NSK Varios 560 LUX User manual

OPERATION MANUAL
Please read this Operation Manual carefully
before use, and file for future reference.
Multi Function Ultrasonic Scaler
OM-E0248E 004
’10.03.01 S
The EU directive 93/42/EEC was applied in the design and production of this medical device.

1
English
Original Operation Manual
Thank you for purchasing the Varios Ultrasonic Scaler (Varios 560 LUX / Varios 560). This product
is used in dental office only. This device generates ultrasonic waves intended for use in dental
applications such as scaling, root canal treatment, periodontal and cavity preparation. Read this
operation manual carefully before use, and keep it within user's reach.
Classifications of equipment
· Type of protection against electric shock :
Class II equipment
· Degree of protection against electric shock :
Type BF applied part
· Method of sterilization or disinfection recommended by the manufacture :
See 8. Sterilization
· Degree of protection against ingress of water as detailed in the current edition of IEC 60529 :
Foot Control · · · IPX1 (Protected against vertically falling water drops)
· Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrousoxide :
EQUIPMENT not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrousoxide.
· Mode of operation :
Continuous operation
Cautions for handling and operation
Read these cautions carefully and use only as intended or instructed.
Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device.
Safety instructions are classified as follows in accordance with the seriousness of the risk.
· Use by medical professional, such as doctor or dental hygienist, is intended.
· Do not unplug the Power Cord with wet hands to avoid electric shock.
· Be sure to prevent water on the Control Unit, because it may result in short circuit and electric shock.
· Do not allow any impacton the Control Unit. Do not drop. This could result in electric shock.
·
Do not touch the handpiece backend, where electrical connections are attached to the cord. It might result in electric
shock.
· Do not disassemble or alter the handpiece/Control Unit.
· Keep away from patients with cardiac pacemakers.
·
Keep away from explosive substances and flammable materials. Do not use for patients anesthetized under laughter gas.
· Use the fuse of specified rating. ( FU201, FU202 120V : T500mA/250V, 230V : T250mA/250V)
· There is the judgment that applies this product to a patient in the user side.
· This product needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information.
· Portable and mobile RF communications equipment can affect this product.
WARNING
Degree of Risk
A hazard that could result in bodily injury or damage to the device if the safety instructions
are not followed.
A hazard that could result in light or moderate bodily injury or damage to the device if the
safety instructions are not followed.
General information needed to operate the device safely.
WARNING
CAUTION
NOTICE
Class

