
17
strongly recommended when selecting the appropriate device size (Table
10.5.1). Appropriate device oversizing has been incorporated into the IFU
sizing guide. Sizing outside of this range can result in endoleak, fracture,
migration, device infolding or compression.
4.5 Implant Procedure
(Refer to Section 11, DIRECTIONS FOR USE)
•Appropriateproceduralimagingisrequiredtosuccessfullypositionthe
Zenith Spiral-Z AAA Iliac Leg and assure accurate apposition to the vessel
wall.
•Donotbendorkinkthedeliverysystem.Doingsomaycausedamageto
the delivery system and the Zenith Spiral-Z AAA Iliac Leg Graft.
•Toavoidanytwistintheendovasculargraft,duringanyrotationof
the delivery system, be careful to rotate all components of the system
together (from outer sheath to inner cannula).
•Donotcontinueadvancinganyportionofthedeliverysystemif
resistance is felt during advancement of the wire guide or delivery system.
Stopandassessthecauseofresistance;vessel,catheterorgraftdamage
may occur. Exercise particular care in areas of stenosis, intravascular
thrombosis, or in calcified or tortuous vessels.
•Inadvertentpartialdeploymentormigrationoftheendoprosthesismay
require surgical removal.
•Unlessmedicallyindicated,donotdeploytheZenithSpiral-ZAAAIliac
Leg in a location that will occlude arteries necessary to supply blood flow
to organs or extremities.
•Donotattempttore-sheaththegraftafterpartialorcomplete
deployment.
•Inaccurateplacementand/orincompletesealingoftheZenithSpiral-Z
AAA Iliac Leg within the vessel may result in increased risk of endoleak,
migration or inadvertent occlusion of the internal iliac arteries.
•InadequateoverlapoftheZenithSpiral-ZAAAIliacLegmayresultin
increased risk of migration of the stent graft. Incorrect deployment or
migration of the endoprosthesis may require surgical intervention.
•Systemicanticoagulationshouldbeusedduringtheimplantprocedure
based on hospital- and physician-preferred protocol. If heparin is
contraindicated, an alternative anticoagulant should be considered.
•ToactivatethehydrophiliccoatingontheoutsideoftheFlexor
introducer sheath, the surface must be wiped with sterile gauze pads
soaked in saline solution. Always keep the sheath hydrated for optimal
performance.
•Minimizehandlingoftheconstrainedendoprosthesisduringpreparation
and insertion to decrease the risk of endoprosthesis contamination and
infection.
•Maintainwireguidepositionduringdeliverysysteminsertion.
•Fluoroscopyshouldbeusedduringintroductionanddeploymentto
confirm proper operation of the delivery system components, proper
placement of the graft, and desired procedural outcome.
•TheuseoftheZenithSpiral-ZAAAIliacLegwiththeZ-TrakIntroduction
System requires administration of intravascular contrast. Patients with
pre-existing renal insufficiency may have an increased risk of renal failure
postoperatively. Care should be taken to limit the amount of contrast
media used during the procedure and to observe preventative methods
of treatment to decrease renal compromise (e.g., adequate hydration).
•Asthesheathand/orwireguideiswithdrawn,anatomyandgraftposition
may change. Constantly monitor graft position and perform angiography
to check position as necessary.
•Usecautionduringmanipulationofcatheters,wiresandsheaths
within an aneurysm. Significant disturbances may dislodge fragments
of thrombus, which can cause distal embolization, or may rupture the
aneurysm.
•Avoiddamagingthegraftordisturbinggraftpositioningafterplacement
in the event reinstrumentation (secondary intervention) of the graft is
necessary.
•Verifythatthepredeterminedcontralateraliliaclegisselectedfor
insertion on the contralateral side of the patient before implantation.
4.6 Molding Balloon Use
•Donotinflatetheballooninthevesseloutsideofthegraft,asdoingso
may cause damage to the vessel. Use the balloon in accordance with its
labeling.
•Usecareininflatingtheballoonwithinthegraftinthepresenceof
calcification, as excessive inflation may cause damage to the vessel.
•Confirmcompletedeflationoftheballoonpriortorepositioning.
•Foraddedhemostasis,theCaptorHemostaticValvecanbeloosenedor
tightened to accommodate the insertion and subsequent withdrawal of a
molding balloon.
4.7 MRI Information
Non-clinical testing has demonstrated that the Zenith Spiral-Z AAA Iliac
Leg Graft is MR Conditional. It can be scanned safely immediately after
placement under the following conditions:
Static Magnetic Field
•Staticmagneticfieldof3.0Teslaorless.
•Highestspatialmagneticgradientfieldof720Gauss/cm
Non-clinical evaluation was conducted in a 3 Tesla MR system (General
Electric Excite) with a maximum spatial magnetic gradient field of
720 Gauss/cm as measured with a gaussmeter in the position of the static
magnetic field pertinent to the patient (i.e., outside of scanner covering,
accessible to a patient or individual).
MRI-Related Heating
1.5 Tesla Systems:
•Staticmagneticfieldof1.5Tesla
•Maximumwhole-body-averagedspecificabsorptionrate(SAR)of2W/kg
for 15 minutes of scanning ( i.e., per scanning sequence)
In non-clinical testing, the Zenith Spiral-Z AAA Iliac Leg produced a
temperature rise of less than or equal to 2.1 °C at a maximum whole-body-
averaged specific absorption rate (SAR) of 2.9 W/kg, for 15 minutes of MR
scanning in a 1.5 Tesla, Siemens Medical Magnetom, Numaris/4 Software,
Version Syngo MR 2002B DHHS MR Scanner. The maximum whole-body-
averaged specific absorption rate (SAR) was 2.9 W/kg, which corresponds to
a calorimetry measured value of 2.1 W/kg.
