GF Signet 5500 User manual

Pg 1 of 12
‡ SIGNET 5500 Flow Monitor
CAUTION!
Never connect 115 VAC or 230 VAC to rear
power terminals. High voltage AC will damage
instrument and void warranty.
CAUTION!
Remove power to unit before wiring input and output
connections.
Follow instructions carefully to avoid personal injury.
-
+
12-24 V
10 W
-
+
12 - 24 VDC
OR
12 - 24 VAC
5500
Terminals
External power
supply
‡ SIGNET 5500 Flow Monitor ENGLISH
Technical Notes:
To reduce the possibility of noise interference, isolate AC power lines from signal lines.
Maximum 4-20 mA loop impedance (sec. 6) is affected by the supply voltage.
1. Power Connections 2. Compatib e Sensor Wiring
Red
Black
Shld.
2000
2507
2536
2540
2550
Vortex*
Red
Black
Shld.
515
525
2517
+GF+SIGNET Sensors:
Open Collector
Sensor
Std. Sensor
Freq. IN
Sen. Pwr.
Freq. IN
Iso. Gnd
5500
Terminals
= Double Insulated
= DC or AC power
Contents
1. Power Connections
2. Compatible Sensor Wiring
3. Sensor Pulse Output Connections
4. Auxiliary Pulse Output Connections
5. Totalizer Reset Connections
6. 4 - 20 mA Current Output Connections
7. Relay Connections
H (6/01) English
3-5500.090-1
8. Output Functions
9. Menu Functions
10. Parts and Accessories
11. Specifications
12. Quick Reference Menu Parameters
13. Troubleshooting
14. Maintenance
Total
reset
AUX
output
PLS
output
12
Gnd
+
+
-
+
Gnd
Freq. IN
Sen. Pwr.
Freq. IN
Iso. Gnd
Std. Sensor
Patent No. D376,328
Flow
Open Collector
Sensor
12-24
10 W
LISTED
77CJ
U
L
1 2 3 4 6 1 2 3 4
1 2
1 2
Total
reset
AUX
output
PLS
output
12
Gnd
+
+
-
+
Gnd
Freq. IN
Sen. Pwr.
Freq. IN
Iso. Gnd
Std. Sensor
Patent No. D376,328
Flow
Open Collector
Sensor
12-24
10 W
LISTED
77CJ
U
L
1 2 3 4 6 1 2 3 4
1 2
1. 2. 3.
Remove termina b ocks for easy wiring

pg 2 of 12 ‡ SIGNET 5500 Flow Monitor
4.2 Totalizer pulse mode (sec. 8.2)
External counter or accumulator inputs without internal open-
collector excitation voltage and pull-up resistor.
+
Gnd
AUX
output
5500
Terminals External power supply
External power supply
10 kΩ
+28 VDC
max.
-
Counter
or
Accumulator
Gnd
Input
NPN sink input
100Ω
5500
Terminals
+
Gnd
AUX
output
10 kΩ
+28 VDC
max.
-
Counter
or
Accumulator
Gnd
Input
PNP source input
100Ω
TTL inputs: An external 5 VDC power supply and 10 kΩ pull-
up resistor is recommended.
CMOS inputs:
An external 3 to 15 VDC power supply and 10 kΩ pull-up
resistor is recommended.
12 VDC Open-collector inputs:
An external 12 VDC power supply and 10 kΩ pull-up resistor is
recommended.
24 VDC Open-collector inputs:
An external 24 VDC power supply and 10 kΩ pull-up resistor is
recommended.
3.2 Pulse output connection options for instruments without internal open-collector excitation voltage and pull-up resistor shown below.
5500
Terminals
+
Gnd
PLS
output
5500
Terminals
External power supply
10 kΩ +28 VDC
max.
-
Other
instrument
Gnd
InputNPN sink input
+
Gnd
PLS
output
10 kΩ
+28 VDC
max.
-Other
instrument
Gnd
Input
PNP source input
100Ω
100Ω
+
Gnd
AUX
output
5500
Terminals
100Ω
10 kΩ
Signal output
Signal ground
+
3 - 28
VDC
-
External power supply
3. Sensor Pu se Output Connections (sec. 8.1)
3.1 +GF+ SIGNET instrument with internal open-collector
excitation voltage and pull-up resistor.
Technical Notes:
Use 2-conductor shielded twisted-pair cable for output lines up
to 30 m (100 ft.) max.
Pull-up resistor NOT REQUIRED when connected to
+GF+ SIGNET instruments.
+
Gnd
PLS
output
Shld
Red
5500
Terminals
+GF+SIGNET
Instrument
Iso. Gnd
Shld
Freq. IN*
Red
*(Open-collector sensor
terminals)
100Ω
4. Auxi iary Pu se Output Connections
4.1 Pulse mode (sec. 8.2) - External device control
The auxiliary output can be configured for either TTL, CMOS,
or open-collector compatible inputs using the proper pull-up
resistor and an external power supply.
Technical Notes:
Open-collector transistor, optically isolated, 5 mA max. sink,
28 VDC max. pull-up voltage.
Programmable pulse width, 0.10 - 999. seconds (sec. 9.2 M)
Selectable Active high or active low pulse logic (sec. 9.3 J).
Use 2-conductor shielded twisted pair cable for AUX output
lines to 30 m (100 ft) max.

