NSK Primado2 User manual


English
English
1
This is a Reprocessing Manual for our total surgical system Primado2.
We recommend that prior to use, you carefully read this document to follow the correct reprocessing procedure so that you can continue
using the Attachments in the future.
In addition, keep this Manual in a place where a user can refer to it at any given time.
●The reprocessing instructions shown in this Manual are created based on the requirements of ISO 17664.
●NSK has verified the effectiveness of reprocessing with the procedure written in this Manual.
●In this Manual Bur, Drill, Blade and Rasp are collectively called Cutting Accessory(ies).
●For Precautions for Handling and Operation, Product Description, Operation, Maintenance, Troubleshooting and After-sales Service, refer to
each Operation Manual (see Table-1).
Table-1
Operation Manual Title Operation Manual No.
Primado2 OPERATION MANUAL Control Unit and Foot Controls OM-SE0021E
Primado2 OPERATION MANUAL Motors OM-SE0023E
Primado2 OPERATION MANUAL Attachments OM-SH0913MA
Cleaning Tool for Primado2 Attachment OM-SZ0904MA
Contents
1. Precautions for Handling and Operation ............................ 2
2. Moving Parts of Each Product ........................................... 3
3. Preparation ..................................................................... 5
3–1 Treatment at the Operating Site ...................................... 5
3–2 Reprocessing of Each Product ........................................ 6
4. Cleaning / Checking the Control Unit ................................ 7
4–1 Wipe Disinfecting .......................................................... 7
4–2 Checking ...................................................................... 7
5. Cleaning / Checking the Foot Control ................................ 7
5–1 Cleaning/Wipe Disinfecting ............................................. 7
5–2 Checking ...................................................................... 7
6. Cleaning / Checking the Motor ......................................... 8
6–1 Cleaning ....................................................................... 8
6–2 Checking ...................................................................... 9
7. Cleaning / Lubrication / Checking of the Attachment, Slim
Attachment Hub and Slim Tube ....................................... 10
7–1 Cleaning ..................................................................... 10
7–2 Lubrication ................................................................. 18
7–3 Checking .................................................................... 19
8. Cleaning / Checking the Hood, Beak, Duraguard,
Rotatable Duraguard, Wrench and Adaptor ...................... 20
8–1 Cleaning ..................................................................... 20
8–2 Checking .................................................................... 22
9. Cleaning / Checking the Irrigation Nozzle ........................ 23
9–1 Cleaning ..................................................................... 23
9–2 Checking .................................................................... 28
10. Packaging / Sterilization ................................................ 29
10–1 Packaging .................................................................. 29
10–2 Sterilization ................................................................. 29
11. Storage ......................................................................... 30
11–1 Storage ...................................................................... 30
11–2 Carrying to the Operating Site ....................................... 30
12. Disposing of Product ...................................................... 30
13. Symbols ........................................................................ 30

