GE Druck PC6-IDOS User manual

Druck PC6-IDOS
Pressure calibrator (Bench/panel mount version)
User manual - K0354
GE
Sensing
A-8010 Graz, Riesstraße 146
Tel.: +43 316 40 28 05, Fax: 40 25 06
Sensorik
Messtechnik
Han elsgesellschaft m.b.H.

Publication
K0354, Issue 1
Copyright
© 2006 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.

CONTENTS
Welcome ( Indicator Note )
Safety Warning
Quick Reference
Specifications
1.0
1.1
1.2
1.3
1.4
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
PC6-IDOS Arrangement
Functions (menu 1)
Functions (menu 2)
1.5
1.6
1.6.1
1.6.3
1.6.5
1.6.6
1.6.7
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6
1.7.7
1.7.8
1.7.9
1.7.10
1.7.11
1.7.12
1.7.13
1.7.14
1.7.15
1.7.16
1.7.17
1.7.18
1.7.19
1.8
1.9
1.10
1.11
1.12
1.13
2.0
2.1
2.2
2.3
2.4
2.5
2.6
2.10
2.12
3.0
3.1
3.2
3.3
Menu 1, Menu 2 & Menu 3
Menu 1
Menu 2
Menu 3
...................... ii
........................................... iii
iv
............................................. v
Introduction ...................................... 2
General ............................................. 2
Keypad ............................................. 2
LCD .................................................. 3
Connection Panels ( .......... 3
4 mm Sockets ............................ . 3
Quick connect coupling ........... .. 3
RS232 ..................................... .... 3
Charger ...................................... . 3
EPM ........................................... . 3
............................... 4
....................................
.................................... 6
Introduction ......................................... 10
Connections
Pressure Input .................................... 10
Voltage/Current Input ...................... 10
RS232 ................................................ 11
Battery charger ................................... 11
Voltage Output .................................. 11
Messages & Parameters
Retained parameters .... .. 12
Battery low state .............................. . 12
Tare Value .......................................... 12
Maximum/Minimum ......................... 12
Void leak test ..................................... 12
Switch Off .......................................... 12
Auto Switch Off .................................. 12
Pressure Overload ............................ 12
Low Battery Detect ............................ 12
Pressure Switch State Change .......... 12
Printer Busy ........................................ 12
Outside User Set Alarm Value ........... 12
Display Hold ...................................... 12
Connection Fault ............................... 13
Short Detected .................................. 13
Voltage / Current Overscale .............. 13
RS232 Output Format ....................... 13
Alarm Flag ......................................... 13
Symbols ............................................. 13
Units ................................................... 15
Max,Min,Switch,Temp.,
Tare .................................................... 15
V / mA ................................................. 15
Leak Test ............................................ 17
File (logged files) ................................ 18
Alarm Settings - High / Low ............... 21
V / mA (Measure mode)............................ 22
V-Out, Display light, Port ................... 26
Resolution ........................................... 26
Event ................................................... 26
Log ..................................................... 27
Zero .................................................... 30
Port
Power_Off ........................................... 31
Span ................................................... 33
History (overload, zero/span, ...)........... 35
Language .................................... 35
Units ................................................... 35
..........................
Unit, Leak ....................................... 38
V / mA ............................................ 39
File ................................................. 40
High & Low .................................... 41
V / mA (measure) ........................... 42
V / mA (source) .............................. 43
V-out, Display light ....... 48
Resolution, Event, RS232 ............. 48
Log ................................................ 49
Date&Time, Language, Zero, Port,
Power_Off ...................................... 50
Span .............................................. 51
History ............................................ 52
Language ........................................ 53
Units ................................................ 54
(Keys, Activating functions, Exit,
Menu selection, Data Entry, Tare) ............................
Maintenance
Functions (menu 3)
1.6.2
1.6.4
2.7
2.8
2.9
................................................ vi
Front/Rear)
........
..........
........
........
........
.................................... 8
......................................... 10
.
EPM ..................................................... 10
Switch Test........................................... 11
.
.
.................... 12
.................. ....
.
Date&Time,% 15
.
V / mA (Source mode)...............................23
..
.
RS232 ................................................. 29
Date & Time ....................................... 30
Language ............................................. 30
....................................................... 31
.
.........
..
.................
........
5
37
1.7
2.11
Description 1
Operating Instructions 9
How to Select Functions (Menu 1) 14
Set-menu (menu 2) 20
Cal-menu (menu 3) 32
Flowcharts 36
Appendices 55
Index 62
1External Pressure Module (EPM) 55
2
3
Calibration Procedures......................... 58
Panel Mount Dimensions...................... 60
Page i
K0354, Iss. 1

