VWR MCP 100 User manual

VWR®
Photometer MCP 100
Chlorine, Chlorinedioxyde
& Ozone
European Catalogue Number:
705 - 1084
Version: 1
Issued: 23. January 2018
INSTRUCTION MANUAL

2VWR ver 1 23.01.2018
Legal Address of Manufacturer
Europe
VWR International bvba
Researchpark Haasrode 2020
Geldenaaksebaan 464
B-3001 Leuven
+ 32 16 385011
http://be.vwr.com
Country of origin
Made in Germany

3
VWR ver 1 23.01.2018
Table of Contents
Warning ...........................................................5
Safety Information ...................................................5
Delivery contens .....................................................6
Unpacking .........................................................6
Set up.............................................................6
Intended use........................................................7
Symbols and conventions ..............................................8
Unpacking .........................................................8
Overview...........................................................9
Description of Buttons and Switches ......................................9
Operational keys ....................................................10
OTZ (One Time Zero) ...............................................10
Display backlight ..................................................11
Recall of stored data ...............................................11
Countdown / reaction period .........................................11
Menu selections...................................................12
Recall of stored data ...............................................12
Transmitting stored data.............................................12
Setting date and time...............................................13
Calibration ........................................................13
User calibration ...................................................13
Factory calibration reset .............................................14
Guidelines for photometric measurements ................................15
Method notes ....................................................15
Correct position of the vial ...........................................16
Correct filling of the vial.............................................16
Methods ..........................................................18
Chlorine LR with Powder Pack (CL 3.5)..................................18
Chlorine HR with Powder Pack (CL 8) ...................................20
Ozone with Powder Pack (O 3)........................................22
Chlorine dioxide with Powder Pack (CLO2)...............................24
Troubleshooting ....................................................26
Operating messages................................................26
Error codes.......................................................26
Repair and maintenance ..............................................27
User replaceable accessories and spare parts ...............................27
Technical service ....................................................27

4VWR ver 1 23.01.2018
Warranty..........................................................28
Compliance with local laws and regulations ...............................28
Equipement disposal .................................................28

5
VWR ver 1 23.01.2018
Warning
CAUTION
The accuracy of the instrument is only valid if the instrument is used in an environment
with controlled electromagnetic disturbances according to DIN 61326. Wireless devices,
e.g. wireless phones, must not be used near the instrument.
Safety Information
Read this entire manual carefully before use!
Carefully read and follow the SAFETY INSTRUCTIONS at the beginning of this manual!
The system may only be used by qualified persons!
Keep this manual for later reference!
Also observe important notices in order to avoid malfunctions and faults.
If possible, save the complete transport packaging for later transport.
Follow the safety instructions listed here for your own safety!
The safety instructions draw your attention to potential dangers. They also contain information
for how you can prevent danger with appropriate conduct.
Use is only permitted for qualified persons.
WARNING!
The device may not be operated in explosion-prone rooms.
Health hazard!
Observe the required protective measures when handling the
sample!
Health hazard!!
Do not handle depleted batteries or leaking material with bare
hands! Wear protective gloves! Avoid contact with the eyes
and skin!
WARNING!
Breakage of glass. Danger of cutting injuries.
Despite protection from the transport packaging, breakage of
glass can occur during transport. The resulting broken shards
may have sharp edges and could cause cutting injuries when
handled carelessly. Look out for broken glass when unpack-
ing, and wear protective gloves.
When handling the system, always ensure a firm grip in order
to avoid breakage of glass!
Do not strike the test bottles against edged during transport
of the system!
WARNING!
Reagents are intended for chemical analysis only and must be
kept out of the hands of children!