2
1 Precautions for Handling and Operation
Please read these precautions carefully and use only as intended or instructed.
Safety instructions are intended to avoid potential hazards that could result in personal injury or damage to the device. Safety instructions
are classified as follows in accordance with the seriousness of the risk.
Class Degree of Risk
WARNING Hazard that could result in serious injury or damage to the device if the safety instructions are not correctly
followed.
CAUTION Hazard that could result in light or moderate injury or damage to the device if the safety instructions are not
correctly followed.
NOTICE General product specification information highlighted to avoid product malfunction and performance reduction.
WARNING
●If this product is used for a patient with Creutzfeldt-Jakob Disease (CJD) or with suspected CJD, reprocess the product appropriately
in accordance with the regulations or guidelines for CJD of each country.
●Follow the regulations or guidelines of each country.
●Reprocess this product after every surgical operation.
●Do not directly spray detergent/liquid on the LCD Touch Panel of the Control Unit when cleaning the Control Unit. It may cause failure.
●The Irrigation Tube and Cutting Accessory are for single use only.
Do not reuse or resterilize.
●The Cutting Accessory is for single use only. DO NOT reuse or reprocess it. If you reuse or reprocess it, the following risks may occur:
– Infection.
– Personal injury due to damaged Cutting Accessory.
– System defect due to damaged/deformed Cutting Accessory.
●Thoroughly clean the Attachment before autoclave sterilization. Autoclaving the Attachment with blood remaining on it may lead to
sterilization failure, resulting in infection to the patient and/or user or malfunction of the Attachment due to blood coagulation.
●Immediately after use, the Attachment should be cleaned, lubricated and sterilized. Remaining blood in the Attachment may cause rust
due to blood coagulation. Failure to properly maintain the Attachment may lead to overheating, causing thermal injuries or product
failure.
●Do not lubricate the electric Motor. It may result in an abnormal heating and failure.
CAUTION
●Wear Personal Protective Equipment (PPE) to prevent infection.
●NSK has verified the effectiveness of reprocessing using the dedicated cleaning tools.
●Be sure to use the dedicated cleaning tools.
●DO NOT use a lint cloth or metal brush for cleaning.
●When cleaning, use disinfecting ethanol or ethanol with isopropanol added as an alcohol disinfectant (e.g. Meliseptol® rapid) to wipe
down the surface of the product.
●DO NOT use a chlorinated cleaner, benzene or thinner for cleaning to prevent damage to the product.
●Do not soak the Control Unit, Foot Control or Motor in water for cleaning. It may cause failure.
●Do not clean the Control Unit or Foot Control under running water. It may cause failure.
●DO NOT clean the product with an ultrasonic cleaning apparatus to prevent damage to the product.
●The Wire Pin Driver includes aluminum, therefore take special precaution to always use a PH neutral detergent compatible with
aluminum, when washing it.
●When using detergents other than the ones we used for validation, it is necessary to verify the effectiveness based on the information
provided by the manufacturer of the detergent and the information obtained from testing.
●Prepare the detergent according to the instructions of the detergent manufacturer when using the detergent.
●Refer to the operation manual of the washer-disinfector before operating it.
●Be sure to use a disinfector compliant with ISO15883.
●Dispose of the Irrigation Tube and Cutting Accessory after use as medical waste
(refer to OM-SE0021E Primado2 OPERATION MANUAL Control Unit and Foot Controls).

3
English
2 Moving Parts of Each Product
Move the moving parts of each product while cleaning.
Be sure to check the moving parts of the product to be cleaned before you start reprocessing (see below).
Motor
Order Code Moving Part Fig.
P200-HMH-HS
P200-SMH-HS
P200-BMH-HS
ON/OFF Switch
Adjustable Lever
P200-WPD Trigger (Upper)
Trigger (Lower)
Mounting/Removal Button
Attachment / Slim Attachment Hub / Slim Tube
Order Code Moving Part Fig.
P200-2SMS, P200-2SES, P200-2SS
P200-2SM, P200-2SL
P200-2AMS, P200-2AES, P200-2AS
P200-2AM, P200-2AL
P200-1A135, P200-1A155, P200-1A155-C
P200-1A175, P200-1A175-C
P200-RA330, P200-RA330-L
P200-2SD, P200-3MC
P200-3GAS, P200-3GAM, P200-3GAL
P300-2S10, P300-2S20, P300-2S40
P300-2S60, P300-2S80
P300-2A10, P300-2A20, P300-2A40
P300-2A60, P300-2A80
PD-1A135, PD-1A155, PD-1A175
PD-2AES, PD-2AS, PD-2AM, PD-2AL
PD-2SES, PD-2SS, PD-2SM
PD-2SD, PD-2SDM, PD-2AD, PD-2ADM
PD-RA330, PD-RA330-L
Twist Collet
P200-CRA, PD-CRA
P300-1AHS, P300-1AHA
P200-CA-4, PD-CA-16, PD-CA-64 Latch

4
Moving Parts of Each Product
Order Code Moving Part Fig.
P200-PER, PD-PER
PD-DA-A
PD-RA-M, PD-RA-H, PD-RA-T, PD-RA-A
Removal Lever
P200-CRD-P, P200-CRD-M, P200-CRD-L
P200-RCRD-P, P200-RCRD-M, P200-RCRD-L
PD-CRD-P, PD-CRD-M, PD-CRD-L
Removal Ring
P200-SAG Fitting/Removal Button
Removal Ring
P200-REC Fitting/Removal Button
Removal Ring
P200-OSC Removal Lever
Removal Ring
PD-DA-W, PD-DA-P Attachment Lever
PD-DA-J4, PD-DA-J6 Jacobs Keyed Drill Chuck

