ADTEK CS2-SG User manual

4 2/3 Strain Gage Meter
CS2-SG
CS2-SG Strain Gauge Indicator has been designed with high accuracy
measurement, display and communication of DC signal 0~1.0/~4.0mV or
0~10.0/~40.0mV as like as Load Cell or Strain Gauge.
The meter supports Field Calibration function. It can be calibrated with
sensor(Load Cell/Strain Gauge) to meet machinery structure.
They are also building in 4 Relay outputs, 3 External Control Inputs, 1
Analogue output and 1 RS-485(Modbus RTU Mode) interface with versatile
functions such as control, alarm, re-transmission and communication for a wide range of machinery and testing
equipments applications.
CS2-SG
Rev 1.1
2023-01-11 CS2-SG 1/11
˙ Measuring load cell, strain gauge signal 0~1.0/~2.0/~3.33/~4.0/~10.0/~20.0/~40.0mV/V (Specified)
˙ Field calibration with load cell or strain gauge to meet the system requirement
˙ 4 relay can be programmed individual to be a Hi / Lo / Hi Hold / Lo Hold / Go, and can be energized with Start
Delay / Hysteresis / Energized & De-energized Delay functions, or to be a remote control.
˙ Analogue output and RS-485 communication port in option
˙ 3 external control inputs can be programmed individual to be Tare (Relative PV) / PV Hold / Maximum or Minimum
Hold / DI (remote monitoring) / Reset for Relay Energized Latch….
˙ CE Approved & RoHS
˙ Testing Equipments for weight/force Measuring, Alarm, Control and Communication with PC/PLC
˙ Leakage testing equipment by tare and relay function.
˙ Weighting control for packing machine, filling machine.
■TechnicalSpecification Response time: ≤ 100 mS (when the AvG = “1”) in standard
Input range: Input High and Low programmable
aIhi: Settable range: 0.00~100.00%
aIlo: Settable range: 0.00~100.00%
Display & Functions
LED: Numeric: 5 digits, 0.8"(20.0mm)
H red high-brightness LED
Relay output indication: 4 square red LED
RS-485 communication: 1 square orange LED
E.C.I. function indication: 3 square green LED
Max/Mini Hold indication: 2 square orange LED
Display range: -19999~29999; 5 Digital Display:-19999~99999
Input
Measuring Range Input Impedance Excitation Voltage
0~1.0/~2.0/~3.33/~4.0 mV/V ≥ 1M ohm DC 5V, 40mA
or DC 10V, 40mA
0~10.0/~20.0/~40.0 mV/V
Calibration: Digital calibration by front key
Field calibration: Calibration with sensor input high & low to
meet system structure. And field calibration
reset is not change the accuracy & linear of
factory calibration.
A/D converter: 16 bits resolution
Accuracy: ≤± 0.04% of F.S. 1C
Sampling rate: 15 cycles/sec
High speed mode: can be 60cycle/sec(scaling:0~6000)
CS2 SG
CODE Aux. Power
AAC 115/230V
ADH AC 85~264V
DC 100~300V
ADL AC/DC 20~56V
CODE Excitation
E05 DC 5V
E10 DC 10V
EO Specify
— — ——
CODE Relay O/P
NNone
R2 2 Relay
R4 4 Relay
— —
CODE Analog
NNone
V0(1) ~ 5 V
0 ~ 10 V
I0 ~ 10 mA
4(0)~20 mA
CODE RS-485 Port
NNone
8RS-485
CODE I/P Range
S1 1.0mV/V
S2 2.0mV/V
S3 4.0mV/V
S6 3.33mV/V
CODE I/P Range
S8 10.0mV/V
S9 20.0mV/V
SA 40.0mV/V
■Description
■Features
■Applications
■OrderingInformation
Input
Signal
RS-485
Port
Relay
Output
Analog
Output
Excitation
Supply
Aux.
Powered *Optional Function
—
CODE Optional
HSM High Speed
Mode
5DG 5 Digital
Display

