Delta OHM HD50R-MB User manual

The quality level of our instruments is the result of the constant development of the
product. This may produce some differences between the information written in this
manual and the instrument you have purchased. We cannot completely exclude the
possibility of errors in the manual, for which we apologize.
The data, images and descriptions included in this manual cannot be legally asserted.
We reserve the right to make changes and corrections with no prior notice.
HD50R-MB
Ethernet Data Logger
ENGLISH
REV. 1.0
20/07/2018

HD50 - 2 - V1.0
TABLE OF CONTENTS
1INTRODUCTION.................................................................................................... 3
2DESCRIPTION....................................................................................................... 4
3CONNECTIONS...................................................................................................... 5
4SOFTWARE ........................................................................................................... 8
5MODBUS ............................................................................................................... 9
6TECHNICAL CHARACTERISTICS .......................................................................... 14
7INSTRUMENT STORAGE ...................................................................................... 15
8SAFETY INSTRUCTIONS...................................................................................... 15
9ORDERING CODES .............................................................................................. 16

HD50 - 3 - V1.0
1INTRODUCTION
The HD50R-MB data logger allows several physical quantities to be monitored in a
large variety of application fields.
To the “Master” RS485 MODBUS-RTU input you can connect a network of sensors for
measuring, for example, temperature, humidity, atmospheric pressure, solar radiation,
wind speed and direction (e.g. an anemometer of the HD52.3D… series can be con-
nected), etc.
As an alternative to the Modbus-RTU protocol, a proprietary protocol can be used to
connect the HD2003 anemometer.
A voltage-free contact input allows connecting a rain gauge with contact output. Rain
gauges with contact output either normally closed or normally open can be connected.
A measurement compensation curve as a function of the rainfall rate can be config-
ured. The data logger calculates the rainfall rate in mm/h (by referring the rainfall
quantity in the last five minutes to an hourly value) and the rainfall quantity in the
last day.
The data logger can be connected to an Ethernet local network and allows the simul-
taneous operation of two communication protocols: proprietary and Modbus TCP/IP.
The data logger manages up to 10 “TCP/IP Client” simultaneously. If the local network
is connected to Internet, the data can be regularly sent to an FTP address, to the
Cloud and via e-mail.
For each detected quantity, two alarm thresholds can be set by the user. Exceeding a
threshold is signaled acoustically, by means of the internal buzzer, and remotely, by
sending alarm e-mails. An alarm hysteresis and a delay in the generation of the alarm
can be configured for each detected quantity.
The PC software HD35AP-S allows configuring the data logger, viewing the real time
measurements, downloading and viewing the data into a database. The HD35AP-S
software allows connecting one data logger at a time. If there are several data loggers
in the local network, the PC software HDServer1 allows detecting automatically all the
data loggers connected to the network and connecting simultaneously to all the data
loggers; it also allows entering the data received by the data loggers into a database,
viewing the data in the database and configuring the main measurement parameters
of the devices (alarms, logging interval, …). The HD35AP-CFR21 software option
(available with both HD35AP-S and HDServer1) allows the protection of recorded data
and configuration in response to FDA 21 CFR part 11 recommendations.
35 mm DIN rail installation.
7…30 Vdc external power supply.

HD50 - 4 - V1.0
2DESCRIPTION
MINI
USB B
NORMAL
NET RST
1
2
3
CLOSE
OPEN
RT
PWR
L2
1
3
7
2
9
8
4 5
6
1. Push-button and short-jumper for restoring the factory LAN configuration.
2. Input for contact rain gauge.
3. Red POWER LED: indicates the presence of the external power supply.
4. Power supply input.
5. RS485 line termination.
6. Input for RS485 Modbus-RTU sensors.
7. RJ45 connector for Ethernet connection.
8. Mini-USB connector.
9. Bicolor NETWORK LED: indicates the status of the connection to the LAN (blinks
green when the unit is in normal operation mode; blinks red to signal problems
in LAN connection).