5
English
3 Preparation
3–1 Treatment at the Operating Site
1 Remove visible contaminants such as blood using a lint-free cloth moistened with distilled water, deionized water or mineral-free water
at the operating site.
2 Disconnect the Motor Cord, Foot Control Cord and Power Cord from the Control Unit.
3 Remove the Cutting Accessory, Motor, Hood, Beak, Duraguard, Rotatable Duraguard, Irrigation Nozzle and other connections from the
Attachment.
4 Remove the Slim Tube from the Slim Attachment Hub (for P300-1AHS/P300-1AHA only).
For removal procedure, refer to the Operation Manual of each product.

6
Preparation
3–2 Reprocessing of Each Product
Check the reprocessing process of each product before performing reprocessing.
Product Order Code
Reprocessing Process / Section No.
Cleaning Lubrication Checking Packaging
Sterilization Storage
Control Unit P200-CU-120, P200-CU-230 4-1
-
4-2 -
11
Foot Control FC-73, FC-74 5-1 5-2
Motor
P200-HMH, P200-HMH-HS
P200-SMH-S, P200-SMH, P200-SMH-HS
P200-BMH, P200-BMH-HS
P200-WPD
Manual:
6-1-1
Automated:
6-1-2
6-2
Packaging:
10-1
Sterilization:
10-2
Attachment
P200-1A135, P200-1A155, P200-1A155-C
P200-1A175, P200-1A175-C
P200-2AMS, P200-2AES, P200-2AS
P200-2AM, P200-2AL
P200-2SMS, P200-2SES, P200-2SS
P200-2SM, P200-2SL
P200-2SD
P200-CRA
P200-3MC
P200-PER
P200-RA330, P200-RA330-L
P200-CA-4
P200-REC, P200-SAG, P200-OSC
P300-2S10, P300-2S20, P300-2S40
P300-2S60, P300-2S80
P300-2A10, P300-2A20, P300-2A40
P300-2A60, P300-2A80
PD-1A135, PD-1A155, PD-1A175
PD-2AES, PD-2AS, PD-2AM, PD-2AL
PD-2SES, PD-2SS, PD-2SM, PD-2SD
PD-2SDM
PD-2AD, PD-2ADM
PD-CRA
PD-PER
P200-3GAS, P200-3GAM, P200-3GAL
PD-RA330, PD-RA330-L
PD-CA-16
PD-CA-64
PD-DA-J4, PD-DA-J6, PD-DA-A
PD-DA-K, PD-DA-W, PD-DA-P
PD-SA-S, PD-SA-R, PD-RA-M
PD-RA-H, PD-RA-T, PD-RA-A
Manual:
7-1-1
Automated:
7-1-2
7-2 7-3
Slim Attachment Hub P300-1AHS, P300-1AHA
Slim Tube
P300-1T110, P300-1T130
P300-1T170, P300-1T240
P300-1T110-C, P300-1T130-C
P300-1T110-PB, P300-1T130-PB
Hood
Beak P300-2-B, P300-1T-H, P300-1T-B
Manual:
8-1-1
Automated:
8-1-2 -
8-2
Duraguard
Rotatable Duraguard
P200-CRD-P, P200-CRD-M, P200-CRD-L
P200-RCRD-P, P200-RCRD-M
P200-RCRD-L
PD-CRD-P, PD-CRD-M, PD-CRD-L
Wrench P200-1ARA-W
PD-TW-J4, PD-TW-J6
SZ114011
Adaptor PD-OSC-A
Irrigation Nozzle
PD-IN-D, PD-IN-2ES, PD-IN-2S, PD-IN-2M
PD-IN-2L, P200-IN-1A155, P200-IN-1A175
PD-IN-3S, PD-IN-3M ,PD-IN-3L
P200-IN-BS
Manual:
9-1-1
Automated:
9-1-2
9-2

