Axis B/P Series User manual

USER MANUAL
B/P SERIES
FILE: 2010-10-BP-113 B-50_0 GB

2
USER MANUAL
_____________________________________________________________________________________________________________________
Table of contents:
1.
General description....................................................................................................4
2.
Completation..............................................................................................................4
3.
Technical data ...........................................................................................................5
4.
Balance keys and indicators ......................................................................................6
5.
Security principles......................................................................................................7
6.
nvironment protection..............................................................................................7
7.
Preparation to work....................................................................................................8
8.
Start-up......................................................................................................................9
9.
General rules ...........................................................................................................10
10.
Balance checking.....................................................................................................11
11.
Scale adjustment .....................................................................................................11
12.
Connecting a computer or a printer (optional)..........................................................12
13.
Basic function ..........................................................................................................16
13.1
Normal weighing ...................................................................................................... 16
13.2
Weighing with tare ...................................................................................................17
13.3
Increased readability................................................................................................17
14.
Special function describes .......................................................................................18
14.1
Menu customization function (ACtIV and d FAUL).................................................. 19
14.2
Autozeroing function (AutoZ )................................................................................20
14.3
Pieces counting (PCS)............................................................................................21
14.4
Printer cooperation settings (PrInt) .........................................................................22
14.5
Serial port parameters setting (Port) .......................................................................23
14.6
Constant tare (tAr ).................................................................................................24
14.7
Average calculation function (AV rAG) .................................................................. 26
14.8
Recipe weighing function (R CIP ) .......................................................................27
14.9
Percentage weighing function (P rC) ..................................................................... 28
14.10
Anti-disturbance filter function (FILt r)..................................................................29
14.11
Function for maximum value indication (UP)............................................................30
14.12
Force measuring function (n Wto) .........................................................................31
14.13
Function for weighing large animals (LOC) .............................................................. 32
14.14
Checkweighing function (thr) ...................................................................................33

USER MANUAL
3
_____________________________________________________________________________________________________________________
14.15
Total weight function (totAL).................................................................................... 36
14.16
Function for setting date and time (dAt ) ................................................................ 38
14.17
Function for selecting label number (LAb L).......................................................... 39
14.18
xternal calibration (CALIb)..................................................................................... 40
14.19
Setting time of stabilization function (Stb)................................................................ 41
14.20
ntering reference zero function (Zero)................................................................... 42
14.21
Automatic switching off the scale function (AutoOF)................................................ 43
14.22 Density determining function (d nSIt) .................................................................... 44
14.23
Calculator for good packaging control (tP)............................................................... 46
14.24
Statistical calculations function (StAt)...................................................................... 49
14.25
Paperweight calculation (PAP) ................................................................................ 51
15. Maintenance and small repairs.................................................................................... 52
Declaration of Conformity................................................................................................... 53

4
USER MANUAL
_____________________________________________________________________________________________________________________
1.
General description
B/P platform scales are designed to weigh parcels.
Scales are produced in two executions: with L D (standard) and LED display ( on
demand).
All scales are metrological tested - calibration or legal verification on demand.
Scales have following verification features:
- a seal protecting scale casing against opening,
- notified body stamps and green metrological marking placed on the balance
name plate.
Renewing of scale legal verification is required when protective seal is violated or
after period of 3 years starting from 1
st
December of year when first legal
verification was performed.
In order to renew legal verification please contact authorized service of AXIS.
NA E classification: 29.24.23.
ertificates:
E Type Axis management System
Approval certificate ertificate DIN EN ISO 9001:2000
No. PL 04 022 No. 78 100 6386
2. Completation
Basic set includes:
1. The scale
2. User manual
3. Guaranty

USER MANUAL
5
_____________________________________________________________________________________________________________________
3. Technical data
Scale type B30P B50P
Max load 30kg 50kg
Mi load 200g 400kg
Readout graduatio (d) 10g 20g
Verificatio graduatio (e)
10g 20g
Tare ra ge -30kg -50kg
Accuracy class III
Worki g temperature -10 ÷ +40°C
Weighi g time <3s
Pa dime sio s 400x400mm 390x390mm
Scale dime sio s 400x480x100mm 390x500x130mm
Mass 11kg
Supply ~230V 50Hz 6VA / =12V 0,16A
EC verificatio