Pg 3 of 12
‡ SIGNET 5500 Flow Monitor
+
-
-
Device
A
External AC/DC
power
+-
+
Device
B
Relay 1Relay 2
NC
C
NO NC
C
NO
5500
Terminals
NC
C
NO NC
C
NO
Relay
De-energized Relay
Energized
7. Relay Connections
Two internal relay contact sets (COM, NO, and NC) may be used for
external device control. Front panel LED annunciators indicate the
activation status of each relay. Each relay can control up to two
devices simultaneously, as shown. Relay operation modes include Low
alarm, High alarm, Pulse, and Totalizer (sec. 8.5).
Common device connections include:
• Pulse mode - metering pump control
• Pulse mode - solenoid valve control
• Low or High mode - warning lamps
• Low or High mode - bells or sirens
• Low or High mode - external heavy-duty relay
Wiring Example Right
Device A IS powered when relay 2 is de-energized (front panel LED
"off"). Power is discontinued when the relay 2 setpoint is reached (front
panel LED "on"). Device B IS NOT powered when the relay 2 is
de-energized. Power is applied after the relay 2 setpoint is reached.
6. 4-20 mA Current Output Connections (sec. 8.4)
+-
4-20
mA 5500
Terminals
-
+420
mA
*
Fuse**
1/8A
Technical Notes:
**1/8A fuse recommended (customer supplied)
* 4-20 mA output is internally powered (non-isolated), maximum loop
impedance 350 Ωwith a 12 V instrument supply voltage, 950 Ω
with a 24 V instrument supply voltage.
To isolate output and prevent ground loop problems:
1. Use monitor device with isolated inputs, or
2. Use separate DC supply for 5500 and monitor device, or
3. Power 5500 with12 - 24 VAC step down transformer
12
Total
reset 5500
Terminals
Momentary pushbutton switch
5. Totalizer Reset Connections (sec. 8.3)
Technical Notes:
•Use 2-conductor shielded twisted pair cable for reset lines up to
30 m (100 ft) max. External reset switch overrides software security
feature (section 9.3 I).
•To reduce the possibility of AC noise interference, isolate AC power
lines from reset lines.
Technical Notes:
•Maximum relay contact ratings: 5 A @ 30 VDC, 5 A @ 125 VAC,
or 3 A @ 250 VAC
•An external heavy-duty relay must be used for devices with surge
currents or operating currents that exceed the above specifications.

