Liko Likorall 200 User manual

Product Descripon
In this document, the person being lied is referred to as the paent, and the person helping them is referred to as the
caregiver.
Likorall overhead li is a staonary li unit oen called an
“overhead li”. Likorall overhead li is mounted in LikoTM
rail system, which is adapted to suit the room/rooms it
will be operated in. The rail system can be built straight,
with or without curves, as a traverse system and also as a
room-to-room system. LikoTM rail system is comprised of
several hundred dierent components, we select the specic
components needed to customize each system to suit the
room it will be mounted in. The rail system can either be xed
or freestanding, such as Liko FreeSpan and Liko FreeStand.
The system must be installed by authorized personnel and in
accordance with LikoTM installaon instrucons.
Likorall overhead li is intended for use in liing and transferring
paents, for example, from bed to a wheelchair, to or from
the oor, for visits to the toilet, for gait, standing and balance
training, when weighing the paent and when liing the
paent with a stretcher.
Likorall 200 overhead li is adapted to the LikoTM Quick-release
Hook System for safe and easy changing of liing accessories.
The LikoTM Room-to-Room system (R2R), enables the paent
to be moved between two rail systems in separate rooms,
without connecng rails and without making holes over doors.
A large range of accessories are available for Likorall overhead
li, including dierent sling models in a range of types and sizes.
Likorall™ 200
overhead li
Instrucons for Use
Likorall 200 overhead li Product No. 3121001
7EN120114 Rev. 10
ENGLISH 2020
IMPORTANT!
Liing and transferring a paent always involves a certain level of risk. Read the instrucons for use for both the paent
li and liing accessories before use. It is important to completely understand the contents of the instrucons for use.
The equipment should only be used by trained personnel. Ensure that the liing accessories are suitable for the li used.
Exercise care and cauon during use. As a caregiver, you are always responsible for the paent’s safety. You must be aware of
the paent’s ability to make it through the liing situaon. If something is unclear, contact the manufacturer or supplier.

2www.hillrom.com
7EN120114 Rev. 10 • 2020
Table of Contents
Symbol Descripon .......................................................................................................................... 3
Safety Instrucons ........................................................................................................................... 4
Denions ....................................................................................................................................... 5
Technical Data ................................................................................................................................. 5
Measurements ................................................................................................................................. 6
EMC Table ......................................................................................................................................... 6
Assembly .......................................................................................................................................... 9
Operaon ......................................................................................................................................... 9
Charging the Baeries .................................................................................................................. 12
Transferring from Room to Room ................................................................................................... 13
Maximum Load ............................................................................................................................... 14
Recommended Liing Accessories ................................................................................................. 14
Simple Troubleshoong .................................................................................................................. 16
Recycling Instrucons ..................................................................................................................... 17
Cleaning and Disinfecon .............................................................................................................. 18
Inspecon and Maintenance ......................................................................................................... 22

3www.hillrom.com
7EN120114 Rev. 10 • 2020
These symbols can be found in this document and/or on the product.
Symbol Descripon
For indoor use only.
The product has extra protecon against electric shock (Insulaon Class II).
Protecon level against electric shock Type B.
Warning. Used were extra care and aenon is needed.
Read instrucon for use before use.
This product complies with EC direcves.
IP N1 N 2 Protecon level against: ingress of solid objects (N1) and ingress of water (N2).
Manufacturer.
Date of manufacture.
Cauon! consult instrucons for use.
Baery.
,
Pb
,
All baeries in this product must be recycled separately.
- Pb underneath the symbol indicate baeries containing lead
- Single Black line underneath the symbol indicate this product have been placed on the market aer 2005.
UL Recognized Component Mark for Canada and the United States.
EFUP, Environmental Friendly Usage Period (years).
Environmentally-friendly product which can be recycled and reused.
The Australian Safety/EMC.
PSE Mark (Japan).
Product Idener.
Serial Number.
Medical Device.
Recyclable.
The safety and essenal performance of medical electrical equipment.
Proof of Product compliance to North American safety standards.
Non-ionizing electromagnec radiaon.
X%
Y% ≤Tmin
Duty cycle for non-connuous operaon.
The maximum acve operaon me X% of any given me unit, followed by a deacvaon me, Y%.
The acve operaon me shall not exceed the specied me in minutes, T.
(01) 0100887761997127
(11) Y Y MM D D
(21) 012345678910
GS1 Data Matrix Barcode that may contain following informaon
(01) Global Trade Item Number
(11) Producon Date
(21) Serial Number
Symbol Descripon