WELCOME
This detailed operating manual will help you to become familiar
with the many features of the PC6-IDOS Calibrator. The simple step by step
instructions will quickly guide you through the procedures for using the
calibrator and other items needed for accurately testing and calibrating
numerous types of instruments. Please t
efore you begin to use the PC6-IDOS.
ake time to carefully read
the whole manual b
Two versions of the PC6-IDOS are available:
PC6-IDOS Calibrator:
PC6-IDOS Indicator:
All of the functions shown in this manual apply to this unit. The Calibrator can be supplied with
a PV211 pneumatic test
system for pressures up to 40 bar (600 psi) or PV212 hydraulic test system for pressures up to 700 bar
(10 psi). Refer to the instructions supplied with the applicable pressure system.
This is a simplified version of the Calibrator and does not include the Voltage/Current
input/output.
PC6-IDOS
PV210 Low pressure pneumatic test system for pressures up to 3 bar (45 psi),
,000
PC6-IDOS
Page ii
K0354, Iss. 1

SAFETY WARNING!
HIGH PRESSURE
RECHARGEABLE Ni-MH BATTERIES
:
Uncontrolled release of high pressure is hazardous to personnel and may cause
damage to equipment. Before connection of any pressure component to the PC6-
IDOS ensure that the component(s) is/are isolated from the pressure supply and any
internal pressure is released slowly.
Rechargeable Ni-MH batteries must be recycled or disposed of properly.
CAUTION: Use charger supplied by GE Sensing only.
May explode if damaged or disposed of in fire. DO NOT short-circuit.
Ni-MH
Page iii
K0354, Iss. 1

QUICK REFERENCE
Keys
Press to turn PC6-IDOS on and off
Press to access the menus
Press to accept functions or settings
Press to move cursor up/left or increase values
Press to move cursor down/right or decrease values
Press to print/log/start event depending on previous
setting
Select menu position cursor press the Enter key
under function
To exit out of any menu select 'Exit' option,
or press key.
Activating Functions
Exit
Menu Selection
Pressure
Display
1st Press for Menu 1
2nd Press for Menu 2
3rd Press back to Pressure Display /
Calibration Menu (see page 33)
Press or to position cursor and press
to select required alphanumeric. Press
key to release cursor.
For next data entry, repeat the above procedures then
press once all data is entered.
Note: To quickly scroll through alphanumeric lists hold
down the or key.
Data Entry
Tare
This is a
temporary function and will be cancelled when the unit
is switched off.
Hold down the key for 4 seconds while in
the pressure display mode. The unit will bleep after two
seconds when the Max. and Min. values reset. After
another two seconds a second bleep will sound as the
displayed pressure now resets to zero.
MENU
MENU
MENU
MENU
MENU
ON/OFF
MENU
MENU
Page iv
K0354, Iss. 1

