PRECAUTIONS
SAFETY
Before
connectingtheinstrumenttothepowersource,
care-
fullyreadthe
following
information,thenverifythatthe
powercordandpowerlinefuseareonesforyourpowerline.
The
instrument'srearpanelhasa fuseholderonthe
left
of
theACinletterminal(thefuseholeralso
serves
asa vol-
tageselector).Thevalueunderthetriangle• markedon
theholderindicatesthelinevoltagesetfortheinstrument.
Ifthepowercordisnotapplied
with
thecorrectvoltage,
thereisa danger
from
electricshock.
Line
voltage
This
instrumentoperatesusingac-power
input
voltagesthat
100/120/220/240
V atfrequencies
from
50Hzto60Hz.
Power
cord
The
groundwireofthe3-wireacpowerplugplacesthe
chassis
andhousingoftheoscilloscopeatearthground.Do
notattempttodefeatthegroundwireconnectionorfloat
theoscilloscope;todosomayposea greatsafetyhazard.
The
appropriatepowercordissuppliedbyan
option
that
is
specifiedwhentheinstrumentisordered.
The
optionalpowercordsareshownasfollowsinFig.1.
Line
fuse
The
fuseholderislocatedontherearpanelandcontains
theline
fuse.
Verifythattheproperfuseisinstalledby
replacingtheline
fuse.
5.
Alwaysusetheprobegroundclipsforbestresults.Do
notuseanexternalgroundwireinlieuoftheprobe
groundclips,asundesiredsignalsmaybeintroduced.
6.
Operationadjacenttoequipmentwhichproducesstrong
ac
magneticfieldsshouldbeavoidedwherepossible.
This
includessuchdevicesaslargepowersupplies,trans-
formers,
electricmotors,etc.,thatare
often
found
inan
industrialenvironment.Strongmagneticshieldscanex-
ceed
thepracticalCRTmagneticshieldinglimitsand
result
interferenceanddistortion.
7.
Probecompensationadjustmentmatchestheprobeto
the
input
ofthescope.Forbestresults,compensation
ofprobeshouldbeadjustedinitially,thenthesamepro-
be
always
used
with
the
input
ofscope.Probecompen-
sationshouldbereadjustedwhenevera probe
from
a
differentscopeis
used.
(Seepage19)
8.
InX-Yoperation,donot
pull
outthe
PULLx
10MAG
switch.
Ifpulledoutit,noisemayappeareonthe
waveform.
EQUIPMENT
PROTECTION
1.
Neverallowa smallspotofhighbrilliancetoremain
sta-
tionaryonthescreenformorethana fewseconds.The
screen
maybecomepermanentlyburned.A spotwilloc-
cur
onlywhenthescopeissetupforX-Yoperationand
nosignalisapplied.Eitherreducetheintensitysothespot
is
barelyvisible,switchbacktonormalsweepoperation
when
nosignalisapplied,orsetupthescopeforspot
blanking.
2.
Nevercovertheventilatingholesonthetopofthe
oscil-
loscope,
asthiswillincreasetheoperatingtemperature
inside
the
case.
3.
Neverapplymorethanthemaximumratingtothe
oscil-
loscope
inputs.
^CH1,
CH2
INPUT
jacks:
500Vp-por250V (DC+ AC
peak)
CH3
INPUT
jack:50V (DC+ ACpeak)
Z
axis
INPUT
jack:50V (DC+ ACpeak)
Never
applyexternalvoltagetotheoscilloscope
output
terminals.
4.
Alwaysconnecta cable
from
theearthground(GND)jack
oftheoscilloscopetothe
chassis
oftheequipmentun-
dertest.
Without
thiscaution,theentirecurrentforthe
equipmentundertestmaybedrawn
through
theprobe
clipleadsundercertaincircumstances.
Such
conditions
couldalsoposea safetyhazard,whichthegroundcable
willprevent.
7