6VWR ver 1 23.01.2018
WARNING!
Ensure proper disposal of reagent solutions.
Material Safety Data Sheets are available on request
Delivery contents
Standard contents for MCP 100:
1 Photometer in plastic case
1 Instruction manual
4 batteries (Type AAA)
1 battery compartment cover
4 screws
1 screwdriver
3 Round vials with cap, height 48 mm, Ø 24 mm
2 Plastic cuvettes (PC) with cap, "Multi-Typ 2, Ø 10 mm
1 Cleaning brush
1 Syringe, 5 ml
Different Refill Packs available on request.
Unpacking
Carefully inspect all items to ensure that every part of the list below is present and no visible
damage has occurred during shipment. If there is any damage or something is missing, please
contact your local distributor immediately.
Set up
Before working with the photometer insert the batteries (delivery contents).

7
VWR ver 1 23.01.2018
CAUTION:
To ensure that the instrument is water proof:
• seal ring (E) must be in position
• battery compartment cover (B) must be fixed with the four screws
If the batteries are removed for more than one minute the date and time menu starts auto-
matically when the photometer is switched on the next time.
Intended use
MCP 100 is a photometer for determining the concentration of water contents. The MCP 100
Photometer is only used for determining 4 parameters. Always adhere to the limits specified
in chapter “Technical Data”. To ensure correct and intended usage, the instructions, especially
the “Important Safety Information”, have to be read and understood.
The instrument is not used as intended if
• it is used in application areas not stated in these instructions.
• it is used in conditions which deviate from the ones described in these instructions.
AAA/LR 03
AAA/LR 03
AAA/LR03
AAA/LR03
(D) batteries
(E) seal ring
(C) notch
(F) instrument back
(B) battery compart-
ment cover
(A) screws
Replacement of batteries:

8VWR ver 1 23.01.2018
The safety instructions in this operating manual must be observed.
The instrument may only be used under the conditions and for the purposes for which it
was designed.
The instrument must be handled with care (do not throw, drop, etc.) and used in accordance
with the technical data. It must be protected against soiling.
Symbols and conventions
CAUTION! This symbol indicates a potential risk and alerts you
to proceed with caution
Product Specifications
Instrument single wavelength, direct reading colorimeter
Light source: LEDs, interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
530 nm = 5 nm
Wavelength accuracy ± 1 nm
Photometric accuracy* 3% FS (T = 20° C – 25° C)
Photometric resolution 0.01 A
Power supply 4 batteries (AAA/LR 03)
Operating time 17hr operating time or 5000 test measurements in
continuous mode when display backlight is off
Auto-OFF automatic switch off
10 minutes after last keypress
Display backlit LCD (on keypress)
Storage internal ring memory for 16 data sets
Interface IR interface for data transfer
Time real time clock und date
Calibration user and factory calibration
resetting to factory calibration possible
Dimensions 155 x 75 x 35 mm (LxWxH)
Weight approx. 260 g (incl. batteries)
Ambient conditions temperature: 5–40°C
rel. humidity: 30–90% (non-condensing)
Waterproof floating; as defined in IP 68 (1 hour at 0.1 meter)
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.

9
VWR ver 1 23.01.2018
Overview
Reagent Quantity cat. no.
Clorine Free-DPD/F10
(blue color marking)
Powder Pack / 100 85601.620
Chlorine Total-DPD/F10
(blue color marking)
Powder Pack / 100 85603.620
GLYCINE TABLETS PACK OF 100 84202.601
Description of Buttons
Switching the photometer on or off
Select the required test using the [MODE] key.
Performing Zero /
performing Test
Press the [!] key to turn
the display backlight on or off
On
Off
Mode
Zero
Test
!
!
Mode
Zero/Test
On/Off
AAA/LR03
AAA/LR03
AAA/LR03
AAA/LR03
(D) batteries
(E) seal ring
(C) notch
(F) instrument back
(B) battery compart-
ment cover
(A) screws
free
total