pg 4 of 12 ‡ SIGNET 5500 Flow Monitor
A. LOW alarm mode
The output is active (relay energized) when the flow drops below the
setpoint, and is inactive (relay de-energized) when the flow rises above
the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K).
•Auxiliary output: ideal for external device
inputs (sec. 4)
•Relay outputs: ideal for external warning
lamps, sirens, etc. (sec. 7)
B. HIGH alarm mode
The output is active (relay energized) when the flow rises above the
setpoint and is inactive (relay de-engergized) when the flow falls below
the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K).
•Auxiliary output: ideal for external device
inputs (sec. 4)
•Relay outputs: ideal for external warning
lamps, sirens, etc. (sec. 7)
Flow
Flow
C. PULSE mode
The output is active (relay energized) for the programmed pulse width each time the programmed pulse volume is measured by the flow sensor
(sec 9.2 H-I, 9.2 L-M).
•Auxiliary output: ideal for external device control (sec. 4)
•Relay outputs: ideal for metering pump or solenoid valve control (sec. 7)
FLOW
volume 1
= Programmed pulse
volume (e.g. 12.5 Gal.)
= Programmed pulse
width (relay energize time)
= Output active
(relay energized)
= Output inactive
(relay de-energized)
volume
volume 2volume 3volume 4volume 5volume 6
Flow
Sensor
Low setpoint=
Adjustable hysteresis=
Output active (relay energized)=
Output inactive (relay de-energized)=
High setpoint=
Adjustable hysteresis=
Output active (relay energized)=
Output inactive (relay de-energized)=
Totalizer function
The output emits an adjustable signal pulse width for each whole engineeing unit measured by the flow sensor (also displayed on totalizer).
To enable the totalizer function, the operator must set the relay or auxiliary pulse volume setpoint to 1.0000 (sec. 9.2 H, 9.2 L).
•Totalizer function: ideal for external counter or accumulator use (sec 4.2)
=
adjustable
pulse width
1234
VIEW menu
totalizer display
8.3 Totalizer reset (sec. 5): This feature allows the operator to reset the totalizer with a dry switch contact closure at any time. The external reset
switch overrides software security feature (sec. 5).
8.4 4-20 mA output (sec. 6): The 4-20 mA output is internally powered by the instrument. 4 mA and 20 mA setpoints are fully adjustable within
the programmed flow range (sec. 11 B). The output can also be reverse spanned.
8.5 Relay 1 and 2 outputs (sec 7): Relay outputs 1 and 2 can be independently configured as Low alarm, High alarm, Pulse or Totalizer output.
Refer to diagrams A - C above for operation details.
8. Output Functions
This section gives a detailed description of all 5500 output functions.
8.1 Sensor pulse output (sec. 3): This output emits an open-collector frequency equal to the sensor input. The output connects directly to a second
+GF+SIGNET instrument for expanded process control with a single sensor input (sec. 3.1). The output is an open-collector type that requires an
external pull up resistor and power supply for other instrument inputs (sec. 3.2).
8.2 Auxiliary pulse output (sec. 4): The auxiliary output can be configured as Low alarm, High alarm, Pulse or Totalizer output. Output logic can
be selected for either active low (factory default) or active high operation (sec. 9.3 J). The auxiliary output is an open-collector type that requires an
external pull-up resistor and power supply to operate. Refer to diagrams A - C below for operation details.

Pg 5 of 12
‡ SIGNET 5500 Flow Monitor
9.1 VIEW Menu
Flow & resettable total
Permanent total
Last calibration
Loop output
Press keys in sequence,
A.
B. B.
C.
A.
OPTIONS = OPTIONS =
To reset
total press:
ENTER
ENTER
To reset
total press:
Range
(display shows: )
OR
Relay 1
Relay 2
Aux output
quick
press
To exit without changes:
(sec. 9.3I) (sec. 9.3I)
A.
B.
C.
D.
E.
F.
G.
H.
Choose: Change: Save:
H↓
A↑
123
Menus:
• VIEW menu (sec. 9.1): The VIEW menu is displayed during
standard operation. The operator can navigate freely through the
menu by pressing either UP or DOWN arrow keys. The VIEW
menu also offers a resettable totalizer feature that allows the
operator to reset the displayed total at the press of a button.
• CALIBRATE Menu (sec. 9.2): The CALIBRATE menu contains all
critical display setup and output parameters. A simple security code
feature prevents unauthorized tampering. The operator is required
to enter a simple access code for menu access. The same code
also unlocks OPTIONS menus.
• OPTIONS Menu (sec. 9.3): The OPTIONS menu contains setup and
display features that are seldom accessed for minor display or
output adjustments.
OPTIONS
(sec. 9.3)
CALIBRATE
(sec. 9.2)
VIEW
(sec. 9.1)
2
seconds 5
seconds
Press & hold ENTER key
for menu
access:
ENTER
Menu Displays A - H:
(Factory default displays shown in menu column 1)
A. Flowrate and resettable total display: can be reset by
the front panel keypad or by remote reset switch (sec.
5). A simple access code feature prevents accidental
total reset, when enabled (sec. 9.3 I). External reset
switch inputs override security feature, when enabled.
B. Permanent total display: records volumetric flow total
for the life of the product and is non-resettable.
C. Loop output display: shows the current output level at
the rear terminals.
D. Flow Range display: shows the programmed min and
max meter range (sec. 9.2 B)
E. Relay 1 display: this display shows the programmed
operation mode and setpoint for relay 1 (sec. 9.2 F-I)
F. Relay 2 display: this display shows the programmed
operation mode and setpoint for relay 2 (sec. 9.2 F-I).
G. Aux output display: this display shows the
programmed operation mode and setpoint for the
auxiliary output (sec. 9.2 J-M).
H. Last calibration: this display shows a user defined
setup date for maintenance records. This feature is not
an internal timer or calender.
9. Menu Functions
To access either CALIBRATE or OPTIONS menus, press and hold the
ENTER key as illustrated below:

