Leica M501 User manual

Leica M501
User manual
10 711 785 – Version D

Leica M501 / Ref. 10 711 785 / Version D

Chapter Overview
1
Leica M501 / Ref. 10 711 785 / Version D
Introduction 3
Controls 7
Preparation for Operation 9
Use 19
Safety Notes 22
Care and Maintenance 32
Messages and Warnings 36
Technical data 37

Table of Contents
2Leica M501 / Ref. 10 711 785 / Version D
Page
Introduction
Design and function 4
Ceiling mounts 5
Controls
Control unit/lamp housing 6
Tilt head/focusing unit 6
Footswitch 6
Handswitch 7
Stand 7
Remote control for Leica Telescope Mount 7
Optics carrier 8
Binocular tube, eyepiece, tubes for second observer 8
Preparation for Operation
Retrofit the surgical microscope and balance
the swinging arm 9
Attach the binocular tube, eyepiece and objective 10
Set the interpupillary distance and eyepoint 12
Adjust for parfocality 12
Attach the adapter for accessories 13
Adjusting second-observer tube 14
Display/change the footswitch type 15
Switch the plus/minus movement directions
of the XY-unit 15
Disable/enable the function of the XY-Reverse button 15
Transport the surgical microscope 16
Position the surgical microscope at the OP stage 17
Sterile components 18
Setting the rest position on the Leica
Telescope Mount 18
Setting the rest position on the Leica Mini Mount 19
Use
Operating the surgical microscope 19
Positioning the microscope 20
Adjusting the focus 20
Adjusting the magnification 20
Adjusting the illumination 21
Setting the illumination type and the working distance
21
Safety Notes
Intended use 22
Directions for the operator of the instrument 22
Directions for the user of the instrument 22
Table 201 according to EN 60601-1-2:2001 23
Table 202 according to EN 60601-1-2:2001 24
Table 204 according to EN 60601-1-2:2001 25
Table 206 according to EN 60601-1-2:2001 26
Hazards associated with instrument use 27
Signs and labels 28
Page
Care and Maintenance
Changing the fuse 32
Changing the bulb 33
Inspecting the functioning 33
Notes on reprocessing of resterilizable products 34
Messages and Warnings
Ventilation 36
General malfunctions 36
Technical data
Electrical data 37
Microscope 37
Stands 38
Swing arm 39
Ambient conditions 39
Limitations of use 39
Standards 39
Dimensions 40

Introduction
3
Leica M501 / Ref. 10 711 785 / Version D
User manual
This user manual contains important safety
precautions as well as information on using the
instrument (see the chapter "Safety notes").
Before attempting to set up the product, carefully
read through the user manual.
Product identification
The model code and serial number of your product are provided
on the nameplate found on the underside of the control unit.
Write this data into your User Manual and always refer to it
when you contact us or the service workshop regarding any
questions you may have.
Model: Serial No.:
Symbols used in this manual
The symbols used in this user manual
have the following meanings:
Warning regarding use hazard or
Warning noncompliant use that can lead to
serious injury or death.
Warning regarding use hazard or non-
Caution compliant use that can lead to minor
injury, but significant article, property
or environmental damage.
Useful information that can help the user
operate the product correctly and effi-
ciently.
Request for action; here, you are
requested to take action.
➩

Introduction
4Leica M501 / Ref. 10 711 785 / Version D
Design and function
1 Base
2 Control unit/lamp housing
3 Swing arm
4 XY-coupling (optional)
5 Tilt head
6 Focusing unit
7 Optics carrier
8 Binocular tube
9 Eyepieces
10 Objective
Floor stand
1
3
2
4
5
8
9
10
6
7

Introduction
5
Leica M501 / Ref. 10 711 785 / Version D
Ceiling mounts
1 Mini Mount ceiling mount
1
2 Leica Telescope Mount
2

