manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. YOKOGAWA
  6. •
  7. Test Equipment
  8. •
  9. YOKOGAWA 30032A User manual

YOKOGAWA 30032A User manual

User’s
Manual
4th Edition: October 2017
All Rights Reserved. Copyright ©
2008, Yokogawa Meters & Instruments Corporation,
2017, Yokogawa Test & Measurement Corporation
Printed in Japan
Store this manual in an easily
accessible place for quick reference.
Safety Precautions
This product is designed to be used by a person with specialized
knowledge. When operating the instrument, be sure to observe
the cautionary notes given below to ensure correct and safe use
of the instrument. If you use the instrument in any way other than
as instructed in this manual, the instrument’s protective measures
may be impaired. This manual is an essential part of the product;
keep it a safe place for future reference.
YOKOGAWAis by no means liable for any damage resulting from
use of the instrument in contradiction to these cautionary notes.
The following safety symbols are used on the instrument
and in this manual:
Danger! Handle with Care
This symbol indicates that the operator must refer to
an explanation in the instruction manual in order to avoid
the risk of injury or death of personnel or damage to
the instrument.
WARNING
Indicates a hazard that may result in the loss of life or
serious injury of the user unless the described instruction is
abided by.
CAUTION
Indicates a hazard that may result in an injury to the user
and/or physical damage to the product or other equipment
unless the described instruction is abided by.
Model 30032A
Leakage Clamp-on Tester
IM 30032A-E
4th Edition:
Oct. 2017 (YMI)
This symbol indicates that this instrument designed to
be applied around or removed from HAZARDOUS LIVE
conductors provided if the RATED circuit-to-earth voltage
dose not exceed the value indicated in the measurement
category
Double Insulation
This symbol indicates double insulation.
Alternating Current
This symbol indicates aiternating current (AC).
Direct Current
This symbol indicates direct current (DC).
Earth TERMINAL
This symbol indicates ground.
■ Strictly observe the following cautionary notes in order to
avoid the risk of injury or death of personnel or damage to
the instrument due to hazards such as electrical shock.
WARNING
• This instrument is a current measuring device (sensor).
Do not use for any other purpose.
• Do not use the instrument if there is a problem with its
physical appearance. (Do not use the instrument if there is any
damage to the casing, battery cover, display and labels
or when the casing is removed.)
• The barrier is there to protect you from touching the conductor.
Be careful not to reach cross the barrier when using the
instrument.
• Disconnect the instrument from the measurable conductors
under test before opening the casing to replace the battery.
• Do not use the product when there are raindrops or droplets
of condensed water on its surface, or if your hands are wet.
• Safety protectors such as rubber-insulated gloves should be
worn to prevent electrical shock when using the instrument.
• Do not use this product in a place where an explosive
gas or vapor is present.
• Do not open the case except when replacing batteries.
Only Yokogawa service personnel are authorized to remove
the casing or disassemble or modify the instrument.
Do not attempt to repair the instrument yourself,
as doing so is extremely dangerous.
When the instrument needs an internal inspection or
calibration, contact Yokogawa or the dealer from whom
you purchased the instrument.
WARNING
The maximum allowed current is 62 Arms and
the RATED circuit-to-earth voltage is 300 Vrms or less.
Do not apply input exceeding this value.
Otherwise, it will not only damage the tester, but also pose
a risk of damage to the human body.
For safety standards, refer to the specications.
CAUTION
The product is for domestic use (Class B) and meets
the electromagnetic compatibility requirenents.
CAUTION
• The jaw section is a delicate, precision sensor.
Do not subject the jaw to unreasonably strong shock,
vibration, or force when using it.
• If dust gets into the tops of the jaws, remove it immediately.
Do not close the jaws when dust is trapped in its joints as
the sensor may break.
CAUTION
• Do not use the instrument near noise-emitting equipment
or where there may be sudden changes in temperature.
Otherwise, the instrument may produce an unstable readings
or errors.
• Do not wipe the instrument using an organic solvent such as