pg 6 of 12 ‡ SIGNET 5500 Flow Monitor
Menu Settings A - N:
(Factory default displays shown in menu column 1)
A. Sets flow units label (gpm) and timebase (gpm).
Flow units label: Aa - Zz, 0 - 9, /
Timebase options: h=hours, m=minutes, s=seconds,
d=days (timebase entry affects flow rate display and
4 to 20 mA output)
B. Sets Min→Max meter and dial range, 00.000 to
99999. (does not affect 4 to 20 mA output)
C. Sets flow K-Factor: 0.0001 to 99999. (see technical
notes below)
D. Sets 8-digit totalizer units label: Aa - Zz, 0 - 9, /
(does not affect totalizer display or outputs)
E. Sets totalizer K-Factor: 0.0001 to 99999. (see
techical notes below)
Menu items F - I repeat for relay 2 setup.
F. Sets relay operation mode Low or High, and flow rate
setpoint, 0.0000 - 99999. (sec. 8.5)
G. Sets relay hysteresis, 0.0000 - 99999. (sec. 8.5)
(Set to zero to disable hysteresis feature)
H. Sets relay pulse volume setpoint, 0.0001 - 99999.
(sec. 8.5). Set to 1.0000 to enable totalizer
function (sec. 8.5).
I. Sets relay pulse width, 0.10 - 999. seconds
(sec. 8.5)
J. Sets auxiliary output mode Low or High, and flow rate
setpoint, 0.0000 - 99999. (sec. 8.2).
K. Sets auxiliary output hysteresis, 0.0000 - 99999.
(Set to zero to disable hysteresis feature)
L. Sets auxiliary output pulse volume setpoint,
0.0001 - 99999. (sec. 8.2). Set to 1.0000 (factory
default) to enable totalizer function (sec. 8.2).
M. Sets auxiliary output pulse width, 0.10 - 999. seconds
N. Sets user defined setup date for maintenance records.
This feature is not an internal timer or calender
Technical Notes:
The flow and total K-Factors are independent of each other.
The K-Factor settings represent the number of pulses
generated by the +GF+SIGNET flow sensor for each
engineering unit of fluid measured (published in flow sensor
manual).
K-Factor
Flow K-Factor
Last calibration
quick
press
To return to
VIEW menu:
quick
press
To restore original value:
Flow units Unit + timebase: h,m,s,d
Flow range
Setup date
Meter Min→Max
- Display label only -
K-Factor
Total units
Total K-Factor
Mode & setpoint
Press keys in sequence
will appear during code entry.
to enter menu,
Mode & setpoint
Relay
Mode & setpoint
Aux output
Mode & setpoint
Aux output
Pulse width
Aux output
Hysteresis
Aux output
Relay
Hysteresis setting
Relay ENTER
Relay Low or High Mode Selected
Aux Output Low or High Mode Selected
Aux Output Pulse mode selected, step J
Relay Pulse Mode Selected, step F
A.
B.
C.
D.
E.
F.
G.
H.
Pulse width
Relay
I.
J.
K.
L.
M.
N.
"SAVING"
briefly
displays
123
Choose: Change: Save:
N↓
A↑
9.2 CALIBRATE Menu