32
· Lightweight and compact device.
· Three types of Power Mode, "G-mode (General)", "E-mode (Endo)", and "P-mode (Perio)" according to each Tip, and also
allows adjusting freely within the recommended power.
· The Control Unit can simply be wiped off so spills are very easy to clean up, and the front control panel is extremely easy
to read.
· This product can be preset operation into the System Memory.
· Handy and small handpiece can lighten user’s fatigue.
· The Handpiece Ring Light is extremely bright for easy viewing of the operating area. (Varios 560 LUX)
· Light, durable, and slippy Handpiece Cord with unique Jacket.
· Tip Wrench has a torque-limiter mechanism to prevent Tips from over-tightening and lack of tightening.
· The Control Unit has provision to operate the handpiece with light or the one without.
· The handpiece is autoclavable at 135°C.
1. Features
2. Component Names
Fig. 1
Control Unit
Power Cord
Foot Control
Water Tube
Water Supply Connector (AC120V Only)
Tip (G1)
Tip (G4)
Tip (G6)
Tip Card
Tip Wrench
Tip Cover S
Tip Holder
Spanner Wrench (5 X 8) · · · · · 2 pcs
Handpiece (Optic)
Handpiece Cord (Optic)
Lamp
Handpiece (Non-Optic)
Handpiece Cord (Non-Optic)
O-ring · · · · · 2 pcs
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
NOTICE
· During vibration, the handpiece and the Handpiece Cord may affect computer and LAN cable. Noise could be heard
during operation near a radio receiver.
· Be sure to turn off the Power Switch after use. Remove the Power Plug and water inside of the Control Unit if not used
for a long time.
· Users are responsible for operational control, maintenance and inspection.
· When trouble is found, send to dealer.
· This product does not consider patient’s age (except infants), gender, weight or nationality.
· No special training is required for this device.
· Applied parts for patient and/or operator are/ is tip and handpiece.
· The use of ACCESSORIES, transducers and cables other than those specified, with the exception of transducers and
cables sold by the manufacturer of the product as replacement parts for internal components, may result in increased
EMISSIONS or decreased IMMUNITY of it.
· This product should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, it should be observed to verify normal operation in the configuration in which it will be used.
· When operating the handpiece always consider the safety of the patient.
· The handpiece is designed only for dental clinical use.
· Check the vibration and overheating outside the patient’s oral cavity before use. If any abnormalities are found, stop using
immediately and contact your dealer.
· Should this products function abnormally, cease operation immediately and returen it to the dealer for repair.
· Do not force or pull on Power Cord and/or Handpiece Cord. It could cause disconnection.
· Be sure to attach NSK genuine tips when using NSK Varios ultrasonic scaler (Varios 560 or Varios 560 LUX). The
problems such as damage, failure and accident of handepieces resulting from use of Non-NSK tips are not included in
the warranty. The following are the possible failure that could happen when using the Non-NSK tips.
· Vibration failure caused by using nonconformable screws
· Patients’ accidental ingestion of damaged tips
· Damage of thread ridge of handpiece
· Do not exceed Maximum Power for Tips. It could damage tooth structure and Tips.
· Check if the damages, bending and rust before use. Use new one.
· Check the Irrigation Hole of the Tip can irrigate adequately water before use. It may not irrigate the water adequately
because of the dirt.
· Check to see if dust does not stick on the screw of the tip before use. If not clean, tips will not generate enough vibration.
· Do not sharpen and/or bend the Tip. Tips may damage and not generate enough vibration during scaling.
· Do not mount or remove the Tip until the vibration completely stopped.
·
Tips are consumables. Check the wear condition with the tip card periodically, and replace it with a new one in good time.
· Always use with pouring enough water, or it may damage tooth plane and overheat the handpiece.
· Always keep Tip connection area clean. Entry of debris into this area could cause Tip run-out or weak holding.
· Do not hit metal or prosthetic crown etc., except for removing them. Tips could break and fall into mouth.
· Do not hit gingiva, mucosa and/or skin directly. It could cause damage and burn.
· Be sure to firmly mount the tip with provided wrench, or the tip will not generate enough vibration.
· Be sure to turn the power switch after mounting the handpiece.
· Do not drop nor give an excessive shock to the control unit.
· Do not autoclave any parts other than those that can be subjected to autoclave sterilization.
· Do not sterilize by ultraviolet light. Handpiece could discolor.
· Remove the handpiece after tip is taken off.
· If chemical, solvent or antiseptic solution is deposited on this product, immediately wipe it away. Discoloration or
deformation may occur if left as it is.
· Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
CAUTION
1
2
76 8 9
3
14 16
15
17 19
18
5
4
10 11 12 13
Varios 560 LUX
Varios 560