Rev 1.1
2023-01-112/11
CS2-SG
CS2-SG
Scaling function lOsc: Low Scale; Settable range: -19999~+29999/99999
hIsc: High Scale; Settable range: -19999~+29999/99999
Decimal point: Programmable from 0 / 0.0 / 0.00 / 0.000 / 0.0000
Over range indication: ovf l, when input is over 20% of input range Hi
Under range indication:
-ovf l, when input is under -20% of input range Lo
Max / Mini recording: Maximum and Minimum value storage during power on.
Display functions: PV / Max(Mini) Hold / RS-485 Programmable
Front key functions: Up and down key can be set to be a function as ECI.
Low cut:Settable range: -19999~29999 counts
Digital fine adjust: pVzro: Settable range: -19999~+29999/99999
pVspn: Settable range: -19999~+29999/99999
Reading Stable Function
Average: Settable range: 1~99 times
Moving average: Settable range: 1(None)~10 times
Digital filter: Settable range: 0(None)/1~99 times
Control Functions (optional)
Set-points: Four set-points
Control relay: Four relays
Relay 2 & Relay 3: Dual FORM-C, 5A/230Vac, 10A/115V
Relay 1 & Relay 4: Dual FORM-A, 1A/230Vac, 3A/115V
Relay energized mode: Energized levels compare with set-points:
Hi / Lo / Go.12 / Go.23 / Hi.HLd / Lo.HLd; programmable
DO function: Energized by RS-485 command of
master.
Energizing functions: Start delay / Energized & De-energized delay /
Hysteresis / Energized Latch
Start band(Minimum level for Energizing): 0~9999counts
Start delay time: 0:00.0~9(Minutes):59.9(Second)
Energized delay time: 0.00.0~9(Minutes):59.9(Second)
De-energized delay time: 0.00.0~9(Minutes):59.9(Second)
Hysteresis: 0~5000 counts
External Control Inputs (ECI)
Input mode: 3 ECI points, Contact or open collector input, Level
trigger
Functions: Relative PV (Tare) / PV Hold / Reset for Max or Mini. Hold /
DI / Reset for Relay Energized latch
Debouncing time: Settable range 5 ~255 x (8m seconds)
Analogue Output (optional)
Accuracy: ≤ 0.1% of F.S.; 16 bits DA converter
Ripple: ≤ 0.1% of F.S.
Response time: ≤ 100 mS (10~90% of input)
Isolation: AC 2.0 KV between input and output
Output range: Specify either Voltage or Current output in ordering
Voltage: 0~5V / 0~10V / 1~5V programmable
Current: 0~10mA / 0~20mA / 4~20mA programmable
Output capability: Voltage: 0~10V: ≥ 1000Ω;
Current: 4(0)~20mA: ≤ 600Ω max
Functions: aOhs: Settable range: -19999~29999/99999
aOls: Settable range: -19999~29999/99999
aOlmt(output High Limit):0.00~110.00% of output High
Digital fine adjust: aOzro: Settable range: -38011~+27524
aOspn: Settable range: -38011~+27524
RS-485 Communication (optional)
Protocol: Modbus RTU mode
Address: 1 ~ 247
Baud rate: 1200/2400/4800/9600/19200/38400 bps
Parity: None / Even / Odd
Data bits: 8 bits
Remote display: To show the value from RS-485 command of
master
Distance: 1200M max
Safety
Isolation: AC 2.5KV, 50/60 Hz, for 1 min
Between Power / Input / Output / Case
Insulation resistance: ≥100M Ω @ 500Vdc
EMC: EN 55011:2002; EN 61326:2003
Safety(LVD): EN 61010-1:2010
Environmental Conditions
Operating temp.: 0~60℃
Operating humidity: 20~95 %RH, Non-condensing
Temp. coefficient: ≤ 100 PPM/℃
Storage temp.: -10~70℃
Enclosure: Front panel: IEC 529 (IP52); Housing: IP20
Mechanical Structure
Dimensions: 96mm(W) x 48mm(H) x 120mm(D)
Panel cutout: 92mm(W) x 44mm(H)
Case material: ABS fire-resistance (UL 94V-0)
Mounting: Panel flush mounting
Terminal block: Plastic NYLON 66 (UL 94V-0)
10A 300Vac, M2.6, 1.3~2.0mm2(16~22AWG)
Weight: 550g / 350g(Aux. Power Code: ADH or ADL)
Power Supply
Power supply: AC115/230V±15%, 50/60Hz;
Optional: AC 85~264V, DC 100~300V, AC/DC 20~56V
Excitation supply: DC 5V/10V, 40mA
Power consumption: 5.0VA maximum
Back up memory: EEPROM

Rev 1.1
2023-01-11 3/11
CS2-SG
CS2-SG
■PinAssignment
Please check the voltage of power supplied first, and then connect to
the specified terminals. It is recommended that power supplied to the
meter be protected by a fuse or circuit breaker.
1A Fuse
AC115/230V
AC85~264V
EMC Filter
L
N
G
L
N
G
789 10
■PowerConnection
■RS-485CommunicationPort
333231
AB
+-
RS-485 Port
mV Input
10Vdc
Excit.S
■LoadCellConnection
11 12 13 14 15
+ - + -
Output
Output
0VE.S.
+VE.S
.
Load
Cell
■FrontPanel
RL1
RL2
2
RL3
RL4
COM
CI11
CI2
CI3
ENT/FN
▲
▲
▼
Kg
M.H
M.H
Relay status
Indication Comm. status
Control Input status
Max. Hold status
Display
screen
Operation
Key
Mini. Hold status
Engineer Unit
■Installation
The meter should be installed in a location that does not exceed the maximum
operating temperature and provides good air circulation.
R
L1
RL
2
2
R
L
3
R
L
4
COM
C
I1
1
C
I
2
C
I
3
ENT/F N
▲
▲
▼
Kg
PANEL CUT-OUT:
92+0.2(W)x44+0.2(H)mm
FIX HOLDER:
104mm(L)/W M3
1.0~8.0mm
■Dimension
Panel Cut Out 44.0+0.2
92.0+0.2
8.0 100.0 12.0
Unit:mm
48.0
96.0
Front View
Dimensions: 96mm x 48mm x 120mm
Panel Cutout: 93mm x 45mm (advise)
Terminal blocks:
10A/300Vac, M2.6,
M2.6, 1.3~2.0mm2
(16~12AWG)
28
21 2922 3023 3124 3225 3326 3427
b c aac acAB
+-
b c a
Relay 2
NC NO
Relay 1
NO
Relay 4
NO
RS-485
port
Analogue
output
Relay 3
NC NO
8
1 92 103
11 4
12 5
13 6
14 15 7
+-
+-
-+
-
-
ECI1 ECI2 ECI3 COM
mv/V
SIGNAL
INPUT
Excit.
Supply
AC115V
AC230V
AUX. POWER
External Control Input
FG
ADH
AC 85~264V
DC 100~300V
AC/DC 20~56V
ADL
Max Distance:1200M
Terminate Resistance
suggested :120Ω/0.5W
(Not standard accessory)
(Not standard accessory)

