CDN THD2FE User manual

Model: THD2FE
NON-CONTACT FOREHEAD THERMOMETER
Forehead mode: 93.2 to 108°F/34 to 42.2°C
Surface mode: -7.6 to +176°F/-22 to +80°C
FOR MEASUREMENT OF HUMAN BODY
TEMPERATURE IN PEOPLE OF ALL AGES
• Response Time: Forehead mode: 1 second;
Surface mode: 0.6 second
• Forehead Measurement Distance: within 1.5 in/4 cm
• Surface mode: Distance:Spot = 1:1, emissivity: 0.95
• Accuracy: ±0.4°F/0.2°C
• Backlit in blue
• Memory recalls 25 readings
• One-button operation
• Battery status indication
• Temperature alert or silent
• On/Off Power button
• Auto-off after 1 minute
• ABS plastic
• Two 1.5V AAA IEC LR03 alkaline batteries, included
• FDA, RoHS, ISO 9001, ISO 13485, ASTM E1965-98,
EN ISO 80601-2-56
Note: Remove label from display before initial use.
Note: In the following instructions, names of the control
buttons are shown in CAPS. Function information that
appears on the display is shown in BOLD CAPS.
USAGE
Indications for Use
The Non-Contact Forehead Thermometer is an
infrared thermometer intended for the intermittent
measurement of human body temperature in people
of all ages.
Intended Operator:
At least 11 years old (5 years intensive reading
experience), no maximum.
Ambient Temperature
If there is any temperature difference between the
places where the device is stored and where it will
be used, allow the thermometer to adjust to the room
temperature where the subject is for at least
15 minutes before measurement.
Points of Attention
1. Forehead temperature is displayed in oral mode.
This mode converts the forehead temperature to
display its “oral-equivalent” value.
2. Before the measurement, the subject should stay
in a stable environment for 5 minutes and avoid
exercise or bath for 30 minutes.
3. Keep the forehead area clean and free from sweat,
cosmetics and scars while taking the temperature.
4. The “Clinical Bias” is -2.5 to -3.1°F (-1.4 to -1.7°C).
5. The “Limits of Agreement” is 0.98.
6. The “Repeatability” is 0.36°F (0.20°C)
BEFORE WE BEGIN
This thermometer has been designed for use in
professional healthcare facility environments and
home healthcare environments (as dened in 3.1 of
IEC 60601-1-11 Edition 2.0 2015-01 and Guidance
for Industry and Food and Drug Administration Staff:
Design Considerations for Devices Intended for
Home Use, Issued November 24, 2014). It is not
meant to replace a visit to the doctor. Compare the
measurement result to your regular
body temperature. Consult with doctor
if you have health concerns.
PO Box 10947 Portland, OR 97296-0947 Tel 800 338-5594 Fax 800 879-2364 Web CDNkitchen.com

BATTERY INSTALLATION
Replace battery when the Low Battery icon ( )
indicates the battery is low. Power off the unit before
installing the batteries. A malfunction may occur if
the power is on when the battery is installed. If a
malfunction occurs, restart the device.
1. Remove battery cover by using the
thumbs to push battery cover out.
2. Install two 1.5V AAA batteries
observing polarity shown in
compartment.
3. Replace the battery cover until it
clicks shut.
OPERATING INSTRUCTIONS
A. On/Off
1. Press the POWER button ( ) to turn the
thermometer on.
2. Forehead is the default mode. The Forehead
icon ( ) appears on the display and two beeps
sound when ready.
3. Press the POWER button ( ) for 5 seconds to
turn the thermometer off.
4. The THD2FE will automatically turn off after
1 minute of inactivity.
B. Temperature Scale
Note: When the temperature scale is changed, the memory is
cleared.
To select temperature reading in Fahrenheit or
Celsius:
1. Power off the unit.
2. Press and hold the START button, then press
and hold the POWER button for 3 seconds.
3. The °F symbol changes to the °C symbol on the
display or vice versa.
C. LCD Backlight
The backlight will turn on automatically for
2 seconds when a reading is taken.
D. Forehead Mode
Forehead is the default mode of the THD2FE.
1. Make sure that the sensor lens is clean and
undamaged and that the forehead is clean.
2. Hold the sensor lens 1.5 inches (4 cm) or less
from the center of the forehead and press
the START button to get the temperature
measurement.
3. Wait for the Forehead icon ( ) to stop ashing
before taking the next measurement.
a. Temperature Alert
If the thermometer detects a temperature
≥99.5°F (37.5°C), three short beeps sound
followed by one long beep.
b. Silent
i. When the THD2FE is on, press and hold
the POWER button for 3 seconds.
ii. The silent icon ( ) ashes on the display.
iii. Release the POWER button to silence the
alert.
iv. Repeat to turn the alert on again.
Note: If the POWER button is pressed for 5 seconds
after the silent icon has begun ashing, the device
will power off without setting the alert to silent.
c. Memory
Recall up to 25 readings.
Note: As the memory gets full, the newer readings will
replace the older readings.
i. When the THD2FE is on, press the POWER
( ) button to see the temperature records
in memory.
ii. The memory icon ( ) appears on the
display.
iii. Press the Power button again to cycle
through all of the previous readings
starting with the most recent.
E. Surface Mode
Note: The surface mode shows the actual and unadjusted
surface temperature which is different from the body
temperature. It can help you monitor if the object temperature
is suitable for the baby or patient, for example the baby’s milk.
1 Turn the thermometer on
2. Press and hold the POWER button and press
the START button once to enter Surface mode.
3. The surface icon ( ) appears on the display.
4. Aim the sensor lens at the target and press
the START button to display the surface
temperature.
5. Press and hold the START button to get
continuous measurements.
Note: Applications include temperature measurements for
water, milk, cloth, skin or other objects.
Note: This mode shows the actual and unadjusted surface
temperature which is different from the body temperature.
Important: HAND WASH AND DRY. DO NOT IMMERSE
IN LIQUID.

