GF 2650 DryLoc®Amperometric Electronics
English
The 2650 Amperometric Electronics provide the polarization voltage and signal
conditioning required by all GF Amperometric Sensors. Sophisticated circuitry
makes system setup and verication easy. The 2650 Amperometric Electronics
processes signals from a chlorine electrode and transmits digital data via a three
wire cable to the 9950-X Chlorine Controller. The DryLoc®electrode connector
quickly forms a robust assembly with the sensor. The Smart Sensor Electronics
also accesses stored factory and real time environmental data stored on the
chlorine sensor.
Features:
• Passes data stored in sensor memory, such as factory calibration data,
service time, range and more.
• GF patented DryLoc®connector provides quick assembly and a secure
connection.
Gold-plated contacts and an O-ring seal ensure a waterproof and reliable
interconnect to the sensor.
• Separate drive electronics from sensor make for easy sensor replacement
without running new cable.
• DryLoc®connectivity allows easy sensor removal for calibration.
* NOTE: The 9950-X Chlorine Controller is not compatible with the standard 9950 controller.
Operating Instructions
Description
• English
• Deutsch
• Français
• Español
• Italiano
• 中文
*3-9900-1P-1.090*
3-2650.090 Rev 12 02/22
WARNING
Refer to the 3-463X Chlorine manual for
additional important information regarding
safety and installation.
2GF 2650 Amperometric Electronics
Refer to your local Georg Fischer Sales o󰀩ce for the most
current warranty statement.
All warranty and non-warranty repairs being returned must
include a fully completed Service Form and goods must be
returned to your local GF Sales o󰀩ce or distributor.
Product returned without a Service Form may not be
warranty replaced or repaired.
GF products with limited shelf-life (e.g. pH, ORP, chlorine
electrodes, calibration solutions; e.g. pH bu󰀨ers, or other
solutions) are warranted out of box but not warranted against
any damage, due to process or application failures (e.g. high
temperature, chemical poisoning, dry-out) or mishandling (e.g.
broken glass, damaged membrane, freezing and/or extreme
temperatures).
Warranty Information
Dimensions
Specications
61.0 mm
(2.40 in.)
94 mm
(3.69 in.)
General
Compatible Electrode .................. All GF Amperometric DryLoc
Sensors
Compatible Instrument ................ 9950-X Chlorine
Controller
Mounting...................................... DryLoc
®
connection
Materials...................................... PC+PBT
Cable ........................................... 4.6 m (15 ft) 3 conductor
shielded, 22 AWG
Shipping Weight .......................... 0.64 kg (1.41 lb)
Performance
Electronics Accuracy ................... < 5 nA or 1% of reading,
whichever is greater @ 25 °C
over full input range
Temperature ................................ ± 1.0 °C (Pt1000) over full
operation range (when calibrated
at ambient temperature)
Resolution.................................... 0.1 nA
System Response ....................... 500 ms (update rate)
Operational Range ...................... ± 450 nA
Electrical
Input Specications
Sensor ......................................... Raw signal
Temperature ................................ Pt1000 RTD
Output Specications
Digital (S³L).................................. Serial ASCII, TTL level 9600 bps
Max. Cable Length ...................... 30 m (100 ft)
Power Supply Input:
Digital (S³L).................................. 5 to 6.5 V ± 10%, 3 mA max
Environmental
Storage Temperature................... –20 °C to 85 °C (–4 °F to 185 °F)
Operating Temperature................ 0 °C to 85 °C (32 °F to 185 °F)
(electronics only)
Relative Humidity......................... 0 to 95%, non-condensing
(no electrode connected)
Enclosure Requirements ............. NEMA 4X/IP65 with electrode
connected
Standards and Approvals
• CE, WEEE
• RoHS Compliant
• Manufactured under ISO 9001, ISO 14001 and ISO 45001
China RoHS (Go to gfsignet.com for details)
Declaration of Conformity according to FCC Part 15
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received,
including interference that may cause undesired operation.
Caution / Warning / Danger
Indicates a potential hazard. Failure to follow all warnings
may lead to equipment damage, injury, or death
Electrocution Danger
Alerts user to risk of
potential of injury or death via
electrocution.
Electrostatic Discharge (ESD)
Alerts user to risk of potential damage to product by ESD
Personal Protective Equipment (PPE)
Always utilize the most appropriate PPE during
installation and service of GF products.
Note / Technical Notes
Highlights additional information or detailed procedure.
• Remove electrical power from the transmitter before wiring
input connections.
• Follow instructions carefully to avoid personal injury or
damage to the electronics.
Safety Information
3
GF 2650 Amperometric Electronics
Do Not Use Lubricant or Sealing Tape on Threads.
Do Not Overtighten.
3610 Flow Tee Installation
1. Lubricate the DryLoc®and the electrode body O-rings
with a non-petroleum lubricant.
2. Insert fully into the tting.
3. Install bottom retaining nut onto the tting and hand
tighten, do not use tools.
1. Lubricate the DryLoc®O-ring with a non-petroleum
lubricant. (DO NOT lubricant the electrode body O-ring.)
2. Insert the sensor into the 2650.
3. Insert into the Chlorine (CI) ow chamber and push
lightly to secure.
pipe tee
pipe tee
Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system.
It is not necessary to lubricate the electrodes body O-ring when using the 463X Flow Cell.
Installation
The bottom yellow nut is not
used when using the 2650
Chlorine electronics with the
463X Flow Cell.
263X Chlorine
Sensor
463X Flow Cell
463X Flow Cell Installation
CHEMICAL COMPATIBILITY WARNING
The retaining nuts of Chlorine, pH and ORP electronics are not designed for prolonged contact with aggressive
substances. Strong acids, caustic substances and solvents or their vapor may lead to failure of the retaining
nut, ejection of the sensor and loss of the process uid with possibly serious consequences, such as damage to
equipment and serious personal injury. Retaining nuts that may have been in contact with such substances, e.g.
due to leakage or spilling, must be replaced.
Wiring
Wiring to the 9950-X Chlorine Controller
To replace the 2650 electronics in a 463X Chlorine panel:
• Measuring from the yellow housing threaded end, cut the cable to approximately 838 mm (33 in.).
• Remove approximately 10 mm (0.4 in.) of insulation and tin each conductor before inserting into terminals.
• For calibration and conguration please refer to the 463X Chlorine Analyzer System and
9950-X Chlorine Controller manuals.
CH1
+V
FREQ
DATA
GND
Black
White
X
No connection
Shield
Red
IN
OUT
pH Cl
9950-X
CH 1
Georg Fischer Signet LLC, 5462 N. Irwindale Avenue, Irwindale, CA 91706 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: www.gfps.com • or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at www.gfps.com
3-2650.090 Rev 12 02/22 English © Georg Fischer Signet LLC 2022
Ordering Information
Mfr. Part No. Code Description
3-2650-7 159 001 670 Ampermetric Electronics
3-2630-1 159 001 746 Free Chlorine electrode, 0.02 to 2 ppm (mg/L)
3-2630-2 159 001 662 Free Chlorine electrode, 0.05 to 5 ppm (mg/L)
3-2630-3 159 001 747 Free Chlorine electrode, 0.1 to 20 ppm (mg/L)
3-2632-1 159 001 767 Chlorine Dioxide electrode, 0.02 to 2 ppm (mg/L)