Rev 1.1
2023-01-114/11
CS2-SG
CS2-SG
■FunctionDescription
Input & Scaling Functions
Input range: Analogue input High and Low programmable
The meter has to be specified and fixed according to ordering code
(ex. 0~2.0mV/V in factory. If the meter has to install in difference range
of input, the meter can be set in function aIlo and aIhi of input
group to meet the input signal.
For example: The meter is 0~2.0mV/V input, and the signal from sensor
is 0~1.5mV/V. Please get into【input group】to set aIhi
(Analogue input High) to be 75.00%(2.0mV/V x 75.00% = 1.5mV/V),
then the meter has been changed the input range to 0~1.5mV/V and
the all relative parameters will work base on 0~1.5mV/V. The meter
doesn’t need re-calibration after change the aIlo and aIhi .
*The setting may course display lower resolution. Please set ower
resolution when the input signal has been high compressed.
Scaling function:
Setting the [ lOsc] (Low scale) and [ hIsc] (High scale) in [input
group] to relative input signal. Reverse scaling will be done too.
Please refer to the figure as below,
Remote Display by RS-485 command rs485 :
The meter will show the value that received from RS-485 sending.
In past, The meter normally receive 4~20mA or 0~10V from AO or
digital output from BCD module of PLC. We support a new solution
that PV shows the value from RS-485 command of master can so
that can be save cost and wiring from PLC.
Other functions :
The meter is also support relative PV (∆PV) and PV hold functions
that set in [ eci group]. Please refer to explain of ECI functions.
Low cut: Settable range from -19999~+99999 counts.
The users can set the value range.
1.If set the positive value (X1) here to display "0" which it
expressed to be low-cut the PV between "+X1 (plus)" &
"-X1(minus)" /absolute value
PV< l Setting value (X1) l, the display will be shown 0
EX: Low Cut is set for 0.50. If the display is from
-0.50~+0.50, that will be 0.
2.If set the negative value (X2) here to display "X2" which it
expressed to be low-cut the PV that it's under the X2 setting
value;
PV< Setting value(X2), the display will be shown X2.
EX: Low Cut is set for -0.01. If the display is < -0.01, and all
the display will be -0.01.
■Display&Functions
Max / Mini recording: The meter wills storage the maximum and
minimum value in [ user level] during power on in
order to review drifting of PV.
Display functions: PV / Max(Mini) Hold / RS 485 programmable in
(Please refer to step A-07) [
dsply] function of [input group]
Present Value pv : The display will show the value that Relative to Input signal.
Maximum Hold maXhd / Minimum Hold miNhd :
The meter will keep display in maximum(minimum)
value during power on, until manual reset by front
key in [User level], rear terminal is close [External
Control Input(ECI) ] or press front down or up key
to reset (according to setting, please refer to the
function of the ECI Group)
Please find the sticker that enclosure the package of the meter
to stick on the right side of square orange LED
M.H
Default Input: aIlo: 0.00%, aIhi: 100.00%
lOsc: 0.00, hIsc: 150.00
Change to aIlo : 0.00%, aIhi: 75.00%
SCALE
INPUT
0.00
199.99
0V
2.0mV/V
1.0mV/V
100.00
0.00%
100.00%
50.00%
1.5mV/V
75.00%
150.00
Default: [lOsc]: )00, [hIsc]: 10)00
Change to [lOsc]: 5)00, [hIsc]: 19(99
SCALE
INPUT
)00
19(99
10)00
5)00
Reverse slop
[hIsc]
[hIsc]
[lOsc]
[lOsc]
%
)00
%
5)00
%
10)00
Low Cut set to be -0.10
-0.10
Low cut
function
Low Cut is set for -0.10, if the PV is
under (< = -0.10), that display will be -0.10.
Present
Value
PV according
to input signal
-)10
[lOcut]
X2
Digital fine adjust: Settable range: -19999~+29999
Users can get Fine Adjustment for Zero & Span of PV by front key
of the meter, and “Just Key In” the value which user want to
show in the current input signals.
Especially, the [pVzro] & [pVspn] are not only in zero & span of
PV, but also any lower point for [pVzro] & higher point for [pVspn].
The meter will be linearization for full scale.
The adjustment can be clear in function [ZSclr]
Max. ( Mini.) Hold & Reset
Reset the Max
(Mini) Hold by
ECI or F. Key
O
N
Level
Trigger
Maximum Hold
Present
Value
Default Input:[ lOsc]: )00, [ hIsc]: 19(99
Digital fine adjustment:
input 2.75V, [pVzro]set to be 8)00
input 9.00V, [pVspn]set to be 17)00
Display: Low point from 55.00 to 80.00
High point from 180.00 to170.00
SCALE
INPUT _
0V
10V
5V
2.75V
9.00V
19(99
[ hIsc]
[ lOsc]
)00
10)00
18)00
5%00
8)00
17)00
Low Cut
function
Low Cut is set for 0.50, if the PV is
from -0.50~+0.50, that display will be 0.
}X1
X1 }
PV according
to input signal
Present
Value
0.50
-0.50
Low Cut set to be +0.50
[lOcut]
)50

