Electrical Connections
Connection by a qualified electrician is recommended
All models covered by this manual may be ordered for single phase A.C. supply, which
may be Live to Neutral non-reversible, Live to Neutral reversible or Live to Live
Check the product rating label before connection. The supply voltage should agree with
the voltage on the label and the supply capacity should be sufficient for the current on the
label
The supply should be fused at the next size equal to, or higher than the current on the
label. This manual contains a table of the most common fuse ratings
When the mains cable is factory fitted, internal fuses are also fitted. It is essential that the
operator ensures that the product is correctly fused
Products with a factory fitted supply cable are designed to be wired directly to an isolator
or fitted with a line plug
Products without a factory fitted supply cable require a permanent connection to a fused
and isolated supply. The product's electrical access panel should be temporarily removed,
and connections made to the internal terminals
When connecting the product to the power supply, the plug or isolating switch should be
accessible, easy to remove, and within reach of the operator
(The supply MUST incorporate an earth (ground
Electrical Connection Details
Supply Terminal Label Cable Colour
Supply Types
Live - Neutral Reversible or Live-
Live
1-phase L1 Brown to live
to either power
conductor
(For USA 200-240V,
connect L1)
N / L2 Blue to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE Green/ Yellow to earth
(ground) to earth (ground)
It is highly recomended that any time working with the oven use high temp. gloves
Explosive Vapours
Unless your product includes the stoving and curing option, this model is not suitable
for drying or heat treatment applications where vapours are released that are
combustible or which can form explosive mixtures with air. Carbolite Gero
manufactures other products suitable for these applications
Atmospheres
When an optional gas inlet is fitted, there is a label near the inlet saying "INERT GAS
ONLY". In practice, inert or oxidising gases may be used, but not combustible or toxic
gases