4www.hillrom.com
7EN120114 Rev. 10 • 2020
Safety Instrucons
Before liing always ensure that:
• the li strap is not twisted or worn and can move in and out of the li unit freely
• the liing accessories are not damaged
• the sling is correctly and securely applied to the paent in order to prevent injuries from occurring
• the liing accessory is correctly applied to the li
• the liing accessories hang vercally and can move freely
• the sling bar latches are intact; missing or damaged latches must always be replaced
• the sling strap loops are correctly connected to the sling bar hooks when the sling´s strap is extended upward, but before
the paent is lied from the underlying surface.
Incorrect aachment of sling to slingbar may cause severe injury to the paent.
If the Likorall is installed in the S65 carriage with single hook, make sure it is resng securely at the boom of the hook
and is not lted.
Never leave a paent unaended during a liing situaon!
Only use the LikorallTM overhead li with a carriage, adapter, sling bar and other accessories approved by Hill-Rom.
Before using the li the rst me make sure that:
• the li is assembled in accordance with the assembly instrucon
• the liing accessories are properly aached to the li
•the baeries are charged for at least 8 hours
• you have read the instrucons for use for the li and liing accessories
• personnel using the li are informed of the correct use of the li and liing accessories
• the liing accessory is selected appropriately, in terms of type, size, material and design with regard to the paent’s needs.
No modicaon of this product is allowed.
Portable RF communicaons equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the li, including cables specied by the manufacturer. Otherwise,
degradaon of the performance of this equipment could result.
Use of the product adjacent to other equipment should be avoided because it could result in improper operaon, if such use
is necessary, observe and verify that the other equipment is operang normally.
Electromagnec disturbance, may aect the liing performance of the product. Modicaon using other parts than original
spare parts (cables etc.) may aect the electromagnec compability of the product.
Parcular care must be observed when using strong sources of electromagnec disturbance, such as diathermy, etc, so that dia-
thermy cables are not posioned on or near the li. If you have quesons, please consult the responsible assisve-device technician
or the supplier.
The li may not be used in areas where ammable mixtures may occur, for example in areas where ammable goods are stored.
Installaon of Likorall overhead li to carriages shall be made by personnel authorized by LikoTM in accordance with the
installaon instrucons and recommendaons for the current li system.
Intended use: The product is intended for use in following environments: Health care, Intensive care, Emergency ward,
Rehabilitaon, Habilitaon environment. This product is not intended to be used by the paent alone. Liing and transferring a
paent shall always be performed with the assistance of at least one caregiver. This product is used as a means to perform the li
but is not in contact with the
paent; therefore we do not go into the various paent condions in this instrucon for use. Contact
your Hill-Rom representave for support and advice.
Likorall 200 is tested by an accredited tesng instute.