Ecom DP3 User manual

Operating Instructions

Page 2 ecom-DP3
Index Page
1. Application and components 3
2. Design ecom-DP3 5
3. Connection of components 7
. Data processing
.1. Data storage 8
.2. Edit customer 8
.3. Select customer 9
5. Control programs
5.1. Setup control programs 11
5.2. Natural gas
5.2.1. Stress test 1
5.2.2. Tightness check 15
5.2.3. Usability check 16
5.3. Liquid gas
5.3.1. Tightness check 18
5.3.2. Pressure test 19
5.3.3. Pressure test 0 – 0 bar 20
5. . Drinking water (option)
5. .1. Drinking water plant (wet)
5. .1.1. Function test 21
5. .1.2. Pressure test 22
5. .1.3. Extended test 23
5. .2. Drinking water plant (dry)
5. .2.1. Tightness check 2
5. .2.2. Stress test 25
5.5. Heating (option) 26
5.6. Sewage (option) 27
6. More measurements
6.1. Single measurement 28
6.2. Heating Check (option) 29
6.3. Pa measurement (option) 32
6. . Pressure check 33
7. Adjustments 3
8. Control (system test) 36
9. Technical Data ecom-DP3 37
10. Maintenance tips 38

ecom-DP3 Page 3
1. Application and components
The ecom-DP3 easies tightness tests. Besides gas pressure measurements,
tightness checks as well as leakage volume determination according to DVG -
TRGI 2018 can be performed. The following components belong to the ecom-
DP3 set:
Standard
By works at gas pipes please respect
the legal safety instructions!
Hose connector
with quick
-fitting
ecom-DP3
Pressure measurement
instrument with automatic
check programmes
Connection hose
2,8 m long, with quick-
fittings
High-
pressure test
stopples
3/8“ auf 1/2“ und
3/8“ auf 3/4“
with qui
ck-fittings
One-pipe counter cap
for gas meters G4/G6
with quick
-fitting
Conic test stopples
1/2“ - 3/4“ and
3/4" – 1 1/4",
with quick
-fittings
Y-Piece
with quick
-fittings
Ball valve
with quick
-fitting
Connecting piece
with
safety valve
with quick
-fitting
Plug
with quick
-fitting
Software „DP3Report“
to create inspection
reports

Page 4 ecom-DP3
Optionen / Zubehör
Gasspürgerät
ecom-LSG
for identification of
leakages at gas pipes
(Item no.: 102145)
External pressure
sensor (0 – 20 bar)
with conne
ction cable
(Item no.: 102158)
2 x Capillary hose
for 4-Pa-Test
2 x Silicone hose
for small pressure
measurement
External room tem-
perature sensor
with connection cable
(Item no.: 102156)
External pipe tempera-
ture sensor
with
connection cable
(Item no.: 12019)
One-pipe counter cap
for gas meters G10/G16
with quick
-fitting
(Item no.: 102705)
Test stopple conic
12 – 22 mm
with quick
-fitting
(Item no.: 102706)
Test stopple cylindrical
diameter 14,5 mm
with quick
-fitting
(Item no.: 102708)
Test stopple cylindrical
diameter 24 mm
with quick
-fitting
(Item no.: 102710)
Test stopple cylindrical
diameter 41 mm
with
quick-fitting
(Item no.: 102712)
External pressure
sensor (0 – 40 bar)
with connection cable
(Item no.: 103932)
Test stopple conic
35 – 65 mm
with
quick-fitting
(Item no.: 102707)
Test stopple cylindrical
diameter 19 mm
with quick
-fitting
(Item no.: 102709)
Test stopple cylindrical
diameter 35 mm
with quick
-fitting
(Item no.: 102711)
Test stopple cylindrical
diameter 52mm
with
quick-fitting
(Item no.: 102713)

ecom-DP3 Page 5
2. Design ecom-DP3
Connection
tightness checks
Pmax = 1500 hPa
Connection small
pressure measurement
(option)
Connection
loading unit
Grafic display Integral printer
Slot for multi-
media card
USB connection
Accu
compartment
Connection
temperature
sensor 2
Connection
external
pr
essure sensor
Connection
AUX
Connection
temperature
sensor 1