54
(1) Water System Setup
1) Connect the Water Filter Case side of the Water Tube deep
into the Water Connector on the Varios Control Unit. (Fig. 2)
2) Connect the Water Tube to the Water Outlet on the dental
unit.
3. Installation and Assembly
Fig. 2
CAUTION
· Insert the Water Tube deep into the connector on the Control Unit.
· If water has not been used at the water outlet of the dental unit for a long time, brownish water may come out
but wait until clean water comes out, and make connection.
NOTICE
Pushing the White Ring, (the Quick Connector Release
Ring) on the Water Connector, gently remove the tube.
(Fig. 3)
Fig. 3
(2) Foot Control Connection
Connect the Foot Control Plug to the Foot Control Connector at
the back of the Control Unit, take care in properly align all the
pins. Push the plug gently into the connector.
(3) Power Cord Connection
Insert the Power Cord into the Inlet at the back of the Control
Unit.
Fig. 4
Fig. 5
White RingWhite Ring
Push
Push
Foot Control Connector
Mark
Foot Control Connector
Mark
InletInlet
Power CordPower Cord
Foot Control PlugFoot Control Plug
Water Filter CaseWater Filter Case
Water Tube
Water Tube
Water ConnectorWater Connector
(4) Handpiece Cord Connection
Align the marks on the Control Unit and the Handpiece Cord
Plug. Gently push the Plug into the Connector.
CAUTION
(1) Water System Setup
Carefully check all Water Supply Connections prior to starting procedure. Open the dental unit's water valve.
Set water pressure between 0.1-0.5MPa (1.0-5.0kgf/cm2).
(2) Tip Connection
1) Mount the Tip by fastening it lightly by hand.
· Detach Tip before removing handpiece.
· Push handpiece against Handpiece Cord Plug inmost to connect.
· When removing the handpiece, grip the plug of the Handpiece Cord.
4. Operating Procedures
(5) Handpiece Connection
Align the Marks on the handpiece and the Handpiece Cord
Plug. Push straight inmost.
· When removing the handpiece, grip the Front and Rear of
the handpiece firmly. Pull to separate.
Do not touch the handpiece backend. (where
Electrical Connections are made to the cord.)
It might result in electrical shock.
WARNING
Fig. 6
Fig. 8
Fig. 7
* Picture shows Varios 560 LUX.
TipTip
Mark
Handpiece Cord Plug
Mark
Handpiece Cord Plug
Handpiece ConnectorHandpiece Connector
MarkMark
Handpiece Backend
Handpiece Backend
Handpiece
Handpiece Handpiece Cord PlugHandpiece Cord Plug

6 7
POWER RANGE
Low High
2) Insert the Tip through the hole of the Tip Wrench, align the
four cornered tip's base into the hole of the Tip Wrench. Turn
it clockwise until it clicks.
· To remove the Tip, turn counterclockwise with the Tip
Wrench.
Fig. 9
CAUTION
CAUTION
(3) Power On
Connect the Power Cord to the wall outlet. Turn on the Power
Switch of the Control Unit. Front Panel will light up.
(4) Power Setting
1) Select the Operating Mode with the Mode Selection Keys on the Front Panel. The light over the selected mode will
illuminate.
Do not use any wall outlets other than those of
AC120V±10% or AC230V±10%. There is a risk that
the Power Cord could be damaged.
CAUTION
· When mounting and/or removing the Tip, always
use groves.
· Tip, Tip Wrench, and Handpiece that have been
sterilized.
· Tip Wrench is consumable For reliable operation
replace annually.
Check the case the selected Tip came in for proper
power.
You can check Maximum Power on the Tip Case Label.
Check your model.
Fig. 10
Output for each mode
Set to
(General)
Mode
Selection
key
(Endo)
Set to
Set to
(Perio)
Tip WrenchTip Wrench
Tighten
Tighten
Tighten
Tighten
Loosen
Loosen Loosen
Loosen
Model
Maximum Power
CAUTION
Always use the water supply unless you are doing
procedure requiring Dry Mode(No Irrigation Mode).
If water supply is insufficient, handpiece will overheat
patient’s tooth surface can be injured.
2) Set the Power Level with the Power Level Key on the Front
Panel. The lamps in the Power Display Panel will indicate the
selected power level. Check to make sure the Power Level is
set in the appropriate range for the selected Tip.
Fig. 12
NOTICE
· Continuous pushing of the Power Level Key will cause the Power Level Display to climb.
· When only water is desired without oscillating the Tip, keep pressing “Decrease” key to extinguish Power Display
Pannel Lamp.
(5) Water Mode Selection
Push the Water Key to select whether or not to supply water.
(When the lamp illuminates, water supply becomes available.)
Fig. 13
CAUTION
(6) Operate Varios 560 LUX, Varios 560
· Depress the Foot Control. Tip will vibrate. (If the lamp stays illuminated by pushing the Water Key, water supply will
also start.)
· The Operating Lamp stays illuminated while the Tip is vibrating. (Fig.14) The handpiece ring light also illuminates.
(Varios 560 LUX)
· Adjust the Power Level to the proper range.
· Turn the Water Adjustment Knob counterclockwise gradually to increase the supply volume. (Fig.15)
· Always use the water supply unless you are doing procedure requiring Dry Mode. If water supply is insufficient,
handpiece will overheat patient´s tooth surface can be injured.
· Prior to starting the procedure, verify that the water spray is clean and at adequate volume.
Fig. 14 Fig. 15
Fig. 11
Front Panel
Power Switch
Front Panel
Power Switch
Water Adjustment KnobWater Adjustment Knob
Operating Lamp
Water Key
Lamp
Power Level Key
Decrease Increase
Power Level Display Panel

