IDEAL 61-494 User manual

IDEAL INDUSTRIES, INC.
TECHNICAL MANUAL
MODEL: 61-494
MODEL: 61-495
The Service Information provides the following information:
•Precautions and safety information
• Specifications
• Basic maintenance (cleaning, replacing the battery and fuses)
• Performance test procedures
• Calibration and calibration adjustment procedures
Form number: TM61494-5
Revision: 2. Date: Feb 2002
Form number TM61494-5 Rev 2 February 2002

TABLE OF CONTENTS
Title Page #
Introduction 1
Precautions and Safety Information 1
Symbols 1
Safety 2
Specifications 3
General Specifications 3
Voltage Specifications 4
dBm / dBv Specifications 5
Current Specifications 5
Resistance Specifications 5/6
Capacitance, Continuity and Diode Specifications 6
Peak Hold, Frequency Counter and Duty Factor, Specifications 6
Temperature Specifications 7
Physical and environment characteristics 7
Certifications and Compliances 8
Required Equipment 9
Basic Maintenance 10
Opening the Meter Case 10
Replacing the Battery 11
Testing Fuses (FS1 and FS2) 11
Replacing Fuses 12
Cleaning 12
Input Terminals 12
Performance Tests 12
A Basic Operability Test 13
Testing the Display 13
Testing the Voltage Function 13/14
Testing the Millivoltage (mV) Function 15
Testing the Peak Hold Function 15
Testing the Resistance Function 15/16
Testing the Capacitance Function 16/17
Testing the Diode 17
Testing the Milliamp Function 17/18
Testing the Amp Function 18
Testing the Frequency Function 18/19
Testing the Duty Factor Function 19
Testing the Temperature Function 19/20
Calibration 20/21
Calibrating the ACV Frequency Response 21
Form number TM61494-5 Rev 2 February 2002

Page 1
Introduction
Warning
To avoid shock or injury, do not perform the verification tests or calibration
procedures described in this manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
Caution
The 61-490 serials contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES, INC. and its products,
and services, visit IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the Meter only as described in the Users Manual. If you do not do so, the
protection provided by the Meter may be impaired. Read the “Safety Information”
page before servicing this product. In this manual, a Warning identifies conditions
and actions that pose hazard (s) to the user; a Caution identifies conditions and
actions that may damage the Meter or the test instruments.
The Symbols
The symbols used on the Meter and in this manual are explained in Table A.
Table A. The Symbols
Symbol Meaning Symbol Meaning
Alternating signal Battery
Direct signal Earth ground
CAT III IEC overvoltage Category III Fuse
Refer to the manual. Important
information. Double insulated
Take appropriate precautions.
Hazardous voltage may be
present
Form number TM61494-5 Rev 2 February 2002

Page 2
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any products
connected to it. To avoid potential hazards, use the product only as specified.
CAUTION. These statements identify conditions or practices that could result in
damage to the equipment or other property.
WARNING. These statements identify conditions or practices that could result in
personal injury or loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating
specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
Form number TM61494-5 Rev 2 February 2002

Page 3
SPECIFICATIONS
All specifications are warranted unless noted typical and apply to the 61-494 and 61-495.
Stated accuracies are at 23°C ±5°C at less than 80% relative humidity.
If Low Battery indicator is active then specification could be effective.
General specifications:
Characteristics Description
LCD display digits 4 ½ or 3 ¾
Bargraph segments 42 Segment Graph
Display count 20,000 or 4,000
Numeric update rate 2 times / sec (20,000 count)
Bargraph 20 times/sec
Polarity display Automatic
Overrange display “OL” is displayed
Low voltage indicator “ ” is indicated
Automatic power-off time Automatic backlit off =15minutes
Power source One 9V dry cell battery
Maximum input voltage 1000V (750V AC) CAT II
between V and COM
Maximum floating voltage 1000V (750V AC) CAT II between
any terminal and earth ground
Maximum input current
200mA between mA and COM
10A continuous between A and
COM (20A for 30 seconds)
Maximum open circuit
Voltage (current inputs)
600V between A and COM and
between mA and COM
Overload protection mA connector 1A (600V) fast blow fuse
A connector 15A (600V) fast blow fuse
V connector
1100 Vp V~ V AC+DC
850 Vp mV~ mV AC+DC
LVΩΩHz% DF °C °F
Temperature Coefficient 0.1 x (Spec. Accuracy) / °C, < 18°C
or > 28°C
Battery Life 100 hours typical (alkaline)
Form number TM61494-5 Rev 2 February 2002