Page 6 ecom-DP3
Keyboard
ESC key
(quit/
exit menu)
In the input mode, the keys are
used for numerical inputs
Enter key
(confirm
selection)
Cursor keys
(Up/Down/Right/
Left/Scroll)
Values
recording
Print key
(access to
printing menu)
ON / OFF
key
Info key
(access to
control menu)
Function keys
(function shown on
display)

ecom-DP3 Page 7
3. Connection of components
Please connect the components as shown on below plan.
Connection hose
for connection on „ P high 1 + “
at the ecom-DP3
ecom-DP3
Y-piece
for the con-
nection of 2
line sections
(f.e. with one-
pipe counter
cap)
Hose
connector
for connection
at measuring
nipple of the
plant
High-pressure
test stopples
for connection
at screw con-
nections
One-pipe coun-
ter cap
for
the connection
in place of the
gas meter
Conic test
stopples
for connection
at line ends
Ball valve
for connection of exter-
nal
devices for pressure
build-up
Connecting piece
with
safety valve
for connection of exter-
nal
devices for pressure
build-up
Plug
to check the tightness
of the ecom-DP3
Never connect external devices for the pressure build-up
without connecting piece with safety valve!

Page 8 ecom-DP3
Option
External pressure sensor
(0 – 20 bar or 0 – 40 bar)
for checking heating and
drinking water plants (con-
nection „ext. Sensor“)
ecom-DP3
2 x Capillary hose
to measure the differ-
ence pressure with the
4-Pa-Test (connection
„P low 2“)
2 x Silicone hose
for small pressure and
flow measurements with
f.e. pitot probe (connec-
tion „P low 2“)
External room tem-
peratur sensor
to measure the room
temperature during the
test (connection „T1“)
External pipe tempera-
ture sensor
to measure the pipe
temperature during the
test (connection „T1“)
External pipe tempera-
ture sensor
to
measure the pipe
temperature during the
test (connection „T2“)

ecom-DP3 Page 9
. Data processing
.1. Data storage
The measurements made with the equipment can be assigned to a customer. For
each customer a file with file name is created on the memory card (e.g.:
DP3_0001.DAT). Into this file all measurements belonging to the customer are
written.
The files can be transferred to the PC using a card reader. ith the software
"DP3Report" a protocol for each check can be made. The following conditions
must be fulfilled for using a multi media card:
- min. card volume 32 MB - max.32 GB (UHC)
- card formatted on 16 bit FAT or FAT32
- SD cards from SanDisk recommended
- PC with card reader
.2. Edit customer
If the instrument gets switched on and a SD or MM
Card with no customer data is inserted the inquiry
follows whether you would like to edit a customer. If
the measurement is to be assigned to a customer,
press the key <F1> (<F > = no: Measurement is
made without allocation). Proceed as follows:
-select line to be filled out with the cursor keys
<up/down> and confirm with <Enter>
-select keyboard <F3>
(4 keyboard are available)
-select a character with the cursor keys
<up/down/right/left>
(selected character is black deposited)
-take over the character with <Enter> (the last
character can be deleted with <F2>)
-repeat the procedure, until the desired designation
is complete
-if a character is to be corrected, proceeded as
follows:
-interrupt the character selection with <F >
No customer
existing!
Create new
YES
NO
Never pull out cards during data record - data loss and damaging
of the data carrier possible!
Customer number
Name
Street
ZIP/City
DP3_ 1.DAT
Tel.
Others

