Novus NP600 User manual

NOVUS AUTOMATION. 1/3
NP
600
RELATIVE
PRESSURE TRANSMITTER – INSTRUCTIONS MANUAL – V1.0x F
PRESENTATION
The NP600 positive relative pressure transmitter series are robust
and reliable equipment suitable for general industrial applications. Its
great advantage is the possibility of a fast and easy configuration of
the operating range. With the aid of a software, the user has access
to the most important information of the transmitter and can set the
most appropriate configuration for the process.
The NP600 model has a ceramic sensor with high mechanical
stability.
PRECAUTION
Before operating the transmitter, carefully read its specifications and
operating instructions. In case of damage caused by incorrect
operation or inappropriate usage, and its consequences, the
guarantee becomes ineffective and null.
A specialized professional shall perform the installation.
A specific electrical power supply network should be provided for
instruments use only.
UNPACKING
After unpacking the product, besides the transmitter it must be
available:
•A quick guide to installation and operation;
•A sealed electrical connector;
•A screw to adjust the transmitter and the connector.
IDENTIFICATION
An identification label is attached to the equipment housing. On this
label, you can found detailed information about the transmitter.
Fig. 1 shows the identification label with their respective information.
Fig. 1 – Transmitter identification
SPECIFICATIONS
Type of Measure:
Positive relative pressure
Maximum Range (Max. Range):
0.2 MPa (2 bar / 29.01 psi)
0.5 MPa (5 bar / 72.52 psi)
1 MPa (10 bar / 145.04 psi)
2 MPa (20 bar / 290.08 psi)
5 MPa (50 bar / 725.19 psi)
10 MPa (100 bar / 1450.38 psi) (*)
Rangeability:
3:1
Measurement Accuracy:
≤ (50 bar / 725.19 psi) ............................... < ±0.5 % Max. Range
100 bar / 1450.38 psi ................................. < ±1.0 % Max. Range
Including linearity, hysteresis and repeatability, 25 °C
Thermal Maximum Deviation:
< ±0.06 % Max. Range / °C
Overpressure Range:
Twice the value specified for the Maximum Measurement Range
Output Signal (Output):
Electric currente, 4-20 mA, 2 wires.
Power Supply (Power):
11 to 33 Vdc
Electrical Connection:
Connector for Type A valves (DIN EN 175301-803), IP65.
Conductor 1.5 mm² (max.) and cables between 6 and 8 mm in
diameter.
Operating temperature:
-20 to 70 °C / -4 a 158 °F
Dinamic Response:
< 30 ms
Process Conection (Thread):
¼ NPT, ½ NPT, ½ BSP, G ¼ (*)
Sensor features:
Piezoresistive – Thick Film on Ceramics (Al2O396 %)
Metal housing of the transmitter:
Stainless 316
Wetted parts:
Sensor Ceramic, Al2O396 %
Sealing Ring, FKM
Metal Connection, stainless steel 316
Compatibility: Any gas or liquid compatible with the constituent
materials of the wetted parts.
(*) Information available on the product identification label.
NP600 Pressure Transmitter
NOVUS AUTOMATION 2/3
ELECTRICAL CONNECTIONS
Fig. 2 – Electrical Connections
Recommendations for installation
•Sensor signals conductors must go through the plant system
separate from power leads (loop). If possible, in grounded
conduits.
•The instruments must be powered from the instrumentation power
supply circuit.
•It is mandatory the use of RC FILTERS (electrical noise
suppressors) in contactor coils, solenoids, etc.
•To improve the stability, immunity and safety of the measurement,
it is recommended to use the system grounding.
CONFIGURATION
When the transmitter is used with the factory setting, no further
action is required and the transmitter is ready to be installed.
When a new configuration is required, you should use the
Transmitter Configuration Kit. This kit includes the TxConfig-II
software, distributed free on the NOVUS website, and the TxConfig-
USB-DIN43650 interface, which can be purchased in our distribution
and reseller network.
To install the TxConfig-II software, run the setup file
TxConfigIISetup.exe and follow the instructions of the installer.
The TxConfig interface must be connected to the transmitter
according to Fig. 3. The interface itself provides the electrical power
necessary for transmitter operation during configuration.
Fig. 3 – Connection of the TxConfig-USB
SOFTWARE CONFIGURATION
Fig. 4 – TxConfig II software main screen
When you run the TxConfig-II software, it will recognize
automatically the transmitter and will display the following
information:
1. General Information:
This field shows information that identifying the transmitter. This
information should be sent to the manufacturer in an eventual
request for technical assistance.
2. Unit:
Setting the pressure unit assumed for the measuring range
setting fields.
3. Measuring Range:
Setting the measurement range assumed by the transmitter.
Lower Limit: Target pressure value for 4 mA current.
Upper Limit: Target pressure value for 20 mA current.
Minimum Range
Do not set a lower band (span) that the Minimum Range
indicated below in this same field.
4. Sensor Failure: Sets the current output behaviour when the
transmitter experiences a fault:
Minimum: Output current goes to < 4 mA.
Maximum: Output current goes to > 20 mA.
5. Zero Correction: It corrects small deviations presented in the
transmitter output, when there is no applied pressure.
6. Send Configuration: It applies the new setup. Once sent, the
setup will be immediately adopted by the transmitter.
7. Read Configuration: Reads the current setup in the transmitter
connected. The screen now presents the current setup that may
be changed by the user.
NP600 Pressure Transmitter
NOVUS AUTOMATION 3/3
DIMENSIONS
Fig. 5 – Transmitter dimensions
THREAD A (mm) B (mm) WEIGHT (g)
NPT ¼ 15.5 51.0 106
NPT ½ 20.0 56.5 159
BSP ½ 14.0 49.5 121
G ¼ 14.0 49.5 105
Table 1 – Dimensions
WARRANTY
Warranty conditions are available on our website
www.novusautomation.com/warranty.