6Leica M501 / Ref. 10 711 785 / Version D
Controls
Control unit/lamp housing
6
2
3
4
51
7
2
1
•XY-Reverse
Tilt head
3 Tilt head fine adjustment
4 Brake knob for rough adjusting
the tilt
Focusing unit
1 XY-unit reset button
2 Focus reset button
5
4
+
-
7
1
2
3
6
5
4
+
-
3
7
1
2
6
Footswitch
Cross pedal variant Vertical pedal variant
1 Main switch
2 Regulating knob for adjusting the travel speed of the
XY-coupling
3 Regulating knob for adjusting the travel speed of the
focus
4 Regulating knob for adjusting the travel speed of the
zoom
5 Regulating knob for adjusting the brightness of the
illumination
6 Rotary knob for changing filters
7 Turn grip for the quick-change lamp mount
1 XY adjustment
2 Focus up and down
3 Zoom up and down
4 Greater illumination
5 Less illumination
6 XY-Reverse
7 Microscope illumination on/off
Tilt head/focusing unit
34

7Leica M501 / Ref. 10 711 785 / Version D
Controls
ZF
A
2
3
6
45
1
7
ML
XY
SL
Handswitch
1 XY adjustment
2 Focus up and down
3 Zoom up and down
4 Greater illumination
5 Less illumination
6 XY-Reverse
7 Microscope illumination on/off
Stand
1 Footbrakes
2 Footbrake release lever
3 Articulation brakes
4 Balance turn knob
5 Safety hook
6 Retaining pin
3
33
3
4
1
2
5
6
Remote control for Leica Telescope Mount
1
1 Buttons for up/down movement

8Leica M501 / Ref. 10 711 785 / Version D
Controls
Optics carrier
1 Turn knob for fading ambient lighting
2 Handles
+5 -5 +5 -5
Binocular tube, eyepiece, tubes for second observer
1 Drive knob for adjusting the interpupillary distance
2 Diopter adjustment
3 Knurled ring for image correction
1
3
3
2

9Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Lock the swing arm.
➩Position the swing arm approximately horizontally.
➩Pull out the retaining pin (3).
➩Move the swing arm slightly up and down until the safety
hook (2) engages.
The swing arm is now locked.
Clean the optical accessories
➩Inspect the eyepieces, objectives and any present photo and
TV-adapters for cleanliness.
➩Remove dust and dirt.
Fitting accessories
➩Equip the microscope ready for use with all necessary
accessories.
Balance the swing arm
➩Hold the microscope firmly.
➩Move the swing arm slightly up and down, at the same time
pushing the counterlever of the safety hook (2) upwards,
until the retaining pin clicks into position.
➩See whether or not the microscope drifts.
Microscope drifts downwards:
➩Turn rotary knob (4) clockwise.
Microscope drifts upwards:
➩Turn rotary knob (4) counter-clockwise.
Warning 1
Risk of injury from the surgical microscope swinging
down!
➩Never rebalance or re-equip with the instrument over
the field of operation.
➩After re-equipping, always rebalance the swinging
arm.
Retrofit the surgical microscope and
balance the swinging arm
2
3

10 Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Attach the binocular tube, eyepiece and
objective
Binocular tube 10° – 50°, ultra low
Binocular tube 10° – 50°, low
Binocular tube with variable 180°
Binocular tube var. 30° – 150°
Attach the binocular tube
➩Unscrew the clamping
screw (1).
➩Push the binocular tube into
the dovetail ring.
➩Tighten the clamping screw.
A range of options enables the Surgical Microscope to be
matched to the requirements of the task in hand.
Inclined binocular tube
Straight binocular tube
Inclined binocular tube 45°
Optional for use on the assistant’s
attachment
(no standard configuration)
1

11Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
1
Fitting the eyepiece
➩Set the eyepiece in place.
➩Tighten the rotary ring (1).
Eyepieces
Eyepiece 10x/21B, adjustable
Eyepiece 12.5x/17B, adjustable
Fitting objectives
Objectives are screwed in on the microscope with right-hand
threading.
Objectives
Objective WD = 175mm APO
Objective f = 175mm
Objective f = 200mm
Objective f = 225mm
Objective f = 250mm
Objective f = 275mm
Objective f = 300mm