4www.hillrom.com
7EN120114 Rev. 10 • 2020
Safety Instrucons
Before liing always ensure that:
• the li strap is not twisted or worn and can move in and out of the li unit freely
• the liing accessories are not damaged
• the sling is correctly and securely applied to the paent in order to prevent injuries from occurring
• the liing accessory is correctly applied to the li
• the liing accessories hang vercally and can move freely
• the sling bar latches are intact; missing or damaged latches must always be replaced
• the sling strap loops are correctly connected to the sling bar hooks when the sling´s strap is extended upward, but before
the paent is lied from the underlying surface.
Incorrect aachment of sling to slingbar may cause severe injury to the paent.
If the Likorall is installed in the S65 carriage with single hook, make sure it is resng securely at the boom of the hook
and is not lted.
Never leave a paent unaended during a liing situaon!
Only use the LikorallTM overhead li with a carriage, adapter, sling bar and other accessories approved by Hill-Rom.
Before using the li the rst me make sure that:
• the li is assembled in accordance with the assembly instrucon
• the liing accessories are properly aached to the li
•the baeries are charged for at least 8 hours
• you have read the instrucons for use for the li and liing accessories
• personnel using the li are informed of the correct use of the li and liing accessories
• the liing accessory is selected appropriately, in terms of type, size, material and design with regard to the paent’s needs.
No modicaon of this product is allowed.
Portable RF communicaons equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the li, including cables specied by the manufacturer. Otherwise,
degradaon of the performance of this equipment could result.
Use of the product adjacent to other equipment should be avoided because it could result in improper operaon, if such use
is necessary, observe and verify that the other equipment is operang normally.
Electromagnec disturbance, may aect the liing performance of the product. Modicaon using other parts than original
spare parts (cables etc.) may aect the electromagnec compability of the product.
Parcular care must be observed when using strong sources of electromagnec disturbance, such as diathermy, etc, so that dia-
thermy cables are not posioned on or near the li. If you have quesons, please consult the responsible assisve-device technician
or the supplier.
The li may not be used in areas where ammable mixtures may occur, for example in areas where ammable goods are stored.
Installaon of Likorall overhead li to carriages shall be made by personnel authorized by LikoTM in accordance with the
installaon instrucons and recommendaons for the current li system.
Intended use: The product is intended for use in following environments: Health care, Intensive care, Emergency ward,
Rehabilitaon, Habilitaon environment. This product is not intended to be used by the paent alone. Liing and transferring a
paent shall always be performed with the assistance of at least one caregiver. This product is used as a means to perform the li
but is not in contact with the
paent; therefore we do not go into the various paent condions in this instrucon for use. Contact
your Hill-Rom representave for support and advice.
Likorall 200 is tested by an accredited tesng instute.

5www.hillrom.com
7EN120114 Rev. 10 • 2020
Technical Data
Maximum load: 200 kg (440 Ibs.)
Baeries: 2 x 12V 2,4-2,6 Ah.
Valve-regulated lead-acid gel-
type baeries. New baeries are
provided by the supplier.
Baery charger: CH01 FW7218M/24; 100-240 V AC,
50-60 Hz, max 500 mA
Liing speed: 50 mm/s (2 inch./s)
Liing interval: 2050 mm (80.7 inch.)
Electrical data: 24 V, 12 A
Li unit dimensions: 340x250x165 mm (LxWxH)
(13.4x9.8x6.5 inch. ,LxWxH)
Li unit weight: 12.5 kg (27.5 lbs.)
Emergency lowering
device: Electrical
Likorall overhead li is equipped with a SFS (Single Fault Safety) safety drum. This patented safety design provides protecon
against uncontrolled lowering.
Intermient power:
Protecon class li
motor:
Sound level:
Protecon class
hand control:
Operang forces
hand control:
Int. Op 10/90, acve operaon max
30 sec.
IP 33
51 dB
IP 54
3 N
Cable length
hand control:
870 mm (34 inch.)
Intended for indoor use.
Type B, in accordance with the electrical shock
protecon class.
Denions
1. Hand control with spring clamp
2. Emergency lowering device (electric)
3. Emergency stop string
4. Emergency stop
5. End cover
6. Charge indicator
7. Li unit
8. SSP Limit Switch
9. Li strap
10. Q-Link II (quick fastener)
11. Quick-release hook
12. Sling bar
13. Latches
14. Q-Link (quick fastener)
3
6
1
8
5
2
7
10
9
11
13
12
4
14

