5 K0237 Issue No. 2
(3) Close the manual let-down valve.
(4) Press the RUN key (F1) and operate the hand-pump, continue until the switch
operates:
(5) Carefully open the release valve and allow altitude to decrease until the switch
operates again. The display shows the operating pressures and hysteresis.
(6) Record the displayed operating data.
(7) Allow the indicated altitude to decrease until it is below 1500ft.
(8) Slowly open the manual let-down valve to return the system to ground level.
(9) To perform another switch test press RUN (F1).
(10) To cancel switch test mode, press TASK.
Method (non-accessible switch contacts)
When the altitude switch contacts cannot be connected to the instrument, an external
indicator or annunciator must be used as the indication of pressure switch operation.
The user, on receipt of an indication of switch operation, presses the MANUAL OPERATION
(F2) key:
(1) Connect the instrument to the static system.
Ensure that the pitot system is vented
to atmosphere before starting the test:
(2) Select negative pressure and close the release valve.
(3) Close the manual let-down valve.
(4) Press the RUN key (F1) and operate the hand-pump, continue until indication of
switch operation and immediately press (F2).
(5) Carefully open the release valve and allow the altitude to decrease until the
indication of switch operation and immediately press (F2).
CONTACT STATE
POD=Off
RUN
ALTITUDE ft
CONTACT STATE
POD=Off
STOP MANUAL
ALTITUDE ft
OPERATION
CLOSED AT 5550 ft
CONTACT STATE
POD=Off
STOP MANUAL
ALTITUDE ft
OPERATION
CLOSED AT 5550 ft
OPENED AT 5325
HYSTERESIS 225
ft
ft
SWITCH TEST SWITCH TEST SWITCH TEST