6
USER MANUAL
_____________________________________________________________________________________________________________________
4. Balance keys and indicators
klawisz
I/
- włącznik / wyłącznik (standby),
klawisz
→
T
←
- tarowanie (wpisywanie masy opakowania odejmowanej
od masy waŜonej),
klawisz B/G - wskazanie masy brutto,
klawisz
→
0
←
- zerowanie,
klawisz MENU - menu funkcji specjalnych,
klawisz - wydruk wyniku,
klawisz HR - zwiększenie rozdzielczości wskazań masy,
wskaźnik
→
0
←
- wskaźnik wyzerowania (przy nieobciąŜonej wadze),
wskaźnik - sygnalizuje ustabilizowanie się wyniku waŜenia,
wskaźnik
NET - masa netto (po uŜyciu klawisza →T←),
wskaźnik
- masa brutto (po uŜyciu klawisza B/G),
wskaźnik
- wskaźnik funkcji liczenia sztuk (wskazania w sztukach)
wersja z wyświetlaczem LCD:
wskaźnik MODE - wskaźnik włączenia funkcji specjalnej,
wskaźnik
słupkowy
- wskaźnik obciąŜenia wagi (0-100%).
wskaźnik O - wyłączenie wagi klawiszem (standby),
wskaźnik B/G - masa brutto (po uŜyciu klawisza B/G),
wskaźnik pcs - wskazania w sztukach

USER MANUAL
7
_____________________________________________________________________________________________________________________
5. Sec rity principles
To avoid electrical shock or damage of the scale or
connected peripheral devices, it is necessary to follow
the security rules below.
• All repairs and necessary regulations can be made by authorised personnel
only.
• To avoid fire risk use a feeder of an appropriate type (supplied with the
scale). Pay attention that supply voltage is compatible with specified
technical data.
• Do not use the scale when its cover is opened.
• Do not use the scale in explosive conditions.
• Do not use the scale in high humidity.
• If the scale seems not to operate properly, unplug it from the mains and do
not use until checked by authorised service.
6. Environment protection
According to legal regulations it if forbidden to dispose
wasted electronic equipment in waste containers.
• Please return wasted scale to the point of purchase or other company
specialised in recycling of wasted electronic components.

8
USER MANUAL
_____________________________________________________________________________________________________________________
7. Preparation to work
1. Take the scale out of the package
removing protective foils.
2. Remove the pan and remove protection
elements from below the pan.
3. Place the scale on a stable ground not
affected by mechanical vibrations and
airflows.
4. Level the scale with the rotating rear legs
so that the air bubble in the water-level at
the back of the scale is in the middle.
5. Put on the pan.
6. Plug the feeder connector to the power
socket at the back of the scale and plug the
feeder to the mains. After self-tests and
result stabilisation, the scale displays zero
indication.
orrect Wrong

USER MANUAL
9
_____________________________________________________________________________________________________________________
8. Start- p
Leave the pan empty, plug the feeder to the mains (~230V/50Hz) and plug the
feeder connector to the 12V power socket at the back of the scale.
The scale proceeds with following start-up actions:
Display test.
Basic electronic components tests: C1,
C-2, ... C-6.
Program version.
The scale is now ready to work..
Attention:
In scales with external feeder, connect the feeder to network socket and then
connect the output feeder joint with 12V joint from the scale.

10
USER MANUAL
_____________________________________________________________________________________________________________________
9. General r les
1. Before each measurement make sure that zero indicator is displayed. If zero
indicator does not displayed or “----“ communicate appears, press →0← key
and wait until zero indication and zero indicator appears.
2. The scale is equipped with a tare equal to its range. To tare the scale press
→T← key (left or right). Storing a tare value does not extend measuring range,
but only subtracts it from a load placed on a pan. To make weight control easier
and to avoid range overdrawing, the scale is equipped with a load indicator
(graduated in percentages).
3. Weighing result should be read when the indicator " " lights, which signalises
stabilisation of a result.
4. When the scale is not used but it is necessary to be ready to work immediately, it
can be switched off by pressing I/ key. The scale reading system is then
switched off to "standby" mode (signalled by the indicator "OFF" in version
with the L D display). To switch the scale on press I/ key. The scale is
immediately ready to operate maximum accuracy (after self tests).
5. Weighed sample should be placed in the centre of the pan.
Protect the scale against dust, aggressive dusts and liquids. To clean the
scale wash it with water with soap and dry it afterwards.
Place the scale on a platform to avoid dropping weighed objects
on the pan.
Do not overload the scale more then 20% of maximum load (Max .