HD50 - 5 - V1.0
3CONNECTIONS
POWER SUPPLY
CLOSE
OPEN
RT
PWR
V+
7...30 Vdc
V-
RS485 MODBUS-RTU CONNECTION
Thanks to RS485 connection, several sensors can be connected in a multi-point net-
work. The instruments are connected in sequence by means of a shielded cable with
twisted pair wires for signals and a third wire for the ground.
The RS485 ground is isolated from the power supply.
DATA+
DATA-
GND
220Ω ShieldShield
Lmax = 1200m
CLOSE
OPEN
RT
PWR
DATA- DATA+
GND
Line terminations must be placed at the ends of the network. If the instrument is at
one end of the line, insert the termination placing the short jumper next to the RS485
connector between the “RT” and “CLOSE” indications. Otherwise, remove the termina-
tion placing the short jumper between the “RT” and “OPEN” indications.
The cable shield must be connected to both ends of the line.
The cable maximum length depends on the transmission speed and on the cable char-
acteristics. Typically, the maximum length is 1200 m. The data line must be kept se-
parated from any power lines to avoid interferences to the transmitted signal.
Each sensor in the RS485 network is univocally identified by an address ranging within
1 and 247. No more than one sensor with the same address can be present in
the network.
Sensor with RS485
MODBUS-RTU output
Termination
Sensor with RS485
MODBUS-RTU output

HD50 - 6 - V1.0
RAIN GAUGE CONNECTION
MINI
USB B
NORMAL
NET RST
1
2
3
L2
USB CONNECTION
The data logger can be connected to a PC via the mini-USB connector and the CP23
cable.
USB connection doesn’t require the installation of drivers: when the data logger is
connected to a PC, Windows® operating system automatically recognizes the instru-
ment as an HID device (Human Interface Device) and uses the drivers already in-
cluded in the operating system.
ETHERNET CONNECTION
Connect the RJ45 connector of the data logger to a socket of the local network via a
standard Ethernet cable.
The data logger is preset to get a dynamic IP address from the network DHCP server.
The IP address can be displayed and changed by connecting the data logger to a PC via
USB and using the HD35AP-S application software. A static IP address can also be set.
The IP address can also be displayed by connecting the data logger to a PC via Ethernet
and using the HDServer1 software, which can automatically detect the data loggers
connected to the network.
Thanks to the availability of two TCP/IP ports, each of which can operate with proprie-
tary (for the connection with HD35AP-S software) or MODBUS TCP/IP protocol, and
ten sockets (in total, to be divided between the two ports), the data logger allows the
Rain gauge contact

HD50 - 7 - V1.0
simultaneous operation of two communication protocols (proprietary and Modbus
TCP/IP) and manages up to 10 “TCP/IP Client” simultaneously. The default setting of
the ports is the following:
•Port number = 5100 for proprietary protocol (8 sockets)
•Port number = 502 for Modbus TCP/IP protocol (2 sockets)
The Ethernet settings can be changed with the HD35AP-S software. It is possible to
restore the factory parameters by placing the short jumper next to the NET RST
push-button between the “2” and “3” indications and then pressing the NET RST
push-button. After the reset, replace the short jumper between the “2” and “1 (NOR-
MAL)” indications.
If the local network is connected to Internet, the data can be regularly sent to an FTP
address, to the Cloud and via e-mail (as attachments).
Note: if the communication with the Cloud is used, the maximum number of “clients”
with proprietary or MODBUS TCP/IP protocol is nine.

HD50 - 8 - V1.0
4SOFTWARE
The parameters of the instrument (logging parameters, alarm thresholds, quantities
to be acquired, network settings, etc.) can be configured by connecting the instru-
ment to the PC via USB or Ethernet local network and using the HD35AP-S applica-
tion software (see the instructions of the software).
Some basic parameters (alarms, logging interval, user code, …) can also be set with
the HDServer1 application software.
To download the data in a database, the HD35AP-S (it allows connecting one data log-
ger at a time) or HDServer1 (it allows connecting several data loggers simultaneously)
application software and the MySQL database management system (included in the
application softwares packages) must be installed.
HD35AP-CFR21 SOFTWARE OPTION
The HD35AP-CFR21 option allows, in addition to the features of the basic software
(both HD35AP-S and HDServer1), the protection of recorded data and instrument con-
figuration in response to FDA 21 CFR part 11 recommendations. In particular be-
come available:
•The traceability of activities (Audit Trail) performed with the software; for exam-
ple, which users connected and what changes were possibly made to the configu-
ration of the instrument.
•The management of users access for the instrument configuration and viewing of
data in the database. Each user can be assigned a different password for using the
software. There are also three levels of access (Administrator, Super-user and stan-
dard User); for each level, the allowed operations can be defined.
The HD35AP-CFR21 option works with USB hardware key to be connected to any PC con-
nected to the same local network of the PC in which the basic software is installed.