Page 4
Measurement Characteristics
(All at 23°C ±5°C, < 80% R.H.) ±([% of reading] + [number of least digits]).
Voltage Specifications:
DCV 61-494 61-495
20mV ±(0.06% + 60) ±(0.06% + 60)
200mV ±(0.06% + 20) ±(0.06% + 20)
2V, 20V, 200V, 1000V ±(0.06% + 10) ±(0.06% + 10)
ACV 61-494 61-495
20mV, 200mV
40Hz ~ 100Hz
100Hz ~ 1KHz
±(0.70% + 80)
±(1.00% + 80)
±(0.70% + 80)
±(1.00% + 80)
2V
40Hz ~ 100Hz
100Hz ~ 1KHz
1KHz ~ 10KHz
10KHz ~ 20KHz
20KHz ~ 50KHz
50KHz ~ 100KHz
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
±(10.00% + 100)
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
±(10.00% + 100)
20V
40Hz ~ 100Hz
100Hz ~ 1KHz
1KHz ~ 10KHz
10KHz ~ 20KHz
20KHz ~ 50KHz
50KHz ~ 100KHz
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
±(10.00% + 100)
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
±(10.00% +100)
200V
40Hz ~ 100Hz
100Hz ~ 1KHz
1KHz ~ 10KHz
10KHz ~ 20KHz
20KHz ~ 50KHz
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
±(0.70% + 50)
±(1.00% + 50)
±(2.00% + 60)
±(3.00% + 70)
±(5.00% + 80)
750V
40Hz ~ 100Hz
100Hz ~ 1KHz
±(0.70% + 50)
±(1.00% + 50)
±(0.70% + 50)
±(1.00% + 50)
Bandwidth 40Hz ~ 100KHz 40Hz ~ 100KHz
AC Conversion Type: AC Coupled True RMS responding.
AC + DC Volts: Same as AC (RMS) + 1.00% + 80.
Crest Factor: +1.5% addition error for C.F. 1.4 to 3
+3.0% addition error for C.F. from 3 to 4
Resolution: 1µV in the 20mV range.
Input Impedance: 10MΩ, <100pF.
Overload Protection: 1000VDC, 750V rms.
Form number TM61494-5 Rev 2 February 2002

Page 5
dBm / dBv Specifications:
dBm (typical): -15 dBm to +55 dBm (0 dBm = 1mW into 600Ω).
dBv (typical): -80 dBv to + 50 dBv (0 dBv = 1 Vrms).
Note: (ACV and AC + DCV)
Add additional 20 counts for reading under 0.5 time of range for
5K ~ 50KHz. Specifications exclude under 0.4 time of range for
50KHz ~ 100KHz.
Current Specifications:
DCA 61-494 61-495
20mA, 200mA ±(0.20% + 40) ±(0.20% + 40)
2A, 10A ±(0.20% + 40) ±(0.20% + 40)
ACA 61-494 61-495
20mA, 200mA, 2A,10A
40Hz ~ 500Hz ±(0.80% + 50) ±(0.80% + 50)
20mA, 200mA, 2A,10A
500Hz ~ 1KHz ±(1.20% + 80) ±(1.20% + 80)
200mA, 10A1KHz ~ 3KHz ±(2.00% + 80) ±(2.00% + 80)
Range: 20mA, 200mA, 2A, 10A.
Resolution: 1µA in the 20mA range.
Burden Voltage: 800mV max. for mA input, 1V max. for A input.
AC Conversion Type: AC Coupled True RMS responding.
Input Protection: Equipped with High Energy Fuse.
1A, 600V, IR 10KV fuse (Bussmann BBS-1 or equivalent) for mA input.
15A, 600V, IR 100KV fuse (Bussmann KTK 15 or equivalent) for A input.
AC + DC Current: Same as AC (RMS) + 1.00% + 80
C.F.: Same as ACV.
Resistance Specifications:
Ohm 61-494 61-495
200Ω, 2KΩ±(0.30% + 30) ±(0.30% + 30)
20KΩ, 200KΩ±(0.30% + 30) ±(0.30% + 30)
2MΩ±(0.30% + 50) ±(0.30% + 50)
20MΩ±(5.00% + 50) ±(5.00% + 50)
200MΩ±(5.00% + 20) ±(5.00% + 20)
2GΩ±(5.00% + 8) ±(5.00% + 8)
LV Ohm 61-494 61-495
2KΩ, 20KΩ, 200KΩ±(0.60% + 20) ±(0.60% + 30)
2MΩ±(0.60% + 50) ±(0.60% + 50)
20MΩ±(7.00% + 50) ±(7.00% + 50)
200MΩ±(7.00% + 20) ±(7.00% + 20)
Form number TM61494-5 Rev 2 February 2002