Pg 7 of 12
‡ SIGNET 5500 Flow Monitor
9.3 OPTIONS Menu
Choose:
1Change:
2Save:
3
quick
press
To return to
VIEW menu:
quick
press
To restore original value:
Display contrast Low -- -- -- -- -- High
4 mA adjust
Loop output
Loop output 20 mA adjust
Flow display Decimal
Press keys in sequence
will appear during code entry.
to enter menu,
Total display Total decimal
Display averaging 4s 8s0s
Loop output 4 mA setpoint
20 mA setpoint
Loop output
J↓
A↑
ENTER
"SAVING"
briefly
displays
Totalalizer reset VIEW menu=
VIEW menu=
A.
B.
C.
D.
E.
F.
G.
H.
I.
J.
Auxiliary out logic Active Low or High
Menu Settings A - J:
(Factory default displays shown in menu column 1)
A. Selects LCD display contrast: 5 levels
B. Selects flow display decimal:
*****.to ** . ***
C. Selects totalizer display decimal:
********.to ****** . **
D. Selects LCD display averaging:
Off = 0 seconds, Low= 4 seconds, High= 8 seconds
(also effects 4 - 20 mA output)
E. Sets 4 mA output setpoint. 4 mA and 20 mA setpoints
are reversible.
F. Sets 20 mA output setpoint. 20 mA and 4 mA
setpoints are reversible.
G. Sets 4 mA current output: 3.0 to 5.0 mA
(overrides 4.00 mA factory calibration)
H. Sets 20 mA current output: 19 to 21 mA
(overrides 20.00 mA factory calibration)
I. Selects totalizer reset options: Lock "Off" disables the
VIEW menu access code feature; Lock "On" enables
the feature (sec. 9.1 A). When enabled, the operator
must enter an access code for totalizer reset privileges.
The totalizer will be reset in either case if a contact
closure is detected at rear reset terminals (sec. 5).
J. Selects Auxiliary output logic:
•Active Low logic (factory default): Output changes from
high to low when active, see diagram below.
•Active high logic: Output changes from low to high
when active, see diagram below.
Inactive (Off)
Active Low Logic (factory default)
Active (On)
Inactive (Off)
Active High Logic
Active (On)

pg 8 of 12 ‡ SIGNET 5500 Flow Monitor
Electrical
Power:
•12 to 24 VDC or 12 to 24 VAC, unregulated, 50-60 Hz,
10 W max.
Relay contacts (2 sets):
•Mechanical SPDT contacts
•Maximum voltage rating: 5 A @ 30 VDC, 5 A @ 125 VAC, or
3 A @ 250 VAC, (power factor = 1.0)
•Hysteresis: User adjustable
Current output:
•4 to 20 mA, non-isolated, internally powered, fully adjustable and
reversible
•Update rate: < 200 ms
•Max loop impedance: 350 Ωmax with a 12 V instrument supply
voltage, 950 Ωmax with a 24 V instrument supply voltage
•Accuracy: ±0.1% of max range
Sensor (PLS) Auxiliary (AUX) pulse outputs:
•Open-collector transistor, optically isolated, 5 mA max. sink,
28 VDC max. pull-up voltage, programmable (AUX) pulse width
Totalizer reset:
•Front keypad or external contact closure, 30 m (100 ft) max. cable
length
•External contact closure overrides security feature (sec. 5)
Noise immunity: EN50082-2
Noise emissions: EN55011
Safety: EN61010-1
11. Specifications
General
Sensor compatibility: All +GF+SIGNET flow sensors
Accuracy: ±0.5% of reading
Input: Optically isolated
Enclosure:
•Rating: NEMA 4X/IP65 front
•Dimensions: 1/4 DIN, 96 x 96 x 88 mm (3.8 x 3.8 x 3.5 in.)
•Case materials: ABS plastic
•Keypad material: Sealed 4-key silicone rubber
•Weight: Approximately 500 g (18 oz.)
Display:
•Type: Microprocessor controlled air-core meter movement and
backlit alphanumeric 2 x 16 LCD
•Update rate: Flow = 1s, totalizer = <200 ms
•Contrast: User selected, 5 levels
•Relay annunciators: 2 LEDs
Totalizers:
•8-digit resettable with security option
•8-digit non-resettable
Environmental
Operating temp.: -10 to 55 °C (14 to 131 °F),
50 °C (122 °F) max. with optional rear
cover
Storage temp.: -15 to 80 °C (5 to 176 °F)
Relative humidity: 0 to 95%, non-condensing
Altitude: 4000 m max.
Pollution degree: 2
Agency Approvals
•CSA, CE, UL listed
•Manufactured under ISO 9001
•Power supply, 115 VAC - 24 VAC, #3-5000.075
•Front snap-on bezel, #3-5000.525 (code 198 840 226)
•Assorted flow unit/multiplier decals, #3-5500.611 (code 198 840 230)
•5500 instruction sheet #3-5500.090-1 (code 198 869 914)
10. Parts and Accessories
Optional surface mount
bracket #3-5000.598
(code 198 840 225)
Splashproof rear cover
#3-5000.395
(code 198 840 227)
5 x 5 inch adapter plate
for +GF+SIGNET retrofit
#3-5000.399
(code 198 840 224)
Internal
PLS/AUX
circuits
+
GND
5500 rear PLS/AUX
output terminals
100Ω