6www.hillrom.com
7EN120114 Rev. 10 • 2020
CSP (Central Suspension Point)
C
A
B
H** D*
L***
Measurements
* Minimum distance from ceiling to CSP at maximum liing height with the standard carriage.
** Built-in dimension: the distance between the aachment point for the li unit on the carriage and the CSP at maximum
liing height.
*** Liing interval: the distance between maximum liing height and minimum liing height measured in CSP.
Measurements in mm.
Measurements in inch.
ABCD* H** L***
165 340 250 304 221 2050
ABCD* H** L***
6.5 13.4 9.8 12.0 8.7 80.7
Overhead view
Lateral view
Guidance and manufacturer’s declaraon – electromagnec emissions
The product is intended for use in the electromagnec environment specied below. The customer or the user of this
product should assure that it is used in such an environment.
Emissions test Compliance Electromagnec environment - guidance
RF emissions
CISPR 11
Group 1 The product uses RF energy only for its internal
fuon. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The product is suitable for use in all establishments
other than domesc and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuaons/ icker emissions
IEC 61000-3-3
Complies
EMC Table
Ceiling

7www.hillrom.com
7EN120114 Rev. 10 • 2020
Guidance and manufacturer’s declaraon – electromagnec immunity
The product is intended for use in the electromagnec environment specied below. The customer or the user of this
product should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnec environment - guidance
Electrostac discharge
(ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Electrical fast transient
/ Burst
IEC 61000-4-4
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 2 kV for power supply
lines
n/a. for input/output lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge
IEC 61000-4-5
+/- 1 kV dierenal mode
+/- 2 kV common mode
+/- 1 kV dierenal mode
n/a. for common mode
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips, short
interrupons and
voltage variaons on
power supply input
lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT))
for 5 sec
Power frequency
(50/60 Hz) magnec
eld
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnec elds should be
at levels characterisc of a typical locaon in
a typical commercial or hospital environment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
10 V/m
80MHz to 2,5GHz
10 V/m
80MHz to 2,5GHz
NOTE UT is the a.c. mains voltage prior to applicaon of the test level.

8www.hillrom.com
7EN120114 Rev. 10 • 2020
Guidance and manufacturer’s declaraon – electromagnec immunity
The product is intended for use in the electromagnec environment specied below. The customer or the user of this
product should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnec environment - guidance
Portable and mobile RF communicaons equipment should
be used no closer to any part of the product, including cables,
than the recommended separaon distance calculated from the
equaon applicable to the frequency of the transmier.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Recommended separaon distance
d P=035,
Radiated RF
IEC 61000-4-3
10 V/m
80MHz to 2,5GHz
d P=029,
80 MHz to 800 MHz
d P=058,
800 MHz to 2,5 GHz
where P is the maximum output power rang of the transmier
in was (W) according to the transmier manufacturer and d is
the recommended separaon distance in meters (m).
Field strengths from xed RF transmiers, as determined by an
electromagnec site survey, a should be less than the compli-
ance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with
the following symbol.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situaons. Electromagnec propagaon is aected by absorpon and reected
from structures, objects and people.
a Field strengths from xed transmiers, such as base staons for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theorecally with accuracy. To assess the
electromagnec environment due to xed RF transmiers, an electromagnec site survey should be considered. If the mea-
sured eld strength in the locaon in which the product is used exceeds the applicable RF compliance level above, the product
should be observed to verify normal operaon. If abnormal performance is observed, addional measures may be necessary,
such as reorienng or relocang the product.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 10 V/m.

