63 22 668 D3561
D3561.021.01.09.02 03.2017 7
Dentsply Sirona 2Safety information
Installation Requirements Sinius / Sinius CS / Sinius TS 2.1Installation by qualified personnel
båÖäáëÜ
2Safety information
2.1 Installation by qualified personnel
The installation of the supply connections must be carried out by qualified
technicians only.
Electrical and water supply installations; reference for INST suctionmachines
2.2 Radiotelephones
Mobile RF communications equipment can affect electro-medical
equipment. Therefore, the use of mobile wireless phones in medical office
or hospital environments must be prohibited.
2.3 Modifications and extensions of the system
Modifications to this unit which might affect the safety of the system
owner, patients or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with
original Dentsply Sirona accessories or third-party accessories expressly
approved by Dentsply Sirona. The user is responsible for any damage
resulting from the use of non-approved accessories.
If any units not approved by Dentsply Sirona are connected, they must
comply with the applicable standards, e.g.:
● IEC 60950-1 or IEC 62368-1 for information technology equipment
(e.g. PC, Ethernet switch, Ethernet hub, Ethernet converter, USB
repeater, Ethernet router, USB switch, HDMI repeater/amplifier,
audio sources, display port repeater/amplifier, etc. and their power
supply units), as well as
● IEC 60601-1 for medical devices.
The treatment center monitor must fulfill the requirements of the
IEC 60950-1 or IEC 62368-1 standard.
The loudspeaker port of the monitor may only be connected to a device
that complies with the IEC 60950-1, IEC 62368-1 (e.g. a PC) or IEC
60601-1 standard. Under no circumstances may it be connected to a
stereo system or similar.
If a system is created during the installation process, the requirements of
IEC 60601-1, 3rd edition, must be fulfilled. The system author assumes
responsibility for conformity of the system with 93/42/EEC Directive, for
example.
WARNING
Professional installation
Comply with the national regulations for electrical installation
(e.g. IEC 60364-1, VDE 0100-100, IEC 60364-7-710, VDE 0100-710,
National Electrical Code).
Comply with the national regulations for water supply installations
(e.g. EN 1717) and sewage installations (e.g. EN 12056-1).