
3.1. Air Supply
3.1.1. Ensure tool valve (or trigger) is in the "off" position before connecting to the air supply.
3.1.2. You will require an air pressure of 90psi, and an air flow according to specification.
3.1.3. WARNING! Ensure the air supply is clean and does not exceed 90psi while operating the
tool. Too high an air pressure and unclean air will shorten the product life due to excessive
wear, and may be dangerous causing damage and/or personal injury.
3.1.4. Drain the air tank daily. Water in the air line will damage the tool.
3.1.5. Clean air inlet filter weekly. Recommended hook-up procedure is shown in fig.1.
3.1.6. Line pressure should be increased to compensate for unusually long air hoses (over 8
metres). The minimum hose diameter should be 1/4” I.D. and fittings must have the same
inside dimensions.
3.1.7. Keep hose away from heat, oil and sharp edges. Check hose for wear, and make certain
that all connections are secure.
3.2. Couplings.
Vibration may cause failure if a quick change coupling is connected directly to the tool. To
overcome this, connect a leader hose to the tool. A quick change coupling may then be used
to connect the leader hose to the air line recoil hose. See fig.1 & 2.
WARNING!Disconnect tool from air supply before changing accessories, servicing or
performing maintenance.Replace or repair damaged parts. Use genuine parts only.
Unauthorised parts may be dangerous and will invalidate the warranty.
Note: Numbers in brackets refer to item numbers in the parts diagram.
5.1. To replace the needles (16), unscrew the needle tube (19) from the barrel sleeve (12).
Remove the needle holder (17) and replace the needles (16). (Use genuine Sealey needles
only.) Refit the needle holder (17) into the needle tube (19), screw it onto the barrel sleeve
(12) and tighten.
5.2. Lubricate the tool daily with a few drops of Sealey air tool oil dripped into the air inlet.
5.3. Clean the tool after use.
5.4. Loss of power or erratic action may be due to the following:
a) Excessive drain on the air line. Moisture or restriction in the air pipe. Incorrect size
or type of hose connectors. To remedy check the air supply and follow instructions in
chapter 3.
b) Grit or gum deposits in the tool may also reduce performance. Flush the tool out
with gum solvent oil or an equal mixture of SAE No 10 oil and paraffin.
Allow to dry before use.
5.5. For a full service contact your local Sealey service agent.
5.6. When not in use, disconnect from air supply, clean tool and store in a safe, dry,
childproof location.
WARNING! Ensure you read, understand and apply safety instructions before use.
Note: Numbers in brackets refer to item numbers in the parts diagram.
4.1. Flux chipper
4.1.1. Fit the spring (4) onto the rounded end of the chisel (15) before fitting the chisel into the
barrel sleeve (12).
4.1.2. Fit the nut (20) over the sharpened end of the chisel (15), screw it onto the barrel sleeve
(12) and tighten.
4.1.3. Connect the tool to the air hose as in chapter 3.
4.1.4. To start the tool, depress the throttle lever.
4.2. Needle scaler
4.2.1. To use as a needle scaler, unscrew the nut (20) and remove the chisel (15) and spring (4).
4.2.2. Fit the needle driver (14) and ensure the spring (18) is located in the needle tube (19).
4.2.3. Ensure the needle holder (17) is fitted with needles (16) such that the heads of the needles
do not protrude above the surface of the needle holder.
4.2.4. Fit the needle holder (17) into the needle tube (19), screw the needle tube (19) onto the
barrel sleeve (12) and tighten.
4.2.5. To start the tool, depress the throttle lever.
DO NOT allow the tool to free run for an extended period of time as this will shorten its life.
fig.1
5. MAINTENANCE
Risk of Hand Arm Vibration Injury.
The SA52, when operated in accordance with these instructions and tested in accordance with EN 28662-1:
1993 results in the following vibration emission declared in accordance with BS EN12096: 1996.
Measured vibration emission value: ......... 6.7m/s2 Uncertainty: .............. 2.68m/s2
These values are suitable for comparison with emission levels of other tools that have been subject to
the same test.
This tool may cause hand-arm vibration syndrome if its use is inadequately managed.
Recommended Measures to reduce risk of hand-arm vibration syndrome:
This tool should not be used by an individual regularly for more than 136 minutes in any 8 hour period.
This duration of use should be reduced if the individual is exposed to hand-arm vibration from other sources.
We, the sole importer into the UK, declare that the product listed here is in conformity with the
following standards and directives. The construction file for this product is held by the Manufacturer
and may be inspected, by a national authority, upon request to Jack Sealey Ltd.
For Jack Sealey Ltd. Sole importer into the
UK of Sealey Power Tools.
AIR FLUX CHIPPER/NEEDLE SCALER
Model: SA52
2006/42/EC Machinery Directive
93/68/EEC Marking Directive
11th September 2009
Declaration of Conformity
Signed by Mark Sweetman
NOTE: It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component
parts without prior notice. IMPORTANT: No liability is accepted for incorrect use of this equipment. WARRANTY: Guarantee is 12
months from purchase date. proof of which will be required for any claim. INFORMATION: For a copy of our latest catalogue and
promotions, call us on 01284 757525 and leave your full name and address including your postcode.
01284 757500
01284 703534
Sole UK Distributor
Sealey Group,
Bury St. Edmunds, Suffolk.
www.sealey.co.uk
Web
email
4. OPERATING INSTRUCTIONS
3. PREPARING TOOL FOR USE
fig.2
SA52 Issue No: 3 - 170909SA52 Issue No: 3 - 170909