Arjo Alpha Active 5 User manual

Alarm Mute – Press the Alarm Mute button to mute an alarm, the alarm will be silenced
f
or 15 minutes. The
indicator remains on but the alarm is muted. Press the button a
g
ain to terminate the mute state manuall
y
.
This
Q
uick Re
f
erence
G
uide does not replace the Alpha Active 5 Instructions For Use
(
IFU
)
or the
g
eneral sa
f
et
y
warnin
g
s,
c
autions and recommendations provided in the IFU. Be
f
ore usin
g
the product read the IFU, part number 648931EN.
Contraindication
Do not use Alpha Active 5 s
y
stem
f
or patients with unstable spinal
f
ractures.
S
ee IFU
f
or
f
ull description.
P
atient
/
Resident assessmen
t
The Alpha Active 5 s
y
stem is
f
or use with patients ran
g
in
g
f
rom 4
0
k
g
(
88 lb
)
to 20
0
k
g
(
441 lb
)
.
S
ee IFU
f
or
f
ull description. I
f
the
p
atient does not meet these criteria an alternative equipmen
t
/
sy
stem shall be used.
$XWR¿UPPRG
H
±
6
HOHFWVWKH$XWR
¿
UPQRQDOWHUQDWLQ
J
PRGH7KH
\
HOORZLQGLFDWRUZLOOLOOXPLQDWHZKHQWKH
S
XPSLVLQWKLVPRGH$XWR
¿
UPODVWVPLQXWHVD
I
WHUZKLFKWKHSXPSZLOOGH
I
DXOWEDFNWRWKHSUHYLRXVVHWWLQ
J

Comfort Control – Press plus
(
+
)
or minus
(
-
)
to ad
j
ust the mattress pressure, actin
g
as a com
f
ort control. The
S
UHVVXUHVHWWLQ
J
LVLQGLFDWHGE
\
WKH
J
UHHQ/('LQGLFDWRUVEHWZHHQWKHSOXV



DQGWKHPLQXV



EXWWRQV)RUVSHFL
¿
F
c
linical needs the semi-automatic s
y
stem
(
Active mode
)
can be converted to a manual-onl
y
s
y
stem
,
see p.2. In
P
DQXDORQO
\
WKHVWDWXVLVLQGLFDWHGE
\

À
DVKLQ
J
/('VZKHQWKHRSHUDWRUPDQXDOO
\
DG
M
XVWVWKHSUHVVXUHVHWWLQ
J

QUICK REFERENCE GUIDE
Alpha Active 5
Lock/Unlock
– Press and hold the button for more than 3 seconds to lock or unlock. Lock mode will engage after
k
3 minutes without manual opperation. The pump needs to be unlocked be
f
ore an
y
operation.
Th
e
Low Pressure indicator
is illuminated when the pump detects low pressure within the mattress
r
r
eplacement. An audible alarm will sound, the audible alarm can be muted
b
y
the mute button. The indicator turns
R
ႇ
RQFHQRUPDOSUHVVXUHLVUHDFKHG
Th
e
Service/Pump fault indicator
will illuminate and remain on if the pump has detected an internal fault.
r
C
ontact a
S
ervice En
g
ineer i
f
the indicator is illuminated.
Th
e
P
ower Fail indicato
r
will illuminate when a mains power failure has been detected. An audible alarm will
r
V
RXQGXQWLOSRZHULVUHVXPHGRUWKHSXPSLVVZLWFKHGR
ႇ
XVLQ
J
WKH
2
Q
2ႇ
EXWWRQ
2Q2ႇEXWWRQ±SUHVVLQ
J
WKH
2
Q
2ႇ
EXWWRQZLOODFWLYDWHWKHSXPS7KHSXPSZLOOSHU
I
RUPDVHO
I
WHVWGXULQ
J
ZKLFK
DOO/('VZLOOLOOXPLQDWHDQGDXGLEOHEHHSVZLOOVRXQG'XULQ
J
DWOHDVWPLQXWHVWKHV
\
VWHPVWDUWVSUHLQ
À
DWLQ
J
the mattress pressure to de
f
ault value
(
Equivalent to patient wei
g
ht 100 k
g)
, indicatin
g
the status on the com
f
ort
c
ontrols
8
LEDs.
Alpha Active 5
M
ode Selec
t
– Press the Mode Select button to chan
g
e the mode of operation. Two modes of operation are
available, Active
(
Alternatin
g)
and Reactive
(
Static
)
.
6
48932EN_02 • 05
/
2022 • Pa
g
e 1
/
3
M
attress Installation
Zip the protective cover over the mattress. Make sure the Ar
jo
l
o
g
o is
uppermost and at the foot end of the mattress.
M
ake sure the
C
PR unit is secured in its closed position.
S
e
e
Fi
g
ur
e
1.
P
u
mp
Install
a
tion
Position the pump,
f
eet down, on a horizontal sur
f
ace or suspend
f
rom the
bed
f
oot rail b
y
means o
f
the inte
g
ral han
g
in
g
brackets.
C
onnect the mattress tube-set to the pump until it clicks into place, check
that tube-set is not twisted. Make sure it is securel
y
connected to the pump.
S
e
e
Fi
g
ur
e
2.
Insert the mains power plu
g
into a suitable mains power socket.
Figure 2
Figure 1
CPR
Quick Star
t
3UHVVWKH
2
Q
2ႇ
EXWWRQRQWKHVLGHR
I
WKHSXPS$OORZ
DSSUR[LPDWHO
\
PLQXWHVIRUWKHPDWWUHVVWRIXOO
\
LQÀDWH
WARNIN
G
D
o not place the patient on the mattress until it is full
y
L
QÀDWHG