10 VWR ver 1 23.01.2018
Operational keys
Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key.
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the in-
strument memorizes the last method used before being switched off.
When the instrument is switched on again, the scroll list comes up with
the last used test method first.
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw
the cap on and place the vial in the sample chamber making sure that
the marks are aligned.
Press the [ZERO/TEST] key (see OTZ).
The "Method" symbol flashes for approx. 8 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the marks are aligned.
Press the [ZERO/TEST] key.
(For Countdown/reaction period see page 11)
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
OTZ (One Time Zero):
The zero setting is held in memory until the instrument is switched
off. It is not necessary to perform a new zero each time, if the water
samples under test are from the same body of water and the conditions
of testing are the same.
The zero setting can be repeated each time if necessary.
Repeating the zero:
Press the [ZERO/TEST] key for 2 seconds.
Zero
Test
METHOD
METHOD
METHOD
RESULT
On
Off
Mode
Zero
Test
Zero
Test
Zero
Test
0.0.0
METHOD

11
VWR ver 1 23.01.2018
Display backlight
Press the [!] key to turn the display backlight on or off. The backlight
is switched off automatically during the measurement.
Recall of stored data
If the instrument is switched on, press the [!] key for more than 4 sec-
onds, then release the [!] key to access the recall menu.
Countdown / reaction period
If a reaction period is included in a method a countdown function
can be used:
Press the [!] key and hold.
Press the [ZERO/TEST] key.
Release the [!] key; the countdown starts.
After the countdown is finished the measurement starts automatically.
It is possible to interrupt the countdown by pressing the [ZERO/TEST]
key. Measurement starts immediately.
Caution:
An incomplete reaction period can lead to incorrect test results.
!
!
!
Zero
Test

12 VWR ver 1 23.01.2018
Menu selections
Press the [MODE] key and hold.
Switch the unit on using the [ON/OFF] key.
Allow the 3 decimal points to be displayed before releasing the
[MODE] key.
The [!] key allows for selection of the following menu points:
1 diS recall stored data
1 Prt printing stored data
2
3setting the date and time
4user calibration
The selected menu is indicated by an arrow in the display.
1diS – Recall of stored data
After confirming the selection with the [MODE] key the photometer
shows the last 16 data sets in the following format (automatically
proceeds every 3 seconds until result is displayed):
Number n xx (xx: 16...1)
Year YYYY (e.g. 2014)
Date mm.dd (monthmonth:dayday)
Time hh:mm (hourhour:minuteminute)
Test Method
Result x,xx
The [ZERO/TEST] key repeats the current data set.
The [MODE] key scrolls through all stored data sets.
Quit the menu by pressing [!] key.
1Prt – Transmitting stored data (to Printer or PC)
The IRIM (infrared interface modul) uses modern infrared technology to
transmit measurement data from the photometer to one of 3 optional
interfaces. These interfaces can be used to connect to a PC, a USB
printer1) or alternatively a serial printer2).
1) USB printer: HP Deskjet 6940 ; 2) All ASCII printer with serial interface
can be used. (or equivalent printer with HP PCL 3 to HP PCL 5)
Note: To print data, or to transmit to a PC, the optional IRIM (Infrared
Interface Module) is required.
The IRIM Module and the connected printer/PC must be ready. Press
the [MODE] key to start the transmitting, the instrument displays
“PrtG” (Printing) for approx. 1 second followed by the number of the
first data set and its transmission. All data sets will be transmitted one
after the other. After finishing the instrument switches to test mode.
The print job can be cancelled by pressing the [On/Off] key. The
On
Off
Mode
!
!
Time Cal
Store Date
2
3 4
1
Mode
Mode
Zero
Test
On
Off
PrtG
diS
1
Time Cal
Store Date
Prt
1
Store Date
Time Cal