HD50 - 9 - V1.0
5MODBUS
The device general information can be read through the function code 0x2B/0x0E:
•Manufacturer (Delta OHM)
•Model
•Firmware version
The list of MODBUS registers is shown below. Depending on the connected sensors,
some of the registers could not be significant for the system.
The following conventions have been used in the tables:
oType: b= bit, B= 8 bits (Byte), W= 16 bits without sign (Word), SW = 16 bits
with sign
o(x10) = decimal value expressed as an integer (e.g., if the content of the register
is 184, the value is to be intended as 18,4).
o(x100) = centesimal value expressed as an integer (e.g., if the content of the reg-
ister is 500, the value is to be intended as 5,00).
The commands for requesting units of measurement return an index according to the
correspondence indicated in the table below:
Indexes of the units of measurement
Index Unit of
meas. Index Unit of
meas. Index Unit of
meas. Index Unit of
meas. Index Unit of
meas.
0 °C 13 inchHg 26 J/m239 inch 52 l/min
1 °F 14 inchH2O 27 µJ/cm240 counts 53 gallon/min
2 %UR 15 kgf/cm228 V 41 mm/h 54 m3/min
3 g/m316 PSI 29 mV 42 inch/h 55 m3/h
4 g/kg 17 m/s 30 mA 43 counts/h 56 µmol/(m2s)
5 mbar 18 km/h 31 ppm 44 mW/m257 mm/day
6 bar 19 ft/s 32 Hz 45 m 58 kV
7 Pa 20 mph 33 % 46 s 59 A
8 hPa 21 knot 34 degrees
47 µW/lumen 60 kA
9 kPa 22 W/m235 lux 48 dB
10 atm 23 µW/cm236 m2/s 49 dBA
11 mmHg 24 Wh/m237 g
(*) 50 kWh
12 mmH2O 25 kWh/m238 mm 51 l/s 255 Not
defined
(*) Gravity acceleration
Discrete Inputs - Read-only parameters
Address Type Discrete Input description
7 b If 1, at least a quantity is in alarm.
Coils - Read/Write parameters
Address Type Coil description
1 b Logging status: 0=active, 1=inactive
2 b Logging mode: 0=non cyclic, 1=cyclic
3 b Set 1 to delete the device logging memory. Bit zeroing is automatic.
4 b Buzzer activation in case of measurement alarm: 0=no, 1=yes

HD50 - 10 - V1.0
Address Type Coil description
9 b
Protection of configuration with password: 0=no, 1=yes
Changing the parameter requires the Administrator password (see Holding
Register 10036).
Input Registers - Read-only parameters
Address Type Input Register description
Measured values and status of measurement alarms
0 SW TEMPERATURE in the set measurement unit (x10).
1 B
Alarm for temperature:
0=OFF, 1= lower threshold alarm, 2= higher threshold alarm
2 SW RELATIVE HUMIDITY in % (x10).
3 B
Relative humidity alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
4 SW DEW POINT in the set measurement unit (x10).
5 B
Dew Point alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
6 SW PARTIAL VAPOR PRESSURE in hPa (x100).
7 B
Partial vapor pressure alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
8 SW MIXING RATIO in g/Kg (x10).
9 B
Mixing ratio alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
10 SW ABSOLUTE HUMIDITY in g/m3(x10).
11 B
Absolute humidity alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
12 SW WET BULB TEMPERATURE in the set measurement unit (x10).
13 B
Wet bulb temperature alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
18 SW SOLAR RADIATION in W/m2.
19 B
Alarm for solar radiation:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
20 SW ILLUMINANCE in lux.
21 B
Illuminance alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
24 SW
ATMOSPHERIC PRESSURE in the set measurement unit (the multiplier de-
pends on the set unit).
25 B
Atmospheric pressure alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
30 SW DAILY SOLAR RADIATION in Wh/m2.
31 B
Alarm for daily solar radiation:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
32 SW CO2in ppm.
33 B
CO2 alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
86 SW RAIN RATE in counts/h.
87 B
Rain rate alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
88 SW DAILY RAIN in counts.
89 B
Daily rain alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.

