manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hioki
  6. •
  7. Test Equipment
  8. •
  9. Hioki 3281 User manual

Hioki 3281 User manual

3281, 3282
DIGITAL CLAMP ON HiTESTER
Instruction Manual
Sept. 2015 Revised edition 19
Printed in Japan
3281A981-19 15-09H
EN
Introduction
Inspection
Safety
DANGER Indicates that incorrect operation presents an extreme hazard that
could result in serious injury or death to the user.
WARNING Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.
CAUTION Indicates that incorrect operation presents a possibility of injury to the
user or damage to the instrument.
NOTE Advisory items related to performance or correct operation of the
instrument.
The symbol printed on the instrument indicates that the user should refer
to a corresponding topic in the manual (marked with the symbol) before
using the relevant function.
In the manual, the symbol indicates particularly important information that
the user should read before using the instrument.
Indicates that dangerous voltage may be present at this terminal
Indicates a double-insulated device.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates a grounding terminal.
Indicates that the instrument may be connected to or disconnected from a live
circuit.
Precautions
DANGER
This instrument is designed to conform to IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. However,
mishandling during use could result in injury or death, as well as damage
to the instrument. Be certain that you understand the instructions and
precautions in the manual before use. We disclaim any responsibility for
accidents or injuries not resulting directly from instrument defects.
Do not use on the voltage
lines exceeding 600 Vrms. Do not use on the primary
side of the breaker.
Do not input voltage in the
resistance measurement,
continuity checking and tem-
perature measurement.
WARNING
To prevent electric shock, when measuring the voltage of a power line use
a test lead that satisfies the following criteria:
Conforms to safety standards IEC61010 or EN61010
Of measurement category III or IV
Its rated voltage is higher than the voltage to be measured
The test leads provided with this instrument conform to the safety standard
EN61010.
Use a test lead in accordance with its defined measurement category and
rated voltage.
WARNING
Do not input voltages ex-
ceeding 600 Vrms. (1000 V
max.)
During current measure-
ment, do not connect the
test leads or temperature
probe to the instrument.
Avoid touching the exposed
metallic parts of the jaw
while measuring voltage.
Do not use when your
hands are wet.
Be sure to insert the battery
with the polarity correct.
Battery
Do not use the unit with the
back casing removed. Do not short-circuit, re-
charge, disassemble or in-
cinerate batteries.
Short-circuiting
Disassembling
Handle and dispose of batteries in accordance with local regulations.
To avoid electric shock when measuring live lines, wear appropriate
protective gear, such as insulated rubber gloves, boots and a safety helmet.
Before using the instrument, make sure that the insulation on the test
leads is undamaged and that no bare conductors are improperly
exposed. Using the instrument in such conditions could cause an
electric shock. Replace the test leads and probes with the specified
Hioki Model L9207-10.
CAUTION
Do not use or store the in-
strument where it is exposed
to direct sunlight, high tem-
peratures, high humidity, or
condensation.
Do not subject the instru-
ment to vibrations or
shocks.
Do not drop the instrument.
Before using the instrument the first time, verify that it operates normally
to ensure that the no damage occurred during storage or shipping. If
you find any damage, contact your dealer or Hioki representative.
Removable sleeves are attached to the metal pins at the ends of the test
leads. 
To prevent a short circuit accident, be sure to use the test leads with the
sleeves attached when performing measurements in the CAT III and
CAT IV measurement categories . In the CATII environment, if the tips
of the test leads do not reach the measurement object, remove the rigid
insulating sleeve before measuring. For details on measurement
categories, see "Measurement categories" in the instruction manual.
When performing measurements with the sleeves attached, be careful to
avoid damaging the sleeves. If the sleeves are inadvertently removed
during measurement, be especially careful in handling the test leads to
avoid electric shock.
To prevent an electric shock accident, confirm that the white or red
portion (insulation layer) inside the cable is not exposed. If a color inside