Page 6
Resolution: 0.01Ωin the 200Ωrange.
Open Circuit Voltage: 3.3V
Open Circuit Low Voltage: 0.6V
Input Protection: 600V rms.
Capacitance Specifications (4000 counts):
Capacitance 61-494 61-495
4nF, 40nF * ±(1.50% + 10d) ±(1.50% + 10d)
400nF, 4µF ±(0.90% + 5d) ±(0.90% + 5d)
40µF, 400µF ±(1.20% + 5d) ±(1.20% + 5d)
4mF, 40mF ±(1.50% + 5d) ±(1.50% + 5d)
Note: For best measurements, with ∆mode on nF ranges.
* With ∆mode.
Range: 4nF, 40nF, 400nF, 4µF, 400µF, 4mF, 40mF
Resolution: 1pF in the 4nF range.
Input Protection: 600V rms.
Continuity Check Specifications:
Continuity Threshold: Approx. 50 Ω.
Continuity Indicator: 2KHz Tone Buzzer.
Input Protection: 600V rms.
Diode Test Specifications:
Test Current: 1.1mA (Typical)
Open Circuit Voltage: 3.3V DC (max).
Input Protection: 600V rms.
Peak Hold Specifications:
+[±(0.7% + 20)] additional error for >20% of full scale
and pulse width greater than 0.5mS; ±(10) more for
50% of full scale on 2V range (2000 counts)
Frequency Counter Specifications:
Range: 20Hz, 200Hz, 2KHz, 20KHz, 200KHz, 1MHz.
Resolution: 0.001Hz in the 20Hz range.
Accuracy: ±(0.01% + 10), ±(0.01% + 50) for 20Hz Range.
Sensitivity: 0.5Vp-p, for 5Hz ~ 1MHz.
Min. Frequency: 5Hz.
Input Protection: 600V rms.
Duty Factor Specifications:
Range: 20% ~ 80%
Resolution: 0.1%.
Accuracy: ±1% (20Hz ~ 10KHz, 5Vp-p), ±2% for 50% ~ 80%
Form number TM61494-5 Rev 2 February 2002

Page 7
Temperature Specifications:
Temperature 61-494 61-495
-100°C ~ 1200°C ±(0.1% + 3°C) ±(0.1% + 3°C)
-200°C ~ -100°C ±(0.1% + 6°C) ± (0.1% + 6°C)
Note: The readings < 360°C on 1°C resolution range when settled as manual range, display reads “Er” to
change to lower range for best measurements.
Multiply the digit accuracy by 2 for °F.
Range: -200°C ~ 1200°C.
Resolution: 0.1°C for –200°C ~ 400°C, 1°C for 400°C ~ 1200°C
Input protection: 600V rms.
Physical and Environmental Characteristics:
Physical Characteristics: Description
Dimensions (H x W x D) 183mm x 85mm x 33mm
202mm x 95mm x 46mm (with holster)
Weight (with battery) 0.4Kg
With holster 0.6Kg
Environmental Characteristics: Description
Temperature operating 0 to + 50°C
Non-Operating -20 to + 60°C
Humidity (operating) < 80% Relative Humidity
Altitude Operating 2,222 m (7290 ft.)
Non-Operating 12,300 m (40354 ft.)
Vibration & shock Operating
MIL-T-28800E TYPE II Class 5
2.66gRMS, 5 to 500 Hz, 3axes
(10 minutes each)
Dust / Water Protection IP Rating IP 64.
Indoor Use
Form number TM61494-5 Rev 2 February 2002