Rev 1.1
2023-01-11 5/11
CS2-SG
CS2-SG
Reading Stable Function
Average display:
Jittery Display caused by the noise or unstable signal. User can set the
times to average the readings, and to get smoothly display.
The meter’s sampling is 15cycle/sec. If the [ avg](Average) set to be
3 to express the display update with 5 times/sec. The meter will
calculate the sampling 1-3 and update the display value. At meantime,
the sampling 4-6 will be processed to calculate.
Control Functions (optional)
Relay energized mode:Hi / Lo / Go-1.2 / Go-2.3 / Hi.HLd / Lo.HLd / DO programmable
Hi hi (Fig.1- ): Relay will energize when PV > Set-Point
Lo lo (Fig.1- ): Relay will energize when PV < Set-Point
Go-1.2 go-!2 : This function is programmable in Relay 4 only.
If the Relay 4 set to be Go function, the relay will
compare with [ry!sp] and [ry@sp].
Go relay energized when the condition is
[
ry!sp] (Hi) > PV > [ry@sp] (Lo)
Go-2.3 go-@3 : This function is programmable in Relay 4 only.
If the Relay 4 set to be Go function, the relay will
compare with [ry@sp] and [ry#sp].
Go relay energized when the condition is
[
ry@sp] (Hi) > PV > [ry#sp] (Lo)
DO function do :The function has been designed not only
a meter but also an I/O interface. In the case
of motor control cabinet can't get the remote
function. It's very easily to get the ON/OFF
status of switch from CS2 series with RS-485
function.
If the [ry_.md] had been set do , the relay
will be energized by RS-485 command directly,
but no longer to compare with set-point.
Moving average:
Jittery Display caused by the reasons as like as noise or unstable
signal. User can set the times to average the readings, and get
smoothly display.
The meter’s sampling is 15cycle/sec. If the [ Mavg](Moving
Average) set to be 3 expressed the display update with 15
times/sec.
In the first updated display value will be same as average function.
In the next updated display value, the function will get the new
fourth sample (sample 4) then throw away the first sample (sample
1) that the newest 3 samples(sample 2,3,4) will be calculated for
the updated display value.
Digital Filter: The digital filter can reduce the magnetic noise in field.
The digital filter can reduce the influence of spark noise caused by
magnetic of coil.
If the values of samples are over digital filter band (fix in firmware
and about 5% of stable reading) 3 times (Digital Filter set to be 3)
continuously, the meter will admit the samples and update the new
reading. Otherwise, it will be as treat as a noise and skip the samples.
Hi / Lo / Go Relay Energized
Hi Setting
Hi Relay Energized
ON
Lo Setting
Lo Relay Energized
ON
ON
Go Relay Energized
ON
ON
Hi.HLd hIhld (Lo.HLd lOhld ) :
The relay energized with latched function is for electrical
safety and human protection.
For example, a current meter relay installed for the over
current alarm of motor. Generally, over current of motor
caused by over load, mechanical dead lock, aging of
insulation and so on.
Above cases will alarm in the meter, if the user doesn't figure
out the real reason and re-start the motor. It may damage
the motor. The functions of Hi.HLd & Lo.HLd are designed
must be manual reset the alarm after checking out and
solving the issue. It's very important idea for electrical safety
and human protection.
As the PV Higher (or lower) than set-point, the relay will be
energized to latch except manual reset by from key in [ user
level] or [ eci](ECI) set to be rYrst rYrst is closed.
Start delay band and Start delay time:
The functions have Been designed for,
To avoid starting current of inductive motor (6 times of rated
current) with alarm.
If the y_.md r relay energized mode had been set to be lo
(Lo) or lOhld (Lo & latch). As the meter is power on and no input
to display the "0" caused the relay will be energized. User can set
a band and delay time to inhibit the energized of relay.
Start band rYsb (Fig.2- ): Settable range from 0~9999 Counts
Start delay time rYsd (Fig.2- ): Settable range from 0.0(s)~9(m)59.9(s);
Hi(Lo) Energized Latch & Reset
Hi.HLd(High Hold)
Relay Energized
ON
Reset the relay Hold
by ECI or Front Key
ON
Level Trigger
Hi Setting
hIhld
[ry_.md]
rYrst
[ ecI_]
[ry_.sp]
Sample1 Sample2 Sample3 Sample4 Sample5 Sample6 ......
Display Update Value=
(Sample 1+Sample 2+Sample 3)/3
Display Update Value=
(Sample 4+Sample 5+Sample 6)/3
Remark: The higher average setting will cause the
response time of Relay and Analogue output slower.
Average set to be 3
Remark: The higher moving average setting wouldn’t cause the response
time of Relay and Analogue output slower after rst 3 samples.
Sample1 Sample2 Sample3 Sample4 Sample5 Sample6 ......
In rst 3 samples, Display Update Value
=(Sample 1+Sample 2+Sample 3)/3
Display Update Value=(Sample 2+Sample 3+Sample 4)/3
Display Update Value=(Sample 3+Sample 4+Sample 5)/3
Display Update Value=(Sample 4+Sample 5+Sample 6)/3
Moving Average set to be 3
Hi / Lo Relay Energized
Fig.1
Hi Setting
Hi Relay Energized
ON
Lo Setting
Lo Relay Energized
ON
hi
lo
[ry_.sp]
[ry_.sp]
[ry_.md]
�
‚
Start Delay
Fig.2
Start Delay Time
Start Band
Hi Setting
Relay
Energized
ON
Inhibit
Inhibit
[ rYsb]
[ry_.sp]
[ rYsd]
�
‚