HD50 - 11 - V1.0
Address Type Input Register description
92 SW WIND SPEED (HD52.3D anemometer) in m/s (x100).
93 B
Wind speed (HD52.3D anemometer) alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
94 SW WIND DIRECTION (HD52.3D anemometer) in degrees (x10).
95 B
Wind direction (HD52.3D anemometer) alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
118 SW AIR SPEED (HD404…SR transmitter) in m/s (x100).
119 B
Air speed (HD404…SR transmitter) alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
122 SW RAINFALL QUANTITY IN THE LAST HOUR in counts.
123 B
Alarm for rainfall quantity in the last hour:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
Measurement units and resolution
5000 W TEMPERATURE unit of measurement: 0=°C, 1=°F.
5004 W DEW POINT unit of measurement: 0=°C, 1=°F.
5012 W WET BULB TEMPERATURE unit of measurement: 0=°C, 1=°F.
5021 SW ILLUMINANCE resolution: -2=100, -1=10, 0=1
5024 W ATMOSPHERIC PRESSURE unit of measurement: see the table of indexes
5025 SW
ATMOSPHERIC PRESSURE resolution:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
5052 W WIND SPEED measurement unit: see TAB 12.1
5053 SW
WIND SPEED resolution:
…, -2=100, -1=10, 0=1, 1=0.1, 2=0.01, …
General information
10000 W Year of last measurement.
10001 W Month of last measurement.
10002 W Day of last measurement.
10003 W Hour of last measurement.
10004 W Minutes of last measurement.
10005 W Seconds of last measurement.
10013 W
Password level for the current connection:
0=no password, 1=user level, 2= administrator level
Holding Registers - Read/Write parameters
Address Type Holding Register description
Measurement alarm thresholds
0 SW TEMPERATURE lower alarm threshold in the set measurement unit (x10).
1 SW Temperature higher alarm threshold in the set measurement unit (x10).
2 SW RH lower alarm threshold in % (x10).
3 SW RH higher alarm threshold in % (x10).
4 SW DEW POINT lower alarm threshold in the set measurement unit (x10).
5 SW Dew point higher alarm threshold in the set measurement unit (x10).
6 SW PARTIAL VAPOR PRESSURE lower alarm threshold in hPa (x100).
7 SW Partial vapor pressure higher alarm threshold in hPa (x100).
8 SW MIXING RATIO lower alarm threshold in g/Kg (x10).
9 SW Mixing ratio higher alarm threshold in g/Kg (x10).
10 SW ABSOLUTE HUMIDITY lower alarm threshold in g/m3(x10).
11 SW Absolute humidity higher alarm threshold in g/m3(x10).

HD50 - 12 - V1.0
Address Type Holding Register description
12 SW
WET BULB TEMPERATURE lower alarm threshold in the set measurement unit
(x10).
13 SW
Wet bulb temperature higher alarm threshold in the set measurement unit
(x10).
18 SW Lower alarm threshold for SOLAR RADIATION in W/m2.
19 SW Higher alarm threshold for solar radiation in W/m2.
20 SW ILLUMINANCE lower alarm threshold in lux.
21 SW Illuminance higher alarm threshold in lux
24 SW
ATMOSPHERIC PRESSURE lower alarm threshold in the set measurement unit
(the multiplier depends on the set unit).
25 SW
Atmospheric pressure higher alarm threshold in the set measurement unit
(the multiplier depends on the set unit).
30 SW Lower alarm threshold for DAILY SOLAR RADIATION in Wh/m2.
31 SW Higher alarm threshold for daily solar radiation in Wh/m2.
32 SW CO2lower alarm threshold in ppm.
33 SW CO2higher alarm threshold in ppm.
86 SW RAIN RATE lower alarm threshold in counts/h.
87 SW
Rain rate higher alarm threshold in counts/h.
88 SW DAILY RAIN lower alarm threshold in counts.
89 SW
Daily rain higher alarm threshold in counts.
92 SW
WIND SPEED (ultrasonic anemometer) lower alarm threshold in m/s
(x100).
93 SW
Wind speed (ultrasonic anemometer) higher alarm threshold in m/s
(x100).
94 SW
WIND DIRECTION (ultrasonic anemometer) lower alarm threshold in de-
grees (x10).
95 SW
Wind direction (ultrasonic anemometer) higher alarm threshold in degrees
(x10).
118 SW AIR SPEED (HD404…SR transmitter) lower alarm threshold in m/s (x100).
119 SW
Air speed (HD404…SR transmitter) higher alarm threshold in m/s (x100).
122 SW Lower alarm threshold for RAINFALL QUANTITY IN THE LAST HOUR in counts.
123 SW
Higher alarm threshold for rainfall quantity in the last hour in counts.
General information
da 10000
a 10019 B User code with ASCII codification.
Acceptable values are in the set {32,…,126}.
10020 W Current year
10021 W Current month
10022 W Current day
10023 W Current hour
10024 W Current minute
10025 W Current second
10026 W
Measurement interval: 0=1s, 1=2s, 2=5s, 3=10s, 4=15s, 5=30s, 6=1min,
7=2min, 8=5min, 9=10min, 10=15min, 11=30min, 12=1h
10027 W
Logging interval: 0=1s, 1=2s, 2=5s, 3=10s, 4=15s, 5=30s, 6=1min, 7=2min,
8=5min, 9=10min, 10=15min, 11=30min, 12=1h
10036 W
Password to be supplied to enable configuration change commands. The
reading provides the fixed value 32768.
da 10037
a 10046 B Device group with ASCII codification.
Acceptable values are in the set {32,…,126}.