Active Mod
e
Su
pp
or
t
Settin
g
Procedur
e
Place a sheet over the mattress replacement and tuck it in loosel
y
.
M
ake sure the
C
PR unit is clearl
y
visible at the head end o
f
the bed
.
1. Lie or sit the patient on the mattress
.
2. Press and hold both the plus
(
+
)
and minus
(
–
)
buttons
f
or at
l
east 3 seconds to reset the mattress pressure to the de
f
ault
c
om
f
ort
(
Equivalent to patient wei
g
ht 100 k
g)
control level
(
4
LEDs li
g
ht up
)
.
3
. The
S
emi-automatic
f
unctionalit
y
o
f
the pump operates onl
y
when in de
f
ault mode, Active
(
Alternatin
g)
. This
f
unctionalit
y
s
enses the patient wei
g
ht and ad
j
usts the pressure in the
c
ells. This function provides onl
y
an assisted ad
j
ustment o
f
mattress cell pressures within a pre-set patient wei
g
ht settin
g
r
an
g
e, operatin
g
in con
j
unction with the care
g
iver’s selection o
f
p
ressure settin
g
. Within each wei
g
ht settin
g
ran
g
e,
f
or individual
ad
j
ustments, care
g
iver selection of pressure settin
g
should
be considered in order to ensure selection of appropriate cell
p
ressures to support the patient
.
4. Within 20 minutes the pump has ad
j
usted to the correct
press
u
re.
Reactive Mod
e
Provides a stable, non-movin
g
support sur
f
ace
f
or instances where
an active support sur
f
ace is contra-indicated. In Reactive mode the
V
XSSRUWVXUIDFHUHPDLQVFRQVWDQW

DOOFHOOVDUHHTXDOO
\
LQÀDWHG


Additional nursin
g
assessment must be undertaken in order to direct
an individualized repositionin
g
pro
g
ramme.
In Reactive mode the pump automaticall
y
reduces the pressure in
the mattress to increase patient com
f
ort.
NOT
E
,WPD
\
EHQHFHVVDU
\
IRUWKHFDUH
J
LYHUWRVHWWKHFRPIRUW
VHWWLQ
J
DFFRUGLQ
J
WRSDWLHQWZHL
J
KWEHIRUHXVLQ
J
WKH
V
\
VWHP

At lower patient wei
g
hts
(
for example, lower than 60 k
g)
, it
P
D
\
EHQHFHVVDU
\
WRFRQVLGHUXVHRIWKHV
\
VWHP¶V¶PDQXDO
RQO
\
¶PRGHWRHQVXUHSUHVVXUHVHWWLQ
J
LVVSHFL¿FDOO
\
VHW
D
FFRUGLQ
J
WRSDWLHQWZHL
J
KW

$WKL
J
KHUSDWLHQWZHL
J
KWVLWPD
\
EHQHFHVVDU
\
WRFRQVLGHU
X
VHRIWKHV
\
VWHP¶V¶PDQXDORQO
\
¶PRGHDQGWRFKHFNIRU
S
DWLHQWERWWRPLQ
J
RXW