SPECIFICATIONS
* Temperature coefficient 0.005% of reading / °C w.r.t. 20°C (68°F)
15% FS
6 Digits
External Pressure Modules. See Appendix 1 for
EPM details.
bar, mbar, MPa, kPa, psi, kg/cm², atm, inH 0,
mH 0, mmH 0, inHg, mmHg, UNIT1, UNIT2 (user
defined)
emperature effects are included in the accuracy.
1011 records into a maximum of 20 files,
64k memory
Backlit LCD, 16 character x 2 line alpha numeric
Overload Pressure
Resolution
EPM
Pressure Units
Data Capacity
Display
2
2 2
Overload Warning
Temperature Display
Pressure Switch Test
Operating Temperature
Calibration Temperature
Calibrated Temp. range
Storage Temperature
Temperature Effects
RS232 Parameters
Zero Reset
Humidity
Resolution Select
Power Supply
Voltage Output
Flashing display and audible tone at 115% of FS
Temperature of pressure sensor
(Accuracy: 1 °C (1.8 °F) )
Status OPEN/CLOSED
0 to 50 °C (32 to 122 °F)
20 2 °C (68 3.6 °F)
5 to 45 °C (41 to 113 °F)
-20 to 70°C (-4 to 158 °F)
T
baud rate 1200, 2400, 4800, 9600, 19200 and
38400. Stop bits 1 or 2, status ON/OFF
Manual keyboard operation
5 to 95% Relative Humidity non condensing
Increase or decrease by a factor of 10
6V Ni-MH rechargeable battery pack (see
warning on page iii)/ mains operation via charger
(supplied).
Regulated DC supply 10V 0.1%
max current 10mA
Unregulated DC supply 24V 10%
max current 50mA
Pressure Range
Bar psi
-0.07 to 0.07 -1.0 1.0
-1 to 1 15
-1 to 2 30
-1 to 3.5 50
-1 to 7 100
-1 to 10 150
-1 to 20 300
-1 to 35 500
0 to 70 1,
0 to 100 1,500
0 to 200 3,
0 to 350 5,
0 to 700
to
-0.2 to 0.2 -3.0 to 3.0
-0.35 to 0.35 -5.0 to 5.0
-0.7 to 0.7 -10 to 10
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
0 to 000
0 to
0 to 000
0 to 000
0 to 10,000
Accuracy (FS)
up to
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.05%
0.05%
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Model
-2-C
-10-C
-20-C
PC6-B-350-C SG
PC6-B-700-C SG
PC6-B-0.07-C G
PC6-B-0.2-C G
PC6-B-0.35-C G
PC6-B-0.7-C G
PC6-B-1-C G
PC6-B G or A
PC6-B-3.5-C G
PC6-B-7-C G or A
PC6-B G
PC6-B G or A
PC6-B-35-C G
PC6-B-70-C G
PC6-B-100-C G
PC6-B-200-C G
Resolution
1 V
0.1mV
1mV
1 A
µ
µ
Measurement
Voltage *
Current *
Auto Range
(Auto Range)
Accuracy
0.025% FS
0.025% FS
0.025% FS
0.025% FS
0.025% FS
Range
0 to 100mV
0 to 5V
0 to 50V
0 to 50mA
0 to 25mA
Input resistance
1 MΩ
1 M
1 M
6
Ω
Ω
Ω
Resolution
0.2mV
1 Aµ
Sourcing
Voltage *
Current *
Accuracy
0.025% FS
0.025% FS
Range
0 to 10V
0 to 50mA
Max. Load
20mA
330 Ohm
†
†
†
Other Pressure ranges are available to special order.
SG = Sealed Gauge,G = Vented Gauge, A = Absolute
° (32 to 122 °F)†0.075% FS accuracy, 0 to 50 C
Page v
K0354, Iss. 1