Rev 1.1
2023-01-116/11
CS2-SG
CS2-SG
Hysteresis ry_.hy (Fig.3- ① ): Settable range from 0~9999 Counts
As the display value is swing near by the set point to cause the
relay on and off frequently. The function is to avoid the relay on and
off frequently such as compressor…….etc.,
Relay energized delay ry_.rd (Fig.3- ② ):Settable range from 0.0(s)~9(m)59.9(s)
The function is to avoid the miss action caused by noise. Sometime,
the display value will swing caused by spark of contactor…etc..
User can set a period to delay the relay energized.
Relay de-energized delay ry_.fd (Fig.3-③ ):Settable range from 0.0(s)~9(m)59.9(s)
DI di : The E.C.I can be set to be di function, when
the meter building in RS-485 port. It is easier to get
remote monitoring a switch status through the
meter as like as DI of PLC.
Reset for Relay Energized Latch rYrst :
If the relay energized mode has been set to be hIhld
(Energized latch), and the [ ecI_] can be set to be
rYrst (Reset the Relay energized latch). When the PV
meets the condition of relay energizing, the relay will be
energized and latch until the ECI is to be closed.
Debouncing time:
The function is for avoiding noise signal to into the meter. And The
basic period is 8mseconds. It means you set the number that has to
multiple 8 m-seconds.
For example: [debnc]set to be 5, it means 5 x 8mseconds = 40mseconds
Please specify the output type either an 0~10V or 4(0) ~ 20mA in ordering.
The programmable output low and high scaling can be based on various
display values. Reverse slope output is possible by reversing point
positions.
Output range: Voltage: 0~5V / 0~10V / 1~5V programmable
Current: 0~10mA / 0~20mA / 4~20mA programmable
Functions: Output High / Low scale, output limit, fine adjustment
Output range high [ aOhs]:
To setting the Display value High to versus output
range High(as like as 20mA in 4~20)
Output range low [ aOls]:
To setting the Display value Low to versus output
range Low(as like as 4mA in 4~20)
Output High Limit [ aOlmt]:
0.00~110.00% of output High User can set the
high limit of output to avoid a damage of receiver
or protection system.
The range between [ aOhs] and [ aOls] should
be over 20% of span at least; otherwise, it will be
got less resolution of analogue output.
Energized / De-energized
Delay & Hysteresis
Fig.3
ON
Hysteresis
De-energized
delay time
Hi Setting
Relay
Energized
[ry_.sp]
[ry_.hy]
[ry_.rd]
[ry_.fd]
Energized delay time
�
‚
ƒ
■ExternalControlInputs(ECI)
The three external control inputs are individually programmable to perform
specific meter control or display functions. All E.C.I. have been designed in
level trigger actions. Please pay attention, the ECI1 or ECI2 input will be
disable while UP or Down Key has been set to be “ yes ”.
Functions: Relative PV / PV Hold / Reset Max or Mini. Hold / DI /
Reset for Relay Energized latch programmable.
Relative PV reLpv or Tare:
The [ ecI_] can be set to be reLpv function. When the E.C.I. is
closed, the reading will show the differential value.
PV Hold pVhld :
The [ ecI_] can be set to be pVhld (PV Hold) function. The display
will be hold when the E CI is closed, until the ECI is to be open. Please
refer to the below figures
Reset for Maximum or Minimum Hold Mrst :
When the [dsply] function in [input group] selected maXhd or
miNhd , the display will show Maximum or Minimum value.
The [ ecI_] function can be set to be Mrst function to reset the
maximum and minimum value in [User Level] by terminals of ECI
(close). Please refer to the figure as below.
Relative PV
Relative
PV(?PV) by
ECIor F.Key
ON
Level
Trigger
Present
Value
reLpv
[ecI_]
ECI_ has closed, and
then PV goes to “0”
ECI_ has opened,
and then display
shows PV
Relative PV
PV Hold & Reset
Present
Value
Reset PV Hold
by E.C.I. or
Front Key
O
N
Level
Trigger
ON
ON
PV Hold
pVhld
[ecI_]
Max. ( Mini.) Hold & Reset
Reset the max storage
when ECI has closed
Reset the Max.
(Mini.) storage
by E.C.I.
O
N
Level
Trigger
Maxi in User Level
Present
Value
Mrst
[ecI_]
Hi(Lo) Energized Latch & Reset
Hi.HLd(High Hold)
Relay Energized
ON
Reset the relay Hold
by ECI or Front Key
ON
Level Trigger
Hi Setting
hIhld
[ry_.md]
rYrst
[ ecI_]
[ry_.sp]
■AnalogueOutput(optional)
Set Scaling: [ lOsc]: 0.00, [ hIsc]: 199.99;
Output: [ aOls]: 50.00(Display Value Low),
[ aOhs]: 150.00(Display Value High)
0.00%
100.00%
50.00%
OUTPUT
199.99
100.00
0.00
50.00
150.00
SCALE
[ aOls]
[aOhs]
Set Scaling: [ lOsc]: 0.00, [ hIsc]: 199.99;
Output: [ aOls]: 50.00(Display Value Low),
[ aOhs]: 150.00(Display Value High)
[aOlmt]: 80.00%( of Output Range)
0.00%
100.00%
50.00%
OUTPUT
199.99
100.00
0.00
50.00
150.00
SCALE
Ao.LMt: 80.00%
80.00%
[aOhs]
[ aOls]
[aOlmt]