the cable is exposed, do not use the cable.
Specification
Function Mode Range Accuracy ( %rdg. dgt.) Muximum
permissible
input
AC current
(A)
RMS
(Effective
value)
30.00 40 to 1 kHz:
1.0%rdg. 0.7%f.s. 3281:
600 AAC
continuous
1000 A max.
3282:
600 AAC
continuous
1000 AAC (5
minutes)
1700 A max.
300.0 45 to 66 Hz: 1.0% 5
40 to 45, 66 to 1 kHz: 1.5% 5
600(1000)
Auto-ranging As per the above range
PEAK
(Peak
value)
30.0 40 to 1 kHz: 5% 5
300 40 to 1 kHz: 3% 5
600 (1000) 40 to 1 kHz: 3% 5
Auto-ranging As per the above range
AC voltage
(V) RMS 300.0/600 45 to 66 Hz: 1.0% 3
40 to 45, 66 to 1 kHz: 1.5% 3600 VAC
continuous
1000 V max.
Auto-ranging
PEAK 300/600 40 to 1 kHz: 3% 5
Crest factor 1.00 to 5.00 10% 5 See the currents
and voltages
above
Frequency (Hz) Auto-ranging
(100.0/1000) 30 to 99.9 Hz: 0.3% 1
95 to 1000 Hz: 1% 1
Resistance (Ω)Auto-ranging
(1000/10.00k) 10 to 10.00 kΩ:1.5% 5Open terminal
voltage:
3 VDC max.
Overload
protection:
600 Vrms
Continuity 1000 ΩBuzzer at approx. 30 Ωor less
Diameter of
measurable conductor3281: 33 mm dia. max. (1.3"), 3282: 46 mm dia. max. (1.8")
Effect of conductor
position At any position based on the center of the jaw
3281: Within 4.0%, 3282: Within 1.0%
Effect of external
magnetic field In an external magnetic field of 400 AAC/m
3281: 1.5 A max., 3282: 0.2 A max.
Functions Record (displays the maximum (MAX), minimum (MIN) and
average (AVE) values in the AC current, AC voltage and
frequency measurements), data hold (holds the display), auto-
power off (approx. 10 minutes, the buzzer alarms just before
the instrument is powered off, can be extended and released),
buzzer (can be turned on or off)
Display LCD, digital (3000 counts), bar graph (35 segments)
Over range display: "O.L."or" " (bar graph input over)
Battery consumption warning: " " (When this mark is lighting,
the accuracy is not guaranteed.)
Data hold display: "HOLD"
Auto power-off display: "APS"
Units (A, V, Hz, Ω,kΩ,*,*)
Zero suppressor: 5 counts max.
*:Temperature probes have been discontinued.
The temperature measurement function is no longer available.
Display update rate Digital display: Approx. twice per second,
SLOW: Approx. once per 3 seconds,
FAST: Approx. 4 times per second
Bar graph display: approx. 4 times per second (fixed)
Response time Current, voltage, frequency: Approx. 2.2 seconds
Resistance, continuity check: Approx. 1.1 seconds
Range selection Auto-ranging/manual ranging (fixed range) selectable (excluding
the frequency, resistance and continuity check)
Circuit dynamic
(Crest factor) 2.5 max. (600 A (3281), 1000 A (3282), 600 V range: 1.7)
Dielectric strength 3281
Between the case and input: AC 8540 V rms /1 minute
Between the case and jaw: AC 5312 V rms /15 sec
3282
Between the case and input terminals: AC 8540 V rms /1 minute
Between the case and jaw: AC 8540 V rms /1 minute
Location for use Altitude up to 2000 m (6562 feet), Indoors
Standards
applying Safety
EMC
EN 61010
3281 (current): 600 VAC (Measurement Category III)
Anticipated transient overvoltage: 6000 V, Pollution Degree 2
3281 (voltage): 600 VAC (Measurement Category IV)
Anticipated transient overvoltage: 8000 V, Pollution Degree 2
3282 (current): 600 VAC (Measurement Category IV)
Anticipated transient overvoltage: 8000 V, Pollution Degree 2
3282 (voltage): 600 VAC (Measurement Category IV)
Anticipated transient overvoltage: 8000 V, Pollution Degree 2
EN 61326
Dust resistance EN 60529 IP40
Operating
temperature and
humidity range
0to40 (32 to 104 ), 80% RH max. (no condensation)
Temperature
characteristics 0.05 x accuracy specifications/ ()at0to40 (32 to 104 )
Warranty
Warranty malfunctions occurring under conditions of normal use in conformity
with the Instruction Manual and Product Precautionary Markings will be
repaired free of charge. This warranty is valid for a period of one (1) year
from the date of purchase. Please contact the distributor from which you
purchased the product for further information on warranty provisions.
Thank you for purchasing the HIOKI "HIOKI 3281, 3282 Digital Clamp-on
HiTester". To obtain maximum performance from the instrument, please read
this manual first, and keep it handy for future reference.
When you receive the instrument, inspect it carefully to ensure that no
damage occurred during shipping. In particular, check the accessories, panel
switches, and connectors. If damage is evident, or if it fails to operate
according to the specifications, contact your dealer or Hioki representative.
This manual contains information and warnings essential for safe operation of
the instrument and for maintaining it in safe operating condition. Before
