Ecom J2KNpro Tech User manual

Operating Instructions
WE PRODUCE QUALITY OF LIFE!

Page 2 ecom-J2KN
pro
Index Page
Important hints 3
Important hints for using IR bench 4
1. Instrument design
1.1. Basic module 5
1.2. Control module 6
2. Gas cooler (option) 7
3. Power supply 8
4. Radio communication basic/control module 9
5. Data record 10
6. Instrument switch on 11
7. Input or selection of combustion plants 13
8. Flue gas analysis
8.1. Gas analysis 16
8.2. CO measurement (gas channel check) 19
8.3. O
2
check 20
8.4. Flow measurement (option) 20
8.5. Draught measurement 21
8.6. Soot dot...Oil trace 22
8.7. Measurement record and printout 24
8.8. Printout ecom-J2KN
pro
25
9. Mean Value Measurement 26
10. Adjustments 28
11. Control 33
12. Data processing
12.1 Communication 34
12.2. Automatic Measurement (option) 35
12.3. Data logger 36
12.4. Data logging with DASNT2 37
12.5. Data processing with DAS5 37
12.6. Data processing with Remote Display App 38
13. Diagnostics
13.1. Fault diagnostic 39
13.2. dT-measrurement 41
13.3. Heating check (option) 41
14. Maintenance tips 45
15. Technical data 48
16. FAQ 49

ecom-J2KN
pro
Page 3
Important hints
The ecom
-
J2KN
pro
meets
the requirements of
DIN EN 50379 part of 2.
The
ecom
-
J2KN
pro
may
not be used for
continuous emission control!
Following minimum times must be kept, in order
to receive correct measured values:
-1 min. to calibrate the sensors at fresh air
-2 min. for stable measured values
The following substances impair the working of
the instrument:
-Cleaning agents
-Degreasers
-Wax polishes
-Adhesives
-Formaldehyde
Adjustments at burners and boilers
should be made
only by specialists, who are familiar with
this installations!
1. Charge the accumulator regularly (Charge not
used instrument min. once a month)!
2. Never store the instrument with unloaded
battery!

Page 4 ecom-J2KN
pro
Important hints for using IR bench
Please start every 15 min. a fresh air calibration by
using the infrared bench (especially with methane) to
get accurate measurements
(Adjustments / Re-Calibration)!
Please switch on the instrument 15 min. before star
t-
ing the fresh air calibration, in order to bring the in-
frared bench on operating temperature!

ecom-J2KN
pro
Page 5
1. Instrument design
1.1. Basic module
Connection
air
temperature
Connection
gas tem-
perature
Condensate
trap with fine
dust filter
Connection
gas hose
Connection
draught
Connection
pressure
Con
nection
condensate
evacuation
Integral
printer
Filter for
IR bench
Control
module
(see next page)
Connection mains
power plug
with fuse
(630 mA / 250 V)
Info display
(see chap-
ter 11.)
Basic
ON / OFF
Analogue
inputs
(Option)
Connection
heatet probe
(Option)
Connection
AUX
SO2/NOx-
filter for
CO sensor
Connections
pitot tube
(Option)
Cable socket
by wire transfer
basic / control
modules

Page 6 ecom-J2KN
pro
1.2. Control module
ESC key
(quit/
escape menu)
In the input mode,
the keys are used
for numerical inputs
OK key
(confirm selec-
tion)
Cursor keys
(Up/Down/Right/
Left/Scroll)
Measurement
values record-
ing
Print key
(access to
printing menu)
Switch ON/OFF
control module
Display
backlit
ON/OFF
Function keys
(function shown on display)
Slot for Multi-
Media-Card
Graphics
display
Connection
USB
Connection
keyboard
Socket for connection
cable by wire
transfer basic
Keyboard control module
Connection air
temperature