HD50 - 13 - V1.0
Address Type Holding Register description
da 20000
a 20011 B User code with ASCII codification of measurement #1.
da 20012
a 20023 B User code with ASCII codification of measurement #2.
da 20024
a 20035 B User code with ASCII codification of measurement #3.
da 20036
a 20047 B User code with ASCII codification of measurement #4.
da 20048
a 20059 B User code with ASCII codification of measurement #5.
da 20060
a 20071 B User code with ASCII codification of measurement #6.
da 20072
a 20083 B User code with ASCII codification of measurement #7.
da 20084
a 20095 B User code with ASCII codification of measurement #8.
da 20096
a 20107 B User code with ASCII codification of measurement #9.
da 20108
a 20119 B User code with ASCII codification of measurement #10.
da 20120
a 20131 B User code with ASCII codification of measurement #11.
da 20132
a 20143 B User code with ASCII codification of measurement #12.

HD50 - 14 - V1.0
6TECHNICAL CHARACTERISTICS
Measuring interval 1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min
(Note: the minimum interval may be greater than 1 s if the instrument
acquires several sensors)
Logging interval 1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30, 60 min
(Note: the minimum interval may be greater than 1 s if the instrument
acquires several sensors)
Internal memory Circular management or stop logging if memory is full.
Number of storable samples from 259,000 a 890,500 depending
on the number of acquired quantities.
Interfaces ETHERNET (RJ45 connector) with proprietary TCP/IP or Modbus
TCP/IP protocol
USB (mini-USB connector)
Master RS485 Modbus-RTU (for acquiring the sensors)
Alarm Acoustic by means of the internal buzzer and sending of e-mails
Power supply External 7…30 Vdc
Consumption 40 mA @ 24 Vdc
LED indicators Power supply and LAN connection
Operating temperature/
humidity -10…+60 °C / 0…85 %RH non-condensing
Housing Plastic material
Weight 200 g approx.
Installation 35 mm DIN rail
Dimensions (mm)
The measurement characteristics depend on the sensors connected.
3 moduli DIN

HD50 - 15 - V1.0
7INSTRUMENT STORAGE
Conditions for storage of the instrument:
•Temperature: -20...+70 °C.
•Humidity: below 90 %RH no condensation.
•When storing, avoid places where:
•humidity is high;
•instrument is exposed to direct solar radiation;
•instrument is exposed to high temperature source;
•there are strong vibrations;
•there is vapor, salt and/or corrosive gas.
8SAFETY INSTRUCTIONS
General safety instructions
The instrument has been manufactured and tested in compliance with the safety stan-
dard EN61010-1:2010 “Safety requirements for electrical equipment for measurement,
control and laboratory use” and left the factory in a safe and secure technical condition.
The proper operation and the operational safety of the instrument can be ensured on-
ly if all the regular security measures are observed as well as the specific measures
described in this operating manual.
The proper operation and the operational safety of the instrument can be ensured on-
ly under the climatic conditions specified in this manual.
Do not use the instrument in places where there are:
•Rapid ambient temperature variations that may cause condensation.
•Corrosive or flammable gases.
•Direct vibrations, shocks to the instrument.
•High-intensity electromagnetic fields, static electricity.
If the instrument is moved from a cold environment to a hot one or vice versa, the
formation of condensation might cause problems to its operation. In this case you
need to wait for the instrument temperature to reach ambient temperature before op-
eration.
User obligations
The user of the instrument must make sure that the following regulations and direc-
tives related to the handling of hazardous materials are fulfilled:
•European directives on safety and health at work.
•National regulations on safety and health at work.
•Accident prevention regulations.

