Sirona Dental Systems GmbH 1Before you begin
Installation Instructions ORTHOPHOS XG 5 / Ceph 1.2 Safety
60 04 902 D 3352
D 3352.031.02.19.02 7
English
Fixed connection!
The installation of a power plug instead of the
prescribed fixed (hard-wired) connection violates
international medical regulations and is prohibited.
In case of a fault, you would thus endanger the life and
limb of the patient, the operator or other persons.
Installation and startup must be carried out in
accordance with the requirements stated in our
Installation Instructions.
Installation and startup may be carried out only by
personnel specifically authorized by SIRONA.
Any person who assembles or modifies a medical
electrical system complying with the standard
IEC 60 601-1-1 (safety requirements for medical
electrical equipment) by combining it with other
equipment is responsible for ensuring that the
requirements of this regulation are met to their full
extent for the safety of the patients, the operators
and the environment.
If any equipment not approved by SIRONA is
connected, it must comply with the applicable
standards:
- IEC 60950-1 for information technology equipment and
- IEC 60 601-1 for medical electrical equipment.
See also “On-site installation, dimensions, technical
data” and the “Compatibility list/Declaration of
conformity” issued by the system integrator.
In case of doubt, contact the manufacturer of the
system components.
Wireless phone interference with medical electrical
equipment:
To ensure safe operation of medical electrical
equipment, the use of mobile wireless phones in
practice or hospital environments is prohibited.
The unit contains class 1 lasers.
A distance of at least 4" (10 cm) between eye and laser must
be observed. Do not stare into the beam.
Do not use the system with any other lasers, and do not
make any changes to settings or processes that are not
described in these operating instructions. This may lead to a
dangerous exposure to radiation.
Use an ESD wrist band during installation.
Connect it to the protective ground wire.
When opening the unit:
Please observe the usual precautionary
measures for handling PCBs (ESD).
Touch a ground point to discharge static
electricity before handling any components.
Extreme fluctuations of temperature may cause
condensation inside the unit. Do not switch the unit on
before it has reached normal room temperature.
1.2 Safety