ecom-J2KN
pro
Page 7
2. Gas cooler (option)
Exhaust gas with a temperature over the steam dew point (35 to
65 °C) is flown spiral via a long gas path thru a surface coated metal
body with good thermal conductivity. The gas radiates its heat to this
metal body. A PELTIER element (semiconductor cooling element)
flown by a continuous current is thermally connected with this body
and with a second metal body with cooling ribs and ventilation slots.
The flow thru the PELTIER element creates a heat transfer from
WARM to COLD, drains the heat of the metal body flown by gas and
conveys it to the outer cooling body. This heat is conveyed thru a
vertical forced ventilation to the surrounding air.
The condensation issued by the heat loss of the gas drops in a re-
ceptacle and is pumped out on request (either by the user or due to
the level monitoring) by a periodically working hose pump.
The sucking capacity of the gas conveying pump avoids a sufficient
dwell time of the gas with the condensate, so that wash out reactions
(NO
2
+H
2
O > H
2
NO
3
) do not take place.
At the cooler outlet the gas has a temperature of ca. 5 °C with a rela-
tive saturation of nearly 100 % relative humidity (corresponds to a
water steam content < 7 g/m
3
).
Fan
Peltier
element
Gas
outlet
Gas
inlet
Level control
Condensate
evacuation
In case of
battery operation the power is reduced
(output temperature 8 – 10 °C)!
COLD

Page 8 ecom-J2KN
pro
3. Power supply
The basic module of the ecom-J2KN
pro
is delivered with internal
loading unit. The instrument can be operated upon a longer time
period with the internal accumulator (6 V; 7,2 Ah). Connecting the
internal charging unit to mains power is only compulsory to recharge
the accumulators and to operate the heated probe system.
The accumulators should be recharged when the instrument shows a
corresponding message (acoustical warning and display infor-
mation). The accumulators loading stand can be checked, looking at
the voltage information on the display (menu "Control"). The battery
warning is activated when the value „ACC.B“ is smaller than 6 V.
By 5,8 V the power operation via battery is no more possible. The
instrument must be further powered via internal charging unit.
The control module of the ecom-J2KN
pro
is powered by 3 nickel-
metal-hydride accumulators (type AA). In case of need, the accumu-
lators can be recharged by docking the control module to the basic
module.
Hereby 2 functions can be selected („Adjustments / „Internal“ /
„Reloading function“ / <OK>):
1. Recharging function ON (<F1> = YES):
- batteries slowly and carefully recharged
- recommended adjustment by frequent use
2. Recharging function OFF (<F4> = NO):
- batteries quickly recharged
- recommended adjustment by occasional use
Used accumulators c
an be returned to us or brought
to recycling stations of public waste disposal compa-
nies respectively accumulators selling stores!
Never use batteries, to operate the control
module of the J2KN
pro
!

ecom-J2KN
pro
Page 9
4. Radio communication basic / control module
Thanks to the detachable control mod-
ule the basic module can be monitored
wireless. The basic module can be un-
locked as follows:
1. Press to unlock.
2. Tip control module forwards
3. Release control module from basic
module
The information exchange between
control and basic module is performed
via radio transmission (868 MHz) with a
coverage of approx. 50 m (by free
sight). The quality of the radio transmis-
sion is documented by a bar indication
in the main menu of the instrument
(long bar = good radio communication).
By interruption of the radio
communication, an error message is
displayed. By persisting disturbances of
the radio communication, a cable
(option) can take over the transfer
(connection betwen socket DATA on
control unit and socket DATA on basic
module).
If the basic module is switched off and
the control module not, so the display
will show an error message inviting to
fix the control module in its docking
station (helps also not to forget the con-
trol module). Observe this order, quit
with <ESC> and finally switch off the
control module.
Gas analysis
Pressure
Soot..Oil trace
Data processing
Adjustments
Control
Diagnostics
Unlocking
Bar indication
radio quality
--
ECOM
-
J2 N
--
Radio connection
interrupted!
Use cable or
switch on basic
module!
Quit with:
-- ECOM-J2 N --
Place J2 N in
basic mod.!
Quit with:

Page 10 ecom-J2KN
pro
5. Data record
The multi-media card enables the storage of both punctual meas-
urements and data logger records.
The values of punctual measurements are written in a text file
(J2KDV.txt). Those of data logger records in a csv file (J2KDL-xx.csv
/ xx = records numbered consecutively).
Both file types have the same structure and can be imported respec-
tively opened in Excel. See chapter „Technical Data“ for data format
information. The files can be transferred on the PC using a card
reader. The following conditions must be fulfilled for using a multi-
media card:
- ecom-J2KN
pro
- min. card volume 32 MB - max.32 GB (UHC)
- card formatted on 16 bit FAT or FAT32
- SD cards from SanDisk recommended
- PC with card reader
Insert the multi-media card as shown.
Take care that the card does not stand
out and hooks on.
Never pull out cards during
data record
-
data loss
and damaging of the data carrier possible!

ecom-J2KN
pro
Page 11
6. Instrument switch on
Once the control module has been switched on (key <I/0>), the main
menu is displayed.
8 sub-menus with the following functions are displayed (non-visible
sub-menus can be called up scrolling the arrow keys):
Always position the probe in the exhaust pipe once
the calibration phase is over!
Always us
e a filtering system as per our
recommendation for combustion plants
firing solid combustibles!
Gas analysis
Mean value
Pressure
Soot..Oil trace
Data processing
Adjustments
Control
Diagnostics
Bar indication
radio quality
Bar indication
transfer quality
accu voltage
(charging status)
Display co
n-
trast adjust-
able with
F1 and F2
At temperatures below 10 °C the gas channel plate will
be heated with mains power connection – connect
instrument to mains power!

Page 12 ecom-J2KN
pro
- Gas analysis : Perform gas analysis
- Mean value : Measurement with mean value calculation
- Draught measure. : Perform draught or pressure measurement
- Soot...Oil trace : Input of soot measurements results
- Data processing : Assign measurements / Load or send data
- Adjustments : Modify instrument adjustments
- Control : Check operation state of instrument
- Diagnosis : Read-out of firing automats
(only in connection with ecom-AK) /
dT-measurement
If measurements are carried out first the
basic module has to be turned on
(switch below the info display). Select
with the cursor keys the sub-menu
"Gas analysis" and confirm with <OK>.
The instrument starts a 1- minute cali-
bration phase and it enquires if you wish
to use the data bank. If you want to
assign the sampled data to a specific
plant, so press <F1> (<F4> = no:
measurement will be performed without
assignment).
Fuel types acc. to 1
st
BImSchV*:
Fuel oil (B)
Natural gas (B)
City gas (B)
Coke oven gas (B)
Liquid gas (B)
Use the cursor keys to select the de-
sired fuel type and confirm with <OK>.
* Country specific fuel types programmable on demand.
Do you wish to
use the data
proce
s
sing?
YES
Quit with:
<
ΟΚ
ΟΚΟΚ
ΟΚ
>
!
NO
Fuel type
Fuel oil (B)
CO2max
A1
B
15.4
0.50
0.007
Select: (↑
↑↑
↑↓)
↓)↓)
↓)
!
!!
!