9www.hillrom.com
7EN120114 Rev. 10 • 2020
Assembly
• the SSP Limit Switch funcons properly
• the baery charger funcons properly and the
indicator lamps are illuminated during charging
• the baeries are charged for at least 8 hours before the
li is used for the rst me.
Aer assembly, ensure that:
• the li’s funcons correspond to the markings of the
hand control
• the electrical emergency lowering device
funcons properly
Liing accessory with quick-release hook,
Push down the red catch and connect the quick-release hook to the Q-Link II or Q-Link. Release
and check that the catch locks in order to prevent involuntary unhooking from the Q-Link II
or Q-Link. Read more about the Liko
TM
Quick-release Hook system on page 15.
Before liing check that the
Quick-release Hook is correctly
aached to the Q-Link II or
Q-Link, see illustraon above.
Quick
Release Hook
Q-Link II
or Q-Link
Manoeuvring
The liing moon in LikorallTM 200 overhead li is controlled
using an aached hand control. LikorallTM 200 overhead li
is manoeuvered by pressing lightly on the hand control
buons. The arrows indicate direcon. Moon ceases when
the buon is released.
Operaon
Up
Down
Hand control with spring clamp
When the li is not in use, the hand control can be aached
to the li strap using the spring clamp on the back side of
the hand control.

10 www.hillrom.com
7EN120114 Rev. 10 • 2020
Electrical Emergency lowering
In the event that the hand control or electronics
malfuncon, the li can be lowered by pulling on
the emergency lowering device.
In the event of emergency lowering, always ensure the
paent is lowered into a bed, wheelchair or other suitable
place.
Emergency stop
To acvate the emergency stop: pull the red emergency
stop string.
To reset the emergency stop: turn the red emergency stop
buon in the direcon shown by the arrows.
The red buon on the li unit end cover is intended
for use if an emergency situaon occurs. When the buon
is pressed, contact between the motor and power source is
broken, which stops the liing moon.
Installaon of Latches
Aer installaon, check that the
latch locks and runs freely in the
sling bar hook.
Li correctly!
Before each li, make sure that:
– the Sling loops at opposite sides of the Sling are at the same height
– all the Sling loops are fastened securely in to the Slingbar hooks
– the Slingbar is level during the li, see Figure 1.
If Slingbar is not level (see Figure 2) or if the sling loops are incorrectly at-
tached to the slingbar (see Figure 3) lower the user to a rm surface and ad-
just according to the Instrucons for use of Sling in use.
An improper li can be uncomfortable for the user and cause damage to
the li equipment! (see Figure 2 and gure 3).
Figure 1. Figure 2. Figure 3.

11 www.hillrom.com
7EN120114 Rev. 10 • 2020
SSP Limit Switch
The liing moon is stopped
by lightly pressing on the SSP
Limit Switch on the underside
of the li unit (see the
illustraon).
SSP Limit Switch acvated by the Q-Link or Q-Link II
When Likorall overhead li strap reaches its highest posion
and physical contact occurs between the SSP Limit Switch
and Q-Link or Q-Link II, the SSP Limit Switch is acvated. Its
funcon stops the liing moon electrically and protects the
li unit against mechanical load. The SSP Limit Switch also
provides squeeze protecon.
Adjustable fricon brake
The amount of drag on the li unit is adjusted
with the fricon brake on the carriage. Turn
the brake clockwise for increased resistance
and counter clockwise to reduce resistance.
The following carriages have a fricon brake:
prod. no. 3126011 and 3126015.
SSP Limit Switch acvated by the Li strap
It is important that the li strap’s liing moon is performed as vercally
as possible to ensure safe operaon. The SSP Limit Switch is intended
to stop the liing moon if the li strap is subjected to harmful strain,
for example if it is pulled sideways or folded over during the liing
moon. If the SSP Limit Switch is acvated so that the liing moon
stops, the li can once again be operated aer the li strap is straight
again (a short delay may occur when re-starng the liing moon is
normal in these cases).
Be aware that the li strap is kept straight and
stretched when it runs in and out of the li unit.
Never move the li by
pulling the hand control!
1
2
A: Losen the emergency lowering device strap.
B: Fasten the emergency lowering device strap.
A B

12 www.hillrom.com
7EN120114 Rev. 10 • 2020
Charging the Baeries
NOTE! When the li will not be in use for a long period of
me, place the hand control in the charger unit in order to
charge the baery.
In order to ensure maximum baery life, it is important to
charge the baeries regularly. We recommend that you charge
the baeries aer use or every night.
A full charge is achieved aer max. 8 hours. Fully charged
baeries will last for approx. 60 li cycles.
1. Check to ensure that the emergency stop buon is not
pressed in during charging.
2. Place the hand control in the charging staon
and connect the charger.
3. Connect the charger to a 100-240 V AC electric socket.
4. A LED on the hand control will illuminate and indicates
that the charger is connected to a power source.
5. Charging starts automacally and a yellow LED on the
li unit indicates that the baeries are charging.
6. When the baery is fully charged, the charger will shut
o automacally and the yellow LED turns o.
Charge indicator
Likorall overhead li has two indicators warning when the
baery has a low charge:
• Buzzer that sounds when liing
• LED that illuminates during liing
When either of these warning signals sound or illuminate,
the unit should be charged as soon as possible.

