Thermal disinfection procedure
1. Disassembly
1. Disassemble the mask according to the instructions in the User Guide.
2. Additionally:
• Firmly pull the elbow clip away from the elbow body
• Pull the elbow ring out of the frame.
2. Cleaning
1. Make a solution of 1% (1¼ oz/gal) Alconox in water1at 86-131°F (30-55°C).
2. Soak all mask components for 5 to 10 minutes. Ensure there are no air bubbles.
3. After the soaking time, while keeping the components in the solution, clean the inside and outside with a soft bristle brush as follows. Pay particular attention to the crevices and cavities.
• Cushion / elbow body 30 seconds brushing
• Elbow clip / elbow ring 10 seconds brushing
• Frame (SLM)2/ headgear 1 minute brushing
4. Rinse the components as follows under running water1at 68-140°F (20-60°C) directly contacting all surfaces.
• Cushion / elbow body / elbow clip / elbow ring / frame (SLM) 30-60 seconds rinsing
• Headgear 1 minute rinsing, while squeezing repeatedly
3. Inspection and drying
1. Inspect and if required, repeat washing until visually clean.3
2. Squeeze the headgear to remove excess water.
3. Allow the mask components to air dry out of direct sunlight.4
4. Disinfection
1. Soak the mask components in a hot water1bath at 194-199°F (90-93°C) for 1 to 10 minutes. Ensure there are no air bubbles.
2. Squeeze the headgear to remove excess water.
3. Allow the mask components to air dry out of direct sunlight.
5. Inspection Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced.
Slight discoloration of the silicone components may occur and are acceptable.
6. Reassembly
1. Push the elbow clip onto the elbow body until it clicks into place.
2. Insert the elbow ring into the top of the frame.
3. Reassemble the mask according to the instructions in the User Guide.
7. Packaging and storage For storage details, check the User Guide.
1For all cleaning, rinsing and disinfection steps, use drinking quality water.
2A soft bristle tube/bottle brush is required to clean the inside of the frame.
3Failure to clean the mask component as indicated may result in inadequate disinfection.
4Drying is not required if thermal disinfection is continued immediately.
GENERAL WARNINGS AND CAUTIONS
• ResMed cannot give any assurance that deviations from the procedures listed in this guide (eg, exceeding the number of reprocessing cycles), and their effect on the
performance or safety of the product, will be acceptable.
• Mask components should not be subjected to autoclave or ethylene-oxide gas sterilization.
• When using detergents or disinfectants, always follow the manufacturer’s instructions. In the event of conflict, this guide takes precedence.
ResMed Ltd
1 Elizabeth Macarthur Drive
Bella Vista NSW 2153 Australia
See ResMed.com for other ResMed locations worldwide. AirFit is a trademark of the ResMed family of companies. Alconox is a trademark of Alconox Inc. For patent and other intellectual property information, see
Resmed.com/ip.© 2018 ResMed Ltd. 608499/1 2018-09
ResMed.com