SPECIFICATIONS / MAINTENANCE
Battery Life
Recharge Time
Low Battery
Dimensions (l: w: h)
Weight
Microprocessor
Recommended Recalibration Period
Electrical Connections
Pressure Media
Pressure Connection
Software (optional)
System requirements:Fully charged: up to 8 hours 1
14-16 hours for full charge
Continuous check, audio and visual warning
x 180 x 85 mm (7.9 x 7.1 x 3.3 in)
Approximately kg (4 lb)
H8S/2138
1 Year (Max)
Ground Standard 4 mm socket
Voltage/Current Input/Output Standard 4 mm sockets
Pressure Switch Input Standard 4 mm sockets
Battery 2 pole miniature round connector
RS232 8 pole miniature round connector
EPM (P2) 5 pole miniature round connector
Water, Mineral Oil, Non-corrosive gases
(for other media, contact GE).
‘quick-fit’ connector / 1/4”BSP or 1/4”NPT female
adaptors.
The ‘ ’ software will present
downloaded data in various graphical forms such
as gauges or a scrolling bar & remotely control
the PC6-IDOS via the RS232 cable linked to a
PC. The downloaded data can be stored in a
variety of file types compatible with most analysis
database and word processing programs.
Calibration procedures can be uploaded to the
PC6-IDOS.
200
1.8
Minimum Intel Pentium with Windows 95 or
higher.
1The battery usage time specified here is heavily dependent upon the
functions selected. For instance, using the back-light function will
reduce the battery usage time, as will source mode functions.
Clean the case with a moist, lint-free cloth and a weak detergent. Do not
use solvents or abrasive materials.
Do not do local repairs. Return the equipment to the
manufacturer or an approved service agent.
Do not dispose of this product as household waste. Use an
approved organisation that collects and/or recycles waste
electrical and electronic equipment.
For more information, contact one of these:
• our customer service department
(Contact us at www.gesensing.com)
• your local government office
SiCal PRO
® ®
Ordering Codes:
Maintenance
Page vi
For Panel Mount version add a 'P' to the order code i.e.
PC6-BP-XXXX-XXXX-C-1. See page 61 for panel dimensions.
PC6 - B - XXXX - XXXX - C - 1
Range 1 (R1)
(R1 > R2)
C Ior (Calibrator or Indicator)
1 2 3 4-UK / -Europe / -USA / -AUS
Range 2 (R2)
(optional)
K0354, Iss. 1

DESCRIPTION
1Page
K0354, Iss. 1

1.0 Introduction
1.1 General
1.2 Keypad
The Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal 6V
rechargeable battery pack. It is portable and capable of accurate pressure measurement, voltage or current
measurements and sourcing when calibrating transducers, pressure switches, etc. This section describes the equipment
from an external viewpoint, allowing the user to become familiar with the various controls and connections provided.
The PC6-IDOS is housed in a painted metal
The Keypad is a non-tactile membrane keyboard with six buttons. The keyboard layout is shown below.
An audible tone accompanies
every key operation.
PC6-IDOS
case with a carry handle/stand, a liquid crystal display unit at the front and a
six button keyboard. The main body contains the circuitry, connections, internal transducer(s) and battery pack.
Note:
PRINT
LOG
ON/OFF
MENU
ENTER
ZERO
2Page
K0354, Iss. 1

1.3 Liquid Crystal Display (LCD)
1.4 Connection Panels (Front/Rear)
The LCD is a 16 character x 2 line alphanumeric LED
backlit display which is capable of displaying special
characters.
Pressure reading Unit of measurement
Function Function Value
Front connection panel: (See Page 4, Fig 1)
This connection panel provides input / output connections as
detailed in Fig.1
on the front panel provide
Pressure input
connections are made to
1.4.1 4 mm sockets.
1.4.3 RS232.
SiCal PRO
1.4.4 Charger.
1.4.5 EPM.
Six standard 4 mm electrical sockets
mounted input or output current
and voltage and switch status.
Voltage outputs of 10 and 24V are available for use with
sensors, transducers, transmitters, etc.
An 8 way circular miniature
connector is used for communication with
RS232 standard communication
protocols/equipment. The user software
‘ ’ (optional) can be used to
remotely monitor or download/upload data
from the PC6-IDOS via this connector.
(Optional RS232 cable - Part No. 01-0625)
The 5 way circular miniature
connector is used to connect to the EPM
(External Pressure Module). These optional
modules cover pressure ranges from 0.07 to
700 bar (1.0 to 10,000 psi).
Rear connection panel:
(See Page 4, Fig 2)
1.4.2 ‘Quick Fit’ coupling.
'Quick-fit'
connectors / 1/4BSP or 1/4NPT female
adaptors.
The 2 way circular miniature
connector is used for the 9V charger/power
supply (provided).
This connector will not be available
on a dual (internal) sensor unit.
Note:
Special
Symbols
3Page
COUPLING
P2
9V DC
RS232
K0354, Iss. 1