13 www.hillrom.com
7EN120114 Rev. 10 • 2020
1. Move Li 1 with the paent as close to the door way as
possible. Lower the li as far as possible, bearing in mind
the paent´s comfort.
2. Move Li 2 as close to the door opening as
possible. Lower the li strap from Li 2 a
sucient length and connect the Q-Link II or
Q-Link to the R2R sling bar.
Check that the catches on the R2R
double hook funcon properly.
NOTE! For transfer between mulple rooms a
Carriage adjustable could be used instead of a
li motor.
Li 2
Li 1
3. Raise Li 2. The paent is successively moved
to the next room and nally suspended in Li 2 only. When the
pressure is relieved from the li strap for Li 1, disconnect the
li strap from the R2R sling bar and the transfer can be carried
out in the next room.
NOTE! To free the Q-Link II or Q-Link from the R2R sling bar, it
may be necessary to let out addional strap from Li 1.
Li 1 Li 2
Mounng a Q-Link II or Q-Link to an R2R Sling Bar
The R2R sling bars with double hooks each t two Q-Links.
The two red catches keep the Q-Link II or Q-Link in place in
the R2R double hook before any load is applied to the li
strap. Open the red catch gently when placing a Q-Link II or
Q-Link in the R2R double hook.
Li 2
Li 1
The LikoTM Room-to-Room (R2R) system is an effective solution for safe transfer of patients between two
or more rooms. The R2R system is mounted without making openings in walls over doors and full isolation
is therefore retained between the rooms supported by the system.
The transfer is performed in a safe manner, with the aid of separate rail systems for each room. The LikoTM R2R system enables
linking together two Likorall overhead lis when transferring from room to room. The actual transfer operaon between two
rooms is performed with a comfortable transion for the paent from one Likorall overhead li to another.
Transferring from room to room
R2R, double hook
Q-Link
II or
Q-Link Catches

14 www.hillrom.com
7EN120114 Rev. 10 • 2020
Recommended Liing Accessories
Maximum Load
Dierent maximum loads may apply to dierent products on the assembled li system: rail system, li, sling bar, sling and
any other accessories used. For the assembled li system, including accessories, the maximum load is always the lowest
maximum load rang for any of the components. For example: a Likorall overhead li that is approved for 200 kg (440 lbs)
can be equipped with a sling bar that is approved for 300 kg (660 Ibs). In this case, the max load of 200 kg (440 lbs) applies to
the assembled li system. Study the markings on the li and liing accessories or contact your Hill-Rom representave if you
have any quesons.
Universal SlingBar 350 R2R Prod. No. 3156094
Max. load 300 kg (660 Ibs)
Universal SlingBar 450 R2R Prod. No. 3156095
Max. load 300 kg (660 Ibs)
Universal SlingBar 350 with Quick-release Hook Prod. No. 3156084
Max. load 300 kg (660 Ibs)
Universal SlingBar 450 with Quick-release Hook Prod. No. 3156085
Max. load 300 kg (660 Ibs)
Universal SlingBar 600 with Quick-release Hook Prod. No. 3156086
Max. load 300 kg (660 Ibs)
Universal SlingBar 670 Twin with Quick-release Hook Prod. No. 3156087
Max. load 300 kg (660 Ibs)
Universal SideBars 450 Prod. No. 3156079
including bag
Max. load 300 kg (660 Ibs)
Sling Cross-bar 450 with Quick-release Hook Prod. No. 3156022
Max. load 300 kg (660 Ibs)
Sling Cross-bar 670 with Quick-release Hook Prod. No. 3156019
Max. load 300 kg (660 Ibs)
Carriage Adjustable
Carriage, adjustable 300-500 mm (12-20 in), R2R Prod. No. 3121660
Carriage, adjustable 500-900 mm (20-35 in), R2R Prod. No. 3121661
Carriage, adjustable 900-1300 mm (35-51 in), R2R Prod. No. 3121662
Using other liing accessories than those recommended below may induce risk.
The LikoTM product range contains a wide assortment of sling bars, slings, stretchers, scales and other accessories to solve most
liing needs. Below is an overview of the liing accessories available for Likorall 200TM overhead li. Some accessories may not
be available for sale.
For addional guidance in selecng a sling, study the instrucons for use for the respecve sling models. There you will also
nd guidance for combining LikoTM sling bars with LikoTM slings.
Contact your Hill-Rom representave for advice and informaon on the LikoTM product range.