7
English
4 Cleaning / Checking the Control Unit
4–1 Wipe Disinfecting
Wipe the product using a cloth moistened with alcohol disinfectant (e.g. Meliseptol®rapid).
CAUTION
●Do not soak the Control Unit in water.
●Prevent water from entering the Connector portion.
●Do not sterilize the Control Unit.
4–2 Checking
Check with the same procedure described in "Check before Operation (Control Unit)".
Proceed to Storage (see “11 Storage”)
NOTICE
●To perform "Check before Operation (Control Unit)", refer to OM-SE0021E Primado2 OPERATION MANUAL Control Unit and Foot
Controls.
5 Cleaning / Checking the Foot Control
5–1 Cleaning/Wipe Disinfecting
1 Wipe the product using a cloth moistened with neutral
enzymatic detergent.
2 Remove the detergent using a cloth moistened with water.
3 Wipe the product using a cloth moistened with alcohol
disinfectant (e.g. Meliseptol®rapid).
CAUTION
●Do not soak the Foot Control into water.
●Prevent water from entering the Connector portion.
●Do not sterilize the Foot Control.
5–2 Checking
Check with the same procedure described in "Check before Operation (Foot Control)".
Proceed to Storage (see “11 Storage”)
NOTICE
●To perform "Check before Operation (Foot Control)", refer to OM-SE0021E Primado2 OPERATION MANUAL Control Unit and Foot
Controls.

8
6 Cleaning / Checking the Motor
6–1 Cleaning
6–1–1 Manual Cleaning
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
_ Cleaning
1 Remove the dirt on the surface of the product using a soft
bristle brush moistened with neutral enzymatic detergent.
≥ 1 min
Water quality: Distilled, deionized or demineralized water.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR.
WEIGERT)
2 Rinse the product under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
3 Rinse the product surface with distilled, deionized or
demineralized water.
_ Drying
1 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see “6–2 Checking”).
6–1–2 Automated Cleaning
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water

9
English
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature Time Water quality Detergent
Pre-cleaning < 25°C 3 min Tap water -
Cleaning 40 - 60°C ≥ 5 min Distilled water
Deionized water
Demineralized water
Neutral enzymatic detergent for medical
purposes (pH 7.0 - 8.0), or alkaline detergent
for medical purposes (pH ≤12)
Rinsing > 10°C ≥ 1 min -
Thermal disinfection 90 - 93°C 5 min -
Drying Set the washer-disinfector according to the instruction of the manufacturer.
* Use neutral enzymatic detergent for medical purposes for P200-WPD.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Checking (see “6–2 Checking”).
6–2 Checking
1 Check for any abnormalities or damage on the outer surface
and for any damage or deformation on the Pins in the Plug.
2 Check that the Motor can be securely connected to the Control
Unit.
3 Depress the Pedal of the Foot Control to check for any abnormal
vibration, noise or heating in the Motor.
Check on the Main Panel of the Control Unit that the rotational
speed can reach the maximum speed you have set.
4 Check that Forward, Reverse, and Oscillate (P200-WPD only)
Modes operate normally.
5 Mount the Attachment to the Motor to check its secure
connection.
Proceed to Packaging / Sterilization (see “10 Packaging /
Sterilization”).
CAUTION
●Do not lubricate the Motor. It may cause failure.