Other GF Measuring Instrument manuals

GF ER 52 User manual

GF

GF ER 52 User manual

GF Signet 5500 User manual

GF

GF Signet 5500 User manual

GF FlowtraMag GF 2581 User manual

GF

GF FlowtraMag GF 2581 User manual

GF 4630 User manual

GF

GF 4630 User manual

GF TL 160/ 225 User manual

GF

GF TL 160/ 225 User manual

GF U1000 V2 User manual

GF

GF U1000 V2 User manual

GF JRG Coral force 8000 User manual

GF

GF JRG Coral force 8000 User manual

GF signet 2551 User manual

GF

GF signet 2551 User manual

GF 4630 User manual

GF

GF 4630 User manual

GF FlowtraMag GF 2581 User manual

GF

GF FlowtraMag GF 2581 User manual

GF Signet 2552 User manual

GF

GF Signet 2552 User manual

GF Signet 2552 User manual

GF

GF Signet 2552 User manual

GF 582 User manual

GF

GF 582 User manual

GF Signet 4150 User manual

GF

GF Signet 4150 User manual

GF 3-2630-X User manual

GF

GF 3-2630-X User manual

GF U1000 V2 User manual

GF

GF U1000 V2 User manual

GF Signet 2750-1 User manual

GF

GF Signet 2750-1 User manual

GF Signet 5500 User manual

GF

GF Signet 5500 User manual

GF signet 2551 User manual

GF

GF signet 2551 User manual

GF U1000 V2 User manual

GF

GF U1000 V2 User manual

GF Signet 4150 User manual

GF

GF Signet 4150 User manual

GF Signet 2537 User manual

GF

GF Signet 2537 User manual

GF DSR 100 User manual

GF

GF DSR 100 User manual

GF CMD User manual

GF

GF CMD User manual

Popular Measuring Instrument manuals by other brands

Demtech Pro-Tester Double-Pull Operator's manual

Demtech

Demtech Pro-Tester Double-Pull Operator's manual

Dinel ULM-70 Series instruction manual

Dinel

Dinel ULM-70 Series instruction manual

Agilent Technologies 85052A Repair manual

Agilent Technologies

Agilent Technologies 85052A Repair manual

ATIKA FM 2 Original instructions

ATIKA

ATIKA FM 2 Original instructions

Seneca S203TA manual

Seneca

Seneca S203TA manual

Novatech Bitronics 50 Series user manual

Novatech

Novatech Bitronics 50 Series user manual

Orion OriLink 23441 manual

Orion

Orion OriLink 23441 manual

ANATEL Novus RHT-Air instruction manual

ANATEL

ANATEL Novus RHT-Air instruction manual

Michell Instruments Easidew Online user manual

Michell Instruments

Michell Instruments Easidew Online user manual

Zoom H1essential quick start guide

Zoom

Zoom H1essential quick start guide

Bove Technology B12 VI Installation & user guide

Bove Technology

Bove Technology B12 VI Installation & user guide

Welch Allyn Propaq LT 802LTAN Service manual

Welch Allyn

Welch Allyn Propaq LT 802LTAN Service manual

Keysight x-series User reference guide

Keysight

Keysight x-series User reference guide

Elma Instruments Elma 35 user manual

Elma Instruments

Elma Instruments Elma 35 user manual

MFJ MFJ-266 instruction manual

MFJ

MFJ MFJ-266 instruction manual

Brauniger 10 Series operating instructions

Brauniger

Brauniger 10 Series operating instructions

AR DC7450M1 Operating and service manual

AR

AR DC7450M1 Operating and service manual

Hach FilterTrak 660 instruction sheet

Hach

Hach FilterTrak 660 instruction sheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.