using the instrument, be sure to carefully read the following safety notes.
The following symbols in this manual indicate the relative importance of
cautions and warnings.
Safety Symbols
Measurement categories
This instrument conforms to the safety requirements for CAT III(3281), CAT
IV(3282) measurement instruments. To ensure safe operation of
measurement instruments, IEC 61010 establishes safety standards for
various electrical environments, categorized as CAT II to CAT IV, and called
measurement categories. These are defined as follows.
CAT II: Primary electrical circuits in equipment connected to an AC electrical
outlet by a power cord (portable tools, household appliances, etc.)
CAT II covers directly measuring electrical outlet receptacles.
CAT III:Primary electrical circuits of heavy equipment (fixed installations)
connected directly to the distribution panel, and feeders from the
distribution panel to outlets.
CAT IV: The circuit from the service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution
panel).
Using a measurement instrument in an environment designated with a
higher-numbered category than that for which the instrument is rated could
result in a severe accident, and must be carefully avoided.
Use of a measurement instrument that is not CAT-rated in CAT II to CAT IV
measurement applications could result in a severe accident, and must be
carefully avoided.
NOTE
Accurate measurement may be impossible in the presence of strong magnetic
fields, such as near transformers and high-current conductors, or in the
presence of strong electromagnetic fields such as near radio transmitters.
The indicator lights up when the remaining battery capacity is low. In
this case, the instrument's reliability is not guaranteed. Replace the battery
immediately.
The 3281 and 3282 are different in the maximum range.(3281: 600 A, 3282: 1000 A)
1. Measurement specification
Temperature and humidity for guaranteed accuracy: 23 5 (73 9 ), 80% RH
or less (This is guaranteed when " " mark is not lighting.)
Guaranteed accuracy period:1 year, or opening and closing of the jaws 10,000 times,
whichever comes first
( ) in the current ranges: 3282
Maximum rated voltage to earth: Max. 600 Vrms
Accuracy is guaranteed for over 10% input of the range in current and voltage.
2. General specifications
Storage temperature
range -10 to 50 (14 to 122 ) (no condensation)
Power source Rated power voltage 9 VDC
6F22 layer-built manganese battery x 1
Maximum rated
power 100 mVA
Battery lifetime Approx. 45 hours (continuous, no load)
External dimensions
and mass Approx. 62W x 216.5H x 39D mm, Approx. 350 g (3281)
Approx. 2.44"W x 8.58"H x 1.54"D, Approx. 12.3 oz. (3281)
Approx. 62W x 231H x 39D mm, Approx. 400 g (3282)
Approx. 2.44"W x 9.06"H x 1.54"D, Approx. 14.1 oz. (3282)
Names and Functions of Parts
Voltagae and resistance
input terminals
3281(600 A)3282(1000 A)
Jaw
Lever
Key switches
Display (LCD)
Back casing
(back)
Alternating current
AUTO
SLO
REC
MAX
MIN
AVE
Auto-ranging
Display update:
approx. once per three seconds
Record function
Maximum value
Minimum value
Average value = (maximum value
+ minimum value) / 2
min
hour
HOLD
APS
*
*
Ω,kΩ
One minute: one segment (bar
graph)
One hour: one segment (bar
graph)
Data hold
Auto power-off
Centigrade
Fahrenheit
Resistance
Continuity
Hz
V
A
RMS
PEAK
C.F.
Frequency
Voltage
Current
True RMS value
Peak value
Crest factor = Peak value /
Effective value
Input over (bar graph)
Battery consumption warning
Measurement Procedure
Preparation
Fresh battery
Battery capacity 0 " " light.
Beep tone sounds 3 timeshe battery
condition.
Measurements taken at this battery level
is not guaranteed for accuracy.
AC current (ACA) measurement A
OK
SLOW PEAK C.F. RMS
Wave
form Peak
S
ine
w
ave Peak
Triangular
wave Peak
Square
wave
C.F. 1.41 1.73 1
Hz RMS
AC voltage measurement V
Black Red
Plug in the test leads
Resistance measurement
Black Red
Plug in the test leads
Continuity check
Data hold function HOLD
Auto power-off function APS
Battery consumption warning
Buzzer
FAST mode
Recording function REC
MAX MIN AVE Instantaneous value
(No display)
Battery Replacement Procedure
WARNING
When replacing the battery, be sure to insert it with the correct polarity.
Otherwise, poor performance or damage from battery leakage could result.
Replace battery only with the specified type.
CAUTION
Do not fix the back casing screws too tightly.
The torque about 0.5N・m is recommended.
Battery holder
How to Attach the Hand Strap
Troubleshooting
Symptom Battery Battery snap Test leads
The instrument cannot be
powered on. -
Power is cut off immediately
after it is turned on.*- -
"" lights. - -