benzine or paint thinner. Otherwise, the front panel may be
damaged or discolored. When cleaning the instrument,
use a dry cloth.
• Do not leave the tester exposed to direct sunlight or in a
hot and humid location such as the inside of a car,
for any prolonged length of time.
• If the instrument will not be used for long periods,
remove the battery.
■ Measurement Category
WARNING
The 30032A is designed for measurement category III.
Do not use the 30032A for measurements in location
that fall under measurement category IV.
Measurement
category Description Remarks
O
(None, Other) For measurements performed on
circuits not directly connected to
MAINS. Circuits not connected to
a mains power source.
CAT II For measurements performed on
circuits directly connected to the
low-voltage installation. Appliances,portable
equipment,etc.
CAT III For measurements performed in
the building installation. Distribution board,
circuit breaker,etc.
CAT IV For measurements performed at
the source of the low-voltage
installation. Overhead wire,
cable systems,etc.
1. Components
1.1 Display (LCD)
1.2 Main unit (30032A)
1) Jaw Section
Is a precision sensor for detecting currents.
2) Open/Close Lever
Opens and closes the jaws.
3) Display
Shows the measured value (digital reading or bar graph),
unit, function and low-battery symbol ( ), etc.
4) mA/A Selector Switch
Selects the unit of AC current (either “mA” or “A”)
to be measured.
5) Data Hold Switch
Retains the measured data.
If you press this switch, the symbol appears and
the data is retained.
If you press this switch once again,
data holding is canceled (the symbol disappears).
6) Power Switch
Turns on the power to the instrument.
7) Filter Switch
Turns on/off the lter.
Press the FILTER switch for more than 2 seconds.
When the lter is on (used), appears.
8) Battery Housing (Battery cover)
Contains the battery.
9) Barrier
Prevents contact with the wires.
Thank you for purchasing our Leakage Clamp-on Tester.
Before using this product, thoroughly read this manual
to understand how to use it properly.
Contact information of Yokogawa ofces worldwide is
provided on thefollowing sheet.
PIM 113-01Z2: Inquiries List of worldwide contacts
1) Jaw
section
2) Open/
close lever
8) Battery
housing
(cover)
4) mA/A
selector
switch 6) Power
switch
5) Data hold
switch
3) Display
Hand strap
9) Barrier
7) Filter
switch
2. AUTO POWER OFF Function
2.1 When the AUTO POWER OFF Function is Used
• AUTO OFF appears.
• The instrument automatically turns off 10 minutes after
the last switch operation. A buzzer starts to beep 15 seconds
before the automatic power-off.
• Pressing any switch after the beep restarts the auto power-off
function.
2.2 When the AUTO POWER OFF Function is not Used
(Cancellation of the AUTO POWER OFF Function)
• Turn off the power.
• With the DATA HOLD switch held down, press the POWER
switch for more than 2 seconds to turn on the power.
This causes the buzzer to sound, canceling the AUTO
POWER OFF function (the AUTO OFF display goes off).
If the instrument is used with the AUTO POWER OFF function
cancelled, take care not to let the battery run down.
2.3 Recovering the AUTO POWER OFF Function
• Turn off the power.
• Turn on the power. This enables the AUTO POWER OFF
function to recover. (AUTO OFF appears.)
3. Measuring Instructions
3.1 Before measurement
a) Examine the casing, battery cover, display, and labels of
the instrument for abnormalities.
b) Make sure that the battery cover is rmly closed.
3.2 AC Current Measurement (unit: mA/A)
a) Press the POWER switch to turn on the power.
b) Squeeze the open/close lever to open the jaws.
Insert a wire from the measurable conductors under test
through the jaws, making sure the tops of the jaws are
tightly shut.
c) When the reading stabilizes, read the value.
If you have difculties in reading the value,
use the DATA HOLD function.
d) When measuring Load Current, press the mA/A switch to
change to A range.
e) When you have nished measurement,
press the POWER switch to turn off the tester.
CAUTION
The correct measured values are not displayed if the jaws
do not t precisely. Make sure that they are shut tightly.
3.3 Using the DATA HOLD switch
Pressing the DATA HOLD switch retains the measured data
and displays .
The mA/A selector switch and the FILTER switch cannot be used
in this situation.
The only switches that can be used are the DATA HOLD switch
(to cancel data holding) and the POWER switch.
4. Range Selection
4.1 Range Selection