8 9
(7) When Treatment is Finished.
· Release the Foot Control and attach the handpiece to the
handpiece holder. (Fig. 16)
· Turn off the Power Switch.
· Close the dental unit's water valve.
· Remove the Tip after use. (Refer. 4. (2) Tip Connection)
When the Control Unit is shut down, the last mode settings in use are automatically retained in memory. When you
power on the system for the next procedure, the system will reactivate in the same configuration as when shut
down.
NOTICE
Fig. 16
Fig. 17
*Memory Function
The Control Unit can be preset for Water Mode, power
setting and operating mode. To set the desired preset
conditions into memory follow the procedure below.
1) Set the desired levels.
2) Push the Memory Key for about 2 seconds or more, the
memory light will illuminate and the control unit will
produce a beep. The memory is set. (Fig. 17)
*Safety Function
· When the operating mode is set to G and the power setting is set to Max the Control Unit will automatically reduce
power if operated at these settings for more than 10 minutes. (The output display moves 3 lamps from "Max" to "4".)
To reset the safety function....
Before a lapse of 10 minutes or when the power weakens after a lapse of 10 minutes, release the Foot Control. The
above safety function is reset. Depressing the Foot Control again repeats the same operation.
· If the Power Switch is turned on while the Foot Control is depressed accidentally ( by an object, etc.), beeping starts to
call attention.
To reset the alarm....
Release the Foot Control (remove the object, etc.).
To enter new settings....
After confirming the settings of Water Mode, Power Setting and Operating Mode, push the memory key for approx. 2
seconds. (The memory light will illuminate.) The old memory is cleared and new settings are programmed.
Memory Key
Memory Lamp
When the Tip is mounted to the handpiece, install the Tip Cover
S for safety.
· Hold the Tip Cover S as shown and insert it all the way to the
Tip-Handpiece joint.
· To remove, hold the Tip Cover S and the handpiece as shown
and pull straight out.
5. How to Use Tip Cover S
Fig. 18
Handpiece
Slit
WARNING
Carefully insert the tip into the Tip Cover S. Avoid injuring the fingers.
Do not use the Tip Cover S as a Tip change tool.
CAUTION
· Use the tip holder for tips removed from the handpiece.
· The tip holder is autoclavable and can hold up to 5 types at
once. To autoclave, tilt the tips in the direction of the arrow in
fig.19.
6. Tip Holder
· Autoclave sterilization is recommended.
· Sterilization is required first time you use and after each patient as noted below.
· The Tip, Handpiece, Tip Wrench, Tip Cover S, and Tip Holder can be autoclaved.
7. Sterilization
Do not autoclave any parts (the control unit, power cord, foot control, water tube, irrigation connector, handpiece
cord including the cover, lamp, O-ring, tip card, and spanner wrench (5 X 8) ) other than those that can be
subjected to autoclave sterilization.
CAUTION
Fig. 19
Slit
Tip
Tip Cover S
Tip-handpiece Joint
Handpiece
Handpiece HolderHandpiece Holder
Tip
Tip Holder