7
KHVHDG
M
XVWPHQWVDQGFKHFNVVKRXOGEHLQFRQ
M
XQFWLRQ
with inde
p
endent clinical determination of a
pp
ro
p
riate
VX
SS
RUW

6
48932EN_02 • 05
/
2022 • Pa
g
e 2
/
3
Chan
g
es in Patient Positi
o
n
When the patient chan
g
es position
f
rom l
y
in
g
to sittin
g
, or reversed,
LWPD
\
EHQHFHVVDU
\
LQRUGHUWRPD[LPLVHWKHEHQH
¿
WR
I
WKHVXSSRUW
s
ur
f
ace, to make ad
j
ustments to the mattress pressure.
From L
y
in
g
to Sittin
g
– increase pressure control.
From Sittin
g
to L
y
in
g
– decrease pressure control.
This ad
j
ustment should be in con
j
unction with independent clinical
determination o
f
appropriate support.
M
anual-onl
y
s
y
ste
m
Convert to manual-onl
y
s
y
stem:
Press and hold the minus
(
-
)
button at least 10 seconds in unlock
m
ode
.
Return to semi-automatic s
y
stem
:
Press and hold the plus
(
+
)
button at least 10 seconds in unlock
m
ode
.
How to unplu
g
the heel zone valves
:
There are 5 Heel Zone cells located in the 16 Active
(
Alternatin
g)
F
HOOVWKDWFDQEHVHOHFWLYHO
\
DQGSHUPDQHQWO
\
GH
À
DWHG
Press the button
(
1
)
on the valve and pull on the tube
(
2
)
to separate
the two parts in order to disconnect and permanentl
y
vent the upper
S
DUWRIDKHHOFHOOWRSURYLGHKHHORႉRDGLQ
J

S
e
e
Fi
g
ure.
3
.
When this
f
unction is no lon
g
er required then reconnect the two parts
RI
WK
H
Y
D
OY
H
W
R
U
H
LQ
ÀD
W
H
WK
H

FH
OO
Figure 3
T
rans
p
ort Mod
e
Transportin
g
the patient on the Alpha Active 5 mattress overla
y
.
1. Disconnect the tube-set
f
rom the pump.
S
e
e
Fi
g
ur
e
4.
2. Place the cap over the end o
f
the tubeset.
S
e
e
Fi
g
ur
e
5
.
With the cap attached the mattress will automaticall
y
switch to
transport mode
.
M
attress support will remain
f
or up to 12 hours in Transport Mode.
To resume normal operation, reconnect the tubeset and run the pump
.
Figure 4 Figure 5
NOT
E
$PD[LPXPRIWZRKHHOFHOOVDUHDOORZHGWREHPDQXDOO
\

GHÀDWHGDWDQ
\
WLPH
CAUTIO
N
7
UDQVSRUWPRGHWHPSRUDULO
\
RႇHUVVXSSRUWIRUXSWR
KRXUV,WLVUHFRPPHQGHGWKDWZKHQLQWUDQV
S
RUWPRGH
WKHSDWLHQWLVIUHTXHQWO
\
PRQLWRUHG

®
and ™ are trademarks belon
g
in
g
to the Ar
j
o
g
roup o
f
companies.
© Ar
j
o 2022.
As our polic
y
is one o
f
continuous improvement, we reserve the ri
g
ht to modi
fy
desi
g
ns without prior notice. The
c
ontent of this publication ma
y
not be copied either whole or in part without the consent of Ar
j
o.
R
APID DEFLAT
E
,17+((9(17
2
)
&
$5',$
&
$55(
67
In the event o
f
that
C
PR needs to be administered
:
7
RDFWLYDWH5D
S
LG'HÀDW
H
:
3XOOWKHUHGVWUDSODEHOOHG
&
35
I
RUWKHPDWWUHVVWRGH
À
DWH
7
RUHVHW5D
S
LG'HÀDWH

R
eplace the stopper securel
y
and reconnect the air tubes
to the pump.
Ar
j
oHuntlei
g
h AB
H
ans Michelsens
g
atan 10
211 20 Malmö,
S
wede
n
www.ar
j
o.com
6
4
893
2E
N
6
48932EN_02 • 05
/
2022 • Pa
g
e 3
/
3