Other Novus Transmitter manuals

Novus NV-202VIP User manual

Novus

Novus NV-202VIP User manual

Novus RHT-WM-485-LCD User manual

Novus

Novus RHT-WM-485-LCD User manual

Novus RHT-P10 User manual

Novus

Novus RHT-P10 User manual

Novus TL400-V User manual

Novus

Novus TL400-V User manual

Novus TxRail-USB User manual

Novus

Novus TxRail-USB User manual

Novus NP640 User manual

Novus

Novus NP640 User manual

Novus TxRail User manual

Novus

Novus TxRail User manual

Novus NP785 User manual

Novus

Novus NP785 User manual

Novus TxMiniBlock User manual

Novus

Novus TxMiniBlock User manual

Novus NP785 User manual

Novus

Novus NP785 User manual

Novus TxRail-USB User manual

Novus

Novus TxRail-USB User manual

Novus NP-430D User manual

Novus

Novus NP-430D User manual

Novus NVPT-131VTP-HD User manual

Novus

Novus NVPT-131VTP-HD User manual

Novus TxIsoRail-HRT User manual

Novus

Novus TxIsoRail-HRT User manual

Novus RHT-WM-485-LCD User manual

Novus

Novus RHT-WM-485-LCD User manual

Novus NP785 User manual

Novus

Novus NP785 User manual

Novus TxRail User manual

Novus

Novus TxRail User manual

Novus TxIsoPack User manual

Novus

Novus TxIsoPack User manual

Novus TxBlock-USB User manual

Novus

Novus TxBlock-USB User manual

Novus NP400 User manual

Novus

Novus NP400 User manual

Novus TXRAIL Series User manual

Novus

Novus TXRAIL Series User manual

Novus RHT-WM User manual

Novus

Novus RHT-WM User manual

Novus NP785-50 User manual

Novus

Novus NP785-50 User manual

Novus TxIsoRail-HRT User manual

Novus

Novus TxIsoRail-HRT User manual

Popular Transmitter manuals by other brands

NVT NV-ET1801 Condensed Installation Guide

NVT

NVT NV-ET1801 Condensed Installation Guide

Servonaut ServoNaut Radio Zwo4 HS16 user manual

Servonaut

Servonaut ServoNaut Radio Zwo4 HS16 user manual

Agilent Technologies 83433A user guide

Agilent Technologies

Agilent Technologies 83433A user guide

Premier PWFMT7 manual

Premier

Premier PWFMT7 manual

MuxLab 500039-2PK Quick installation guide

MuxLab

MuxLab 500039-2PK Quick installation guide

Kenco KT Series Interface manual

Kenco

Kenco KT Series Interface manual

AUREL TX-SAW/433 instruction manual

AUREL

AUREL TX-SAW/433 instruction manual

Alba FM76969 instruction manual

Alba

Alba FM76969 instruction manual

Savant SmartLink HDBaseT HTX-SLN501 Quick reference guide

Savant

Savant SmartLink HDBaseT HTX-SLN501 Quick reference guide

Sennheiser evolution wireless SR 300 IEM Specifications

Sennheiser

Sennheiser evolution wireless SR 300 IEM Specifications

Maytel Co., Ltd Telme-It user manual

Maytel Co., Ltd

Maytel Co., Ltd Telme-It user manual

Jamara SCX 2,4 GHz instructions

Jamara

Jamara SCX 2,4 GHz instructions

Sanrio Hello Kitty KT4510 operating instructions

Sanrio

Sanrio Hello Kitty KT4510 operating instructions

Lightware HDMI-3D-OPT-TX210A quick start guide

Lightware

Lightware HDMI-3D-OPT-TX210A quick start guide

FUTABA 9CAP SUPER instructions

FUTABA

FUTABA 9CAP SUPER instructions

Hycontrol REFLEX RF Series Handbook

Hycontrol

Hycontrol REFLEX RF Series Handbook

Spectron IT-12-Exd operating manual

Spectron

Spectron IT-12-Exd operating manual

newins NT 3400P Series quick start guide

newins

newins NT 3400P Series quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.