+10V
V
+24V
mA GND
SWITCH
Front View:
LCD, Keypad & Connections
Fig 1
Front Connection Panel
Display.
Keypad
See Page 2
ARRANGEMENT OF PC6-IDOS
4Page
+24 V Volts output.
+10 V Volts output.
GND Common Ground Socket
mA Milliamp Input (controls input current
in source mode).
V Volt Input (output in source mode).
SWITCH Pressure Switch Input.
(See Page 10
for V/mA
connections)
EPM
Rear View:
Pressure / RS232 / Power connections
Fig 2
9V DC / Battery Charger
Input Socket
See Page 11
9V DC
RS232
P1
PRESSURE INPUT
External Pressure Module Socket
See Note on page 3
RS232 Communication
Socket See Page 11
Pressure Input connections
See Page 3
P2
Druck PC6-IDOS
K0354, Iss. 1

PRESSURE DISPLAY -
Selection of pressure unit.
inches of mercury (at 0°C) of mercury
user defined unit second user defined unit
Returns to menu 1
Displays the maximum pressure measured since the last reset (60 samples per second).
Displays the pressure as a percentage between user pre-set high and low alarm values.
Displays the pressure at the time of tare.
Volts / milliamps functions
Displays the following input measurements.
Displays actual input measurements.
Displays converted input measurements.
Returns to Pressure Display.
Indicates the status of the pressure switch under test (OPEN/CLOSED).
Allows the PC6-IDOS to measure pressure changes over time.
Starts the leak test.
Sets the leak test time.
Returns to Pressure Display.
Allows the user to monitor the temperature around the selected pressure transducer in pressure display mode.
Allows the user to monitor the date & time of the on board real time clock in pressure display mode.
Allows the user to view or manipulate a file previously stored in the filing system.
Allows the user to see logged file.
Outputs a logged file
Imports a file (refer to ‘ instructions for exporting files to PC6-IDOS)
Deletes a single file or all logged files.
Returns to Pressure Display.
Returns to Pressure Display mode.
Allows the user to operate the PC6-IDOS for pressure measurements. (Display Functions)
SiCal PRO’
(SHORTCUT: To abort any menu level, simply press the MENU key).
Menu 1
Unit
inHg mmHg
UNIT1 UNIT2
Exit
Max
%
Tare
V / mA
Measure
Actual
Converted
Exit
Switch
Leak
Start
Set-Time
Exit
Temp C
Date & Time
File
View
Export
Import
Delete
Exit
Exit
bar millibar
megapascals kilopascals
pounds force per square inch kilograms force per square centimetre
atmospheres i inches of water (at 4°C)
metres of water (at 4°C) millimetres of water (at 4°C)
millimetres (at 0°C)
Displays the minimum pressure measured since the last reset (60 samples per second).
Displays the following outputs (status must be ON - see ‘Setup’ in Menu 2)
Displays actual output measurements.
Displays converted output measurements.
Displays Pressure to Volts / milliamps conversion.
Returns to Pressure Display.
Returns to pressure display
bar mbar
MPa kPa
psi kg/cm
atm nH O
mH O mmH O
Min
Source
Actual
Converted
P to V / P to mA
Exit
Exit
2
2 2
FUNCTIONS (MENU 1)
5Page
K0354, Iss. 1