10
7 Cleaning / Lubrication / Checking of the
Attachment, Slim Attachment Hub and Slim Tube
7–1 Cleaning
Be sure to use the Cleaning Tools for cleaning the following products.
Standard Attachment 300 P300-2S10, P300-2S20, P300-2S40, P300-2S60, P300-2S80
P300-2A10, P300-2A20, P300-2A40, P300-2A60, P300-2A80
Slim Attachment 300
P300-1AHS, P300-1AHA
P300-1T110, P300-1T130, P300-1T170, P300-1T240
P300-1T110-C, P300-1T130-C, P300-1T110-PB, P300-1T130-PB
Check the compatible Cleaning Adaptor before using the Cleaning Tools.
Product Order Code Compatible Cleaning Adaptor
Slim Tube
P300-1T110, P300-1T130, P300-1T170, P300-1T240
P300-1T110-C, P300-1T130-C
P300-1T110-PB, P300-1T130-PB
Cleaning Adaptor 1T
PD-CTAD-1T
Slim Attachment
Hub P300-1AHS, P300-1AHA
Cleaning Adaptor SM
PD-CTAD-SM
Attachment
P200-2SMS, P200-2SES, P200-2SS, P200-2SM, P200-2SL
P200-2AMS, P200-2AES, P200-2AS, P200-2AM, P200-2AL
P200-1A135, P200-1A155, P200-1A155-C
P200-1A175, P200-1A175-C
P200-RA330, P200-RA330-L
P200-2SD, P200-CA-4, P200-3MC
P200-PER, P200-CRA
P300-2S10, P300-2S20, P300-2S40, P300-2S60, P300-2S80
P300-2A10, P300-2A20, P300-2A40, P300-2A60, P300-2A80
P200-3GAS, P200-3GAM, P200-3GAL
PD-1A135, PD-1A155, PD-1A175
PD-2AES, PD-2AS, PD-2AM, PD-2AL
PD-2SES, PD-2SS, PD-2SM, PD-2SD, PD-2SDM
PD-2AD, PD-2ADM
PD-CRA, PD-PER
PD-RA330, PD-RA330-L, PD-CA-16, PD-CA-64
Cleaning Adaptor HM
PD-CTAD-HM
PD-DA-J4, PD-DA-J6, PD-DA-A, PD-DA-K, PD-DA-W
PD-DA-P, PD-RA-M, PD-RA-H, PD-RA-T, PD-RA-A
Cleaning Adaptor WPD
PD-CTAD-WPD
The Cleaning Tools are not applicable for the following products.
P200-SAG, P200-REC, P200-OSC, PD-SA-S, PD-SA-R

11
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
7–1–1 Manual Cleaning
Manual Cleaning using the Cleaning Tools
Use the Cleaning Tools according to the following system figure.
*1: With O-ring (Order Code: D0312074080)
*2: With O-ring (Order Code: D0312150150)
Attachment
Cleaning Adaptor SM
Slim Attachment Hub
Cleaning Adaptor 1TSlim Tube
Cleaning Adaptor HM*1
Attachment
Cleaning Adaptor WPD*2
Attachment
Cleaning Drying
Luer-lock syringe
Cleaning Adaptor G Air Gun
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
For Slim Attachment Hub, perform this step with the Slim Tube removed.
_ Cleaning
1 Turn the Twist Collet to the closed position (if applicable).
2 Mount the Cleaning Adaptor.
Pull the Cleaning Adaptor to make sure that it is securely
connected.
3 Screw the disposable luer-lock syringe (hereafter referred to as
syringe) into each Cleaning Adaptor of the Attachment.
4 Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature Water quality Detergent
45°C
Distilled water
Deionized water
Demineralized
water
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)

12
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
5 With the product immersed in the cleaning tank, pull the syringe
plunger to suck and discharge the cleaning solution. Perform
this step 5 times to retain the cleaning solution inside the
product.
Immerse the product in the cleaning tank for 20 - 30 min.
x5
Push and pull the syringe plunger to discharge the cleaning
solution remaining in the product.
6 Immerse the product in the soak tank containing tap water, then
operate the syringe to suck and discharge water 10 times to
rinse inside the product.
≤ 38°C
Water quality: Drinking water
x10
7 Rinse the product under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
8 Put distilled, deionized or demineralized water into the syringe,
screw the syringe into the product, and discharge water from
the product tip.
Perform this step 10 times.
x10
9 Rinse the product surface with distilled, deionized or
demineralized water.
_ Drying
1 Screw the Cleaning Adaptor G into each Cleaning Adaptor.
2 Push the air gun into the Cleaning Adaptor G and remove
moisture inside the product.
Allowable air pressure < 500 kPa
Purging Time ≥ 5 sec.
3 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
4 Remove the Cleaning Adaptor from the Attachment by
depressing the Release button.
5 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see “7–2 Lubrication”)