Rev 1.1
2023-01-11 7/11
CS2-SG
CS2-SG
Fine zero & span adjustment:
Users can get Fine Adjustment of analogue output by front key of
the meter. Please connect standard meter to the terminal of
analogue output. To press the front key(up or down key) of meter to
adjust and check the output.
Zero adjust [aOzro]: Fine Zero Adjustment for Analog Output;
Settable range: -38011~27524;
Span adjust [aOspn]: Fine Span Adjustment for Analog Output;
Settable range: -38011~27524;
Remote Display:
The meter will show the value that received from RS-485 command.
In past, The meter normally receive 4~20mA or 0~10V from AO or
digital output from BCD module of PLC .We support a new solution
that PV shows the value from RS-485 command of master so that
can be save cost and wiring from PLC.
When the [ dsply] set to be RS-485, it means, the PV screen will
show the number from RS-485 command & data. The data(number)
will be same as PV that will compare with set-point, analogue
output and ECI functions so that is to control analogue output, relay
energized and so on.
Field Calibration:
In pass time, engineers have take a lot of time to adjust meters
or converter to meet the structure of machinery zero and span for
the Load Cell measuring. Now, our CS2-SG support easier process
to do it called “Field Calibration”
High Speed Mode: Code: -HSM
According the scaling, the controller can be specify higher
sampling rate up to 60times/second(Average set to be 1). The relay
trip, analogue output will be quicker response according to update
of Present Value.
Calibration System calibration by front key. The process of
calibration, please refer to the operating manual.
CS2 series supports Modbus RTU mode protocol to be used as
Remote Terminal Unit (RTU) for monitoring and controlling in a
SCADA (Supervisor Control And Data Acquisition) system. The
baud rate can be up to 38400 bps. It's not only can be read the
measured value and DI (external control inputs) status but also
controls the relays output (DO) by RS-485 communication ports.
CS2-SG & SC2-RL APPLICATION MEASURING & RS485 COMMUNICATION
RS 485 Modbus RTU Mode (up to 38400bps)
RS485 wiring 1.2KM maximum
Terminate Resistor:
Analog Output 0~10V/0(4)~20mA
3 External Control Inputs:
Functions: Relative PV / PV hold / Reset for
Max(Mini) / Reset for relay latch / DI
Application: Leakage test / Thermal Switch /
Circuit Breaker / Push Button / …
4 Relay outputs:
Functions: Hi(Lo) alarm / Hi(Lo)
energized hold / Go
function / DO function
RL1
RL2
RL3
RL4
COM
ECI1
ECI2
ECI3
ENT/FN
◄
▲
▼
Kg/cm
2
RL1
RL2
RL3
RL4
COM
ECI1
ECI2
ECI3
ENT/FN
◄
▲
▼
Kg/cm
2
■RS-485Communication(optional)
CS2 APPLICATION FOR REMOTE DISPLAY FROM RS485 COMMAND
RS 485 Modbus RTU Mode (up to 38400bps)
RS485 wiring 1.2KM maximum
Terminate Resistor:
RL1
RL2
RL3
RL4
COM
ECI1
ECI2
ECI3
ENT/FN
◄
▲
▼
Kg/cm
2
RL1
RL2
RL3
RL4
COM
ECI1
ECI2
ECI3
ENT/FN
◄
▲
▼
Kg/cm
2
Please accord to the numbers to do the field calibration
(ððððð)
RL1
RL2
2
RL3
RL4
COM
CI1
1
CI2
CI3
ENT/FN
▲
▲
▼
Kg
2
caLlo
caLls
3
5
caLhi
caLhs
6
4
Calibration
Low
Calibration
High
1
2
3
4
5
6
1

Rev 1.1
2023-01-118/11
CS2-SG
CS2-SG
■ ErrorMessage
Before power on, please check the specification and connection again.
Self-diagnosis and error code:
DISPLAY DESCRIPTION REMARK
ovfl Display is positive-overflow (Signal is over display range) (Please check the input signal)
-ovfl Display is negative-overflow (Signal is under display range) (Please check the input signal)
ovfl ADC is positive-overflow (Signal is higher than input range high 20%) (Please check the input signal)
-ovfl ADC is negative-overflow (Signal is lower than input range low -20%) (Please check the input signal)
eep fail EEPROM occurs error (Please send back to manufactory for repaired)
aiCng pv Calibrating Input Signal do not process (Please process Calibrating Input Signal)
aic fail Calibrating Input Signal error (Please check Calibrating Input Signal)
aoCng pv Calibrating Output Signal do not process (Please process Calibrating Output Signal)
aoc fail Calibrating Output Signal error (Please check Calibrating Output Signal)
■ Numeric Screens
0.8"(20.0mm) red high-brightness LED for 4 2/3 or 5 digital present
value.
• I/O Status Indication
• Relay Energized: 4 square red LED
RL1 display when Relay 1 energized
RL2 display when Relay 2 energized
RL3 display when Relay 3 energized
RL4 display when Relay 4 energized
• External Control Input Energized: 3 square green LED
ECI1
display when E.C.I. 1 close(dry contact)
ECI2
display when E.C.I. 2 close(dry contact)
ECI3 display when E.C.I. 3 close(dry contact)
• RS-485 Communication: 1 square orange LED
• COM will flash when the meter is receive or send data, and COM flash
quickly means the data transient quicker.
• Max/Mini Hold indication: 2 square orange LEDs
• M.H
displayed: When the display function has been selected in
Maximum or Minimum Hold function.
■FrontPanel
Stickers
Each meter has a sticker what are functions and engineer label
enclosure.
• Relay energized mode:
• E.C.I. functions mode:
PV.H(PV Hold) / Tare / DI(Digital Input)
M.RS(Maximum or Minimum Reset) /
R.RS(Reset for Relay Latch)
• Engineer Label: over 80 types
HN
PV.H
M.RS
R.RS
Tare DI
Hi Lo LL DO
■OperatingKey
4 keys for Enter(Function) / Shift(Escape) / Up key / Down key
Setting Status Function Index
Up key Increase number Go back to previous function
index
Down key Decrease number Go to next
function index
Shift key Shift the setting
position
Go back to this function
index, and abort the setting
Enter/Fun key Setting Confirmed and
save to EEProm
From the function index to
get into setting status
RL1
RL2
2
RL3
RL4
COM
CI11
CI2
CI3
ENT/FN
▲
▲
▼
Kg
M.H
M.H
Relay status
Indication Comm. status
Control Input status
Max. Hold status
Display
screen
Operation
Key
Mini. Hold status
Engineer Unit
Pass Code Pcode : Settable range: 0000~9999;
User has to key in the right pass word so that get into【Programming
Level】. Otherwise, the meter will go back to measuring page. If
user forgets the password, please contact with the service window.
Function Lock
• None none : no lock all.
• User Level user : User Level lock. User can get into
User Level for checking but setting.
• Programming Level eng : Programming level lock.
User can get into programming level for checking but setting.
• ALL all : All lock. User can get into all level for checking but setting.
Front Key Function
• The Key can be set to be the same function as the setting of
ECI1.
Ex. The ECI1 set to be pVhld and the function [E1=up] set to
be yes in [ eci group]. When user presses Key, the PV
will hold as like as ECI1 close.
• The Key can be set to be the same function as the setting of
ECI2.
Ex. The ECI2 set to be reLpv and the function [E2=dn] set to be
yes in [ eci group]. When user presses Key, the PV will
show relative value as like as ECI2 close.
If the front key function has been set, the terminal input for ECI will
be disabling.