F. Trouble Shooting
Error
Message
Problem Solution
Error 5-9, the system is not
functioning properly.
Unload the battery, wait for 1
minute and repower it. If the
message reappears, contact
CDN for further assistance.
Measurement before device
stabilization.
Wait for “Er1” to disappear.
The ambient temperature is not
within the range between 50 to
104°F (10 to 40°C).
Allow the thermometer to rest in
a room for at least 15 minutes at
room temperature: 50 to 104°F
(10 to 40°C).
Forehead mode: Temperature
taken is higher than 108°F
(+42.2°C)
Surface mode: Temperature
taken is higher than 176°F
(+80°C) Select the target within
specications. If a malfunction
still exists, contact CDN for
further assistance.
Forehead mode: Temperature
taken is lower than 93.2°F
(+34°C)
Surface mode: Temperature
taken is lower than -7.6°F
(-22°C)
Device cannot be powered on to
the ready stage.
Replace batteries with new
batteries.
E. Battery Status
The thermometer incorporates visual battery
status indication:
1. Battery OK: measurements are possible
2. Battery Low: replace battery with two 1.5V
AAA alkaline cells; measurements are possible
3. Battery Exhausted: replace battery;
measurements are not possible
CARE OF YOUR PRODUCT
• The sensor lens is the most delicate part of the
thermometer and should be kept clean at all times.
Take care when cleaning the lens. Use only a
soft cloth or cotton swab with water or rubbing
alcohol. Allow the lens to dry fully for at least
1 minute before using the thermometer.
• Do not submerge any part of the thermometer in
liquids. Keep it dry and away from any liquids and
direct sunlight. Wipe clean with a damp cloth.
• Store the thermometer at room temperature
between -4 to +122°F/–20 to +50°C, RH ≤85%.
• Avoid holding the thermometer too long. This could
cause the body temperature measurement to be
lower than usual.
PRECAUTIONS
• Dispose of used batteries promptly and keep away
from children.
• Keep the thermometer and batteries away from
children.
• Do not clean the case with abrasive or corrosive
compound, which may scratch the plastic and
corrode the electronic circuits.
• Do not subject the unit to excessive force shock,
dust, temperature or humidity, which may result in
malfunction, shorter electronic life span, damaged
battery and distorted parts.
• Do not tamper with the unit’s internal components.
Doing so will invalidate the warranty on the unit
and may cause unnecessary battery damage and
distorted parts.
• Do not subject the unit to excessive exposure to
direct sunlight. The unit is not waterproof — do
not immerse it into water or expose to heavy rain.
• To avoid deformation, do not place the unit in
extreme temperatures.
• Do not use the thermometer in a microwave oven.
• Always read the user manual thoroughly
before operating.
SPECIFICATIONS
Measurement Range Forehead mode: 93.2 to 108°F/34
to 42.2°C; Surface mode: -7.6 to
+176°F/-22 to +80°C
Unit of Measure °F/°C
Resolution 0.1°F/0.1°C
Water Resilience IP22: protected from solids bigger
than 12.5 mm (e.g. ngers); dripping
water when enclosure tilted up to 15°
Accuracy Forehead mode: 95 to 107.6°F/35
to 42°C: ±0.4°F/0.2°C, otherwise
±0.5°F/0.3°C;
Surface mode: 71.6 to 108°F/22 to
42.2°C: ±0.5°F/0.3°C, otherwise
±4% or ±4°F/2°C whichever is
greater
Measurement Distance
Within 1.5 in/4 cm
Distance:Spot Surface mode: 1:1
Emissivity Surface mode: 0.95
Operating Range 50 to 104°F/10 to 40°C, 15% to 85%
RH
Storage Range -4 to +122°F/-20 to 50°C, RH ≤85%
Transportation
Temperature
Shall be less than 158°F/70°C,
RH≤95%
Power Supply Two AAA 1.5V IEC LR03 alkaline
batteries, included
Product Dimensions 1.89 W x 6.22 H x 1.58 D (in)/
4.8 W x 15.8 H x 4.02 D (cm)
Weight 3.5 oz / 100 g (including battery)