10 11
Autoclave Procedure
1) Remove the Tip after use. (Refer. 4. (2) Tip Connection)
2) Remove the handpiece from the Handpiece Cord. (Refer. 3. (5) Handpiece Connection)
3) Wipe off dirt and debris on the handpiece with an alcohol soaked cloth.
4) Insert into an autoclave pouch. Seal the pouch.
5) Autoclavable up to max. 135°C.
ex.) Autoclave for 20 min. at 121°C, or 15 min. at 132°C.
6) Keep the products in the autoclave pouch to keep it clean until you us.
Sterilization at 121°C for more than 15 minutes is recommended by EN13060 or EN ISO17665-1.
WARNING ON STERILIZATION
· This products can not be cleaned and disinfected with a Thermo-Disinfector.
· Do not sterilize by ultraviolet ray. The handpiece could discolor.
· If autoclaved with other instruments stained with chemical solution, it could strip the plating and make the
surface black.
· Since the bottom shelf of the sterilizer is close to the heat source, it can be too hot. Place instruments in the
middle or top shelves.
· Do not autoclave any parts (the Control Unit, Power Cord, Foot Control, Water Tube, Water Connector, Handpiece
Cord including the Cover, Lamp, O-ring, and Spanner Wrench (5 X 8) ). Other than those that can be subjected to
autoclave sterilization. Perform alcohol disinfection to the Control Unit, Power Cord, Foot Control, Handpiece Cord
including with reference after every patient.
· Do not sterilize diamond coated Scaler Tips for the reason of single use.
(1) How to use the Tip Card
1) Place the neck of the Tip in the cut out.
2) Check wear of the Tip.
3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we
recommend to replace a Tip when the Tip meets the yellow line (wear of 1mm) to guarantee safe and effective use.
8. Care and Maintenance
Green: No wear - Tip is OK
Tip replacement is not necessary.
Yellow: Wear of 1mm - Tip is showing some wear
Tip replacement is recommended.
Red: Wear of 2mm - Tip is badly worn
Tip replacement is necessary.
Fig. 20
OK
NO
Tip Card
Handpiece
Neck
*The Tip Card can be used to check the following tips : G1, G4,
G5, G6, G8, P1/P1D, P10, and P20
Efficiency
1mm 2mm
25%
50%
Decrease
Decrease
(2) Changing Water Filter
1) Close the water valve of the dental unit to which Varios
560 LUX or Varios 560 is connected. Mount two
Spanner Wrenches(5 X 8) as shown in Fig. 21, and
turn in the directions shown.
Fig. 21
2) When the Water Filter Case is separated, the Water Filter
can be removed as shown in Fig. 22. Replace with a new
one and reassemble the Filter in the reverse order.
Optional Water Filter : Order No.U387042
Fig. 22
When replacing the water filter, be careful about its
direction. (Point the O-ring on the water filter toward
the control unit side.) If it is assembled in the opposite
direction, the effect of the water filter will be lost.
CAUTION
CAUTION
NOTICE
Tips are consumables. The efficiency of dental scaling
decreases approximately 25% when the top of the Tip
wears 1 mm and approximately 50% when it wears 2
mm. In addition, the vibration condition changes owing
to the wear, which may damage a patient’s tooth
surface. Check the Tip wear condition with the Tip Card
periodically, and replace the Tip with a new one in good
time.
When Water Tube is getting twisted by this, relieve the twist by turning it, as its end on the Control Unit side is free
to turn.
CAUTION
(3) Cleaning of Donut-Shape Light (Varios 560 LUX)
Wipe the debris off the End Face Optic Fibers at the
handpiece with alcohol soaked cotton swab.
Fig. 23
Do not use any sharp pointed tools to clean the End
Face Optic Fiber. In case the light becomes dim,
contact your dealer.
Water Filter CaseWater Filter Case
Water FilterWater FilterO-ringO-ring
Control Unit SideControl Unit Side
Cotton SwabCotton Swab
End Face Optic FiberEnd Face Optic Fiber