15 www.hillrom.com
7EN120114 Rev. 10 • 2020
Quick-release Hook
The LikoTM Quick-release Hooks form a system providing safe and easy
changing of liing accessories. The LikoTM Quick-release Hooks protect
against unintenonal detachment.Likorall 200 overhead li is used solely
with liing accessories equipped with Quick-release Hooks.
Quick-release Hook Universal ts Universal SlingBar 350, 450 and 600
(Prod. No. 3156074-3156076).
Quick-release Hook TDM ts Sling Cross-bar 450 and 670 (Prod. No.
3156021 and 3156018) and Universal TwinBar 670 (Prod. No. 3156077).
Contact your Hill-Rom representave for more informaon
Baery Charger Likorall 200 overhead li
CH01 EU 24V/0.5A Prod. No. 3126131
CH01 UK 24V/0.5A Prod. No. 3126132
CH01 US/CA 24V/0.5A Prod. No. 3126133
CH01 AU/NZ 24V/0.5A Prod. No. 3126134
Stretcher
Likorall can be used for horizontal liing with:
Liko™ FlexoStretch Prod .No. 3156057
Liko™ OctoStretch with Leveller Prod. No. 3156056
Liko™ Stretch Mod 600 IC, wide Prod. No. 3156065B
Contact your Hill-Rom representave for more informaon
Scale
For weighing persons in combinaon with the use of Likorall overhead li,
we recommend LikoScaleTM 350, Max 400 kg (880 lbs) Prod. No 3156228
This can easily be mounted with the LikoScaleTM adapter kit.
LikoScaleTM 350 is cered according to the European Direcve
NAWI 2014/31/EU (Non Automac Weighing Instruments)
For United States and Canada only:
LikoScaleTM 200, Max. 200 kg (440 Ibs.) Prod. No. 3156225
LikoScaleTM 400, Max. 400 kg (880 lbs.) Prod. No. 3156226
Contact your Hill-Rom representave for more informaon.
SlingBar Cover Paddy 30 Prod. No. 3607001
(ts Universal SlingBars 350, 450 and 600,
as well as SlingBar Slim 350)
Parking Panel 600, LR/MR Prod. no. 3126075
Parking Panel 1500, LR/MR Prod. no. 3126080
Can be completed with the following accessories:
Hook for SlingBar Prod. no. 3126070
Hook for Accessories Prod. no. 3126071
Quick Reference Guide (see respecve product)
Carriage Adapter Likorall for S65. Prod. no. 3126030
Quick-release Hook
Universal
Prod. No. 3156508
Quick-release Hook
TDM
Prod. No. 3156502
Likorall
Användaranvisning
Modell242,243 & 250
Svensk

16 www.hillrom.com
7EN120114 Rev. 10 • 2020
Simple Troubleshoong
The li doesn’t work.
A repeated signal can be heard from the li.
1. Make sure that the emergency stop buon has
not been acvated (shall not be pressed in).
2. Check that the handcontrol cable is connected
correctly.
3. Charge the baery.
4. If the problem persists, please contact Hill-Rom.
The li emits a repeated LED signal.
1. Charge the baery immediately.
2. If the problem persists, please contact Hill-Rom.
1. Charge the baery immediately.
2. If the problem persists, please contact Hill-Rom.
The li is stuck in the high posion.
The li does not achieve maximum liing
capacity.
1. Make sure that the emergency stop buon has
not been acvated (shall not be pressed in).
2. Check that the handcontrol cable is connected
correctly.
3. Use the selected electrical emergency lowering
device to lower the paent onto a rm surface.
4. Charge the baery.
5. If the problem persists, please contact Hill-Rom.
1. Charge the baery.
2. If the problem persists, please contact Hill-Rom.
Please contact Hill-Rom.
In event of unusual sounds or any
leakage from the li