Rev 1.1
2023-01-11 9/11
CS2-SG
CS2-SG
0
aOspn
0
aOzro
10)00
aOhs
)0
aOls
A4-20
aOtyp
no
E1=up
12
debnc
Mrst
ecI3
pVhld
ecI2
reLpv
ecI1
)0)0
ry!fd
)0)0
ry!rd
0
ry!hy
hi
ry!md
)0)0
rYsd
0
rYsb
eci
group
ao
group
relay
group
input
group
ver!4
no
29999
00)00
no
10000
10000
10000
10000
Next Page Next Page Next Page
■ OperatingDiagram
Power ON
Self-diagnosis
*****
ver!4
cs@sg Show the model
number
Show
the firmware
version
MEASURING PAGE
Press 1 sec
Press 1 sec
168*8
ENT
User Level (5Digi-Display Max range:99999 )
Press for 1 sec. can back to Measuring
min
min: the Minimum
value of PV saving
Checking only
0-6
max
max: the Maximum
value of PV saving
Checking only
0-7
29999
pv
If the [dsply]
function has been
set RS-485
pv:Present Value
0-0-1
ry!sp
ry!sp: Relay 1
Set-point
-19999~+29999
/99999
0-1
ry@sp
ry@sp: Relay 2
Set-point
-19999~+29999
/99999
0-2
ry#sp
ry#sp: Relay 3
Set-point
-19999~+29999
/99999
0-3
ry$sp
ry$sp: Relay 4
Set-point
-19999~+29999
/99999
0-4
rYrst
rYrst: Reset for
energized Latch of
Relay
yes / no
0-5
Mrst
Mrst: Reset the
saved value of
Maximum &
Minimum
yes / no
0-8 0-9
cs@pr
model number
firmware
version
Checking only
299(9
pv
Cyclic to first page
0-10
enter
Pcode
Enter the
password to
access Engineer
Level
Default:1000
ENT
Pass Code
NO
YES Programming Level(5Digi-Display Max range:99999 )
Press for 1 sec. can back to Measuring
INPUT
GROUP
ENT
RELAY
GROUP
ENT
rYsb: Start band
for Relay energized
0~9999counts
B-1
ecI1: External
Control Input 1
none / reLpv /
pVhld / Mrst /
rYrst / di /
C-1
ecI3: External
Control Input 3
none / reLpv /
pVhld / Mrst /
rYrst / di /
C-3
ecI2: External
Control Input 2
none / reLpv /
pVhld / Mrst /
rYrst / di /
C-2
debnc:
Debouching of
external control
Input
5~255(x8ms)
C-4
EXTERNAL
CONTROL INPUT
(E.C.I) GROUP
ENT
E1=up: ECI.1
set to be UP Key
function
yes / no
C-5
aOtyp: Analogue
Output type
selection
V0-10 / V 0-5 /
V 1-5 / A0-10 /
A0-20 / A4-20
D-1
aOls: Analogue
Low Output versus
Low Scale
-19999~+29999
/99999
D-2
aOhs: Analogue
High Output
versus High Scale
-19999~+29999
/99999
D-3
aOzro: Fine Zero
Adjustment for
Analogue Low
Output
-38011~+27524
D-4
aOspn: Fine
Span Adjust. for
Analogue High
Output
-38011~+27524
D-5
ANALOGUE
OUTPUT
GROUP
ENT
rYsd: Start delay
time for Relay
energized
0.00.0~
9(M).59.9(S)
B-2
ry!md: Relay 1
energized mode
off / lo /
hi / lOhld /
hIhld / do
B-3
ry!hy: Relay 1
Hysteresis
0~5000 counts
B-4
ry!rd: Relay 1
energized delay
time
0.00.0~
9(M).59.9(S)
B-5
ry!fd: Relay 1
de-energized
delay time
0.00.0~
9(M).59.9(S)
B-6
10)00
aIhi
A-2
ai.hi:Analogue
input low
0.00~100.00%
Next Page
)00
aIlo
A-1
aIlo: Analogue
input low
0.00~100.00%
0
pVdp
pv.dp: Analogue
input High
0 / )0 / )00 /
)000 / )0000
A-3
0
lOsc
lOsc: Decimal
Point
-19999~+29999
/99999
A-4
10000
hIsc
hIsc: High scale
of PV
-19999~+29999
/99999
A-5 A-6
0
pVzro
pVzro: Fine Low
point Adjustment
for PV display
-19999~+29999
/99999