13
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
Manual Cleaning without Cleaning Tools
_ Pre-cleaning
Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
_ Cleaning
1 Prepare cleaning solution in the cleaning tank under the
following conditions (prepare the cleaning solution according to
the instructions of the detergent manufacturer).
Temperature Water quality Detergent
45°C
Distilled water
Deionized water
Demineralized
water
Neutral enzymatic
detergent for medical
purposes (pH 7.0 - 8.0)
* Detergent used for validation:
neodisher MediZym (DR. WEIGERT)
2 Immerse the product in the cleaning tank for 20 - 30 min.
3 Thoroughly rinse the dirt released in immersion as well as the
detergent under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 10 sec.
Water quality: Drinking water
4 Rinse the product surface with distilled, deionized or
demineralized water.

14
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
_ Drying
1 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see “7–2 Lubrication”)
7–1–2 Automated Cleaning
Automated Cleaning with Cleaning Tools
Use the Cleaning Tools according to the following system figure.
*2: With O-ring (Order Code: D0312074080)
*3: With O-ring (Order Code: D0312150150)
Ⓒ
Ⓓ
Attachment
Cleaning Adaptor SM
Slim Attachment Hub
Cleaning Adaptor 1TSlim Tube
Cleaning Adaptor HM*2
Attachment
Cleaning Adaptor WPD*3
Attachment
Cleaning Drying
Cleaning Adaptor G Air Gun
Ⓐ Ⓑ
ⒶCleaning Adaptor Luer M ⒷCleaning Tube 6.0/L50 ⒸCleaning Tube 6.0/L650 ⒹCleaning Adaptor Luer F*1
Cleaning Holder
*1 Select the adaptor which matches the shape of the connection of the washer-disinfector.

15
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
_ Pre-cleaning
1 Remove contaminants on the surface with a cloth or soft bristle brush under running tap water.
For products with moving parts, perform cleaning while moving all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
2 Turn the Twist Collet to the closed position (if applicable).
3 Mount the Cleaning Adaptor.
Pull the Cleaning Adaptor to make sure that it is securely
connected.
4 Assemble the product as shown in the figure.
Pull each connection part to make sure that it is securely connected.
* If the Cleaning Tube is deteriorated, replace it with a new one.

16
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
5 Place the Cleaning Holder in the cleaning basket and set the cleaning basket in the washer-disinfector.
6 Mount the Cleaning Tube 6.0/L650 to the washer-disinfector.
If necessary, insert Cleaning Adaptor Luer F, etc.
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature Time Water quality Detergent
Pre-cleaning < 25°C 3 min Tap water -
Cleaning 40 - 60°C ≥ 5 min Distilled water
Deionized water
Demineralized water
Neutral enzymatic detergent for medical purposes
(pH 7.0 - 8.0), or alkaline detergent for medical
purposes (pH ≤12)
Rinsing > 10°C ≥ 1 min -
Thermal
disinfection 90 - 93°C 5 min -
Drying Set the washer-disinfector according to the instruction of the manufacturer.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1 Screw the Cleaning Adaptor G into each Cleaning Adaptor.
2 Push the air gun into the Cleaning Adaptor G and remove
moisture inside the product.
Allowable air pressure < 500 kPa
Purging Time ≥ 5 sec.
3 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
4 Remove the Cleaning Adaptor from the Attachment by
depressing the Release button.
5 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see “7–2 Lubrication”)

17
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
Automated Cleaning without Cleaning Tools
_ Pre-cleaning
1 Remove contaminants on the surface with a cloth or soft bristle
brush under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see “2 Moving Parts of Each Product”).
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
2 Turn the Twist Collet to the closed position (if applicable).
3 Place the product in the cleaning basket and set the cleaning basket in the washer-disinfector.
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature Time Water quality Detergent
Pre-cleaning < 25°C 3 min Tap water -
Cleaning 40 - 60°C ≥ 5 min Distilled water
Deionized water
Demineralized water
Neutral enzymatic detergent for medical
purposes (pH 7.0 - 8.0), or alkaline detergent
for medical purposes (pH ≤12)
Rinsing > 10°C ≥ 1 min -
Thermal disinfection 90 - 93°C 5 min -
Drying Set the washer-disinfector according to the instruction of the manufacturer.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1 Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2 Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see “7–2 Lubrication”)