Page 8
Certifications and compliances
Safety Designed to IEC 1010-1, UL3111-1 and CSA specifications
Input rating 1000V DC Category II
600V DC Category III
750V AC Category II
600V AC Category III
Over voltage category CAT III: Distribution level mains, fixed installation.
CAT II: Local level mains, appliances, portable equipment.
CAT I: Signal level, special equipment or parts of
equipment, telecommunication, electronics.
Pollution Degree 2 Do not operate in environments where conductive Pollutants may be
present.
EC Declaration of
Conformity
Meets the intent of Directive 89/336/EEC for
Electromagnetic Compatibility and Low Voltage
Directive 73/23/EEC for Product Safety. Compliance
was demonstrated to the following specifications
as listed in the official Journal of the European
Communities:
En 55011 Class A: Radiated and Conducted Emissions.
EN 50082-1 Immunity:
IEC 801-2 Electrostatic Discharge
IEC 801-3 RF Radiated
EN 61010-1 Safety requirements for electrical equipment
for measurement, control, and laboratory use.
Form number TM61494-5 Rev 2 February 2002

Page 9
Required Equipment
Required equipment is listed in Table B. If the recommended
models are not available, equipment with equivalent specifications may be used.
Repairs or servicing should be performed only by qualified personnel.
Table B. Required Equipment
Equipment Required Characteristics Recommended
Model
Calibrator AC Voltage Range: 0-750V ac
Accuracy: ±0.07% (Basic)
Frequency Range: 40 ~ 100KHz
Accuracy: ±2%
DC Voltage Range: 0-1000V dc
Accuracy: ±0.006% (Basic)
Current Range: 0 ~ 10A
Accuracy:
AC (40Hz to 1KHz): ±0.08% (Basic)
DC: ±0.02% (Basic)
Frequency Source:
5.00Hz ~ 1.0000MHz
Accuracy: ±0.001%
Amplitude:
0.5V p-p ~ 1.0V p-p (square wave)
Accuracy: ±5%
Resistance Range: 1Ω~ 2GΩ
Accuracy: ±0.03% (Basic)
Capacitance Range: 1pF ~ 40mF
Accuracy: ±0.10% (Basic)
Temperature Range: -200°C ~ 1200°C
Accuracy: ±0.3°C (Basic)
Fluke 5500 or
Wavetek 9100
Calibrator or
equivalent
Form number TM61494-5 Rev 2 February 2002

Page 10
Basic Maintenance
Warning
To avoid shock, remove the test leads and any input signals before opening
the case or replacing the battery or fuses.
Opening Meter Case
Caution
To avoid unintentional short circuit, always place the uncovered Meter
assembly on a protective surface. When the case of the Meter is open,
circuit connections are exposed.
To open the Meter case, refer to Figure 1.1 and do the following:
1. Disconnect test leads from any live source, turn the rotary switch to OFF, and remove
the test leads from the front terminals.
2. Remove the battery door by using a flat-blade screwdriver to turn the battery door
screws 1/4 turn counterclockwise.
3. The case bottom is secured to the case top by three screws and two internal snaps
(at the LCD end). Using a Phillips-head screwdriver, remove the three screws.
4. Hold the Meter display side up.
5. Lifting up on the input terminal end, disengage the case top from the gasket.
6. Gently unsnap the case top at the display end.
Note
The gasket between the two case halves is sealed to, and must remain with, the case bottom. The case top
lifts away from the gasket easily. Do not damage the gasket or attempt to separate the case bottom from the
gasket.
Figure 1.1
Form number TM61494-5 Rev 2 February 2002

