CS2 SERIES –OPERATION MANUAL(EN)-A-2017-11-16
EX: Low Cut is set for -0.01. If the display is < -0.01, and
all the display will be -0.01.
Low Cut is set for -0.10, if the PV is
under (< = -0.10), that display will be -0.10.
PV according
to input signal
Digital Fine Adjustment [pVzro] &[pVspn]
Settable range from -19999~+29999 ;
Users can get "Fine Adjustment PV" by front key on
the meter for lower and/or higher points. “Just Key-In”
the value, if user wants to show the value in input
signals currently.
Especially, the [pVzro]&[pVspn]are not only in
zero & span of PV, but also randomly lower point in
function [pVzro]& randomly higher point in function
[pVspn]. The meter will be auto-linearization for full
scale.
The adjustment can be cleared in function [ZSclr].
Default Input: [lOsc]: )00, [hIsc]: 19(99
Digital fine adjustment:
input 2.75V, [pVzro]set to be 8)00
input 9.00V, [pVspn]set to be 17)00
Display: Low point from 55.00 to 80.00
High point from 180.00 to 170.00
*Please make sure that the [pVzro]point must be less
than [pVspn]during the process of digital fine
adjustment. Generally, the interval should be over 50%
of input range. If the interval is too narrow, that may be
made bigger error between zero and span,
Square root function [sQrot]
Selectable no or yes ;
The function can be set no or yes in【input
group】to measure the signal from differential
pressure flow-meter. The formula = √(Pv/HS) x HS
The meter will calculate the sampling 1-3 and update
the display value. At meantime, the sampling 4-6 will be
processed to calculate.
Display Update Value =
(Sample 1 + Sample 2 + Sample 3)/3
Display Update Value =
(Sample 4 + Sample 5 + Sample 6)/3
Moving Average update [mVavg]
Settable range: 0(no function)/1~10 times;
Jittery Display caused by the reasons as like as
noise or unstable signal. User can set the times to
average the readings, and get smoothly display.
Remark: To set the moving average times with higher
wouldn’t affect the response time of Relay and
Analogue output except the 1st updated display value
will be slower.
The meter’s sampling is 15cycle/sec. If the
[mVavg](Moving Average) set to be 3expressed
the display update with 15 times/sec.,
In the first updated display value will be same as
average function. In the next updated display value, the
function will get the new fourth sample (sample 4) then
throw away the first sample (sample 1) that the newest
3 samples(sample 2,3,4) will be calculated for the
updated display value.
Moving Average set to be 3
In first 3 samples, Display Update Value
= (Sample 1 + Sample 2 + Sample 3)/3
Display Update Value
= (Sample 2 + Sample 3 + Sample 4)/3
Display Update Value
= (Sample 3 + Sample 4 + Sample 5)/3
Display Update Value
= (Sample 4 + Sample 5 + Sample 6)/3
Digital filter [Dfilt]
Settable range from 0(None)/1~99 times.
The digital filter can reduce the influence of spark
noise by magnetic of coil.
If the values of samples are over digital filter
band(fixed in firmware and about 5% of stable reading)
3 times (Digital Filter set to be 3) continuously, the meter
will admit the samples and update the new reading.
Otherwise, it will be as treat as a noise and skip the
samples.
Digital Filter set to be 3
The values of samples are
two times only, so that the
reading will admit the
samples.
The values of samples are
over the filter band and three
times, so that the reading will
accept the samples, and
update the reading.
■Reading Stable Functions
Average Display update [avg]
Settable range: 1~99 times;
Jittery Display caused by the noise or unstable
signal. User can set the times to average the readings,
and to get smoothly display.
Remark: To set the average times with higher will make
the response time slower of Relay and
Analogue output.
The meter’s sampling is 15cycle/sec
If the [avg](Average) set to be 3to express the
display update with 5 times/sec.
Amend 2009/11/14: Add Square root function