The instrument is powered
off during operation.*-
Voltage measurement does
not function. - -
Resistance measurement
does not function. - -
Remedy:
If the trouble cannot be
remedied, send the
instrument for repair.
Replace with a
new battery. The terminals of
the battery snap
are poorly
contact.
Check the test
leads wiring.
An indication E.001 to E.005 appears. Send the instrument for repair.
Service
3. Accessories
Model L9207-10 Test Lead (black and red set), Instruction manual, Model 9399 Carrying
Case, Hand strap, 6F22 (006P) battery
*: Temperature probes have been discontinued.
The temperature measurement function is no longer available.
1. Remove the rear cover and insert a battery. (Refer to Battery Replacement Procedure.)
2. Press POWER to turn the unit on. Verify that all segments of the isplay light up briefly.
Then the model name is shown, and the bar graph indicates the battery condition.
3. The AC current measurement mode is activated.
Low battery voltage detection function
After the mark lights and battery voltage drops below a certain level, the power goes
off automatically. When this occurs, bAtt and Lo are displayed.
When power goes off after display of these marks, replace the exhausted battery with a
new one.
1. Press the key.
2. Clamp only one of the conductors and
place it in the center of the jaw.
The effective value (RMS) of the current is
displayed in the digital display and bar
graph.
A suitable measurement range is selected
automatically (AUTO).
NOTE
Use data hold function when you abolish indication and want to read it.
Please note that waveforms that include elements outside the frequency characteristic
range may not be measured correctly.
Current measurements exceeding 600 A AC should be of short duration. Heat builds
up in the jaw proportionate to the current value, and will reach a dangerous level over
a long period of time.
The unit cannot read zero with no input at low temperature. Even then, the accuracy is
guaranteed when a current of 3 A or more is measured.
Range selection
Press the RANGE key repeatedly cycles through the 30 A, 300 A, 600 A (1000 A) and
AUTO ranges.
Changing the display update SLOW
When the readings fluctuate and are difficult to take, it is possible to make the display
update slow (approx. once per three seconds), and the readings easy to take. The
screen-updating speed cannot be changed for the bar-graph display.
Pressing the SLOW/PEAK key repeatedly changes the display as follows.
Peak value display PEAK
The peak value is displayed. The effective value is displayed in the bar graph.
NOTE
Mode displaying the PEAK (peak value) of a continuous wave which lasts for more
than 250 ms.
To keep the displayed value, use the recording function in the PEAK display mode
(refer to recording function REC 1.).
As there is a period whereby no sampling is done on this instrument, it may not be
possible to measure an instantaneous peak current that does not reach 250 ms, such
as the motor starting current, even when the recording function is used.
To accurately measure an instantaneous peak current such as an inrush current,
please use HIOKI CM4371 and CM4373.
Crest factor display C.F.
The crest factor (peak-to-rms ratio) of a
waveform is displayed.
Crest factor = Peak value / Effective value
The crest factor of an undistorted sine wave
is 1.41.
A crest factor of other than 1.41 indicates
that a waveform is distorted, i.e., contains
harmonic components.
When a crest factor of current is displayed,
the indicator "A" flashes.
The effective value is displayed in the bar graph.
Frequency display Hz
1. Press the Hz key.
2. Pressing the Hz key changes the display as shown in the figure.
3. The frequency of the current being measured is displayed.
When no input is applied, "----" is displayed. When measuring the current frequency,
"A" flashes. The effective value is displayed in the bar graph.
NOTE
When the frequency is lower than 30 Hz, "----" is displayed.
The AUTO range display indicates the current range.
1. Press the key.
2. The effective value (RMS) of voltage is displayed in
the digital display and bar graph.
The display update changing, and the peak value,
crest factor and frequency displays are possible as
well as in the AC current measurement.
NOTE
Be sure to use the test leads with the sleeves
attached when performing measurements in the CAT
III and CAT IV measurement categories. In the
CATII environment, if the tips of the test leads do not
reach the measurement object, remove the rigid insulating sleeve before measuring.
Please note that waveforms that include elements outside the frequency characteristic
range may not be measured correctly.