Page 11
Figure 1.2
Replacing the Battery
The Meter is powered by a single 9V battery (NEDA 1604, 6F22, or 006P).
To replace the battery, refer to Figure 1.2 and do the following:
1. Turn the rotary switch to OFF and remove the test leads from the terminals.
2. Remove the battery door by using a flat-blade screwdriver to the turn the battery
door screws 1/4-turn counterclockwise.
3. Remove the battery and replace it with a new one. Dress the battery leads so that
they will not be pinched between the battery door and case bottom.
4. Replace the battery door and secure the door by turning the screws 1/4-turn clockwise.
Testing Fuses (FS1 and FS2)
To test the internal fuses of the Meter, refer to Figure 1 and do the following:
1. Turn the rotary selector switch to the mA position for 1A fuse test FS2 or
A position for 15A fuse test FS1.
2. To test FS2, plug a test lead into mA input terminal, and turn the rotary selector to
the mA position.
2.1 Fuse is OK if the Display shows normal functional graphics.
2.2 Fuse is defective if “Prob” is displayed and the built-in beeper alarm is on.
3. To test FS1, plug a test lead into A input terminal, and turn the rotary selector to
the A position.
3.1 Fuse is OK if the Display shows normal functional graphics.
3.2 Fuse is Defective if “Prob” is displayed and the built-in beeper alarm is on.
Form number TM61494-5 Rev 2 February 2002

Page 12
Replacing Fuses
Warning
To avoid electrical shock, remove the test leads and any input signals before
replacing the battery or fuses. To prevent damage or injury, INSTALL ONLY
quick acting fuses with the following Amp/Volt current interrupt rating:
FS1 Fuse: 15A, 600V, FAST. Minimum interrupt rating 100,000A
FS2 Fuse: 1A, 600V, FAST. Minimum interrupt rating 10,000A
1. Open the case as described in “Open Meter Case” section, steps 1 – 6
2. Gently lift the case top to separate the two halves of the case.
3. Remove the fuse by gently prying one end loose, then sliding the fuse out of its
bracket.
4. Replace the fuse only with one specified above.
5. Verify that the rotary switch and the circuit board switch are in the OFF position.
6. Place the case top on the case bottom, ensuring that the gasket is properly seated and
the case halves snap together (See Figure 1).
7. Reinsert the three case bottom screws and the battery door.
Cleaning
Warning
To avoid electrical shock or damage to the Meter, never allow water inside
the case. To avoid damaging the Meter’s housing, never apply solvents to
the Meter.
Input Terminals
Water, dirt, or other contamination in the Aor mA input terminals may
activate the Input Alert beeper even though test leads are not inserted.
Such contamination might be dislodged by turning the Meter over and,
with all test leads removed, gently tapping on the case.
To clean the input terminals more effectively, do the following:
1. Turn the Meter off and remove all test leads from the terminals.
2. Use a clean swab in each of the four terminals to dislodge and clean out the
contamination.
3. Moisten a new swab with a cleaning and oiling agent. Work this swab around in each
of the four terminals. The oiling agent insulates the terminals from moisture-related
shorting and ensures against false Input Alerts.
Performance Tests
The following performance tests verify the complete operability of the Meter and checks the accuracy of
each Meter function against the Meter’s specifications.
Accuracy specifications are valid for a period of one year after calibration, when measured at an operating
temperature of 18°C to 28°C and at a maximum of 80% relative humidity.
To perform the following tests, it is not necessary to open the case; no adjustments are necessary. Merely
make the required connections, apply the designated inputs, and determine if the reading on the Meter display
falls within the acceptable range indicated.
If the Meter fails any of these tests, it needs calibration adjustment or repair.
Form number TM61494-5 Rev 2 February 2002