18
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
7–2 Lubrication
Lubricate the product using separately sold EZ Spray Adaptor (Order Code PD-EZ-M) and PANA SPRAY Plus (Order Code Z182100).
Use the Spray Adaptor according to the following system figure.
Attachment
Spray Adaptor for Slim
Slim Attachment Hub
Attachment (connect directly to EZ Spray Grip)
EZ Spray Grip
Spray Adaptor for Bone Saw
Attachment Spray Adaptor
Attachment
Slim Tube
Check the compatible Spray Adaptor before lubrication.
Order Code Spray Adaptor
P200-2SMS, P200-2SES, P200-2SS, P200-2SM, P200-2SL
P200-2AMS, P200-2AES, P200-2AS, P200-2AM, P200-2AL
P200-1A135, P200-1A155, P200-1A155-C
P200-1A175, P200-1A175-C
P200-RA330, P200-RA330-L
P200-2SD, P200-CA-4, P200-3MC
P200-PER, P200-CRA
P300-2S10, P300-2S20, P300-2S40, P300-2S60, P300-2S80
P300-2A10, P300-2A20, P300-2A40, P300-2A60, P300-2A80
P300-1AHS, P300-1AHA
Spray Adaptor for Slim
P200-EZ-SM
P200-REC, P200-SAG, P200-OSC Spray Adaptor for Bone Saw
P200-EZ-BM
PD-DA-J4, PD-DA-J6, PD-DA-A, PD-DA-K, PD-DA-W
PD-DA-P, PD-RA-M, PD-RA-H, PD-RA-T, PD-RA-A
Spray Adaptor
P200-EZ-WPD
P200-3GAS, P200-3GAM, P200-3GAL
PD-1A135, PD-1A155, PD-1A175
PD-2AES, PD-2AS, PD-2AM, PD-2AL
PD-2SES, PD-2SS, PD-2SM, PD-2SD, PD-2SDM
PD-2AD, PD-2ADM
PD-CRA, PD-PER
PD-RA330, PD-RA330-L, PD-CA-16, PD-CA-64
-

19
Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
English
1 Connect each part of PANA SPRAY Plus, EZ Spray Grip, and
each Spray Adaptor.
2 Mount the Slim Tube to the Slim Attachment Hub.
(for Slim Attachment 300 only)
3 Mount the Attachment to the Spray Adaptor.
4 Pull the product to make sure that it is securely connected.
5 Hold the Attachment firmly to prevent it from moving by the
pressure of PANA SPRAY Plus. Spray PANA SPRAY Plus for 3 - 5
sec.
6 Remove the Attachment from each Spray Adaptor by depressing
the Release Button Ⓐon each Spray Adaptor.
Ⓐ
7 Remove each Spray Adaptor from the EZ Spray Grip by
depressing the Release Button Ⓑon the EZ Spray Grip.
Proceed to Checking (see “7–3 Checking”).
Ⓑ
CAUTION
●DO NOT use anything other than PANA SPRAY Plus for lubrication.
●Shake PANA SPRAY Plus well before use.
●Due to its structure, the lubrication oil of PANA SPRAY Plus comes out from the tip of the Attachment and the Twist Collet part. To
prevent dispersal of the lubrication oil in the surrounding area, it is recommended to cover the Attachment with a dry cloth or use the
Spray Mist Absorber.
●After lubrication, there could be high pressure inside the Spray Adaptor. When removing the Spray Adaptor, be careful because it may
pop out with force.
●Do not lubricate the Motor, Duraguard or Rotatable Duraguard. It may cause failure.
7–3 Checking
1 Confirm that there is no damage on the outer surface and
connection parts of the product or missing parts.
2 Check with the same procedure described in "Check before
Operation".
Proceed to Packaging / Sterilization (see “10 Packaging /
Sterilization”).
NOTICE
●To perform "Check before Operation", refer to OM-SH0913MA Primado2 OPERATION MANUAL Attachments.
Other manuals for Primado2
5
Table of contents
Languages:
Other NSK Medical Equipment manuals