ecom-J2KN
pro
Page 13
Input number
25.11.2007
Please use the
numeral keys!
7. Input or select plant specific data
To call up plant data recorded in the
ecom-J2KN
pro
or to create a new file,
the following possibilities are available:
Create new (is automatically selected
by first use of a MM card): To create a
new file, a numerical number can be
assigned.
-Select „Create new“
-Confirm with <OK>.
-Input a number (max. 16 numbers):
Example: "25.11.2007"
Tip: We suggest a date-related input to easily find the data record
later on via the search function (search per date).
After confirming with <OK> it is possible to enter a text (max. 6 lines
with 20 indications) with a software keyboard (for print out or data
processing). Proceed as follows:
-choose Text line 1 with the cursor keys <up/down> and confirm
with <OK>
-select keyboard (4 keyboards are available) with <F3>
-select with the cursor keys <up/down/right/left> a character
(selected character is black deposited)
-choose the character with <OK> (the last character can be deleted
by pressing <F2>)
-repeat procedure, until line is complete
-if you want to correct a character, proceed as follows:
-interrupt choosing characters with <F4>
-select character with the cursor keys <right/left>
-activate choosing characters with <F4> and set a new character
-select the next line after pressing <F1>
You can close the text input with <ESC>. Activate the next free rec-
ord number with <OK> and start flue gas measurement with <F1>.
Selection upon:
Search word
Record number
Create new
Quit with:

Page 14 ecom-J2KN
pro
Record number: For check of the plant
already stored in the instrument, the
selection upon record number is most
appropriate.
- Select „Record number“
- Confirm with <OK>.
- Input a random record number:
Example: "1" for record number 1
- Press <OK> once the input is completed to call up said data
record number. The cursor keys <Up/Down> enable the check of
the record numbers.
- Press <F1> to select the first record number and <F2> for the last.
- Press <F4> to delete the content of the selected record number.
- Press <OK> to activate the record number.
- Finally press <F1> to start the gas analysis.
Record number
1
Please use the
numeral keys!
Selection upon:
Search word
Record number
Create new
Quit with:

ecom-J2KN
pro
Page 15
Search word: If the plant code is known, it is possible to find the
plant data stored with help of a search machine.
-Select "Search word" and press <OK>.
-Input with software keyboard 4 related
figures of the plant code:
Example: "25.11"
for plant code 25.11.2007
-Press <F1> after input to start the
searching process. All possible
correspondences with this figures
sequence will be filtered. The selection
can be stepped thru with the arrow keys
(<F1> for selection beginning, <F2> for
selection end)
-Press <OK> to activate once the desired
data block is found
-Press / „View“ / <OK> to view the
previous analysis at this plant
All measured and calculated values can
be called up on 4 display pages using the
arrow keys to step thru.
Press <F1> to quit the previous measurement and start recording of
the current measurement values.
Search word
25.11
Please use the
numeral keys!
Measurement
available
Memory n
umber 1
25.11.2007
End with : <OK> !
F1:First record
F2:Last record F4:Delete
Memory number 1
25.11.2007
Data record 12:15:53 25.11.07
Further pages:<↑↓> Memory number 1
Gas analysis 12:15:53 25.11.07
Further pages:<↑↓> Memory number 1
O2 3.2 %
CO2 13.1 %
CO 0 ppm
Eff. 92.5 %
Losses 7.5 %
Exc. air 1.18
T.Gas 184 °C
T.Air 20 °C
Draught -0.03 hPa
CO measurement 12:15:53 25.11.07
Further pages:<↑↓> Memory number 1
O2
17.5
%
CO 0%
738
ppm
CO
123
ppm
Lambda
7.00
O2 value in air
O2 value in air 12:15:53 25.11.07
Further pages:<↑↓> Memory number 1
O2
19.5
%
CO
3
ppm
Zug
0.01
hPa
Soot..Oil trace
Mean value : 0.5
Boile
r temp. : 65°C
1st Soot meas.: 0.5
2nd Soot meas.: 0.3
3rd Soot meas.: 0.7
Oil trace : NEIN
dT measurement
dT measurement 12:15:53 25.11.07
Further pages:<↑↓> Memory number 1
T1
70.4
°C
T2
56.3
°C
dT
14.1
°C