Page 13
A Basic Operability Test
To check the basic operability, do the following:
1. Turn the rotary switch to Aand plug a test lead into Ainput terminal, display is of
functional graphics if the fuse is good.
2. Then test the mA input fuse (1A) by inserting a test lead into mA input terminal and
turn the rotary switch to mA. (Ainput must be open).
The display is of normal functional graphics if the fuse is good.
If the Meter fails to operate properly:
• Check the battery and fuses and replace as needed.
• Verify that you are operating the Meter correctly by reviewing the operating instructions
found in the Users Manual.
To complete a comprehensive performance test and verify the accuracy of each Meter function and
operation, perform the remainder of the tests under “Performance Tests”.
Testing the Display
Press Bar key and turn the meter on from the OFF position to hold the Meter in Display Test Mode.
Compare the display with the appropriate example in Figure 2. Turn off the meter to escape the test mode.
LCD Display of the 61-490 Series
Testing the Voltage Function
To verify accuracy in the AC and DC voltage ranges, do the following:
1. Connect the Calibrator to the VΩand COM inputs on the Meter.
2. Set the Calibrator for the voltage and frequency from step 1 to 19 in Table 1.
3. Compare the reading on the Meter display with the display reading shown in Table 1.
4. If the display reading falls outside of the range shown in Table 1, the Meter does not
meet specification.
Form number TM61494-5 Rev 2 February 2002

Page 14
Table 1. AC Voltage Test
Input Reading
Step Voltage Frequency
1 1.0000V 60Hz 0.9880 to 1.0120
2 1.0000V 1KHz 0.9850 to 1.0150
3 1.0000V 10KHz 0.9740 to 1.0260
4 1.0000V 20KHz 0.9630 to 1.0370
5 1.0000V 50KHz 0.9420 to 1.0580
6 1.0000V 100KHz 0.8900 to 1.1100
7 10.000V 60Hz 9.880 to 10.120
8 10.000V 1KHz 9.850 to 10.150
9 10.000V 10KHz 9.740 to 10.260
10 10.000V 20KHz 9.630 to 10.370
11 10.000V 50KHz 9.420 to 10.580
12 10.000V 100KHz 8.900 to 11.100
13 100.00V 60Hz 98.80 to 101.20
14 100.00V 1KHz 98.50 to 101.50
15 100.00V 10KHz 97.40 to 102.60
16 100.00V 20KHz 96.30 to 103.70
17 100.00V 50KHz 94.20 to 105.80
18 700.0V 60Hz 690.1 to 709.9
19 700.0V 1KHz 688.0 to 712.0
5. Press the blue button on the Meter to toggle to DCV function.
6. Set the calibration for the voltage from step 1 to 4 in Table 2.
7. Compare the reading on the Meter display with the display reading shown in Table 2.
8. If the display reading falls outside of the range shown in Table 2, the meter does not
meet specification.
Table 2. DC Voltage Test
Input Reading
Step Voltage
1 1.0000V 0.9984 to 1.0016
2 10.000V 9.981 to 10.016
3 100.00V 99.84 to 100.16
4 900.0V 898.5 to 901.5
Form number TM61494-5 Rev 2 February 2002