17 www.hillrom.com
7EN120114 Rev. 10 • 2020
Metals
Pb-baeries
Waste of Electrical and
Electronic Equipment (WEEE)
Recycling Instrucons
Old baeries are to be deposited at the nearest recycling staon or given to personnel authorized by Hill-Rom.
Likorall overhead li comply with the Direcve 2012/19/EEC on waste electrical and electronic equipment.
Hill-Rom evaluates and provides guidance to its users on the safe handling and disposal of its devices to aid in the
prevenon of injury, including, but not limited to: cuts, punctures of the skin, abrasions, and any required cleaning
and disinfecon of the medical device aer use and prior to its disposal. Customers should adhere to all federal, state,
regional, and/or local laws and regulaons as it pertains to the safe disposal of medical devices and accessories.
If in doubt, the user of the device shall rst contact Hill-Rom Technical Support for guidance on safe disposal protocols.

18 www.hillrom.com
7EN120114 Rev. 10 • 2020
Cleaning and Disinfecon
These instrucons do not replace the facility’s own cleaning and disinfecon policies.
Warnings:
To help prevent injury and/or equipment damage, obey these warnings:
• Warning—The potenal for electrical shock exists with electrical equipment. Failure to follow facility protocol could cause
death or serious injury.
• Warning—Do not reuse wiping material for mulple steps or on mulple products.
• Warning—Harmful cleaning soluons may cause skin rash and/or irritaon upon contact.
Follow the manufacturer’s instrucons found on the product label and Safety Data Sheet (SDS).
• Warning—Li and move items correctly. Do not twist, and seek assistance when necessary.
• Warning—Fluid spills on to the li electronics could cause a hazard. If this happens do not put the li back into service
unl it is completely dry, tested, and found to be safe to operate.
Cauons:
To help prevent equipment damage, obey these cauons:
• Cauon—Do not steam clean or power wash the li. Pressure and excessive moisture can damage
the protecve surfaces of the li and its electrical components.
• Cauon—Do not use harsh cleansers/detergents, heavy duty grease removers, solvents such as toluene,
xylene, or acetone, and do not use scouring pads (you may use a so bristle brush).
• Cauon—Fully extend the li strap prior to the cleaning and disinfecon process.
Safety Recommendaons
• Wear protecve equipment according to manufacturer’s instrucon and per facility protocol
throughout the cleaning operaons, such as: gloves, eye protecon, apron, face mask and shoe covers.
• Unplug mains (AC power source) before cleaning and disinfecon.
• Never clean the li by pouring water on it, steam cleaning it, or by using a high-pressure jet.
• Refer to the recommendaons made by the cleaning and disinfecng product manufacturer.
Process Recommendaons:
For proper cleaning and disinfecon, sta members should be trained.
The trainer should carefully read the instrucons and follow them when the trainee is being trained.
The trainee should:
• Be given me to read the instrucons and to ask any quesons.
• Clean and disinfect the product while the trainer supervises. During, and/or aer this process,
the trainer should correct the trainee about any dierences from the instrucons for use.
The trainer should supervise the trainee unl the trainee can clean and disinfect the li as instructed.
Hill-Rom recommends to clean and disinfect the li between paent use and regularly during extended paent stays.
Some uids used in the hospital environment, such as iodophor and zinc oxide creams, can cause
permanent stains. Remove temporary stains by wiping vigorously with a lightly-dampened wiping
cloth.
Cleaning and Disinfecon Overview:
Cleaning and disinfecon are disnctly dierent processes. Cleaning is the physical removal of visible and
non-visible soil and contaminants. Disinfecon is intended to kill microorganisms.
When you perform the detailed cleaning steps, please note the following:
• Do not use harsh cleansers/detergents, heavy duty grease removers, solvents such as toluene, xylene,
or acetone, and do not use scouring pads (you may use a so bristle brush).
• A microber cloth is recommended as the wiping cloth.
• Always replace the wiping cloth when visibly soiled.
• Always replace the wiping cloth between steps (spot clean, clean, and disinfect)
• Always use Personal Protecve Equipment (PPE) such as gloves, eye proecon, apron, face mask,
and shoe covers, as recommended by the facility protocol and manufacturers instrucons