Page 10 ecom-DP3
-select the character to be corrected with the
cursor keys <right/left>
-activate the character selection with <F > and
correct the character
-return to line selection with <F1> and choose
the next line
If all necessary lines are filled out, a file is created
with <F1>. ith <ESC> you can leave to select a
control program.
.3. Select customer
ith the help of the software „DP3Report“ customer
data can be stored for the use at the DP3 on SD or
MM Card. If the instrument gets switched on and a
SD or MM Card with customer data is inserted a
customer for allocation can be selected with "Search
for/Scroll" / <Enter>:
Scroll:
Select „Scroll“, if you want to find the customer with the help of the cursor keys
<up/down>. Further functions are:
<F1> = to the first customer
<F2> = 10 customers forward
<F3> = 10 customers backward
Search for:
Select „Search for“, if you want to find the customer by the input of at least 3
connected characters. Edit 3 characters as described in the chapter before and
start the search with <Enter>. All agreements with the character sequence are
picked out. The selection can be paged through with <F3> (with <F1> to the be-
ginning of the selection).
If the desired customer is found, select with <Enter>. ith „View/Modify“ the
customer data can be viewed and modified.
Format:
This function is usually needed by initial adjustment of the instrument at our facto-
ry (preparation of SD or MM Card for data recording).
Caution: All stored values will be cancelled! Please switch instrument after
formatting off and on again!
Search for/Scroll
Create new
View/Modify
Format
Fr 2 . 1.12 9:5 :33
Cus
tomer data
DP3_ 1.DAT
Customer is selected
Customer number
123456789
Name
Customer 1
Street
Street 1
ZIP/City
11111 City 1
DP3_ 1.DAT
Tel.
123456789
Others
Customer 1 Others

ecom-DP3 Page 11
5. Control programs
After switching on of the instrument the following pre-programmed control pro-
grams are available:
Natural gas:
-Stress test acc. to TRGI 2018
-Tightness check acc. to TRGI 2018
-Usability check acc. to TRGI 2018
Liquid gas:
-Tightness check
-Pressure test
Drinking water:
-Drinking water plant (wet) acc. to DIN EN 806-4:
-Function test
-Pressure test
-Extended test
-Drinking water plant (dry) acc. to DIN EN 806-4:
-Tightness test
-Stress test
Heating:
-Heating installation acc. to DIN 18380
Sewage:
-Sewage installation acc. to DIN 1610
More measurements:
-Pressure check (arbitrary pressure)
In menu „Adjustments“ / „Setup checks“ the parameters for the control pro-
grams can be adjusted:
5.1. Setup control programs
Scroll with the cursor keys <up/down> to the menu
„Adjustments“. Activate the menu with <Enter>.
Scroll with the cursor keys <up/down> to the menu
point „Setup checks“ and activate with <Enter>.
The following parameters can be adjusted:
Pressure check (arbitrary pressure):
-Stabilisation time (0 - 10 min, default: 0 min)
-Measurement time (0 - 6000 min, default: 0 min)
Respect the „Technical Rules for Gas installations“!
Natural gas
Liquid gas
Drinking water
Heating
Sewage
More measurement
djustments
Data storage
Fr 2 . 1.12 9:5 :33
Setup checks
Set clock
Paper feeding
Print contrast
Displ. contrast
Key beep
Language: English
Pitot factor
Scan data logger
Printout
Select:
Adjustments