6Page
Continued on
Page 7
}
FUNCTIONS (MENU 2)
SETTING MENU
Menu 2
- To set user specific functions. (Set Functions)
High
Low
Measure
Zero
FS
Text
Root
Source
Setup
Status
Root
Step
Ramp
Status
Rate
Dwell_time
Exit
Exit
Exit
Allow the user to set a High or Low Pressure Alarm Limit.
(by using the increase/decrease facility above which an alarm sounds.)
Voltage / Current input settings for Converted Measurement.
Set Zero.
Set Full scale value.
User defined measurement text.
Switch measurement root function ‘on’ or ‘off’.
Voltage / Current output settings.
Sets the parameters for voltage/current sourcing.
Switches output on or off.
Switch sourcing root function ‘on’ or ‘off’.
Set step value (%FS)
Settings for Ramp mode
Switches ramp mode on or off.
Sets the ramp rate (V or mA/Sec).
Sets the time for which the output remains at the same level.
Returns to Source sub-menu.
Returns to Source sub-menu.
Returns to Pressure Display.
V / mA
Exit
Range
Zero
FS
Exit
Convert
Zero
FS
Text
Exit
P to V / mA
Zero
FS
Exit
Exit
Reverts back to Pressure Display.
Sets the following values:-
Sets Zero value
Sets Full Scale value
Returns to Source sub-menu.
Sets the following converted values:-
Sets converted Zero values.
Sets converted Full Scale values.
User defined converted text.
Returns to Source sub-menu.
Sets Pressure to V / mA conversion factors.
Set zero values.
Set Full Scale values.
Returns to Source sub-menu.
K0354, Iss. 1

FUNCTIONS (MENU 2) Continued.
Page 7
Switches the Power Supply outputs +10V & +24V ON or OFF
(separate pc software utility needed -contact supplier).
Vout
Switches ON or OFF the LED back light.
Selects low, normal or high resolution.
Sets up the event status for RS232 output and for logging functions.
Sets key driven output or logging.
Sets timed output defined by the user.
Reverts back to Pressure Display.
Allows the user to setup, amend and view files for logging.
To select a log file and start logging.
To look at a file.
Deletes a file.
To enable or disable logging.
To export logged files.
To import files from a PC. (see ‘ ’ software instructions for exporting files to the PC6-IDOS)
Reverts back to Pressure Display.
Settings for RS232 communications.
To enable RS232 communication.
To select the baud rate.
To select the number of stop bits.
Reverts back to Pressure Display.
Sets up Date & Time parameters
Sets the date & time.
To look at date & time.
Reverts back to Pressure Display.
Allows user to change language
Will reset pressure reading to Zero. ( This will remove Zero offset permanently / NOT available for absolute
units ).
Selects current pressure module (P1, P2)
Allows the user to set Power_Off time or to switch the function on or off.
Switches the auto-power function off.
Allows the user to set the auto-power off time in minutes.
Reverts back to Pressure Display.
Returns to Pressure Display
Display Light
Resolution
Event
Key
Timed
Exit
Log
Log
View
Delete
Status
Export
Import
Exit
RS232
Status
Baud-rate
Stop-bits
Exit
Date/Time
Set
View
Exit
Language
Zero
Port
Power_Off
Key
Timed
Exit
Exit
SiCal PRO
K0354, Iss. 1

FUNCTIONS (MENU 3)
8Page
CAL MENU
Menu 3
Span
Continue
Exit
History
Overload
Zero-Span
Exit
Language
Units
View
Alter
Exit
Exit
Allows user to set span.
Starts span calibration procedure.
Returns to Cal Menu.
Allows user to see overload and zero-span history.
Displays period of time over which overload occurred.
Displays date last calibrated, last span set, range and overload value.
Returns to Cal Menu.
Allows user to download different languages.
Displays the two current user-defined units.
Allows user to modify the conversion factor and/or text for the two user-defined units.
Returns to Cal Menu
Returns to Pressure Display.
PC6S/No Displays the unit’s serial number.
Displays the serial number of the currently selected pressure module.PM S/No
K0354, Iss. 1