Page 15
Testing Millivoltage (mV) Function
To verify accuracy of the mV function, do the following.
1. Connect the Calibrator to the VΩand COM inputs on the meter.
2. Turn the rotary switch to mV
3. Set the calibrator for the voltage and frequency from step 1 to 4 in Table 3.
4. Compare the reading on the Meter display with the display reading in Table 3.
5. If the display reading falls outside of the range shown in Table 3, the Meter does not
meet specification.
Table 3. AC mV Test
Input Reading
Step Voltage Frequency
1 10.000mV 60Hz 9.850 to 10.150
2 10.000mV 1KHz 9.820 to 10.180
3 100.00mV 60Hz 98.50 to 100.15
4 100.00mV 1KHz 98.20 to 100.20
6. Press the blue button on the Meter to toggle to DC mV function.
7. Set the calibrator for the voltage from step 1 to 2 in Table 4.
8. Compare the reading on the Meter display with the display reading in Table 4.
9. If the display reading falls outside of the range shown in Table 3, the Meter does not
meet specification.
Table 4. DC mV Test
Input Readings
Step Voltage
1 10.00mV 9.934 to 10.066
2 100.00mV 99.74 to 100.26
Testing the PEAK HOLD Function
To check PEAK HOLD feature, do the following:
1. Connect the Calibrator to the VΩand COM inputs on the Meter.
2. Apply 1.0V AC at 60Hz from the Calibrator to the VΩand COM inputs on the Meter.
3. Turn the rotary switch to V
Note: The RMS converter is not used in Peak Hold mode. The digital display represents
the actual peak value of the input.
4. Press PEAK H and then M/M/A
5. The reading on sub-display should be within 1.441 (Max) and -1.441 (MIN).
6. Press PEAK H to escape.
Testing the Resistance Function
To verify the accuracy of the resistance function, do the following:
1. Connect the Calibrator to VΩand COM on the Meter.
2. Turn the rotary switch to Ω
3. Apply the inputs for steps 1 - 7 in Table 5.
4. Compare the Meter display readings to the readings in Table 5.
Note: Press RANGE on the Meter to enter the 200MΩand 2GΩranges. Then
proceed with steps 8 - 9 in Table 5. Compare the meter display readings to the
readings in Table 5.
5. If the display reading falls outside of the range shown in Table 5, the meter does not meet specification.
Form number TM61494-5 Rev 2 February 2002

Page 16
Table 5. Resistance Test
Input Reading
Step Resistance (Ω)
1 0.00 0 to 0.30
2 100.00 99.40 to 100.60
3 1.0000K 0.9984 to 1.0016
4 10.000K 9.984 to 10.016
5 100.00K 99.84 to 100.16
6 1.0000M 0.9920 to 1.0080
7 10.000M 9.450 to 10.550
8 100M 75 to 125
9 1.0G 0 to 1.3
6. Press the blue button on the meter to toggle to LVΩfunction.
7. Apply the inputs for step 1 - 6 in Table 6.
8. Compare the Meter display readings to the readings in Table 6
Note: Press RANGE on the Meter to enter 200MΩrange.
9. If the display reading falls outside of the reading shown in Table 6, the meter does not
meet specification.
Table 6 LVΩTest
LVΩInput Reading
Step Resistance (Ω)
1 0.00 0 to 0.0030
2 1.0000K 0.9910 to 1.0090
3 10.000K 9.910 to 10.090
4 100.00K 99.10 to 100.90
5 1.0000M 0.9890 to 1.0110
6 10.000M 9.250 to 10.750
7 100M 43 to 157
Testing the Capacitance Function
The Meter measures capacitance by charging the capacitor with a known Direct current, measuring the
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an impedance
bridge, a different reading may result. This variance is likely to be Greater at higher frequencies.
To verify the accuracy of the capacitance measuring function, do the following:
1. Apply the Capacitor to the VΩand COM inputs to the Meter. For steps 1 through 8
in Table 7.
2. Turn the rotary switch to
3. Set the VΩand COM inputs un-connected
Note: Press REL∆for steps 1 and 2 in Table 7, after steps 1 and 2 are completed, then
escape REL mode, then press RANGE until AUTO is displayed.
{upper left corner of display}
Form number TM61494-5 Rev 2 February 2002