Pg 9 of 12
‡ SIGNET 5500 Flow Monitor
96 mm
(3.8 in.)
Front
View
96 mm (3.8 in.)
RELAY 2
RELAY 1
ENTER
4
0
12356789
10
GPM
+GF+SIGNET
Dimensions
SIDE VIEW
Mounting Clamp (2 ea.)
Optional Splashproof Rear Cover
88 mm
(3.5 in.)
Mounting Panel
Panel
Gasket
91 mm
(3.6 in.)
76 mm (3 in.)
88 mm (3.5 in.)
96 mm (3.8 in.)
VIEW Menu Setup Parameters (sec. 9.1)
92 x 92 mm
(3.62 x 3.62 in.)
Panel
cutout
88 mm
(3.5 in.)
88 mm (3.5 in.)
4-20
mA
Total
reset
AUX
output
PLS
output
12
Gnd
-
+
+
+
NC
C
NO NC
C
NO
-
+
Gnd
Freq. IN
Sen. Pwr.
Freq. IN
Iso. Gnd
Std. Sensor
Relay 1Relay 2
Patent No. D376,328
Flow
Open Collector
Sensor
12-24 V
10 W
+GF+SIGNET
Model #3-5500
®
LISTED
77CJ
UL
®
Rear
View
Menu
Parameters Factory
Default
Range
Display
Description
•Flow rate and units
•Resettable total
Permanent
total
Current loop
output
Min →Max meter
and dial range
•Relay 1 mode
•Relay 1 setpoint
•Relay 2 mode
•Relay 2 setpoint
•Aux output mode
•Aux output setpoint
Last calibration
date
00.000 - 99999. GPM
000000.00 - 99999999.
000000.00 - 99999999.
3 - 21 mA
00.000 - 99999.
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
00 - 00 - 00 to
39 - 39 - 99
0000.0
0000000.0
0000000.0
n/a
0.000 - 100.00
Low
10.000 GPM
High
90.000 GPM
Pulse
1.0000 Gallons
01 - 01 - 98
A.
B.
C.
D.
E.
F.
G.
H.
12. Quick Reference Menu Parameters

pg 10 of 12 ‡ SIGNET 5500 Flow Monitor
Menu
Parameters Factory
Default
Range
Display
Description
flow unit= digits 1-3
flow timebase= digits 4
Min →max
range
Flow K-Factor
Total units
Total K-Factor
•Relay 1 mode
•Relay 1 setpoint
•Relay 1 Hysteresis
•Hysteresis volume
•Relay 1 mode
•Relay 1 pulse volume
Relay 1 pulse width
•Aux output mode
•Aux output setpoint
Aux output hysteresis
volume
•Aux output mode
•Aux output setpoint
Aux output pulse
width
Last calibration
date
Aa - Zz, 0 - 9, /, _;
Mm, Ss, Hh, Dd
00.000 - 99999.
0.0001 - 99999.
Aa - Zz, 0 - 9, /, _
(8-digits max.)
0.0001 - 99999.
Low or High
0.0000 - 99999.
0.0000 - 99999.
Pulse
0.0001 - 99999.
0.10 - 999. seconds
Low or High
0.0000 - 99999.
0.0000 - 99999.
Pulse
0.0001 - 99999.
0.10 - 999. seconds
00 - 00 - 00 to
39 - 39 - 99
_GPM
00.000 - 100.00
GPM
60.000
_Gallons
60.000
Low
10.000 GPM
2.0000
GPM
10.000
Gallons
10.0
seconds
Low
1.0000
0.0000
1.000
Gallons
0.10
seconds
01 - 01 - 98
A.
B.
C.
D.
E.
F.
G.
H.
I.
J.
K.
L.
M.
N.
OPTIONS Menu Setup Parameters (sec. 9.3)
Relay mode
and setpoint
displays repeat
for relay 2
setup
CALIBRATE Menu Setup Parameters (sec. 9.2)
Menu
Parameters Factory
Default
Range
Display
Description
Display contrast
Flow decimal
Totalizer decimal
Display averaging
4 mA setpoint
0 to 5
**.*** to *****.
******.** to ********.
Off= 0 sec., Low= 4sec.,
High= 8 sec.
Unlimited
3
****.*
*******.*
Off
0.0000 GPM
A.
B.
C.
D.
E.
_ = blank
display digit
option
_ = blank
display digit
option