Page 12 ecom-DP3
Stress test acc. to TRGI 2018:
-Stabilisation time (1 - 10 min, default: 1 min)
-Measurement time (10 - 120 min, default: 10 min)
-Test pressure (900 - 1500 hPa, default: 1000 hPa)
Tightness check acc. to TRGI 2018:
-Stabilisation time installation lower 100 Liter (1 - 20 min, default: 10 min)
-Measurement time installation lower 100 Liter (9 - 120 min, default: 10 min)
-Stabilisation time installation 100 bis 200 Liter (20 - 40 min, default: 30 min)
-Measurement time installation 100 bis 200 Liter (19 - 120 min, default: 20 min)
-Stabilisation time installation higher 200 Liter (50 - 70 min, default: 60 min)
-Measurement time installation higher 200 Liter (29 - 120 min, default: 30 min)
-Test pressure (15 - 500 hPa, default: 150 hPa)
Usability check acc. to TRGI 2018:
-Stabilisation time (5 - 240 min, default: 10 min)
-Measurement time (2 - 30 min, default: 5 min)
-Reference pressure (10 - 100 hPa, default: 23 hPa)
Function test drinking water plant (wet) acc. to DIN EN 806- :
-Stabilisation time (0 - 120 min, default: 30 min)
-Measurement time (1 - 120 min, default: 15 min)
-Test pressure (0.01 – 20.00 bar, default: 6.00 bar)
Pressure test drinking water plant (wet) acc. to DIN EN 806- :
-Stabilisation time (0 - 120 min, default: 0 min)
-Measurement time (1 - 120 min, default: 30 min)
-Test pressure (0.01 – 20.00 bar, default: 11.00 bar)
Extended test drinking water plant (wet) acc. to DIN EN 806- :
-Stabilisation time (0 - 120 min, default: 0 min)
-Measurement time (1 - 240 min, default: 120 min)
-Test pressure (0.01 – 20.00 bar, default: 5.50 bar)
Tightness test drinking water plant (dry) acc. to DIN EN 806- :
-Stabilisation time (0 - 120 min, default: 0 min)
-Measurement time (1 - 240 min, default: 120 min)
-Test pressure (1 - 2000 hPa, default: 150 hPa)
Stress test drinking water plant (dry) acc. to DIN EN 806- :
-Stabilisation time (0 - 120 min, default: 0 min)
-Measurement time (1 - 120 min, default: 10 min)
-Test pressure (0.01 – 5.00 bar, default: 1.00 bar)
Heating plant acc. to DIN 18380:
-Stabilisation time (0 - 120 min, default: 0 min)
-Measurement time (1 - 240 min, default: 120 min)
-Test pressure (0.01 – 20.00 bar, default: 10.00 bar)
Sewage plant acc. to DIN 1610:
-Stabilisation time (0 - 100 min, default: 0 min)
-Measurement time (1 - 120 min, default: 30 min)
-Test pressure (1 - 2000 hPa, default: 200 hPa)

ecom-DP3 Page 13
Tightness check liquid gas:
-Stabilisation time (0 - 60 min, default: 10 min)
-Measurement time (0 - 60 min, default: 10 min)
Test pressure (0 - 1000 hPa, default: 100 hPa)
Pressure test liquid gas:
-Stabilisation time (0 - 60 min, default: 10 min)
-Measurement time (0 - 60 min, default: 20 min)
-Test pressure (0 - 1500 hPa, default: 1000 hPa)
Pressure test 0 – 0 bar liquid gas:
-Stabilisation time (0 - 120 min, default: 10 min)
-Measurement time (1 - 240 min, default: 10 min)
-Test pressure (0 - 40 bar, default: 10 bar)
Each parameter can be adjusted the same way:
1. Select the parameter with cursor keys <up/down>
2. Press <Enter> to activate
3. Use numerical keys to input the value
To adjust the standard value press <F2>
Pressure check
min
1 min
Stabilisation
time
Please use the
Numeral keys!
2
Pressure check
Standard value : F2
1 min
Stabilisation time
Select:

Page 14 ecom-DP3
5.2. Natural gas
5.2.1. Stress test
The stress test acc. to DVG – TRGI Process
Instructions G 600 at gas pipes (operation pressure
up to 100 hPa) is registered as a control pragram in
the ecom-DP3. Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu "Natural gas" to the menu point „Stress
test“. Activate the menu point with <Enter>.
2. Check if necessary the values of stabilisation time,
measurement time and test pressure with <F1>
(back with <ESC>).
3. Lock the gas pipe with a suitable adapter (test
stopple, high pressure stopple or one-pipe counter
cap).
4. Connect the components as described in
chapter 3.
5. Start the control program with <Enter> (the pump
of the instrument is building up the adjusted pres-
sure).
6. ait for stabilisation time (if the pressure remains
in the range „test pressure - 10%" during stabilisation
time, so the measurement time will start). If neces-
sary the stabilisation time can be terminated with
<F1>.
7. Once the measurement time is over, the result is
displayed and can be printed with <Print>.
8. If the menu „Stress test“ is selected again, so
the result can be called up with <F > (as long as the
ecom-DP3 is switched on) or a new measurement
can be started with <F1>.
Stress test
1Bar/1 hPa TRGI 2 18
Stabilisation time
Res. time: 58sec
Pressu 1 3.1hPa - .8hPa
Temp. 22.6°C .3°C
Start Difference
1 2.3hPa
Stress test
1Bar/1 hPa TRGI 2 18
Measur. in process
Res. time: 9:25min
Pressu 1 2.3hPa . hPa
Temp. 22.6°C . °C
Start Difference
1 2.3hPa
Stress test
1Bar/1 hPa TRGI 2 18
Set value 1 hPa
is built up!
9 2.3hPa
1min
1 min
Stress test
1Bar/1 hPa TRGI 2 18
Connect!
Forward w.:
Set value 1
hPa
. hPa
Stress test
1Bar/1 hPa TRGI 2 18
Measurement 1 : min
Ini. Press 1 2.3hPa
Fin. Press 1 2.3hPa
Difference . hPa
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Stress test
1Bar/1 hPa TRGI 2 18
Start new
measurement?
Select:
YES
NO

ecom-DP3 Page 15
5.2.2. Tightness check
The tightness check acc. to DVG – TRGI Process
Instructions G 600 at gas pipes (operation pressure
up to 100 hPa) is registered as a control pragram in
the ecom-DP3. Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu "Natural gas" to the menu point „Tightness
check“. Activate the menu point with <Enter>.
2. Adjust the gas pipe volume with <F2> (lower
100L, 100L..200L or higher 200L).
3. Check if necessary the values of stabilisation time,
measurement time and test pressure with <F1>
(back with <ESC>).
4. Lock the gas pipe with a suitable adapter (test
stopple, high pressure stopple or one-pipe counter
cap).
5. Connect the components as described in
chapter 3.
6. Start the control program with <Enter> (the pump
of the instrument is building up the adjusted pres-
sure).
7. ait for stabilisation time (if the pressure remains
in the range „test pressure - 10%" during stabilisation
time, so the measurement time will start). If neces-
sary the stabilisation time can be terminated with
<F1>.
8. Once the measurement time is over, the result is
displayed and can be printed with <Print>.
9. If the menu „Tightness check“ is selected again,
so the result can be called up with <F > (as long as
the ecom-DP3 is on) or a new measurement can be
started with <F1>.
Tightness check
15 hPa (acc. TRGI 2 18)
Stabilisation time
Res. time: 9:58min
Pressu 153.1hPa - .8hPa
Temp. 22.6°C .3°C
Start Difference
152.3hPa
Tightness check
15 hPa (acc. TRGI 2 18)
Measur. in process
Res. time: 9:25min
Pressu 152.3hPa . hPa
Temp. 22.6°C . °C
Start Difference
152.3hPa
Tightness
check
15 hPa (acc. TRGI 2 18)
Set value 15 hPa
is built up!
14 .3hPa
1 min
1 min
Tightness check
15 hPa (acc. TRGI 2 18)
Measurement
1 : min
Ini. Press 152.3hPa
Fin. Press 152.3hPa
Difference . hPa
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Tightness check
15 hPa (acc. TRGI 2 18)
Start new
measurement?
Select:
YES
NO
Tightness check
15 hPa (acc. TRGI 2 18)
Connect!
Forward w.:
Set value
15 hPa
lower 1 L
. hPa