OPERATING
INSTRUCTIONS
9Page
K0354, Iss. 1

10Page
Fig.3
Input
0V
Output
Three Wire
Transmitter
0 to 5V
+10V
V
V
+24V
+24V
mA
mA
GND
GND
SWITCH
SWITCH
+10V
EPM
+V
-V
Two Wire
Transmitter
4 to 20 mA
1.5 Introduction
The PC6-IDOS enables the user to carry out checks on
pressure transducers, transmitters, pressure switches, pressure
gauges and process indicators quickly and accurately. It is
intended for use as a portable, on-site calibrator, used in close
proximity with the equipment under test and without the need to
remove the unit from its parent system. This section describes
the connections to be made when checking switches,
transmitters, etc. It also lists menu options, selection procedures
and purpose of options available.
1.6 Connections
1.6.1 Pressure Input The PC6-IDOS measures pressure
directly
Alternatively, connect the item under test to the PC6-
IDOS with the adaptors and seals provided.
via the 'Quick-Fit' connectors found at the rear of the
case.
1.6.2 EPM (External Pressure Module)
Different pressure ranges can be read by attaching an
External Pressure Module to the 5-way socket (only available
on single range instruments - ).See Fig.2
Warning: The PC6-IDOS must be switched off prior to
connecting or disconnecting the EPM .
1.6.3 Voltage / Current Input / Output Voltage and Current
readings can be taken by connecting the outputs from the item
under test to the 'V' or 'mA' and 'GND' input sockets on the
connection panel using the connection leads provided.
Warning: When measuring voltage or current, make sure
'GND' is connected (where applicable, see fig. 3)
P2
K0354, Iss. 1

1.6.4 Switch Test Activate ‘Switch’ function in menu 1.
The switching pressure of a pressure switch can now be
monitored by connecting the contacts of the pressure switch
between the INPUT 'GND' and 'SWITCH' sockets on the
connection panel and pressurizing the pressure switch.
1.6.5 RS232
If the RS232 facility is to be used then connections to the 8-
way socket must be made to the specifications given in the
table below. Alternatively an RS232 communication cable
can be purchased from GE suitable for most RS232
applications as shown below.
To RS232 Standard 9-way
socket on PC6- 'D'type female
IDOS connector
Pin 6 and pin 8 should not be used.
RS232 Connections Table
Note:
1.6.5.1 RS232 Output
Warning:
1.6.6 Battery Charger
CAUTION:
1.6.7 Voltage Output
Connect the PC6-IDOS to a printer or computer using a suitable
cable (not supplied) as detailed above and set the parameters as
detailed in Para. 2.5 & 2.7. The output can now be obtained as
‘single shot' or ‘timed' as follows:
- This output is available in Display mode when
RS232 status is ON & Log status is OFF. The output cycle is
initiated by pressing the key. Pressing the key cancels
the output. Once initiated, the output is repeated at the selected
time interval (see ‘Event' function, page 26 ) until the key is
pressed.
The internal battery pack may be
recharged (when indicated by the battery low message) by
plugging the charger (supplied) into the '9V' socket.
When recharging internal batteries, only the charging
unit supplied with the PC6-IDOS should be used.
A voltage of +10V or +24V may be
sourced from the PC6-IDOS by connecting to the 4 mm sockets.
A typical application for this voltage output would be to power a
pressure transmitter. ( see Fig. 3. page 10 )
To switch on voltage output function:-
a) Press the key twice to activate 'Set' menu.
b) Select 'V-out' and press the key.
c) Select 'ON' and press the key.
1.6.5.2 Single Shot - This output is always available in Display
mode when RS232 status & Log status are OFF. The output is
obtained each time the key is pressed (provided any
previous output is complete).
The PC6-IDOS must be switched off prior to
connecting or disconnecting the RS232 cable.
1.6.5.3 Timed
SIGNAL
Data Terminal Ready (DTR)
Transmit Data (TxD)
Receive Data (RxD)
Request To Send (RTS)
Clear To Send (CTS)
Do not use
Ground
Do not use
STATE
Held High
Active High
INPUT/OUTPUT
Output
Output
Input
Output
Input
Output
Common
Input
PIN No.
1
2
3
4
5
6
7
8
PRINT
LOG
PRINT
LOG MENU
MENU
MENU
Cable Part No: 01-0625
11Page
ENTER
ZERO
ENTER
ZERO
K0354, Iss. 1