Pg 11 of 12
‡ SIGNET 5500 Flow Monitor
13. Troubleshooting
OPTIONS Menu Setup Parameters Continued...
14. Maintenance
Clean the instrument case and front panel with a soft cloth and mild liquid soap solution.
Menu
Parameters Factory
Default
Range
Display
Description
20 mA setpoint
4 mA adjust
20 mA adjust
VIEW menu total
reset access code
Auxiliary (AUX) output
pulse logic
Unlimited
3.0 to 5.0 mA
19 to 21 mA
On or Off
Active Low or
Active High
100.00 GPM
4.00 mA
20.00 mA
Off
Low
F.
G.
H.
I.
J.
Display Problem Solution
-- -- -- -- Display timebase too large Change flow timebase (S=Seconds, M=Minutes, H=Hours, D=Days) in
CALIBRATE menu to a smaller value (e.g. GPDto GPM)
Check settings Relay 1 pulse width too large Reduce Relay 1 pulse width setting (sec. 9.2 I) or increase pulse volume
for Relay 1 for frequency input or pulse setting (sec. 9.2 H)
volume too small
Check settings Relay 2 pulse width too large Reduce Relay 2 pulse width setting (sec. 9.2 I) or increase pulse volume
for Relay 2 for frequency input or pulse setting (sec. 9.2 H)
volume too small
Check settings Auxiliary output pulse width too Reduce auxiliary output pulse width setting (sec. 9.2 M) or increase pulse
for AUX output large for frequency input or pulse volume setting (sec. 9.2 L).
volume too small
Must be between Relay 1, Relay 2, or Auxiliary output Enter volumetric setpoint within 0.0001 - 99999. engineering units
0.0001 - 99999 (AUX) pulse volume setpoint must be (sec. 9.2 H or 9.2 L)
within 0.0001 - 99999.
Must be between Relay 1, Relay 2, or Auxiliary output Enter pulse width within 0.10 - 999 seconds (sec. 9.2 I or 9.2 M)
0.10 - 999 (AUX) pulse width must be within
0.10 - 999 seconds
Flow K-Factor Flow display K-Factor cannot be Enter K-Factor greater than zero (sec. 9.2 C)
can not be zero zero
Total K-Factor Totalizer K-Factor cannot be Enter K-Factor greater than zero (sec. 9.2 E)
can not be zero zero
SETUP READ ERROR Power fault occurred while saving Press any key to reload factory defaults then reprogram flow system setup
Press any Key setup menu entry parameters. Note: totalizer displays remain undistrubed and will
resume after instrument reset.

pg 12 of 12 ‡ SIGNET 5500 Flow Monitor
‡ SIGNET
Signet Scientific Company, 3401 Aerojet Avenue, E Monte, CA 91731-2882 U.S.A. • Te . (626) 571-2770 • Fax (626) 573-2057
For Wor dwide Sa es and Service, visit our website: www.gfsignet.com • Or ca (in the U.S.): (800) 854-4090
GEORGE FISCHER ‡ Piping Systems
3-5500.090-1 (H-6/01) English © Signet Scientific Company 2001 Printed in U.S.A. on recycled paper
Other manuals for Signet 5500
1
Table of contents
Other GF Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

SATAM
SATAM 7887 Series Description, Installation, Operations and Maintenance Manual

SIGEYI
SIGEYI AXO PM01 user guide

Keysight
Keysight U8903B user guide

Fieldpiece
Fieldpiece EH4W Operator's manual

Jain Technology
Jain Technology XONIC-100P user manual

Cyber Sciences
Cyber Sciences CyTime SER-32e Instruction bulletin