HD50 - 16 - V1.0
9ORDERING CODES
HD50R_MB Data logger with Master RS485 Modbus-RTU interface for acquiring
sensors with RS485 Modbus-RTU output. A proprietary protocol is also
available for connecting anemometers of the HD2003 series. A voltage-
free contact input allows connecting a tipping bucket rain gauge. USB
and Ethernet (proprietary TCP/IP or Modbus TCP/IP protocol) connec-
tion. Measurements are transmitted to an FTP address, to the Cloud and
via e-mail. Acoustic alarm with internal buzzer and sending of alarm e-
mails. External power supply 7…30 Vdc. 35 mm DIN rail installation.
HD35AP-S and HDServer1 softwares downloadable from Delta OHM
web site are included.
The probes and the CP23 USB cable have to be ordered separately.
The Ethernet cable is not included.
Accessories
HD35AP-CFR21 Software option that adds to the features of the basic software (both
HD35AP-S and HDServer1), the management of the data logging
system in accordance with the FDA 21 CFR part 11 recommenda-
tions. For Windows®operating systems.
CP23 Direct USB connection cable with mini-USB connector on the instru-
ment side and A type USB connector on the PC side. Cable length
1.5 m.
DELTA OHM metrology laboratories LAT N° 124 are ISO/IEC 17025 accredited by AC-
CREDIA for Temperature, Humidity, Pressure, Photometry / Radiometry, Acoustics and
Air Velocity. They can supply calibration certificates for the accredited quantities.

HD50 - 17 - V1.0
NOTES

HD50 - 18 - V1.0
NOTES


GUARANTEE
TERMS OF GUARANTEE
All DELTA OHM instruments are subject to accurate testing, and are guaranteed for 24 months from the
date of purchase. DELTA OHM will repair or replace free of charge the parts that, within the warranty
period, shall be deemed non efficient according to its own judgement. Complete replacement is excluded
and no damage claims are accepted. The DELTA OHM guarantee only covers instrument repair. The
guarantee is void in case of incidental breakage during transport, negligence, misuse, connection to a
different voltage than that required for the appliance by the operator. Finally, a product repaired or
tampered by unauthorized third parties is excluded from the guarantee. The instrument shall be returned
FREE OF SHIPMENT CHARGES to your dealer. The jurisdiction of Padua applies in any dispute.
The electrical and electronic equipment marked with this symbol cannot be disposed of in public
landfills. According to the Directive 2011/65/EU, the european users of electrical and electronic
equipment can return it to the dealer or manufacturer upon purchase of a new one. The illegal
disposal of electrical and electronic equipment is punished with an administrative fine.
This guarantee must be sent together with the instrument to our service centre.
IMPORTANT: Guarantee is valid only if coupon has been correctly filled in all details.
Instrument Code: HD50R-MB
Serial Number
RENEWALS
Date Date
Inspector Inspector
Date Date
Inspector Inspector
Date Date
Inspector Inspector
Table of contents
Other Delta OHM Data Logger manuals

Delta OHM
Delta OHM HD30.1 User manual

Delta OHM
Delta OHM HD35 Series User manual

Delta OHM
Delta OHM HD2013-DB User manual

Delta OHM
Delta OHM HD32MT.3 User manual

Delta OHM
Delta OHM HD32MT.1 User manual

Delta OHM
Delta OHM HD 2013 Series User manual

Delta OHM
Delta OHM DO2003 User manual

Delta OHM
Delta OHM HD32.1 User manual

Delta OHM
Delta OHM HD32.8.8 User manual

Delta OHM
Delta OHM HD37B17D User manual