3.0 Tesla Systems:
•Staticmagneticfieldof3.0Tesla
•Maximumwhole-body-averagedspecificabsorptionrate(SAR)of2W/kg
for 15 minutes of scanning( i.e., per scanning sequence)
In non-clinical testing, the Zenith Spiral-Z AAA Iliac Leg produced a
temperature rise of less than or equal to 2.6 °C at a maximum whole-body-
averaged specific absorption rate (SAR) of 2.9 W/kg, for 15 minutes of MR
scanning in a 3.0 Tesla Excite, GE Electric Healthcare, 14X.M5 Software,
MR Scanner. The maximum whole-body-averaged specific absorption rate
(SAR) was 2.9 W/kg, which corresponds to a calorimetry measured value of
2.7 W/kg.
Image Artifact
The image artifact extends throughout the anatomical region containing the
device, obscuring the view of immediately adjacent anatomical structures
within approximately 5 cm of the device, as well as the entire device and its
lumen, when scanned in nonclinical testing using the sequence: Fast spin
echo, in a 3.0 Tesla, Excite, GE Electric Healthcare, with G3.0-052B Software,
MR system with body radiofrequency coil.
For all scanners, the image artifact dissipates as the distance from the device
to the area of interest increases. MR scans of the head and neck and lower
extremities may be obtained without image artifact. Image artifact may be
present in scans of the abdominal region, depending on distance from the
device to the area of interest.
Cook recommends that the patient register the MR conditions disclosed in
this IFU with the MedicAlert Foundation. The MedicAlert Foundation can be
contacted in the following manners.
Mail: MedicAlert Foundation International
2323 Colorado Avenue
Turlock, CA 95382
Phone: 888-633-4298 (toll free)
209-668-3333 from outside the US
Fax: 209-669-2450
Web: www.medicalert.org
5 ADVERSE EVENTS
5.1 Observed Adverse Events
For information on observed adverse events in patients receiving Zenith
AAA Endovascular Grafts, please refer to the appropriate IFU from the Zenith
AAA Endovascular Graft family of products. A copy is available online at
www.cookmedical.com.
5.2 Potential Adverse Events
Adverse events that may occur and/or require intervention include, but are
not limited to:
•Amputation
•Anestheticcomplicationsandsubsequentattendantproblems(e.g.,
aspiration)
•Aneurysmenlargement
•Aneurysmruptureanddeath
•Aorticdamage,includingperforation,dissection,bleeding,ruptureand
death
•Arterialorvenousthrombosisand/orpseudoaneurysm
•Arteriovenousfistula
•Bleeding,hematomaorcoagulopathy
•Bowelcomplications(e.g.,ileus,transientischemia,infarction,necrosis)
•Cardiaccomplicationsandsubsequentattendantproblems(e.g.,
arrhythmia, myocardial infarction, congestive heart failure, hypotension,
hypertension)
•Claudication(e.g.,buttock,lowerlimb)
•Death
•Edema
•Embolization(microandmacro)withtransientorpermanentischemiaor
infarction
•Endoleak
•Endoprosthesis:impropercomponentplacement;incompletecomponent
deployment;componentmigration;componentseparationfromanother
graftcomponent;suturebreak;occlusion;infection;stentfracture;graft
materialwear;dilatation;erosion;puncture;perigraftflow;andcorrosion
•Feverandlocalizedinflammation
•Genitourinarycomplicationsandsubsequentattendantproblems(e.g.,
ischemia, erosion, fistula, incontinence, hematuria, infection)
•Graftornativevesselocclusion
•Hepaticfailure
•Impotence
•Infectionoftheaneurysm,deviceoraccesssite,includingabscess
formation, transient fever and pain
•Lymphaticcomplicationsandsubsequentattendantproblems(e.g.,
lymph fistula)
•Neurologiclocalorsystemiccomplicationsandsubsequentattendant
problems (e.g., stroke, transient ischemic attack, paraplegia, paraparesis,
paralysis)
•Pulmonary/respiratorycomplicationsandsubsequentattendant
problems (e.g., pneumonia, respiratory failure, prolonged intubation)
•Renalcomplicationsandsubsequentattendantproblems(e.g.,artery
occlusion, contrast toxicity, insufficiency, failure)
•Surgicalconversiontoopenrepair
•Vascularaccesssitecomplications,includinginfection,pain,hematoma,
pseudoaneurysm, arteriovenous fistula
•Vesseldamage
•Woundcomplicationsandsubsequentattendantproblems(e.g.,
dehiscence, infection)
•Vascularspasmorvasculartrauma(e.g.,iliofemoralvesseldissection,
bleeding, rupture, death)
Device-Related Adverse Event Reporting
Any adverse event (clinical incident) involving the Zenith Spiral-Z AAA Iliac
Leg should be reported to Cook immediately. For customers inside the
United States, to report an incident, call the Customer Relations Department
at 1-800-457-4500 (24 hour) or 1-812-339-2235. For customers outside the
United States, please call your distributor.
6 SUMMARY OF CLINICAL STUDIES
For clinical study information regarding patients receiving Zenith AAA