1. Insert the test leads in the instrument as shown in
the figure.
2. Attach or remove the rigid insulating sleeve as
required by the measurement object.
3. Press the Ω//TEMP key to display Ω, changes the
display as follows.
Ω* *
(Centigrade) (Fahrenheit)
*: Temperature probes have been discontinued.
The temperature measurement function is no longer available.
4.The resistance value is displayed in the digital display and bar graph. Ranging is
automatic (AUTO).
NOTE
If a voltage is input, a warning beep will sound. Stop measurement immediately. (The
internal circuit is protected against up to AC 600 V.)
In some cases, the alarm does not beep for DC or DC weighted components.
1. Display " " in the same way as in the resistance measurement.
2. The buzzer beeps at less than approximately 30 Ω, and " " flashes.
NOTE
The digital display indicates the measured resistance value.
If a voltage is input, a warning beep will sound. Stop measurement immediately. (The
internal circuit is protected against up to AC 600 V.)
In some cases, the alarm does not beep for DC or DC weighted components.
Data hold functions to "stop" the display at its present reading.
Press the HOLD key. "HOLD" appears, and the digital and bar graph displays are
held. This function is effective for all measurement functions and modes.
To release this function, press the HOLD key again.
When "APS" is being displayed, the auto power-off function is effective.
The unit is powered off in approx. 10 minutes unless any key is pressed.
"APS" flashes and the alarm beeps for approx. 30 seconds just before the unit is
powered off.
Pressing a key other than the POWER key prolongs the auto power-off function for 10
minutes.
To release the auto power-off function, press the POWER key while holding down the
HOLD key to power on the unit. In this case, "APS" does not appear.
When using the record function, the auto power-off function is ineffective.
If is indicated, the battery power is running low and accuracy cannot be guaranteed.
Replace with a new battery. Refer to "Preparation" for the confirmation of the capacity of
the battery.
To turn off the buzzer, press the POWER key while holding down the RANGE key to
power on the instrument. The alarm and continuity buzzers cannot be turned off.
Make it FAST mode when you measure load currents with variations.
The digital display update can be set to approx. 4 times per second.
1. Press the key twice to set to the FAST mode.
"F" appears for an instance, and the unit enters the FAST mode.
Then "F" appears each time the or key is pressed.
2. Press the RANGE key to fix the current range.
3. It is convenient for taking readings to hold the maximum value (MAX) by using the
record function.
4. To release the FAST mode, press the key twice again.
NOTE
The stable measurement cannot be made unless the waveform lasts for more than
250 ms.
Press the key in the case of the voltage measurement as well after it is made FAST
mode.
This mode is not effective for the resistance, continuity and temperature
measurements.
If setting to the SLOW display in the FAST mode, the display update is the same as
in the normal mode (approx. twice per second).
Use the recording function to hold the maximum and minimum measured values and
maximum/minimum averages.
1. Measurement indicated value
Pressing the MAX/MIN key during measurements of current, voltage, or frequency
activates the recording function. REC flashes and the instrument saves the
maximum value (MAX), minimum value (MIN), and average value (AVE) in internal
memory from the instant you press the MAX/MIN key. Pressing the MAX/MIN key
with the recording function activated switches the display as shown below. If MAX,
MIN, or AVE is not displayed, an instantaneous value is displayed.
Data (MAX, MIN, AVE) remains saved while the display is switched. If maximum or
minimum data is updated in the meantime, however, the data values will change.
With the recording function activated, the auto power-off function remains disabled.
(APS off.)
The average value (AVE) displayed is calculated by: Average Value = [(Maximum
value + Minimum Value)/2].
After pressing the SLOW/PEAK key to display the peak value, activate the recording
function and select MAX. The peak hold function will be activated.
2. Display of Elapsed Time
When you press the MAX/MIN key to activate the recording function, the bar graph
segments flash and the elapsed time appears.
When "min" is shown in the right-hand corner of the bar graph, each segment of the
bar graph corresponds to one minute. Every time one minute elapses, one segment