Page 16 ecom-J2KN
pro
8. Gas measurement
8.1. Gas analysis
After the 1-minute calibration phase, the instrument switches over to
the measurement mode. The gas measurement values can be
viewed on different display pages. Use the cursor keys to scroll the
pages.
With <F1> you can switch from the measured value display to a
menu or to „Standby“ selected before (see chapter “Adjustments”).
Possible menu options are: Soot... Oil trace, Data processing, Ad-
justments, Control, Fuel type, Eff. (C), Memory -> M, Pressure. Fur-
ther you can switch with <F1> from any menu back to the measured
value display.
With <F2> you can print and store the measured values into the in-
termediate memory at the same time.
With <F3> you can switch off the CO sensor, in order to protect the
sensor against too high concentrations. The automatic disconnection
will switch off the CO sensor at approx. 4000 ppm.
With <F4> you can switch from the measured value display to a
menu or to „Standby“ selected before (see chapter “Adjustments”).
Possible menu options are: Soot... Oil trace, Data processing, Ad-
justments, Control, Fuel type, Eff. (C), Memory -> M, Display values,
Pressure. Further you can switch with <F1> from any menu back to
the measured value display.
Re
-
calibrate the instru
ment after each
measurement (after one hour at the latest)
to get correct results!
Hotkey
<F1>
Hotkey
<F4>
Store and print
measured
values <F2>
Switch off
CO-Sensor
<F3>
O2
3.2
%
CO2
13.1
%
T.Gas
184
°C
T.Luft
20
°C
Gas analysis
25.11.07

ecom-J2KN
pro
Page 17
The position of the measured and calculated values (gas analysis)
on the display pages is free selectable (choose „Display values“ for
<F4> Hotkey). For alteration of the existing succession respectively
personal listing, proceed as follows:
-Press <F4> to activate the function.
-select the line with the cursor keys <up/down>,
-select the measured or calculated value with the cursor
keys <right/left>,
-repeat this procedure until all modifications are completed.
-Press <F4> to deactivate the function.
Position the sampling probe in the exhaust channel so that the ther-
mocouple is fully surrounded with the gas (see picture).
Perform the measurement in the core stream of the exhaust gas
channel (probe placed in the highest gas temperature area). A trend
indication for T.Gas easies the core stream search. As long as the
display shows a + symbol, the measured temperature increases, it
means the probe tip moves towards the core stream. If a - symbol is
displayed, pull the probe out of the core stream and the temperature
sinks. If no temperature change is shown for at least 3 seconds, so
the trend indication will be deleted.
Core stream search
O2
3.2
%
CO2
13.1
%
T.Gas
184
°C
T.Air
20
°C
Gas analysis
25.11.07
+
Gas stream
Probe tip
Protection bow

Page 18 ecom-J2KN
pro
CO
2
, efficiency, losses, excess air and dew point are calculated val-
ues. They can only be calculated if realistic values for the basic pa-
rameters O
2
and temperatures are available. It must be ascertained
that:
O
2
< 20,5 % and T.Gas - T.Air > + 5 °C
are given. The dew point can only be calculated accurately if, in the
menu "Adjustments", the current barometric air pressure value has
been entered. This value cannot be determined by the ecom-
J2KN
pro
. If the gas temperature falls below the dew point (between
25 and 65 °C), ETA will be calculated with condensation. In the dis-
play (C) appears behind ETA.
Correct measurement values are displayed first after a short delay,
necessary for the gas transport and the build-up of a stable electro-
chemical reaction at the sensors. This time period lasts approx. be-
tween 1 and 1.5 minute. For recording, printout and evaluation wait
until the values do not change anymore. If deviations higher than 2
ppm still occur by the gas values, they can be due to unstable pres-
sure conditions in the exhaust channel.
If the measurement values are stable
and the results can be printed out, press
the key <Record> (disc symbol) to
transfer the values in the intermediate
memory (caution: store gas analysis
and CO measurement values separate-
ly). The values are stored for a later
printout and, if need be, for a final data
record storage.
If a printout of the values should be made simultaneously to the in-
termediate recording, so press <F2> (the complete content of the
intermediate memory will be printed).
Measurement stored in
intermediate memory
O2
3.2
%
CO2
13.1
%
T.Gas
184
°C
T.Air
20
°C
Gas analysis 25.11.07
recorded!