Switching between the 3 mA and 30 mA ranges and between
the 30 A and 60 A ranges is performed by auto-ranging
(automatic).
For switching between the 3/30 mA and 30/60 A ranges,
the range must be switched using the mA/A selector switch.
4.2 Measurement Ranges
0.000 mA
0.010 mA<I
I≤2.88 mA
I≥3.270 mA
I≥32.70 mA
OL
3 mA range
30 mA range
I≤28.8 A
I≥32.70 A
I≥60.6 A 62.0 A MAX
OL
0.05 A<I
0.00 A
30 A range
60 A range
Zero correction
Zero correction
Range down
Range down
Range up
Range up
Overrange
Overrange
5. Battery Replacement
5.1 Battery Voltage
If the battery runs down and its voltage falls below the
operating voltage (effective range), the symbol appears in
the LCD display; promptly replace the battery with a new one.
(Battery: CR2032, 1 piece)
CAUTION
• The specied accuracy is assured when the battery voltage
is in the effective range.
If the symbol is displayed, promptly replace the battery.
• If the power does not come on even after pressing the
POWER switch, the battery voltage may have fallen too low.
Replace the battery with a new one.
5.2 Replacing the Battery
CAUTION
• Before replacing the battery, always disconnect the tester from
the measurable conductor under test because there is a risk
of electrical shock. After replacement, close the battery cover
securely and conduct measurements.
• Use the specied lithium battery (CR2032).
To replace the battery:
a) Press the POWER switch to turn off the power.
b) Turn the battery cover on the backside of the instrument in
the direction of the arrow using a coin, etc.
c) Remove the cover and the battery.
d) Insert a new battery, making sure that the polarities are
correct.
e) Close the cover back in place by turning it in the reverse
direction of the arrow.
Inserting a Battery Closing the Battery Cover
Battery cover
Insert the battery being careful
that the polarities are correct.
Close the battery cover,
aligning the arrow's tail with
the dot and then turn clockwise.
6. Examples of Measurement
6.1 Examples of Leakage Current Measurement
Load
Load
Single-phase
Two-wire Circuit
Load
Grounding Wire
(1) (3)
(2)
Single-phase Three-
wire or Three-phase
Three-wire Circuit
6.2 Example of Load Current Measurement
Load
Insert only one of the wires through the jaws.
7. Filtering Function
The instrument has a function for turning the lter on (use)/off.
Press the FILTER switch for more than 2 seconds to turn it on/off
(default: ON). When the lter is on, the symbol appears.
When the FILTER switch is turned on, the sharp lter is enabled.
This lter attenuates 2nd-order and higher harmonic currents and
measures only the fundamental harmonics.
See the Filtering Characteristics below.
30032A Filtering characteristics
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Ampritude ratio [dB]
0 50 100 150
Frequency [Hz]
■ To Ensure Accurate Measurement
The maximum input current of the harmonic components in
the 3 mA and 30 mA ranges is 150 mA.
If the input exceeds 150 mA, the correct value is not displayed.
In some cases, “0L” does not appear and an error becomes
larger or readings do not change.
Therefore, follow the procedure below to check if the current
under measurement including 2nd-order and higher harmonic
components exceeds 150 mA.
Checking procedure:
1.Press the mA/A selector switch to choose the 30 A range.
2.Press the FILTER switch to select FILTER OFF and conduct
measurement.
3.Check that the measured value (reading) in the 30A range is
0.15 A* or less.
4.When the measured value is 0.15A or less, switch the range to
3 mA/30 mA and conduct measurement.
*: The accuracy of the 30 Arange is 1.0% of reading +5 digits
(0.05A < I ≤ 50 A).
Five digits are equivalent to 0.05 A; when checking the current,
also consider the accuracy. IM 30032A-E <P2>
8. Specications (30032A)
8.1 Specications (Accuracy)
Condition:
Temperature and humidity: 23 ˚C ±5 ˚C, 80% RH or less
Frequency: 50 Hz ±1.0 Hz, 60 Hz ±1.0 Hz
Battery voltage: Within the effective range
AC current measurement
Accuracy: ± (% of reading + digits)
Filter function OFF
Range Resolution Accuracy Maximum
Allowable Current
3 mA 0.001 mA 0.010 < I ≤ 32.70 mA:
1.0%+5 3.270 mA
30 mA 0.01 mA 32.70 mA
30 A 0.01 A 0.05 < I ≤ 50.0 A: 1.0% +5 32.70 A
60 A 0.1 A 50.0 < I ≤ 60.6 A: 5.0% +5 60.6 A
Filter function ON
Range Resolution Accuracy Maximum
Allowable Current
3 mA 0.001 mA 0.010 < I ≤ 32.70 mA:
1.5%+5 3.270 mA
30 mA 0.01 mA 32.70 mA
30 A 0.01 A 0.05 < I ≤ 50.0 A: 1.5% +5 32.70 A
60 A 0.1 A 50.0 < I ≤ 60.6 A: 5.5% +5 60.6 A