Page 17
4. Compare the readings on the Meter display to readings in Table. 7
Note: The meter selects the proper range automatically. Each measurement takes
about one second per range, 5mF takes about 15 seconds.
5. If the display reading falls outside of the range shown in Table 7, the Meter does not
meet specification.
Table 7: Capacitance Test
Input Reading
Step Capacitance
I 1 .000nF 0.975 to 1.025
2 10.00nF 9.750 to 10.25
3 100.0nF 98.60 to 101.4
4 1 .000uiF 0.986 to 1.014
5 10.00pF 9.830 to 10.17
6 100.0pF 98.30 to 101.7
7 1.000mF 0.980 to 1.020
8 5.00mF 4.880 to 5.120
Checking the Diode Test Function
To check the diode test function, do the following:
1. Connect the Calibrator to the VΩand COM inputs on the Meter.
2. Turn the rotary switch to .
3. Apply 3.000V.
The meter display should read approx. 3.000V DC.
4. Press the blue button to test the function.
5. Apply a 50Ωresistor to the meter. The built-in beeper alarms.
Testing the Milliamp (mA) Function
To verify the accuracy of AC and DC current measurement functions, do the following:
1. Connect the Calibrator to the mA and COM inputs on the Meter.
2. Turn the rotary switch to mA
3. Apply the inputs for steps 1 - 5 in Table 8.
4. For each input, compare the readings on the Meter display to the readings
in Table 8.
5. If the display reading falls outside of the range shown in the Tables 8, the meter does
not meet specification.
Table 8: AC mA Test
Input Reading
Step Current Frequency
1 10.000mA 60Hz 9.870 to 10.130
2 10.000mA 1KHz 9.800 to 10.200
3 100.00mA 60Hz 98.70 to 101.3
4 100.00mA 1KHz 98.00 to 102.00
5 100.00mA 3KHz 97.20 to 102.80
6. Press the blue button on the Meter to toggle to DC measurement function.
7. Apply the inputs for steps 1 - 2 in Table 9.
8. For each input, compare the readings on the Meter display to the readings
in Table 9.
9. If the display reading falls outside of the range shown in the Tables, the meter does
not meet specification.
Form number TM61494-5 Rev 2 February 2002

Page 18
Table 9: DC mA Test
Input Reading
Step Current
1 10.000mA 9.940 to 10.060
2 100.00mA 99.40 to 100.60
Testing the Amp (A) Function
To verify the accuracy in the ampere (A) measurement function, do the
Following:
1. Connect the Calibrator to the Aand COM inputs of the Meter.
2. Turn the rotary switch to A
3. Apply the inputs for steps 1 - 6 in Table 10.
4. For each input, compare the readings on the Meter display to the readings
in Table 10.
5. If the display reading falls outside of the range shown in Table 10, the meter does not
meet specification.
Table 10. ACA Test
Input Reading
Step Current Frequency
1 1.0000A 60Hz 0.9870 to 1.0130
2 1.0000A 1KHz 0.9800 to 1.0200
3 1.0000A 3KHz 0.9720 to 1.0280
4 9.000A 60Hz 8.878 to 9.122
5 9.000A 1KHz 8.812 to 9.188
6 9.000A 3KHz 8.740 to 9.260
6. Press the blue button on the Meter to toggle to the DCA measurement function.
7. Apply the inputs for steps 1 - 2 in Table 11.
8. For each input, compare the reading on the Meter display to the readings for
your meter in Table 11.
9. If the display reading falls outside of the range shown in Table 11, the meter does not
meet specification.
Table 11. DCA Test
Input Reading
Step Current
1 1.0000A 0.9940 to 1.0060
2 9.000A 8.942 to 10.058
Testing the Frequency Function
To verify the accuracy of the Meter’s frequency function, do the following:
1. Connect the Calibrator to the VΩand COM inputs on the Meter.
Note The accuracy of the Calibrator’s frequency function must be appropriate
for the specified accuracy of the Meter.
2. Set the rotary switch to Hz.
3. Set the Function Generator for the square wave voltage and frequency
for steps 1 - 6 of Table 12.
Form number TM61494-5 Rev 2 February 2002
This manual suits for next models
1
Table of contents
Other IDEAL Multimeter manuals

IDEAL
IDEAL 61-312 User manual

IDEAL
IDEAL ND 2351-1 User manual

IDEAL
IDEAL 61-764 User manual

IDEAL
IDEAL ND 1075-2 User manual

IDEAL
IDEAL 61-310 User manual

IDEAL
IDEAL 480 Series User manual

IDEAL
IDEAL 61-797 User manual

IDEAL
IDEAL 602 User manual

IDEAL
IDEAL 61-603 User manual

IDEAL
IDEAL 61-354 (TRMS) User manual