1.7 Messages & Parameters
The PC6-IDOS registers pressure input at switch on after the initial start-
up message. The pressure reading is displayed on the top line of the
LCD (see Page 3). Pressure values are monitored, alarm values are
compared, alarm warnings and other information messages displayed
until the unit is switched off.
Any read / write errors to memory detected at switch on will be reported
as: Memory Error ‘Error No.’
When switched on, the unit registers the
same state as when last switched off. The display mode (Operating/Set),
units of measurement, function and parameter settings, alarm limits and
RS232 output control, Port (P1 or P2) selection, logged parameters are
all retained at switch off, with the exception of the following conditions:
A battery low state occurring prior to the switch
off. When switched on, the unit registers the latest retained set of
parameters.
Any TARE value set during use is not retained and
therefore should be set, if required, each time the unit is switched on.
Maximum and Minimum values are not
retained. Current values reflect pressure monitored from the start of the
pressure input.
If the unit is switched off during a LEAK test then the
test will be void. When switched on again the unit will start up in LEAK
test mode.
When the key is pressed the 'switch off' message
is displayed for two seconds accompanied by an audible bleep. Normal
operation may be resumed by pressing any key during switch off.
The PC6-IDOS will automatically turn itself of
after a set period (See Page 31 for instructions)
(Bleep rate 4 Hz) ‘OVERLOAD' occurs when
input pressure exceeds overload pressure level.
(Bleep rate 1 Hz). ‘Low Battery' occurs when
internal battery pack drops below 5.5V. Message flashes for 4 seconds,
after a further 30 seconds, low battery detection is re-enabled.
(Bleep rate 1 Hz)
If the receiving device is not ready for input (e.g. device not connected or
handshake LOW), 10 seconds after starting an output cycle (single shot
or timed output) the message 'Printer Busy' is flashed on the display,
together with an audible warning. If the device does not accept input
within the next 5 seconds then the output cycle is cancelled and output is
stopped until the key is pressed again.
(Bleep rate 2 Hz)
Occurs when measured pressure is lower than the user set Low alarm
value or greater than the user set High alarm value and the
message is flashed. Indication ceases when pressure is equal to
either alarm value or between the values.
(Bleep rate 2 Hz)
Occurs in pressure mode. If the key is pressed once (less than 1
second) the pressure value (upper line) is frozen and the lower line is
flashed with the ‘Display Hold’ message at 1 Hz. Normal display update
resumes when any top row key is pressed.
Memory Errors:
Retained parameters
Battery low state
Tare value
Maximum/Minimum
Void leak test
Error Numbers:-
0 8 9- Onboard eeprom; - RTC; - 2nd Onboard eeprom.
Errors with Pressure Modules will be indicated by the message: "No P1
(or P2) Module" preceeded by "PM EEPROM ERROR" at power on.
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6 Switch off
1.7.7 Auto Switch off
1.7.8 Pressure Overload
1.7.9 Low Battery Detect
1.7.11 Printer Busy
1.7.12 Outside User Set Alarm Values
1.7.13 Display Hold
1.7.10 Pressure Switch State Change (Bleep rate 2 Hz)
Occurs in pressure mode with the pressure switch state display being
selected. When the connected pressure switch changes state, the
pressure value (upper line) is frozen and the pressure switch state (lower
line) display is flashed at 1 Hz unless a log file is open. Normal display
update resumes when the key is pressed.
’outside
limits’
ON/OFF
12Page
ENTER
ZERO
ENTER
ZERO
PRINT
LOG
K0354, Iss. 1
Table of contents
Other GE Test Equipment manuals