of the flashing bar graph goes on. When all segments of the bar graph go on, the
elapsed time is 30 minutes.
When the elapsed time exceeds 30 minutes, one segment of the flashing bar graph
goes off every time one minute elapses.
When the segments left of a flashing segment remain on: the number of "on"
segments represents the elapsed time (0 to 29).
The illustration below shows when 20 minutes have elapsed:
When the segments right of a flashing segment remain on: the number of "off"
segments (+30) represents the elapsed time (30 to 59).
The illustration below shows when 50 minutes have elapsed:
When digital display switches the average value (AVE) to a instantaneous value when
you press the MAX/MIN key, the right corner of the bar graph indicates hours. In this
mode, each segment of the bar graph corresponds to one hour. The way to read the
bar graph here is similar to reading it in minutes. When all bar graph segments
remain on, the elapsed time is 30 hours.
The illustration below shows when one hour, 40 minutes have elapsed.
3. Deactivation of Recording Function
Pressing the HOLD key deactivates the recording function. HOLD goes on, REC
stops flashing and goes on, and the elapsed time stops incrementing.
While the recording function is being deactivated, data are not updated, even if the
jaw is disconnected from the conductor.
Pressing the HOLD key again cancels HOLD display and activates the recording
function again, with REC flashing again.
4. Cancellation and Resetting of Recording Function
To cancel the recording function, press the related function key (A, V or Hz) for the
measurement in progress. Once the recording function is canceled, the auto power-
off function becomes effective. (APS goes on.)
To restart the unit after resetting the data, temporarily cancel the recording function,
then activate it again by pressing the MAX/MIN key.
Note
An instantaneous power failure and a surge cannot be detected.
The recording function is not effective for the resistance and temperature
measurements.
Use the recording function after having confirmed a battery residual quantity.
(Shown on the bar graph at power on.)
The lowest possible frequency that can be displayed is 30 Hz.
If changing the range when "O.L." is being displayed in any of the displays, the held
data and elapsed time are cleared.
Activate the recording function during measurement to obtain minimum value or
average value data. If the function is activated with no input, the minimum value will
remain zero. To deactivate the recording function, press the HOLD key first, and then
terminate the measurement. If you disconnect the jaw or test lead from the circuit
under measurement without deactivating the recording function beforehand, the
minimum value will be zero.
When the unit is turned off, accumulated data are lost.
1. Remove the two fastening screws of the rear cover,using a
Phillips screwdriver.
2. Remove the rear cover.
3. Remove the old battery without pulling the codes of the
snap.
4. Attach the new battery onto the battery snaps, paying
attention to the polarity, and then install the battery in the
battery holder.
5. Fasten the rear cover.
6. Screw in the fastening screws to fasten the rear cover.
Note
A 6LR61 battery is also usable.
Each of positive and negative terminals of nine-volt layered-
type dry batteries differ slightly in shape and size according
manufactures and types. When attaching the battery onto the battery snap, you may
notice them fastened each together tightly or loosely. Even then, the instrument will work
if the battery with the battery snap attached is installed properly in the battery holder.
The hand strap improves the operation.
Although the instrument seems to be out of order in the following cases, there may be
the causes of the troubles. Check it again before you send it for repair.
*:When the battery is drained, the relay may be operated immediately after the power
is turned on or when the measurement function is changed, and the power may
suddenly be cut off. Replace the battery with a new one when this arises.
To clean the instrument, wipe it gently with a soft cloth moistened with water or mild
detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones,
thinners or gasoline, as they can deform and discolor the case.
The shortest period for possession of the repair parts is 5 years after stopping the
production.
For inquiries about service, contact your dealer or Hioki representative.
Pack the instrument carefully so that it will not be damaged during shipment, and
include a detailed written description of the problem. Hioki cannot be responsible for
damage that occurs during shipment.