Page 16 ecom-DP3
5.2.3. Usability check
The usability check acc. to DVG – TRGI Process
Instructions G 600 at gas pipes (operation pressure
up to 100 hPa) is registered as a control pragram in
the ecom-DP3. Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu “Natural gas" to the menu point „Usability
check“. Activate the menu point with <Enter>.
2. The measurement becomes prepared. The piston
is brought into its initial position.
3. Connect the components as described in
chapter 3.
4. Open gas shut-off valve of the installation. As
soon as the operating pressure of the plant is recog-
nized, the system is aired out and the instrument
starts automatically the stabilisation time. If neces-
sary the stabilisation time can be terminated with
<F1>.
5. Close gas shut-off valve after stabilisation time.
The measurement of the leak rate begins with <En-
ter>.
Usability check
acc. to TRGI 2 18
Connect tubes!
Open gas shut-off valve!
Measuring starts once ope.
pressure is identifed
. hPa
Usability check
acc. to TRGI 2 18
Stabilisation time
Res. time: 9:58min
24.15hPa
Usability check
acc. to TRGI 2 18
Measur. in process
Time: 1:25min
Leak rate 1.52l/h
Pressu 24.11hPa
Temp. 24.3°C
Forward w.:
Usability check
acc. to TRGI 2 18
Lose gas shut-off
valve now!
Continue with <Enter>
24.11hPa
U
sability check
acc. to TRGI 2 18
Air is
getting removed
from system!
Measuring then starts
automatically!
Quit with:
Usability check
acc. to TRGI 2 18
Measuring in preparation
Do not connect
any tubing!
Quit with:

ecom-DP3 Page 17
6. After measuring time the installation must be
brought back into its original condition (remove hos-
es and open gas shut-off valve).
7. After the pressure in the instrument reduced itself
(or after pressing <Enter>), the result is displayed
and can be printed with <Print>.
8. If the menu „Usability check“ is selected again,
so the result can be called up with <F > (as long as
the ecom-DP3 is on) or a new measurement can be
started with <F1>.
Usability check
acc. to TRGI 2 18
Measurement 5: min
Leak rate 1.51l/h
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Usability che
ck
acc. to TRGI 2 18
Start new
measurement?
Select:
YES
NO
Usability check
acc. to TRGI 2 18
Measurement completed!
Remove tubes!
Continue with <Enter>
Forward w.:

Page 18 ecom-DP3
5.3. Liquid gas
5.3.1. Tightness check
The tightness check acc. to TRF at liquid gas pipes
is registered as a control pragram in the ecom-DP3.
Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu "Liquid gas" to the menu point „Tightness
check“. Activate the menu point with <Enter>.
2. Check if necessary the values of stabilisation time,
measurement time and test pressure with <F1>
(back with <ESC>).
3. Lock the gas pipe with a suitable adapter (test
stopple, high pressure stopple or one-pipe counter
cap).
4. Connect the components as described in
chapter 3.
5. Start the control program with <Enter> (the pump
of the instrument is building up the adjusted pres-
sure).
6. ait for stabilisation time (if the pressure remains
in the range „test pressure - 10%" during stabilisation
time, so the measurement time will start). If neces-
sary the stabilisation time can be terminated with
<F1>.
7. Once the measurement time is over, the result is
displayed and can be printed with <Print>.
8. If the menu „Tightness check“ is selected again,
so the result can be called up with <F > (as long as
the ecom-DP3 is on) or a new measurement can be
started with <F1>.
Liquid gas
Tightness check
Connect!
Forward w.:
Set value 1
hPa
. hPa
Liquid gas
Tightness check
Stabilisation time
Res. time: 9:58min
Pressu 1 3.1hPa - .8hPa
Temp. 22.6°C .3°C
Start Difference
1 2.3hPa
Liquid gas
Tightness check
Measur. in process
Res. time: 9:25min
Pressu 1 2.3hPa . hPa
Temp. 22.6°C . °C
Start Difference
1 2.3hPa
Liquid gas
Tightness check
Set value 1 hPa
is built up!
92.3hPa
1 min
1 min
Liquid gas
Tightness check
Measurement 1 : min
Ini. Press 1 2.3hPa
Fin. Press 1 2.3hPa
Difference . hPa
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Liquid gas
Tightness check
Start new
measurement?
Select:
YES
NO