Note: Input current of 2nd-order and higher harmonics
: 150 mA rms maximum in the 3 mA/30 mA range
: 62 A rms maximum in the 30 A/60 A range
Filter specications
(3 mA and 30 mA ranges and 30 A and 60 A ranges)
Amplitude ratio at 100 Hz: -38 dB (1.26%) or less (typical: -41 dB)
Amplitude ratio at 120 Hz: -53 dB (0.22%) or less (typical: -56 dB)
Zero correction
3 mA range: Displays 0.000 mA (zero) when 0.010 mA < I
30 A range: Displays 0.00 A (zero) when 0.05 A< I
8.2 General Specications
• Method: Mean-value detection and rms-value calibration
• Display (LCD): Digital reading............... 3200 counts
Bar graph...................... 32 segments
“OL” over-range indication
low-battery symbol
data hold symbol
“AUTO OFF” auto power off indication
lter ON
• Sampling: Digital reading............... 2 times/sec
Bar graph...................... 12 times/sec
• Range: 3 mA, 30 mA, 30 A, 60 A
• Range selection:
Automatic (between 3 mA and 30 mA ranges and
between 30 A and 60 A ranges)
Manual (between 3 mA and 30 mA ranges and
between 30 A and 60 A ranges)
• Additional functions: Data hold and Auto power-off
• Operating temperature and humidity range:
0 to 50°C, with a maximum humidity of 80% RH or less
(no condensation)
• Storage temperature and humidity range:
-20 to 60°C, with humidity range is 20 to 70%RH
(no condensation)
• Temperature coefcient:
The following values must be added in the temperature
range of either 0 to 18°C or 28 to 50°C.
0 ≤ I ≤ 50.0A: ± (0.08% of reading/˚C + 0.5 digits/˚C)
50.0 < I ≤ 60.6A: ± (0.3% of reading/˚C + 0.5 digits/˚C)
• Effect of external magnetic elds: 0.0005% typical value
(in terms of the magnitude of current in adjacent wires)
• Diameter of measurable conductors: 40 mm (Maximum)
• The RATED circuit-to-earth voltage: 300 Vrms or less
• Maximum allowed current: 62 Arms
• Withstanding voltage: 3.7 kV AC for one minute
(Tested between the cores of the jaw section and the case)
• Power consumption: 6 mW maximum
• Automatic power-off:
Automatically turns of the power about 10 minutes
after the last switch operation.
(The alarm buzzer begins beeping 15 seconds
before power-off.)
• Power supply: CR2032 lithium battery (cell) .............1
• Battery life: Approx. 90 hours (of continuous operation)
• Dimensions: Approx. 70 (W) x 178 (H) x 25 (D) mm
(excluding protrusions)
• Weight: Approx. 200 g (including the battery)
• Accessories: Battery (housed in the instrument).....1
RB057 soft carrying case...................1
User’s manual
Safety standards:
EN 61010-1, EN 61010-2-032
Measurement category III
(The RATED circuit-to-earth voltage: 300 Vrms AC)
Indoor use, 2000 m max. above sea level,
pollution degree 2
EMC standards:
EN 61326-1 Class B, EN 61326-2-2
EMC Regulatory Arrangement in Australia and
New Zealand
EN 55011 Class B Group 1
Korea Electromagnetic Conformity Standard
 (한국전자파적합성기준)
Environmental standard:
EN 50581
Monitoring and control instruments including
industrial monitoring and control instruments
9. Sales in Each Country or Region
9.1 Disposing the Product
Waste Electrical and Electronic Equipment (WEEE), Directive
(This directive is valid only in the EU.)
This product complies with the WEEE directive
marking requirement.
This marking indicates that you must not discard
this electrical/electronic product in domestic
household waste.
Product Category
With reference to the equipment types in the WEEE directive,
this product is classied as a “Monitoring and control
instruments” product.
When disposing products in the EU, contact your local
Yokogawa Europe B.V. ofce.
Do not dispose in domestic household waste.
9.2 How to Replace and Dispose the Batteries
EU Battery Directive
(This directive is valid only in the EU.)
Batteries are included in this product.
When you remove batteries from this product and dispose them,
discard them in accordance with domestic law concerning
disposal.
Take a right action on waste batteries, because the collection
system in the EU on waste batteries are regulated.
Battery type: Lithium battery (cell)
Notice:
This marking indicates they shall be sorted out and
collected as ordained in the EU battery directive.
How to remove batteries safely:
For further details, see “5.2 Replacing the Battery.”
9.3 Authorized Representative in the EEA
Yokogawa Europe B.V. is the authorized representative of
Yokogawa Test & Measurement Corporation for this product in
the EEA. (EEA: European Economic Area)
To contact Yokogawa Europe B.V., see the separate list of
worldwide contacts, PIM 113-01Z2.