12 Leica M501 / Ref. 10 711 785 / Version D12
Preparation for Operation
The distance between both pupils and the desired contact with
the eyepieces are adjusted individually.
You can prepare the surgical microscope before the operation
using the data from a user table.
Regulate the interpupillary distance
➩Turn the eyelenses on the eyepieces to "0" or set the dioptric
settings, if known.
➩Set the magnification changer to step 10.
➩Look through the eyepieces and adjust the tubes by hand
(for binocular tubes without drive knobs) or using the drive
knob (1) until you see a concentrically circular image field.
Set the interpupillary distance and
eyepoint
Adjust for parfocality
1
Adjusting the dioptric settings
Adjust the dioptric settings accurately for each eye separately;
only this method will ensure that the image will stay in
focus throughout the entire zoom range (parfocal).
1. Prepare the microscope
➩Switch on the microscope at the control unit and place a flat
test object such as a piece of paper beneath the objective.
➩Engage the highest brightness.
➩Engage the lowest magnification.
➩Displace the microscope so that the test object is visible in
the center of the field of view.
2. Focus on the test object
➩Engage the highest magnification.
➩Bring the test object into focus.
➩Engage the lowest magnification.
3. Adjust the dioptric settings
➩Adjust the dioptric settings for each eye in turn (2), so that
the image is seen in sharp focus.
➩Engage the highest magnification.
➩Refocus using motor-driven focusing.
➩Engage the lowest magnification.
➩Inspect the dioptric settings, readjusting them if necessary
so that both images are sharp.
4. Inspect the parfocality
➩Zoom through the whole range, observing the test object.
The image sharpness must remain constant at all magnifica-
tions. If it does not, then repeat points 2 to 4 of this proce-
dure.
+5 -5 +5 -5
2

13Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Attach the adapter for accessories
Beam splitter with 50/50% observance
Stereo adapter for accessories
Adapter for Zeiss accessories on the Leica M501
Adapter for M600 accessories for the Leica M501
1
Attach the beam splitter/stereo adapter
➩Unscrew clamping screw (1).
➩Push the beam splitter/stereo adapter into the dovetail ring.
➩Tighten the clamping screw.
Attach the adapter
➩Insert adapter into beam splitter.
➩Tighten the rotary ring (1).

14 Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Stereo attachment for second observer
The dual stereo attachments can be attached to the left or right
on the beam splitter and turned in any direction.
Tubes for second observer
The assistant turns the monocular tube for second observer in
the desired direction, adjusts his or her diopter on the eyepiece
and centers the image using the knurled ring.
Stereo attachment for second observer
The assistant turns his or her diopters in the desired direction,
centers the binocular tube horizontally, adjusts his or her
diopters and centers the image using the knurled ring.
Attach the stereo attachment/tube for second observer.
➩Feed the stereo attachment/tube for second observer into
the beam splitter,
➩Tighten the rotary ring (1).
1
Adjusting second-observer tube

15Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Display/change the footswitch type
43
2
You can choose between cross pedal or vertical pedal variants
for the footswitch.
Display the footswitch type:
➩Press the brightness regulating knob (3) for approx. 5 sec-
onds and release it.
The display bar (4) displays a small light for about three sec-
onds beneath the symbol (2) of the footswitch variant that is
currently selected
( or ).
Change the footswitch type:
➩While the footswitch type is displayed (see above), briefly
press the brightness regulating knob.
This switches to the other footswitch variant and the type
that is now enabled is displayed for about 3 seconds.
Switch the + and – movement directions of the XY-unit (figure 1)
-Y
+Y
-X
+X
+Y
-Y
+X
-X
11
3
1
2
Use the "XY-Reverse" button (1) on the footswitch to reverse the
+ and – movement directions for the XY-unit to – and +.
➩Press button (1), the small light (2) next to the zoom
indication (Figure 3) turns on.
➩Pressing button (1) again reverses the XY function back to
normal and the small light (2) turns off.
This requires that this function is enabled (see below).
Disable/enable the function of the XY-Reverse button (figures 1 and 2)
To prevent this function from being triggered inadvertently, the
"XY-Reverse" button can be disabled.
Disable the function of the "XY-Reverse" button:
➩Switch off the control unit/lamp housing at the power
switch (3).
➩Press and hold button (1) of the footswitch.
➩Switch on the control unit/lamp housing at the power switch
(3) and release button (1).
The "XY-Reverse" button is now disabled.
Enable the XY-Reverse button:
➩Repeat the above steps.
2
•XY-Reverse
3

