .1/ 2. 1/ 3. 1 Leist ung s bes chr eib ung @ 3\mod_1234258595211_79.doc@ 23951@ 2
5.1. Specifications
Pos: 37 /T D/ L ei st ung s bes c hr ei b ung/ V er we nd u ng /M UF 63x x/M UF 63x x P2A @ 3\mod_1234258967326_79.doc @ 24065@ 355
The P2A software is used for the parameterization, adjustment and
analysis of Testo transmitters. The following applies:
• If a Testo transmitter is bought at a later stage (and is therefore
more recent than the existing P2A software version), a software
update is required.
• You can find the free upgrade on the Testo website
www.testo.com under the product-specific downloads.
The software therefore only needs to be bought once, even for
owners of several Testo transmitters.
5.1.1. Functions and use
In the P2A software, two different file types are used: The
instrument and the parameter file.
Instrument file
The parameters of a particular transmitter are stored in its so-called
instrument file. Using this file, the parameters can be edited and the
instrument can be tested and adjusted.
Instrument files also contain the respective histories in addition to
the parameter data, i.e. "log books" are kept for the previous
parameterizations, adjustments and messages (see 5.4.5
Transmitter history).
Instrument files are ".cfm" format files.

5 Parameterizing, adjusting and analyzing software (P2A software)
19
Parameter file
Parameter files are not tied to a specific individual transmitter and
contain only parameter data/no history data.
If you use various instruments of the same type, you can create
parameter files once (e.g. by saving the appropriate instrument file
as the parameter file) and transmit these onto the other
instruments.
Parameter files are ".cfp" format files.
Pos: 38 /T D/ L eist ung s bes chr eib ung / Sys tem v ora uss etz u ng e n/MU F 63xx @ 3\m od_1234260654399_79.doc@ 24084@ 3555
5.1.2. System requirements
Operating system
• Windows®7
• Windows®8
• Windows®10
Computer
The computer must meet the requirements of the respective
operating system. In addition, the following requirements must be
met:
• Interface USB 1.1 or higher
• Graphics resolution min. 1024 x 768
Date and time settings are automatically transferred from
the PC. The administrator must ensure that the system time
is regularly synchronized with a reliable time source and, if
necessary, adjusted to ensure the authenticity of the data.
Pos: 40 /TD/Ü berschr iften/MU F/3.2 Er ste Schr itte @ 3\mod_1234258633304_79.doc @ 23970@2

5 Parameterizing, adjusting and analyzing software (P2A software)
20
5.2. First steps
Pos: 41 /T D/ E rs t e Sc hri t te /M UF 63xx /P 2 A/ So ftw ar e/T r ei ber ins t all ier e n @ 3 \mod_1234261192065_79.doc@ 24123@ 3444
5.2.1. Installing the software/driver
Administrator rights are required.
5.2.1.1. Installing P2A software
Without the input of a licence key, the software will run only
as a demo version (time limit 30 days).
1. You can download the software under the following link:
https://www.testo.com/download-center
If the installation program does not start automatically:
> Open download folder and start P2A.exe.
2. Follow the instructions of the installation wizard.
3. Click on [Finish] to complete the software installation
5.2.1.2. Installing USB driver
1. You can download the software under the following link:
https://www.testo.com/download-center (Testo USB driver)
If the installation program does not start automatically:
> Open download folder and start USBDriver.exe.
2. Follow the instructions of the installation wizard.
3. Click on [Finish] to complete the software installation
5.2.1.3. P2A software upgrade
1. You can find the P2A software upgrade on the Testo website
www.testo.com under the product-specific downloads.
Download and save P2A software upgrade.
2. Start P2A upgrade.exe file.
3. Follow the instructions of the installation wizard.
Pos: 42 /T D/ E rs t e Sc hri t te /M UF 63xx /P 2 A/ So ftw ar e s tar t e n _6 3 21 @ 3\m od_1242285508828_79.doc@ 32883@ 3444
Other manuals for 6321
2
Table of contents
Other TESTO Transmitter manuals
Popular Transmitter manuals by other brands

sielco
sielco EXC GX Series User and maintenance manual

Tait
Tait TMAA11-01 installation instructions

Vaisala
Vaisala Indigo 201 quick guide

Sentera Controls
Sentera Controls RSMFX-3 Mounting and operating instructions

Comnet
Comnet FVR1031 series Installation and operation

Suunto
Suunto Cadence Pod user guide