ecom-DP3 Page 19
5.3.2. Pressure test
The pressure test acc. to TRF at liquid gas pipes is
registered as a control pragram in the ecom-DP3.
Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu "Liquid gas" to the menu point „Pressure
test“. Activate the menu point with <Enter>.
2. Check if necessary the values of stabilisation time,
measurement time and test pressure with <F1>
(back with <ESC>).
3. Lock the gas pipe with a suitable adapter (test
stopple, high pressure stopple or one-pipe counter
cap).
4. Connect the components as described in
chapter 3.
5. Start the control program with <Enter> (the pump
of the instrument is building up the adjusted pres-
sure).
6. ait for stabilisation time (if the pressure remains
in the range „test pressure - 10%" during stabilisation
time, so the measurement time will start). If neces-
sary the stabilisation time can be terminated with
<F1>.
7. Once the measurement time is over, the result is
displayed and can be printed with <Print>.
8. If the menu „Pressure test“ is selected again, so
the result can be called up with <F > (as long as the
ecom-DP3 is on) or a new measurement can be
started with <F1>.
Liquid gas
Pressure test
Stabilisation time
Res. time: 9:58min
Pressu 1 3.1hPa - .8hPa
Temp. 22.6°C .3°C
Start Difference
1 2.3hPa
Liquid gas
Pressure test
Measur. in process
Res. time: 9:25min
Pressu 1 2.3hPa . hPa
Temp. 22.6°C . °C
Start Difference
1 2.3hPa
Liquid gas
Pressure test
Set
value 1 hPa
is built up!
9 2.3hPa
1min
1 min
Liquid gas
Pressure test
Measurement 1 : min
Ini. Press 1 2.3hPa
Fin. Press 1 2.3hPa
Difference . hPa
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Liquid gas
Pressure test
Start new
measurement?
Select:
YES
NO
Liquid gas
Pressure test
Connect!
Forward w.:
Set value
1
hPa
. hPa

Page 20 ecom-DP3
5.3.3. Pressure test 0 – 0 bar
The pressure test 0 - 40 bar at liquid gas pipes is
registered as a control pragram in the ecom-DP3.
Proceed as follows:
1. Scroll with the cursor keys <up/down> in the
menu "Liquid gas" to the menu point „Pressure
test 0 – 0 bar“. Activate the menu point with <En-
ter>.
2. Check if necessary the values of stabilisation time,
measurement time and test pressure with <F1>
(back with <ESC>).
3. Connect the external pressure sensor to the liquid
gas pipe.
4. Connect the external pressure sensor as de-
scribed in chapter 3.
5. Start the control program with <Enter> (the pump
of the instrument is building up the adjusted pres-
sure).
6. ait for stabilisation time (if the pressure remains
in the range „test pressure - 10%" during stabilisation
time, so the measurement time will start). If neces-
sary the stabilisation time can be terminated with
<F1>.
7. Once the measurement time is over, the result is
displayed and can be printed with <Print>.
8. If the menu „Pressure test 0 – 0 bar“ is select-
ed again, so the result can be called up with <F >
(as long as the ecom-DP3 is on) or a new meas-
urement can be started with <F1>.
Liquid gas
Pressure test – 4 bar
Stabilisation time
Res. time: 9:58min
Pressu 11.13bar - . 8bar
Temp. 22.6°C .3°C
Start Difference
11. 5bar
L
iquid gas
Pressure test – 4 bar
Measur. in process
Res. time: 9:25min
Pressu 11. 5bar . bar
Temp. 22.6°C . °C
Start Difference
11. 5bar
Liquid gas
Pressure test – 4 bar
Set value 11. bar
Built up pressure!
1 .23bar
1 min
3 min
Liquid gas
Pressure test – 4 bar
Measurement 1 : min
Ini. Press 11. 5bar
Fin. Press 11. 5bar
Difference . bar
Ini. temp. 22.6°C
End temp. 22.6°C
Difference
. °C
Forward w.:
Liquid gas
Pressure test – 4 bar
Start new
measurement?
Select:
YES
NO
Table of contents
Other Ecom Measuring Instrument manuals