16 Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
Caution 3
Surgical microscope can move without warning!
➩Always lock the footbrakes when you are not moving
the system.
Caution 2
The swinging arm can swing out uncontrollably!
➩Before transport, always set the instrument to the
transport configuration.
Transport position
Lock the swing arm:
➩Position the swing arm approximately horizontally.
➩Pull out the retaining pin (1).
➩Move the swing arm slightly up and down until the safety
hook (2) engages.
The swing arm is now locked.
➩Release the articulation brakes (3) and fold up the swing
arm.
➩Set the surgical microscope in its transport position and lock
the articulation brakes.
Caution 4
Feet in lightweight shoes could become trapped
beneath the casing of the base!
➩Always push the instrument to displace it; never pull it.
Transport the surgical microscope and secure it in its setup
position
➩Pull the power cable out of the socket.
➩Step on the footbrake release lever (5) to release the foot-
brakes (6).
➩Push the surgical microscope to its setup position using the
handle (4).
➩Lock the footbrakes at the setup position.
Transport the surgical microscope
(floor stand only)
56
6
33
4
2
1

17Leica M501 / Ref. 10 711 785 / Version D
Preparation for Operation
➩Using the handle, transport the surgical microscope to the OP
stage and position it as needed for the operation.
➩Apply footbrake.
➩
Connect the power cable to the control unit/lamp housing (13).
➩Connect the potential equalization to the control unit/lamp
housing (9).
➩Connect the foot or handswitch to the control unit/lamp
housing (11).
CAN-Bus interface (10).
Service interface socket (12).
Optional Leica Telescope Mount
Position the surgical microscope at the
OP stage Warning 5
Danger of fatal electrical shock!
➩Plug the surgical microscope only into a grounded
socket.
1211109
13
1
Caution 9
Danger of collision!
The surgical microscope can collide with nearby com-
ponents, light fixtures, or the ceiling.
➩Check the danger area before moving the swing arm.
➩Move the ceiling mount upwards carefully, and note
the position of the ceiling and light fixtures.
Warning 8
Risk of injury from the surgical microscope being
lowered!
➩Never position the surgical microscope over the
patient while it is moving downwards.
➩To avoid collisions, watch the surgical microscope
while it is being lowered.
Move the telescope arm to the desired height:
➩Press the button (1).
Raise the telescope arm
Lower the telescope arm
In continuous operation, the telescope may be oper-
ated for a maximum of 1 minute out of every 10.
After 2 minutes of uninterrupted operation, the built-in
temperature switch switches off the motor of the
Leica Telescope Mount.

18 Leica M501 / Ref. 10 711 785 / Version D
Setting the rest position on the Leica Telescope Mount
➩Swing the microscope away.
➩Remove the sterile components.
➩Switch off the power switch.
➩Move the swing arm all the way upwards.
➩Release the articulation brakes.
➩Press the raise lift arm (2) button.
The Leica Telescope Mount moves upwards.
2
Caution 9
Danger of collision!
The surgical microscope can collide with nearby com-
ponents, light fixtures, or the ceiling.
➩Check the danger area before moving the swing arm.
➩Move the ceiling mount upwards carefully, and note
the position of the ceiling and light fixtures.
Preparation for Operation
Attach the sterile components Set for one operation
• 3 handles
• 3 drive knobs
The handles and drive knobs can be sterilized via steam or gas.
➩Sterilize the handles and drive knobs.
➩Place the sterile drive knobs onto the binocular tube and the
turn knob for fading ambient lighting.
➩Place the sterile handles onto the optics carrier and the tilt
head.
11
11
➩Adjust the tilt of the footswitch. Screw the adjustable sup-
ports in or out.
➩Check that all accessories are correctly connected and
securely attached.
➩Release the articulation brakes and apply them lightly.
Make the joint easier to move:
➩Release the joint brake (1).
Make the joint stiffer:
➩Tighten the joint brake (1).
➩Extend the swinging arm.
➩Release the retaining hook on the swinging arm.
➩Check the weight setting on the swinging arm by raising and
lowering the microscope and correct as necessary.
Caution 7
Danger of infection!
➩Leave sufficient space around the stand to ensure
that the sterile drape does not come into contact with
non-sterile components.
Table of contents
Other Leica Medical Equipment manuals