Other manuals for Alpha Active 5

1

Other Arjo Medical Equipment manuals

Arjo Lifeguard LG55 User manual

Arjo

Arjo Lifeguard LG55 User manual

Arjo AtmosAir User manual

Arjo

Arjo AtmosAir User manual

Arjo MAA8000 User manual

Arjo

Arjo MAA8000 User manual

Arjo Calypso User manual

Arjo

Arjo Calypso User manual

Arjo MAXI MOVE Operating instructions

Arjo

Arjo MAXI MOVE Operating instructions

Arjo MAA1150 User manual

Arjo

Arjo MAA1150 User manual

Arjo Auralis User manual

Arjo

Arjo Auralis User manual

Arjo Auralis User manual

Arjo

Arjo Auralis User manual

Arjo Enterprise 9000X User manual

Arjo

Arjo Enterprise 9000X User manual

Arjo Pentaflex User manual

Arjo

Arjo Pentaflex User manual

Arjo Flowtron ACS900 User manual

Arjo

Arjo Flowtron ACS900 User manual

Arjo Sara 3000 User manual

Arjo

Arjo Sara 3000 User manual

Arjo Dual-Loop Attachment Straps User manual

Arjo

Arjo Dual-Loop Attachment Straps User manual

Arjo MAA2000-XS User manual

Arjo

Arjo MAA2000-XS User manual

Arjo AtmosAir Velaris - Plus Flex User manual

Arjo

Arjo AtmosAir Velaris - Plus Flex User manual

Arjo Prioma Semi Electric Couch Series User manual

Arjo

Arjo Prioma Semi Electric Couch Series User manual

Arjo Maxi Air User manual

Arjo

Arjo Maxi Air User manual

Arjo Sapphire Comfort Turn User manual

Arjo

Arjo Sapphire Comfort Turn User manual

Arjo Century User manual

Arjo

Arjo Century User manual

Arjo Miranti User manual

Arjo

Arjo Miranti User manual

Arjo Flowtron Excel User manual

Arjo

Arjo Flowtron Excel User manual

Arjo TheraKair Visio User manual

Arjo

Arjo TheraKair Visio User manual

Arjo Maxi 500 User manual

Arjo

Arjo Maxi 500 User manual

Arjo Miranti User manual

Arjo

Arjo Miranti User manual

Popular Medical Equipment manuals by other brands

cobas b 101 Quick reference guide

cobas

cobas b 101 Quick reference guide

HoMedics MySpa instruction manual

HoMedics

HoMedics MySpa instruction manual

mychway MS-11R5 user manual

mychway

mychway MS-11R5 user manual

Ossur Unloader One Plus Instructions for use

Ossur

Ossur Unloader One Plus Instructions for use

Lloyd COSMIC III Operation manual

Lloyd

Lloyd COSMIC III Operation manual

Riester ri-focus LED instructions

Riester

Riester ri-focus LED instructions

Freedom Innovations Kinnex 2.0 Instructions for use

Freedom Innovations

Freedom Innovations Kinnex 2.0 Instructions for use

Aesculap Acculan 4 Instructions for use/Technical description

Aesculap

Aesculap Acculan 4 Instructions for use/Technical description

Esogetics synapsis home 2 instruction manual

Esogetics

Esogetics synapsis home 2 instruction manual

Braemar SEER Event Recorder instructions

Braemar

Braemar SEER Event Recorder instructions

Boston Scientific Vercise DBS Remote Control 3 Handbook

Boston Scientific

Boston Scientific Vercise DBS Remote Control 3 Handbook

Joerns Oxford User instruction manual

Joerns

Joerns Oxford User instruction manual

DiaMedical INFINITY 1200 Quick reference guide

DiaMedical

DiaMedical INFINITY 1200 Quick reference guide

Gima QV-600 quick start guide

Gima

Gima QV-600 quick start guide

Provizio SEM Scanner quick start guide

Provizio

Provizio SEM Scanner quick start guide

Leisegang 1E LED Instructions for use

Leisegang

Leisegang 1E LED Instructions for use

BEMIS BemisSentinel 4008 manual

BEMIS

BEMIS BemisSentinel 4008 manual

Chattanooga Artromot Elbow Patient instructions

Chattanooga

Chattanooga Artromot Elbow Patient instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.