Contraindications:
These devices should never be used when:
• There is visible evidence of damage to the exterior of the device such as cracked plastic
or connector damage.
• These devices fail the inspection described herein.
• In the presence of flammable gases, liquids, and/or oxygen enriched environments.
Safety Tips:
• Use lowest possible power setting on electrosurgical unit capable of achieving desired
surgical effect.
• Never allow the cables connected to these devices to be in contact with skin of
the patient or operator during electrosurgical activations.
• Do not permit the cables connected to these devices to be parallel and in close
proximity to the leads of other electrical devices.
• Always place unused electrosurgical accessories in a safe insulated location such as
a holster when not in use.
• Inspect and test each device before each use.
• Discard devices that have reached their life expectancy.
• Activation of an electrosurgical device when not in contact with target tissue or in
position to deliver energy to target tissue (fulguration) may cause capacitive coupling.
Inspection:
These devices should be inspected before and after each use. Visually examine the devices
for obvious physical damage including:
• Cracked, broken or otherwise distorted plastic parts.
• Broken or significantly bent connector contacts.
• Damage including cuts, punctures, nicks, abrasion, unusual lumps, significant discoloration.
• Tips for damage, corrosion or misalignment condition.
Storage & Handling:
These devices must be stored in a clean, cool and dry area. Protect from mechanical damage
and direct sunlight. Handle with extreme care.
GIMA WARRANTY CONDITIONS
Congratulations for purchasing a GIMAproduct. This product meets high qualitative standards
both as regards the material and the production. The warranty is valid for 12 months from the
date of supply of GIMA. During the period of validity of the warranty, GIMA will repair and/or
replace free of charge all the defected parts due to production reasons. Labor costs and
personnel traveling expenses and packaging not included. All components subject to wear are
not included in the warranty. The repair or replacement performed during the warranty period
shall not extend the warranty. The warranty is void in the following cases: repairs performed
by unauthorized personnel or with non-original spare parts, defects caused by negligence or
incorrect use. The warranty is void if the above regulations are not observed and if the serial
code (if available) has been removed, cancelled or changed. The defected products must be
returned only to the dealer the product was purchased from. Products sent to GIMA will be