ecom-J2KN
pro
Page 19
8.2. CO measurement (gas channel check)
For the technical check of gas-fired plants in regards of safety as-
pects the gas channel check called also CO measurement is used.
Hereby the CO concentration in the gas channel is measured after
the flow safety device and calculated on an undiluted value (oxygen
rest content in flue gas = 0 %).
As the gas conditions after the flow safety device are no more ho-
mogeneous because of the flow in of secondarily air and conse-
quently the core stream measurement can be erratic, the analysis of
the exhaust gas is performed along the totality of the exhaust pipe
diameter. A multi-hole probe (optional accessory) is hereby used as
sampling probe. The calculated value shown on the line CO 0 %
corresponds to the measured CO concentration supposed the oxy-
gen content would amount 0% by the same exhaust gas volume.
It is consequently the undiluted CO
content in exhaust gas. If the value indi-
cation is stable, press the key
<Memory> (disk symbol) to store the
result in the intermediate memory. If a
printout of the values should occur sim-
ultaneously to the recording in the in-
termediate memory, press <F2> (the
complete content of the intermediate
memory will be printed out).
O2
17.5
%
CO 0%
738
ppm
CO
123
ppm
Exc. air
7.00
CO measurement 25.11.07
recorded!
Measurement stored in
intermediate memory

Page 20 ecom-J2KN
pro
8.3. O
2
check
This measurement is performed by room-independent plants like
gross calorific value plants. It is determined if exhaust gas flows into
the combustion air (O
2
content drops down / CO content may be rise)
and herewith influence on the combustion quality.
For this analysis a special multi-hole
probe (optional accessory) should be
used. If the value indicated is stable,
press the key <Memory> (disk symbol)
to store the value in the intermediate
memory. If a printout of the values
should occur simultaneously to the re-
cording in the intermediate memory,
press <F2> (the complete content of the
intermediate memory will be printed
out).
8.4. Flow measurement (option)
This measurement can be done with a
pitot tube. At first the pitot factor of the
pitot tube must be entered („Adjust-
ments“ / „Internal“ / „Pitot factor“).
After connecting the pitot tube to the
instrument, the zero point of the sensor
can be set with <F4>. With <F1> the
cross section of the flow channel can be
entered (needed for calculation of the
flow rate). After the pitot tube is posi-
tioned in the flow channel, the display
shows the speed (m/s), the flow rate
(Nm
3
/h) and the differential pressure
(Pa). If the value indicated is stable,
press <Record> to store the value in
the intermediate memory. If a printout of
the values should occur simultaneously
to the recording in the intermediate
memory, press <F2> (the complete
content of the intermediate memory will
be printed out).
O2
19.5
%
CO
3
ppm
Draught
0.01
hPa
O2 value in air 25.11.07
recorded!
O2 value in air
V.Gas
0.4
m/sec
M.Flow
44
Nm3/h
dP
0.1
Pa
Flow measurem. 25.11.07
recorded!
Flow measur
e
ment
Connections for
pitot tube
Measurement stored in
intermediate memory
Measurement stored in
intermediate memory
Other manuals for J2KNpro Tech
5
Table of contents
Other Ecom Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Agilent Technologies
Agilent Technologies E8363A Installation note

Dräger
Dräger FG7000 Short operating instructions

Aquametro Oil & Marine
Aquametro Oil & Marine CONTOIL VZF II DN 15 Spare Parts - Repair - Maintenance

VOLTCRAFT
VOLTCRAFT DO-600 operating instructions

Signal Hound
Signal Hound USB-SA124B user manual

EVC
EVC EVC63A 1P-Mod quick start guide