This manual suits for next models

1

Other Hioki Test Equipment manuals

Hioki L2020 User manual

Hioki

Hioki L2020 User manual

Hioki 3281 User manual

Hioki

Hioki 3281 User manual

Hioki 3802-50 User manual

Hioki

Hioki 3802-50 User manual

Hioki 3504-40 C HiTester User manual

Hioki

Hioki 3504-40 C HiTester User manual

Hioki 3444 User manual

Hioki

Hioki 3444 User manual

Hioki 3561 User manual

Hioki

Hioki 3561 User manual

Hioki 3244-60 User manual

Hioki

Hioki 3244-60 User manual

Hioki IR4055 User manual

Hioki

Hioki IR4055 User manual

Hioki 3008 User manual

Hioki

Hioki 3008 User manual

Hioki 3665-20 User manual

Hioki

Hioki 3665-20 User manual

Hioki BT3554-50 User manual

Hioki

Hioki BT3554-50 User manual

Hioki 3157 User manual

Hioki

Hioki 3157 User manual

Hioki FT6031-50 User manual

Hioki

Hioki FT6031-50 User manual

Hioki 3159 User manual

Hioki

Hioki 3159 User manual

Hioki 3145 User manual

Hioki

Hioki 3145 User manual

Hioki RM3542 User manual

Hioki

Hioki RM3542 User manual

Hioki 3274 User manual

Hioki

Hioki 3274 User manual

Hioki 3403 User manual

Hioki

Hioki 3403 User manual

Hioki 9261 User manual

Hioki

Hioki 9261 User manual

Hioki 8807-01 User manual

Hioki

Hioki 8807-01 User manual

Hioki RM3543 User manual

Hioki

Hioki RM3543 User manual

Hioki 3803 User manual

Hioki

Hioki 3803 User manual

Hioki RM3543 User manual

Hioki

Hioki RM3543 User manual

Hioki ST4030 User manual

Hioki

Hioki ST4030 User manual

Popular Test Equipment manuals by other brands

Velleman DVM1356 instruction manual

Velleman

Velleman DVM1356 instruction manual

Apera Instruments 28APE190027 manual

Apera Instruments

Apera Instruments 28APE190027 manual

JPI FS-450 Quick reference guide

JPI

JPI FS-450 Quick reference guide

DOM WENS 800 Series Operator's manual

DOM

DOM WENS 800 Series Operator's manual

Ametek JOFRA HPC400 Reference manual

Ametek

Ametek JOFRA HPC400 Reference manual

JYE Tech DSO 068 Assembly guide

JYE Tech

JYE Tech DSO 068 Assembly guide

Testboy 15 operating instructions

Testboy

Testboy 15 operating instructions

Schumacher Electric PTI900X manual

Schumacher Electric

Schumacher Electric PTI900X manual

Ocean Optics LS-1-CAL operating instructions

Ocean Optics

Ocean Optics LS-1-CAL operating instructions

Proceq DYNA ESTRICH operating instructions

Proceq

Proceq DYNA ESTRICH operating instructions

Baso 120-2 manual

Baso

Baso 120-2 manual

Calnex Sentinel user guide

Calnex

Calnex Sentinel user guide

Fluke CombiScope PM3370B user manual

Fluke

Fluke CombiScope PM3370B user manual

pico Technology PicoVNA 106 Programmer's guide

pico Technology

pico Technology PicoVNA 106 Programmer's guide

RS PRO RS-33S instruction manual

RS PRO

RS PRO RS-33S instruction manual

DRUCK & TEMPERATUR Leitenberger LR-Cal LPC 300 operating manual

DRUCK & TEMPERATUR Leitenberger

DRUCK & TEMPERATUR Leitenberger LR-Cal LPC 300 operating manual

Benning IT 130 Short instructions

Benning

Benning IT 130 Short instructions

Jamara ACCU MASTER 8 BAL Instruction

Jamara

Jamara ACCU MASTER 8 BAL Instruction

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.