sampling dish. Correct operation can make sure there is no cross infection of the samples in the
cabinet.
(h) Note: The operator’s arm and hand maybe the pollution source, so the operator must wear
sterilized gloves or wash hands using sterilization soap. The used gloves should be gotten off from
the sleeve of lab-gown.
(i) When the room air or adjacent working area air is serious polluted, we should take measures to
prevent the particle from entering into the clean working area.
(j) The most basic is placing the instruments to the right place. If don’t know the place, using
airflow meter to check, to make sure the right airflow pattern.
(k) Don’t block the air grill, especially after we used the sterile rag.
2.1.5 Circuit of digital control
(a) Our Biosafety Cabinet use the latest microprocessor control technology to monitor your safety
cabinet’s function, easy to operate, with friendly interface and automatic diagnostic function.
(b) Connect safety cabinet to the recommended power interface.
(c) After power on, the LCD display will display the cabinet’s working time, and filter’s working
time and the life time. When push the power button, the LCD display background lights up, the
system will be in standby mode.
(d) After power on, the equipment will check itself automatically. When the front window is too
high, it will alarm buzzer, and the LCD display will show “!” and flicker, now you can adjust the
height of the front window. If the differential pressure of the supply filter and exhaust filter are too
big, there will be audio and visual alarm, then we need to replace the filter, to ensure the safety of
operator.
(e) After self-examination, the equipment will be in standby state.
(f) The fan run by activating the switch on the control panel, the Fan sign on the LCD display will
light, meanwhile, the motor will start, and it will reach the normal speed after 10-15 seconds.
(g) When activate the fan switch again, the motor and the sign of fan will close, and the motor will
shut down after 10-15 seconds.
(h) Three-pin socket is controlled by the relevant key on the control panel, the max current is 5A. If
there are two three-pin sockets, they will be controlled at the same time.
(i) The digital controller can adjust all velocity of the fan, and it can provide the linear compensation
for voltage fluctuation.
(j) The airflow volume will be displayed and monitored through the pressure, and will be sampled
by the microprocessor. When the airflow volume exceed the normal value, the buzzer will alarm
continuously.
(k) Before seeking the maintenance service help, pull the plug off the wall, then plug in to restart the
motor after about 2 minutes, the safety cabinet will restart and run well. But if the cabinet can’t start,
then seek the service.
2.1.6 Airflow velocity indication.
Airflow velocity indication is inflow velocity and down flow velocity.
2.1.7 Differential pressure indication
Differential pressure indication is the differential pressure between upper and lower filter, and it will
increase when the filter loading increase. The value after the Exhaust filter is the exhaust filter
differential pressure, the value after the downflow filter is the supply filter differential pressure.