PREPARATIONFORUSE
SAFETY
Beforeconnectingtheinstrumenttothepowersource,
carefullyreadthefollowinginformation,thenverifythat
thepowercordandpowerlinefuseareonesforyourpower
line.
Theinstrument'srearpanelhasa fuseholderontheleftof
theACinletterminal(thefuseholderalsoservesasa
voltageselector).Thevalueunderthetriangle• marked
ontheholderindicatesthelinevoltagesetfortheinstru-
ment.Ifthepowercordisnotappliedwiththecorrect
voltage,thereisa dangerfromelectricshock.
Linevoltage
Thisinstrumentoperatesusingac-powerinputvoltages
that100/120/220/240V atfrequenciesfrom50Hzto
60Hz.
Powercord
Thegroundwireofthe3-wireacpowerplugplacesthe
chassisandhousingoftheoscilloscopeatearthground.Do
notattempttodefeatthegroundwireconnectionorfloat
theoscilloscope;todosomayposea greatsafetyhazard.
Theappropriatepowercordissuppliedbyanoptionthatis
specifiedwhentheinstrumentisordered.
TheoptionalpowercordsareshownasfollowsinFig.1.
Linefuse
Thefuseholderislocatedontherearpanelandcontains
thelinefuse.Verifythattheproperfuseisinstalledby
replacingthelinefuse.
EQUIPMENTPROTECTION
1.
Neverallowa smallspotofhighbrilliancetoremainsta-
tionaryonthescreenformorethana fewseconds.The
screenmaybecomepermanentlyburned.A spotwilloc-
curonlywhenthescopeissetupforX-Yoperationand
nosignalisapplied.Eitherreducetheintensitysothe
spotisbarelyvisible,switchbacktonormalsweep
operationwhennosignalisapplied,orsetupthescope
forspotblanking.
2.
Nevercovertheventilatingholesonthetopofthe
oscilloscope,asthiswillincreasetheoperating
temperatureinsidethecase.
3.Neverapplymorethanthemaximumratingtothe
oscilloscopeinputs.
CH1,
CH2INPUTjacks:
500Vp-por250V (dc+ acpeak)
EXTTRIGINPUTandZ AXISINPUTjacks:
50V (dc+ acpeak)
Neverapplyexternalvoltagetotheoscilloscopeoutput
terminals.
4.
Alwaysconnecta cablefromtheearthground(GND)
jackoftheoscilloscopetothechassisoftheequipment
undertest.Withoutthiscaution,theentirecurrentfor
theequipmentundertestmaybedrawnthroughthe
probeclipleadsundercertaincircumstances.Suchcon-
ditionscouldalsoposea safetyhazard,whichthe
groundcablewillprevent.
5.Alwaysusetheprobegroundclipsforbestresults.Do
notuseanexternalgroundwireinlieuoftheprobe
groundclips,asundesiredsignalsmaybeintroduced.
6.Operationadjacenttoequipmentwhichproducesstrong
acmagneticfieldsshouldbeavoidedwherepossible.
Thisincludessuchdevicesaslargepowersupplies,
transformers,electricmotors,etc.,thatareoftenfound
inanindustrialenvironment.Strongmagneticshields
canexceedthepracticalCRTmagneticshieldinglimits
andresultinterferenceanddistortion.
7.Probecompensationadjustmentmatchestheprobeto
theinputofthescope.Forbestresults,compensation
ofprobeshouldbeadjustedinitially,thenthesamepro-
bealwaysusedwiththeinputofscope.Probecompen-
sationshouldbereadjustedwhenevera probefroma
differentscopeisused.(Seepage17)
8.InX-Yoperation,donotpulloutthePULLx10MAG
switch.
Ifpulledoutit,noisemayappeareonthe
waveform.
7