12 13
(4) Changing Lamp (Varios 560 LUX)
1) Refer to 3. (5) Handpiece Connection. Disassemble the handpiece from the Handpiece Code. Remove the cover.
Use a precision screw driver and push lamp out.
2) Align the Lamp Pins of a new lamp with the holes, and push the lamp into.
* Optional Lamp : VA Bulb Set (A Set consist of 3 pieces) Order No. Y900107
Fig. 24
Cover Lamp
Socket
View from A
Lamp Pin Lamp
A
3) Align the Mark on the cover with the same on the Handpiece
Cord. Push the cover to the cord unit it clicks. Assemble the
handpiece and the Handpiece Cord. (Refer. 3. (5) Handpiece
Connection)
Fig. 25
Mark
Cover Handpiece Cord
Mark
CAUTION
· Make sure that the bulb is sufficient cooled, wheen it is
replaced.
· Do not touch glass part of new lamp.
· Care needs to be execised when mounting a new lamp
to avoid the O-ring off from the groove or being twisted.
(5) Maintenance of the Control Unit etc
If dirt sticks on the Control Unit, Foot Control, or Handpiece Cord, remove the Power Cord first. Then, soak a clean
cloth in water and squeeze cloth tightly to drain water. Use the cloth to wipe away dirt. After that, wipe off with an
alcohol soaked clean cloth.
When trouble is found, check the following items again before asking repairs. If none of these are applicable or the trouble
is not remedied even after action has been taken, a failure of this product is suspected. Contact your dealer.
9. Troubleshooting
No / Poor
vibration.
The tip is bent
or broken.
The tip walks
out.
Noise from the
handpiece.
The handpiece
is overheating.
The Power Lamp
does not light, even
if the Power Switch
is ON.
The tip does not
generate vibration,
in spite of
depressing the Foot
Control.
The Power Cord or the Jack is
disconnected.
The internal Fuse has burned out.
The tip is not tightened firmly.
Worn Tip.
Output has not been correctly
adjusted for the Tip.
The Foot Control is disconnected.
Failure of vibrator in the handpiece.
Failure of the Foot Control.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the
handpiece or the Control Unit.
Output has not been properly
adjusted for the Tip.
The Tip is not tightened firmly.
Failure of vibration in the
handpiece or the Control Unit.
Correctly insert the Power Cord or the Jack.
Contact your dealer.
Tighten the Tip until the Tip Wrench rotates
freely with clicks.
Replace the Tip.
Adjust the Power on the Tip Case. Do not
exceed.
Connect the Foot Control correctly.
Contact your dealer.
Contact your dealer.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the tip until the Tip Wrench rotates
freely with clicks.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the Tip until the Tip Wrench rotates
freely with clicks.
Contact your dealer.
Adjust the power on the Tip Case. Do not
exceed.
Tighten the tip until the Tip Wrench rotates
freely with clicks.
Contact your dealer.
Problem Probable Cause Cause Solution