Other YOKOGAWA Test Equipment manuals

YOKOGAWA 30031A User manual

YOKOGAWA

YOKOGAWA 30031A User manual

YOKOGAWA 701921 User manual

YOKOGAWA

YOKOGAWA 701921 User manual

YOKOGAWA 709830 User manual

YOKOGAWA

YOKOGAWA 709830 User manual

YOKOGAWA SL1400 ScopeCorder LITE User manual

YOKOGAWA

YOKOGAWA SL1400 ScopeCorder LITE User manual

YOKOGAWA 702906 User manual

YOKOGAWA

YOKOGAWA 702906 User manual

YOKOGAWA AQ1215A User manual

YOKOGAWA

YOKOGAWA AQ1215A User manual

YOKOGAWA 701928 User manual

YOKOGAWA

YOKOGAWA 701928 User manual

YOKOGAWA US300FM User manual

YOKOGAWA

YOKOGAWA US300FM User manual

YOKOGAWA CA700 User manual

YOKOGAWA

YOKOGAWA CA700 User manual

YOKOGAWA CA330 User manual

YOKOGAWA

YOKOGAWA CA330 User manual

YOKOGAWA CA11E User manual

YOKOGAWA

YOKOGAWA CA11E User manual

YOKOGAWA DL7440 User manual

YOKOGAWA

YOKOGAWA DL7440 User manual

YOKOGAWA AQ7280 OTDR User manual

YOKOGAWA

YOKOGAWA AQ7280 OTDR User manual

YOKOGAWA CA11E HANDY CAL User manual

YOKOGAWA

YOKOGAWA CA11E HANDY CAL User manual

YOKOGAWA ScopeCorder DL350 User manual

YOKOGAWA

YOKOGAWA ScopeCorder DL350 User manual

YOKOGAWA AQ1300 Series User manual

YOKOGAWA

YOKOGAWA AQ1300 Series User manual

YOKOGAWA DL6000 Series User manual

YOKOGAWA

YOKOGAWA DL6000 Series User manual

YOKOGAWA DL9040 Series User manual

YOKOGAWA

YOKOGAWA DL9040 Series User manual

YOKOGAWA AQ1100A User manual

YOKOGAWA

YOKOGAWA AQ1100A User manual

YOKOGAWA MY40 User manual

YOKOGAWA

YOKOGAWA MY40 User manual

YOKOGAWA DLM2000 Series User manual

YOKOGAWA

YOKOGAWA DLM2000 Series User manual

YOKOGAWA MY600 User manual

YOKOGAWA

YOKOGAWA MY600 User manual

YOKOGAWA AQ1200A User manual

YOKOGAWA

YOKOGAWA AQ1200A User manual

YOKOGAWA AQ1200 User manual

YOKOGAWA

YOKOGAWA AQ1200 User manual

Popular Test Equipment manuals by other brands

MSG Equipment MS300 user manual

MSG Equipment

MSG Equipment MS300 user manual

Altrad BELLE GPX Operator's manual

Altrad

Altrad BELLE GPX Operator's manual

WHEELY-SAFE PC-1 operating instructions

WHEELY-SAFE

WHEELY-SAFE PC-1 operating instructions

Tohnichi DOTE4-G Operating	 instruction

Tohnichi

Tohnichi DOTE4-G Operating instruction

Uniks Link LAN manual

Uniks

Uniks Link LAN manual

Dräger Alcotest 5510 Instructions for use

Dräger

Dräger Alcotest 5510 Instructions for use

Alcovisor BAC-100 instruction manual

Alcovisor

Alcovisor BAC-100 instruction manual

Leader LBO-5880 instruction manual

Leader

Leader LBO-5880 instruction manual

Obdstar MT006 user manual

Obdstar

Obdstar MT006 user manual

Sanwa CP-7D instruction manual

Sanwa

Sanwa CP-7D instruction manual

GW Instek GDS-1000-U Series Freewave Installation & User Guide

GW Instek

GW Instek GDS-1000-U Series Freewave Installation & User Guide

Rigol RP1010H user guide

Rigol

Rigol RP1010H user guide

Tektronix TBS1000B Series user manual

Tektronix

Tektronix TBS1000B Series user manual

DingLicom Pilot Scout 3.0 user manual

DingLicom

DingLicom Pilot Scout 3.0 user manual

Rohde & Schwarz RTP044B Getting started

Rohde & Schwarz

Rohde & Schwarz RTP044B Getting started

Reed Instruments R5820 instruction manual

Reed Instruments

Reed Instruments R5820 instruction manual

ACS ALCOSIM instruction manual

ACS

ACS ALCOSIM instruction manual

Beta 1498TB/12-24 Instructions for use

Beta

Beta 1498TB/12-24 Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.