13
VWR ver 1 23.01.2018
E 132
CAL
METHOD
0.0.0
METHOD
CAL
CAL
cAL
Zero
Test
Zero
Test
METHOD
23Setting date and time (24-hour-format)
After confirming the selection with the [MODE] key the value to be
edited will be shown for 2 sec.
The setting starts with the year (YYYY) followed by the actual value
to be edited. The same applies for month (mm), day (dd), hour (hh)
and minutes (mm). Set the minutes first in steps of 10, press the [!]
key to continue setting the minutes in steps of 1.
Increase the value by pressing the [MODE] key.
Decrease the value by pressing [ZERO/TEST] key.
Proceed to the next value to be edited by pressing [!] key.
After setting the minutes and pressing the [!] key the display will show
“IS SET” and the instrument returns to the measurement mode.
User calibration
Note
:
user calibration (Display in calibration mode)
factory calibration (Display in calibration mode)
After confirming the selection with the [MODE] key the instrument will
show CAL/"Method".
Scroll through methods using the [MODE] key.
Fill a clean vial with the standard up to the 10 ml mark, screw the cap
on and place the vial in the sample chamber making sure that the
marks are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 8 seconds.
The display shows the following in alternating mode:
Perform calibration with a standard of known concentration (see
"Operation").
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
Mode
!
SET
DATE
YYYY
(2 sec.)
Mode
Zero
Test
Time Cal
Store Date
2
3
Time Cal
Store Date
4
Cal
instrument switches off.
If the instrument is not able to communicate with the IRIM, a timeout
occurs after approx. 2 minutes. The error E 132 is displayed for approx.
4 seconds. Subsequently, the instrument switches to test mode (see
also IRIM manual).
705-1086 Infrared Data Transfer Modul IRIM

14 VWR ver 1 23.01.2018
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
CAL
RESULT
+ x
Mode
On
Off
Zero
Test
: :
cAL
SEL
SEL
CAL
CAL
SEL
Factory calibration reset
A user calibrated method is indicated by an arrow while the test result
is displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST] key
and hold.
Switch the unit on using the [ON/OFF] key.
Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.
The following messages will appear in turn on the display:
The factory setting is active.
(SEL stands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messages will appear in turn on the display:
Switch the unit off using the [ON/OFF] key.
Mode
On
Off
Mode
On
Off
Zero
Test
Time Cal
Store Date
RESULT
Cal
CAL
RESULT

15
VWR ver 1 23.01.2018
Guidelines for photometric measurements
1. Vials, caps and stirring rods should be cleaned thoroughly after each analysis to prevent
interference. Even minor reagent residues can cause errors in the test result.
2. The outside of the vial must be clean and dry before starting the analysis. Clean the
outside of the vials with a towel to remove fingerprints or other marks.
3. Zero calibration and test must be carried out with the same vial as there may be slight
differences in optical performance between vials.
4. The vials must be positioned in the sample chamber for zeroing and test with the mark
on the vial aligned with the mark on the instrument.
5. Always perform zeroing and test with the vial cap tightly closed. Only use the cap with
a sealing ring.
6. Bubbles on the inside wall of the vial lead to incorrect measurements. To prevent this,
remove the bubbles by swirling the vial before performing the test.
7. Avoid spillage of water into the sample chamber because this can lead to incorrect test
results.
8. Contamination of the transparent cell chamber can result in wrong readings. Check at
regular intervals and – if necessary – clean the transparent cell chamber using a moist
cloth or cotton buds.
9. Large temperature differences between the instrument and the environment can lead
to errors – e.g. due to the formation of condensation in the cell chamber or on the vial.
10. To avoid errors caused by stray light do not use the instrument in bright sunlight.
11. The reagents must be added in the correct sequence.
Method notes
• Prior to measurement ensure that the sample is suitable for analysis (no major interferences)
and does not require any preparation i.e. pH adjustment, filtration etc.
• Different Refill Packs available on request.
• Reagents are designed for use in chemical analysis only and should be kept well out of
the reach of children.
• Ensure proper disposal of reagent solutions.
• Material Safety Data Sheets are available on request
(Internet: www.vwr.com)

16 VWR ver 1 23.01.2018
Correct filling of the vial:
Correct position of the vial (Ø 24 mm):

17
VWR ver 1 23.01.2018

18 VWR ver 1 23.01.2018
Methods
Chlorine LR with Powder Pack
0.02 – 3.5 mg/l
Mode CL 3.5
a) free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see “Operation”).
Remove the vial from the sample chamber.
Add the contents of one Chlorine Free-DPD/F10 Powder Pack
(blue color marking ) straight from the foil into the water
sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Remove the vial from the sample chamber.
Rinse the vial and the cap several times and then fill the vial with 10 ml
of water sample (Note 2 and 3).
Add the contents of one Chlorine Total-DPD/F10 Powder Pack
(blue color marking ) straight from the foil into the water
sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the marks
are aligned.
Wait for a reaction period of 3 minutes.
(Countdown can be activated, see page 11)
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
!
Zero
Test
Zero
Test
CL 3.5
CL 3.5
0.0.0
RESULT
RESULT
CL 3.5