14 15
Manufacturer warrants its products to the original purchaser against defects in material and workmanship under normal
practices of installation, use and servicing. Such expendable items as O-rings and lamps are not covered by this warranty.
10. Warranty
Specifications
TUV Rhineland of North America is a Nationally Recognized Testing Laboratory (NRTL) in the United States and is
accredited by the Standards Council of Canada to certify electro-medical products with Canadian National Standards.
Follow the waste of electric and electronic equipment (WEEE) Directive (2002/96/EC) to dispose of the product and
accessories.
Consult operation instructions.
This conforms to CE European Directive of “Medical equipment directive 93/42/EEC.”
Symbols
Protected against vertically falling water drops.
Manufacturer.
Type BF applied part. Authorised representative in the European community.
Class II equipment. Autoclavable up to Max.135°C. *for detail see Sterilization.
Guidance and manufacturer's declaration - electromagnetic emissions
The Varios560/ Varios560LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios560 / Varios560LUX
should assure that is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR11
RF emmissions
CISPR11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/flicker emissions
IEC61000-3-3
Group 1
class B
class A
Complies
The Varios560 / Varios560LUX uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The Varios560 / Varios560LUX is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that
supply network that supplies buildings used for domestic purposes.
No / Poor
water.
Water leakage.
Donut light
does not
illuminate.
(Varios 560
LUX)
Loss of power
output without
operation.
The operating
lamp flashes.
Beeping during
operation,
automatic mode
change, or
vibration stop.
Start Beeping.
The water does not
reach to the Control
Unit.
Check to see if
water reaches the
Control Unit.
Water is leaking from
the joint between the
Irrigation Tube and the
Irrigation Connector.
Water is leaking from
the Control Unit.
Tip oscillates, but
donut light turns on
and off.
Tip oscillates, but
light does not turn
on.
Lamp is correctly
and securely
mounted in the
socket, but light
does not turn on.
Power output is set
at maximum at G
mode.
·
Continuous
operation for a long
time at high mode
(G: General).
·
Operation in a
high-temperature
room.
Beeping while
power on.
Beeping while stopping
vibration of Tips.
The Water Adjustment Dial is
closed.
Disconnected irrigation supply at low
volume range. (less than 10cc/min.)
The water Filter is clogged.
The Irrigation Tube is not
connected correctly.
The water circuitry in the Control
Unit is damaged.
The handpiece is not connected
into the Handpiece Cord correctly.
The Lamp Pins are not correctly
engaged in the socket.
Has the lamp burned out.
Disconnection in the handpiece, or
failure in the Control Unit.
Safety function is activated.
Safety function is activated.
Depress Foot Control.
Abnormal heating of the Control
Unit.
Check the water circuitry and supply to the
Control Unit. Water Pressure : 0.1-0.5MPa
(1.0-5.0kgf/cm2)
Turn the Water Adjustment Dial and adjust to
the appropriate volume.
No problem. Turn the Water Adjustment Dial
and increase the irrigation volume.
Replace with new Water Filter. (Refer. 8. (2)
Changing Water Filter)
Firmly insert the Irrigation Tube into the
Irrigation Connector inmost.
Contact your dealer.
Firmly insert the handpiece into the
Handpiece Cord inmost.
Mount the lamp correctly and securely.
Replace with new lamp. (Refer. 8. (4)
Changing Lamp (Varios 560 LUX))
Contact your dealer.
Power output will weaken automatically while
continuous operation is over 10min at the
setting of Maximum power at G mode.
Releasing the foot from the Foot Control will
reset the safety function. Stepping on the Foot
Control then will resume the previous function.
(Refer. 4. * Safety Function)
Turn off the Power Switch and leave until
Control Unit becomes cool. Turn on the Power
Switch after it cools off completely.
Turn the Power Switch on after release of the
Foot Control.
Stop the operation and leave until Control Unit
becomes cool.
Problem Probable Cause Cause Solution
Type
Power Source
Vibration Frequency
Maximum Output
Power Consumption
Water Pressure
Lighting
NE171
AC120V10% 50/60Hz
AC230V10% 50/60Hz
28-32kHz
8W
Varios 560 LUX : Max.35VA
Varios 560 : Max.30VA
0.1-0.5MPa (1.0-5.0kgf/cm2)
Varios 560 LUX : Yes
Varios 560 : No
Dimensions
Use Environment
Store Environment
Transport
Environment
W109 x D194 x H95mm
(Control Unit)
Temperature 0-40˚C
(The liquid must not freeze up)
Humidity 30-75 %
Atmospheric pressure
700-1060 hPa
Temperature -10-60˚C
Humidity 10-85 %
Atmospheric pressure
500-1060 hPa

