
16
To enter or change any of the system’s calibration values,
turn all used Sections OFF or put the console in HOLD and
press and hold the CAL button until the “CAL” icon appears
(approximately one second).
NOTE: Calibration may be entered while moving, but
it is not recommended, for safety reasons, to attempt
calibration while the vehicle is moving.
The console will remain in calibration mode, with the RED
warning light illuminated until you exit calibration or turn
power OFF.
NOTE: The Console must be in Hold or all used Sections
OFF in order to toggle Calibrate Mode on, however, it can
be in Run, Hold or Sections ON or OFF to toggle Calibrate
Mode off.
Once in calibration mode, you may change any one, all, or
none of the values, in any order.*To select a calibration
position, simply turn the rotary selector to the desired
position. Calibration positions are identified by the White
labeling on each side of the rotary selector. All values are
entered and adjusted using the “+” and “-” buttons on the
front panel.
*Test speed must be last.
Hold the “CAL key again for 1 second to exit calibration.
“CAL” will disappear from the display.
NOTE: You MUST exit CAL to save changes.
Calibration (cont)
Entering Calibration Values
ROWS/SECTION: This displays the Rows Per Section from
0 to 255 rows for the Section selected. Unused Sections must
be programmed
to zero rows. The
DrillMaster can have
up to three Sections. The total width of
the Drill (sum of all Section widths) must
not exceed 65,535 inches or 65.535 meters
(Number Rows X Row Spacing).
To adjust the number of Rows of a
particular Section simply turn that Section on and
all others Off. The corresponding Section Icon
1 - 3 will turn on and the number of rows in that
section can be adjusted. Actually the first Section
turned on, from left to right, will be selected.
For example if Section 1 is off and Sections 2
and 3 are on, then Section 2 is selected since
it is the “first” Section on, and Section 2 Rows
can be adjusted. If no Sections are turned on, it
will display “no SECtn” (alternating “no” and
“Section”) to remind the user to turn a Section on,
and/or select RUN to enable the Sections.
ROWS/
SECTION
AREA TOTALS
(1) (2) (3)
SPEED CAL: This position is used to calibrate the speed
sensor for accurate speed and distance measurement. When
thispositionisselected,
the display will show
the SPEED CAL value.
The SPEED CAL value
is the number shown along with “CAL” on
the display. See details for determining
SPEED CAL on page 17.
SPEED
CAL
DISTANCE
CAL TEST: This is not a true “Calibrate Factor” but rather
a method of testing the Drill to determine if the correct
Meter Cal factor is being used and to test if a particular Row
is dispensing the correct amount of seed.
When RUN is selected it will use Test Speed
and AUTO control to dispense Sample Size
amount of Seed per row
and then automatically
stop.
The “Cal Test” is recommended for a quick spot check to
ensure proper operation. It can be used to:
• Test the accuracy of ‘Meter Cal’ (make sure it matches
what the Drill is actually planting).
• Test one Row, or many Rows, or the entire Drill at once.
• Test the consistency of the Rows (see variation from one
row to another).
• Find the minimum and maximum Row (lowest and
highest planting rate).
• Find the AVERAGE planting rate for the entire Drill.
This procedure is different than the “Fine Tuning” found in
the Meter Cal section in the following ways:
1. This procedure is a quick method to test the accuracy
of Meter Cal. It DOES NOT require weighing entire Drill.
Instead only a small sample from one Row is needed
and it can be performed in the field. But it can also test
multiple Rows at once if desired.
2. This procedure uses AUTO to make sure the planter is
running at the Target Speed. This is important because
the Seed Meter planting mechanism is speed sensitive
so it must be tested at the proper speed. In the Meter
Cal Fine Tuning procedure, the user can use AUTO
or MANUAL mode and use a lower or higher speed if
desired.
3. This procedure uses a fixed Sample Size (another cal
factor) making it more repeatable. In the Meter Cal Fine
Tuning procedure the sample size is up to the operator
and can be any size.
4. This procedure automatically stops when the sample
size is reached. In the Meter Cal Fine Tuning procedure
the operator must stop the test at the appropriate time.
5. This procedure will not affect or change any of the three
Weight Totals counters. In the Meter Cal Fine Tuning
procedure one of the three Weight Totals counters must
be cleared.
NOTE: See the Cal Test Fine Tuning procedure in Appendix C.
CAL
TEST
METER
RPM