19
VWR ver 1 23.01.2018
Tolerances:
0 – 1 mg/l: ± 0.05 mg/l
> 1 – 2 mg/l: ± 0.10 mg/l
> 2 – 3 mg/l: ± 0.20 mg/l
> 3 – 3.5 mg/l: ± 0.30 mg/l
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Chlorine may show lower results. To avoid any meas-
urement errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionised water.
2. For individual testing of free and total Chlorine, the use of different sets of glassware is
recommended (EN ISO 7393-2, 5.3).
3. Do not use the same sample vial for free and total chlorine without thoroughly rinsing
the vial between the two different tests.
4. Preparing the sample:
When preparing the sample, the lost of Chlorine, e.g. by pipetting or shaking, must be
avoided. The analysis must take place immediately after taking the sample.
5. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
6. Exceeding the measuring range:
Concentrations above 2 mg/l Chlorine can lead to results showing 0 mg/l. In this case,
the water sample must be diluted with water free of Chlorine and the measurement
repeated.
7. Oxidising agents such as Bromine, Ozone etc. interfere as they react in the same way as
Chlorine.
30144.294 SULPHURIC ACID 0.5 MOL/L 1N AVS TITRINORM
31627.290 SODIUM HYDROXIDE 1 MOL/L AVS R.PE/USP/NF
85413.601 PH TEST STRIPS 4.5-10.0 DOSATEST
84208.600 REFERENCE STANDARD KIT RSK CHLORINE MCP 100

20 VWR ver 1 23.01.2018
Chlorine HR with Powder Pack
(plastic vial type 2, 10 mm)
0.1 – 8.0 mg/l
Mode CL 8
a) free Chlorine
Fill a clean vial (10 mm ) with 5 ml of the water sample and perform
zero calibration (see “Operation”).
Add the contents of two Chlorine Free-DPD/F10 Powder Pack
(blue color marking ) straight from the foil into the water
sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the marks
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l free Chlorine.
b) total Chlorine
Fill a clean vial (10 mm ) with 5 ml of the water sample and perform
zero calibration (see “Operation”).
Add the contents of two Chlorine Total-DPD/F10 Powder Pack
(blue color marking ) straight from the foil into the water
sample.
Close the vial tightly with the cap and invert several times to mix the
contents (20 sec.).
Place the vial in the sample chamber making sure that the marks
are aligned.
Wait for a reaction period of 3 – 6 minutes.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l total Chlorine.
c) combined Chlorine
combined Chlorine = total Chlorine – free Chlorine
Zero
Test
Zero
Test
CL 8
CL 8
CL 8
0.0.0
0.0.0
RESULT
RESULT
10 ml
5 ml
2
Table of contents
Languages:
Other VWR Measuring Instrument manuals

VWR
VWR MCT 100 User manual

VWR
VWR avantor CO-2500L User manual

VWR
VWR DO-230 User manual

VWR
VWR avantor 76359-754 User manual

VWR
VWR Peqlab peqPETTE eType User manual

VWR
VWR DO-230 User manual

VWR
VWR avantor CO-2500L User manual

VWR
VWR DO210 User manual

VWR
VWR MD 8000 L User manual

VWR
VWR 304 Series User manual
Popular Measuring Instrument manuals by other brands

Macnaught
Macnaught MX Series manual

PCB Piezotronics
PCB Piezotronics 356A14 Installation and operating manual

gefran
gefran KH-SIL2 Series Manual of Electrical Installation

Keysight Technologies
Keysight Technologies 16034G Operation manual

BRONKHORST
BRONKHORST FLEXI-FLOW Compact quick start guide

Elvaco
Elvaco CMi4170 user manual