19 www.hillrom.com
7EN120114 Rev. 10 • 2020
Cleaning and Disinfecon Equipment:
• Protecve equipment (such as: gloves, eye protecon, apron, face mask and shoe covers) as recommended by the facility
protocol and manufacturers instrucons
• Disposable microber cloths recommended
• So bristle brush (Hill-Rom recommends a so bristle brush for cleaning.)
• Warm water
• To nd Cleaning / Disinfectants compable or not compable for use on LikoTM products, follow the
”Applicaon of commonly used Cleaning / Disinfectants on Liko products” in this document.
Prepare the Unit for Cleaning and Disinfecng:
1. Unplug mains (AC power source) before cleaning and disinfecon.
2. Fully extend the li strap by using the emergency lowering.
Step 1: Cleaning
1. Unplug the mains (AC power source) before cleaning and disinfecon.
2. As necessary, rst remove visible soil from the li with a cloth moistened with warm water and a neutral, approved cleaner/
disinfectant. See "Applicaon of commonly used Cleaning / Disinfectants on Liko products." Do not use a cloth that is dripping wet.
• A so bristle brush may be used for hard-to-clean areas to remove stains and resistant dirt and to loosen hardened soil.
• Use as many wiping cloths as needed to remove the soil. Replace cloth when soiled.
• Aer cleaning the li strap, make sure it is dry before you raise the sling bar.
3. Wipe down the enre li starng from the top down. Give special aenon to seams, cracks and other areas where soil
may accumulate. In parcular, pay special aenon to the following areas:
• Li strap
• Electrical emergency
lowering/raising
• Emergency stop
• Emegency stop cord
• Sling bar
• Hand control
Cleaner/Disinfecon:
NOTE:
It is important to remove all visible soil from all areas before connuing to remove non-visible soil.
With a new wiping cloth soaked in an approved cleaner/disinfectant, use rm pressure to wipe all surfaces of the li. Use a new
or clean wiping cloth as oen as necessary. Make sure the following items are cleaned:
• Hand Control • Sling (refer to the specic sling Instrucons for use and 7EN160884 Care and Maintenance
of
• Li Motor Liko Slings)
• Li Strap • Power cord
• Scale (if applicable)
• Connecon points
• Any part of the rail that may be soiled
Damaged items should be replaced!
Li strap
Sling bar
Emergency stop
Electrical emergency
lowering/raising
Hand control
Emergency stop cord

20 www.hillrom.com
7EN120114 Rev. 10 • 2020
Step 2: Disinfecon:
1. For the use of suitable disinfectants see "Applicaon of commonly used Cleaning / Disinfectants on Liko products” in this document.
2. Follow the manufacturer's instrucons.
3. Make sure all surfaces remain wet with the cleaner/disinfectant for the specied contact me. Rewet surfaces with a new
wiping cloth as necessary and per the manufacturer's instrucons. Addional disinfecon may need to be applied to maintain
the required "wet" me.
NOTE:
If bleach is used with another cleaner/disinfectant, use a new or clean cloth/wipe soaked in tap
water to remove any disinfectant residue prior to and aer the bleach applicaon.
The li may not be cleaned with CSI or equivalent.
The hand control may not be cleaned with Viraguard or equivalent.
The li strap may not be cleaned with Oxivir Tb, Dispatch, Chlor-Clean, Dismozon Pur or equivalent.
Other manuals for Likorall 200
5
This manual suits for next models
1
Table of contents
Other Liko Lifting System manuals

Liko
Liko M220 User manual

Liko
Liko 91 Manual

Liko
Liko Sabina User manual

Liko
Liko Multirall 200 Manual

Liko
Liko 3EN401101-04 Operating manual

Liko
Liko Viking XL 2040003 User manual

Liko
Liko Solo SupportVest Manual

Liko
Liko SlingGuard 450 User manual

Liko
Liko Likorall 200 User manual

Liko
Liko LikoGuard L User manual
Popular Lifting System manuals by other brands

morse
morse 400A-72SS-120 Operator's manual

HydroHoist
HydroHoist LevelLift L Series Operational instructions

Avanti
Avanti SWP L02 CE-AECO Service manual

U-Line
U-Line Presto Lifts P3 All Around Installation, operation and service manual

Vestil
Vestil EPC Series Use and maintenance manual

KITO
KITO ER2M owner's manual