Rev 1.1
2023-01-1110/11
CS2-SG
CS2-SG
NstB2
9600
1
adres
baud
prity
rs485
group
11)00
aOlnt
none
ZSclrZSclr
no
E2=dn
)0)0
ry$fd
)0)0
ry$rd
0
ry$hy
0
ry$hy
hi
ry$md
)0)0
ry#fd
)0)0
ry#rd
0
ry#hy
hi
ry#md
)0)0
ry@fd
)0)0
ry@rd
0
ry@hy
hi
ry@md
0
lOcut
A-10
lOcut: Low Cut
Function
-19999~+29999
/99999
5
avg
A-11
avg: Average
update for PV
1(None)~99 times
1
mVavg
A-12
Mavg: Moving
Average
update for PV
1(None)~
10 times
0
Dfilt
A-13
Dfilt: Digital filter
0(None)/
1~99 times
0000
Pcode
A-14
Pcode: Pass Code
for enter
Engineer Level
0000~9999
none
Flock
A-15
Flock: Function
Level Lock
none / user /
eng / all
0
pVspn
A-7
pVspn: Fine High
point
Adjustment for PV
display
-19999~+29999
/99999
none
ZSclr
A-8
ZSclr: Clear
Fine Zero & Span
Adjustment for PV
display
none / pVzro /
pVspn / both
pv
dsply
dsply: Display
Function
pv / minIh /
maXh / rs485
A-9
Plesae refer to operating manual for detail description
ry@md: Relay 2
energized mode
off / lo /
hi / lOhld /
hIhld / do
B-7
ry@hy: Relay 2
Hysteresis
0~5000 counts
B-8
ry@rd: Relay 2
energized delay
time
0.00.0~
9(M).59.9(S)
B-9
ry@fd: Relay
2 de-energized
delay time
0.00.0~
9(M).59.9(S)
B-10
ry#md: Relay 3
energized mode
off / lo /
hi / lOhld /
hIhld / do
B-11
ry#hy: Relay 3
Hysteresis
0~5000 counts
B-12
ry#rd: Relay 3
energized delay time
0.00.0~
9(M).59.9(S)
B-13
ry#fd: Relay
3 de-energized
delay time
0.00.0~
9(M).59.9(S)
B-14
ry$md: Relay 4
energized mode
off / lo /
hi / lOhld /
hIhld / do /
go-!2 / go-23
B-15
ry$hy: Relay 4
Hysteresis
0~5000 counts
B-16
ry$rd: Relay 4
energized delay
time
0.00.0~
9(M).59.9(S)
B-17
ry$fd: Relay
4 de-energized
delay time
0.00.0~
9(M).59.9(S)
B-18
E2=dn: ECI.2 set
to be Down Key
function
yes / no
C-6
ZSclr: Zero &
Span Clear for
Adjustment
none / aOzro /
aOspn / both
D-6
aOlmt: Analog
Output High Limit
0.00~110.00%
D-7
adres:Device
number of the
meter
1~255
E-1
baud: Baud rate
1200 / 2400 /
4800 / 9600 /
19200 / 38400
E-2
prity: Parity
n.Stb.1 / n.Stb.2 /
odd / EvEn
E-3
RS-485
GROUP
ENT

Rev 1.1
2023-01-11 11/11
CS2-SG
CS2-SG
MEASURING
PAGE
Enter the password
to access
Programming Level
Enter the password
to access Field
Calibration Level
16*88
■ FieldCalibration
Pcode
enter
enter
Fcode
Pass Code
NO YES
default=2000
Press
caLls: the value to be set is relative to Field Calibration lower poin
»Press
to set the value of lower scale.
caLls
00000
F-2
Adjust the structure to be a lower signal output status (or any lower status)
and keep it in stable.
caLlo: Field Calibration Low
»Press
to read signal of the lower status.
»Press
again to finish the calibration lower point, and go to next page.
F-1
caLlo
00000
Adjust the structure to be a higher signal output status (or any higher
status) and keep it in stable.
caLhi: Field Calibration High
»Press
to read signal of the higher status.
»Press
again to finish the calibration higher point, and go to next page.
F-3
caLhi
00000
caLhs: the value to be set is relative to Field Calibration higher point
»Press
so set the value of Higher scale
caLhs
F-4
00000
Csel: Calibration parameter selection
»Press
to access the function and stand by selection
»Press or to select
(default: defld);
Settable: defld / field
defld (default calibration) field (Field calibration)
Csel
F-5
de f lt
◎Once the user select field calibration, the [lOsc](step A-2) and [hIsc](A-3)
will be instead of [caLls] and [caLhs], and can not to be change. If user has to
change the scaling, it's the only way to access field calibration level to set in
[
caLls](step F-2) and [caLhs](step F-4).
◎Please double check the [lOsc](step A-2) and [hIsc](A-3) are correct after
selection the deflt or field .
&
Other manuals for CS2-SG
1
Table of contents
Other ADTEK Measuring Instrument manuals

ADTEK
ADTEK AEM-DRB User manual

ADTEK
ADTEK CS2-SG User manual

ADTEK
ADTEK AFM-8A User manual

ADTEK
ADTEK CPM-20 Series User manual

ADTEK
ADTEK CS3S-TR User manual

ADTEK
ADTEK CM1 Series User manual

ADTEK
ADTEK AEM-DRK User manual

ADTEK
ADTEK MWH-10A User manual

ADTEK
ADTEK CS2-T User manual

ADTEK
ADTEK NED User manual
Popular Measuring Instrument manuals by other brands

HP
HP 436A Operating and service manual

Ludlum Measurements
Ludlum Measurements 30 Series manual

Absolute Process Instruments
Absolute Process Instruments Cecomp DPG2000B Series instructions

Keit
Keit IRmadillo DOC1048R user manual

Emerson
Emerson Rosemount Analytical X-STREAM X2 Series instruction manual

Camlogic
Camlogic PFG05 Series Use and maintenance manual