SYMBOL DESCRIPTIONS
Caution
BF type applied part
IP22 Classication for water ingress and
particulate matter
Please read the instructions for use
Power, standby
Paper recycling
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specied below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Immunity Test
IEC 60601
Test Level Compliance Level
Electromagnetic Environment –
Guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Recommended separation
distance:
d = 1,2 √P
d = 1,2 √P80MHz to 800 MHz
d = 2,3 √P800MHz to 2,7 GHz
Where Pis the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by
an electromagnetic site survey,a
should be less than the compliance
level in each frequency range.b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reection from structures, objects and people.
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to xed RF transmitters, an electromagnetic site survey should be considered. If the measured
eld strength in the location in which the THD2FE is used exceeds the applicable RF compliance
level above, the THD2FE should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the THD2FE.
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC EMISSIONS
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specied below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Emissions Test Compliance
Electromagnetic Environment – Guidance
(for home healthcare environment)
RF emissions
CISPR 11 Group 1
The THD2FE uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11 Class B
The THD2FE is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specied below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic
Environment –
Guidance
(for home healthcare
environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood,
concrete or ceramic
tile. If oors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Power frequency
(50, 60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power
frequency magnetic
elds should be at
levels characteristic
of a typical
location in a typical
home healthcare
environment.
RECOMMENDED SEPARATION DISTANCES
BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE THD2FE
The THD2FE is intended for use in an electromagnetic
environment (for home healthcare) in which radiated RF
disturbances are controlled. The customer or the user of the
THD2FE can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the THD2FE
as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d= 1,2 √P
80 MHz to 800 MHz
d= 1,2 √P
800 MHz to 2,7 GHz
d= 2,3 √P
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1,2 2,3
10 N/A 3,8 7,3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance din meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where Pis the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reection from structures, objects and people.

MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATIONS EQUIPMENT
The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specied below. The customer or the user
of the THD2FE should assure that it is used in such an environment.
Test frequency
(MHz)
Band a)
(MHz) Service a) Modulation b)
Maximum power
(W)
Distance
(m)
Immunity Test Level
(V/m)
Compliance Level
(V/m)
(for home healthcare)
385 380–390 TETRA 400 Pulse modulation b) 18 Hz 1,8 0,3 27 27
450 430–470 GMRS 460,
FRS 460
FM c)
±5 kHz deviation
1 kHz sine
20,3 28 28
710
704–787 LTE Band 13,17 Pulse modulation b)
217 Hz 0,2 0,3 9 9745
780
810
800–960
GSM 800/900, TETRA
800, iDEN 820, CDMA
850, LTE Band 5
Pulse modulation b)
18 Hz 20,3 28 28
870
930
1720
1700–1990
GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE
Band 1, 3, 4, 25; UMTS
Pulse modulation b)
217 Hz 20,3 28 28
1845
1970
2450 2400–2570
Bluetooth, WLAN, 802.11
b/g/n, RFID 2450, LTE
Band 7
Pulse modulation b)
217 Hz 20,3 28 28
5240
5100–5800 WLAN 802.11
a/n
Pulse modulation b)
217 Hz 0,2 0,3 9 95500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50 % duty cycle square wave signal. c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz
may be used because while it does not represent actual modulation, it would be worst case.

© 05 -2020 CDN Made in China CD9999397en - 5/20 L-DESIGN 614.525.1472
The information in this document has been reviewed and is believed to
be accurate. However, neither the manufacturer nor its afliates assume
any responsibility for inaccuracies, errors or omissions that may be
contained herein. In no event will the manufacturer or its afliates be
liable for direct, indirect, special, incidental or consequential damages
arisen by using this product or resulting from any defect/omission in
this document, even if advised of the possibility of such damages.
The manufacturer and its afliates reserve the right to make
improvements or changes to this document and the products and
services described at any time, without notice or obligation.
1-Year Limited Warranty: Any instrument that proves to be
defective in material or workmanship (excluding batteries)
within one year of original purchase will be repaired or
replaced without charge upon receipt of the unit prepaid at:
CDN, PO Box 10947, Portland, OR 97296-0947 USA.
This warranty does not cover damage in shipment or
failure caused by failure to adhere to the accompanying
instructions, inadequate maintenance, normal wear
and tear, tampering, accident, misuse, unauthorized
modication, obvious carelessness or abuse. CDN shall
not be liable for any consequential or incidental damages
whatsoever.
Distributed by:
Component Design Northwest, Inc. dba CDN
2355 NW Vaughn Street
Portland, OR 97210-2311
For more detailed information on our
products, please visit CDNkitchen.com
or call 800-338-5594.
Other manuals for THD2FE
1
Table of contents
Other CDN Thermometer manuals

CDN
CDN BT482 User manual

CDN
CDN ProAccurate DTW450L User manual

CDN
CDN DTC450 User manual

CDN
CDN ProAccurate TCT572 User manual

CDN
CDN WT2 User manual

CDN
CDN DTTW572 Instruction manual

CDN
CDN THD2FE User manual

CDN
CDN IRT220 User manual

CDN
CDN ProAccurate Insta-Read IRB220-F-6.5 User manual

CDN
CDN ProAccurate DTT450 User manual