16 17
Guidance and manufacturer's declaration - electromagnetic immunity
The Varios560 / Varios560LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios560 / Varios560LUX
should assure that it is used in such an environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Immunity test IEC60601 test level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
IEC61000-4-2
Electrical fast transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut) for 0.5 cycle
40% Ut
(60% dip in Ut) for 5 cycles
70% Ut
(30% dip in Ut) for 25 cycles
<5% Ut
(>95% dip in Ut) for 5 secs
3 A/m
Compliance level
±6kV contact
±8kV air
±2kV for power supply lines
±1kV for input/output
±1kV line(s) to line(s)
±2kV line(s) to earth
<5% Ut
(>95% dip in Ut) for 0.5 cycle
40% Ut
(60% dip in Ut) for 5 cycles
70% Ut
(30% dip in Ut) for 25 cycles
<5% Ut
(>95% dip in Ut) for 5 sec
3 A/m
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the
Varios560 / Varios560LUX requires continued operation
during power mains interruptions, it is recommended that the
Varios560 / Varios560LUX be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Guidance and manufacturer's declaration - electromagnetic immunity
The Varios560 / Varios560LUX is intended for use in the electromagnetic environment specified below. The customer or the user of the Varios560 / Varios560LUX
should assure that it is used in such an environment.
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobiles radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the Varios560 / Varios560LUX is used exceeds the
applicable RF compliance level above, the Varios560 / Varios560LUX should be observed to verity normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Varios560 / Varios560LUX.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.
Immunity test IEC60601 test level Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3Vrms
150 kHz to 80MHz
3V/m
80MHz to 2.5 GHz
Compliance level
3Vrms
3V/m
Portable and mobile RF communications equipment should be used no
closer to any part of the Varios560 / Varios560LUX, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P 80MHz to 800MHz
d = 2.3 P 800MHz to 2.5GHz
Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters as determined by an
electromagnetic site survey, should be less than the compliance level
in each frequency range.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Marking on the outside of Equipment or Equipment parts that include RF transmitters or that apply RF electromagnetic energy for diagnosis or treatment.
150kHz to 80MHz
d =1.2 P
Recommended separation distances between portable and mobile RF communications equipment and the Varios560 / Varios560LUX
Complies with
RF emissions, CISPR11,
Harmonic emissions,
Voltage fluctuations/ flicker emission,
Electrostatic discharge (ESD)
Electric fast transient / burst
Surge
Voltage dips, short interruptions and voltage variations on power
supply input lines
Power frequency(50/60Hz) magnetic field
Conducted RF
Radiated RF
Maximum length
2.0 m
2.5 m
Shield
Unshielded
Unshielded
Cables and accessories
Handpiece cord
Foot Controler
The Varios560 / Varios560LUX is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the Varios560 / Varios560LUX can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Varios560 / Varios560LUX as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Rated maximum output power of transmitter
W
0.01
800MHz to 2.5GHz
d =2.3 P
80MHz to 800MHz
d =1.2 P
0.12 0.230.12
0.1 0.38 0.730.38
1 1.2 2.31.2
10 3.8 7.33.8
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
100 12 2312
Separation distance according to frequency of transmitter
m
Class B/ Group 1
IEC61000-3-2
IEC61000-3-3
IEC61000-4-2
IEC61000-4-4
IEC61000-4-5
IEC61000-4-11
IEC61000-4-8
IEC61000-4-6
IEC61000-4-3
This manual suits for next models
1
Table of contents
Other NSK Medical Equipment manuals
Popular Medical Equipment manuals by other brands

LEADING CHAIR
LEADING CHAIR E-400 Instructions for use

Nellcor
Nellcor Oximax NPB-40 Operator's manual

Orliman
Orliman GONARTEC PLUS OCR300D INSTRUCTIONS FOR USE AND PRESERVATION

Flight Medical Innovations
Flight Medical Innovations Flight 60 iO2 Quick reference guide

AGFA
AGFA DR 100e Getting started

FISIOTECH
FISIOTECH ERMES Use and maintenance manual