3,
Hagavish
st.
,
Holon,Israel
58817
Te
l:
972
3
5595252,
Fax:
972
3
5594529
[email protected] ⑩
Indicator
Over-Heat Temperature. Lights on when the set temperature is
over 5℃, and turns off the heater.
⑪OUT 1 2
Control output lamp 1.
Light on when heater is activated
MANU Lamp flashes during Manual Mode operates
COMM Communication indicator
TUNE Lamp flashes during auto-tuning operates
Lock indicator
Setting the Temperature
Turn on the temperature controller/power switch. The temperature displays light up and after a
few seconds, the upper display shows the present value (PV) or chamber temperature. The lower
display shows the set value (SV) or set temperature.
Press to increase temperature gradually.
Press to decrease temperature gradually.
Setting temperature upon what you need.
Setting low Temperature
For a set point that requires the use of the refrigeration system, set the temperature and then flip
the refrigeration switch up to turn it ON. Because of a delay timer, the compressor will not turn
on to begin cooling until after several minutes. Similarly, when the switch is turned off, or if
the compressor shuts off in case of a power failure, the compressor will not restart for a few