USER MANUAL
11
_____________________________________________________________________________________________________________________
10. Balance checking
It is advised to check scale indication accuracy before and after series of
measurement using any load with known weight.
To check the scale with legal verification use a calibration weight with valid
calibration certificate. In case permissible error is exceeded it is advised to contact
the nearest service to calibrate the scale.
11. Scale adj stment
Scale adjustment (calibration with external standard of mass) should be made, if
the scale precision is unsatisfactory. In that situation M1 standard of mass should
be used (according to OIML) with the value equal to scale Max range (technical
data) and a special function CALIb detailed in 14.18 chapter.
CALIb function is only available in scales not designed for legalization.
To adjust a balance it is necessary to break protective marks, therefore
to calibrate the scale please contact the nearest service.

12
USER MANUAL
_____________________________________________________________________________________________________________________
12. Connecting a comp ter or a printer (optional)
The scale is equipped with RS232 or
RS485 (optional), which can be used to
connect external devices such as a computer
or a printer. Detailed way of communication
data is defined by one of protocol:
- LonG – protocol to printer and computer
- ELtron – protocol to label printer
When cooperating with a printer data is send
automatically after result stabilisation, but
next transmission is possible after removing
previously weighted sample.
When protocol Long is chose scale sends
result of weighting on an initialize signal
from computer, or after pressed a key.
When automatic transmission mode is chose
(function LPt), a printer data is send
automatically after result stabilisation, but
next transmission is possible after removing
previously weighted sample. The scale sends
following information: successive number of
weighing, date and time (if clock module is
installed )and weighing result. During
transmission communicate Print is displayed
.
Scale with Eltron protocol sending record of data after pressed key.
The scale sends following information: number of label, date and time (if clock
module is installed ) and weighing result. During transmission communicate Label
is displayed.
In special features scale can be equipped in second RS232 interface for additional
functions, ex. continues transmit of indicate to second display.
omputer must have a special program for cooperation with a data from a scale.
This programs are procurement by Axis.
RS
-
2
32C
RS485
(option)
R D (receiver)
T D (transmitter)
GND
Case
1
2
3
4
5
6
7
8
9
A
Input
Output
A
B
B
Case
1
2
3
4
5
6
7
8
9
}
}

USER MANUAL
13
_____________________________________________________________________________________________________________________
Detailed protocol description
LonG protocol
ommunication parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps,
initialising signal (data send order) – correspond to press key
omputer→Scale: S I R LF (53h 49h 0Dh 0Ah),
Scale→omputer: scale response according to description below (16 bytes):
Byte
1 -
sign „-” or space
Byte
2 -
space
Byte
3÷4 -
digit or space
Byte
5÷9 -
digit, decimal piont or space
Byte
10 -
digit
Byte
11 -
space
Byte
12 -
k, l, c, p or space
Byte
13 -
g, b, t, c or %
Byte
14 -
space
Byte
15 -
R
Byte
16 -
LF
Attention:
Network number different than zero (F..-rS / nr function) changes scale working
mode: communication with a computer is possible after logging the scale in with
02h scale number command. To log the scale out use 03h command.
For example: Using a program to test RS232 interface ( program is available in
www.axis.pl / programy komputerowe ) for scale number 1 please write: $0201 to
log in, then SI, and write: $03 to close communication.
Scale tarring (calling
→
T
←
key press) :
omputer→Scale: S T R LF (53h 54h 0Dh 0Ah),
Scale→omputer: without response,
Scale zeroing (calling
→
0
←
key press):
omputer→ Scale: S Z R LF (53h 5Ah 0Dh 0Ah),
Scale →omputer: without response,
Scale turning on / off (calling
I/
key press):
omputer→ Scale: S S R LF (53h 53h 0Dh 0Ah),

14
USER MANUAL
_____________________________________________________________________________________________________________________
Scale →omputer: without response,
Entering to special function menu (calling MENU key press):
omputer→ Scale: S F R LF (53h 46h 0Dh 0Ah),
Scale →omputer: without response,
Setting low threshold value (option):
omputer→ Scale: S L D1...DN R LF (53h 4 h D1...DN 0Dh 0Ah)
D1...DN – threshold value, maximum 8 characters („-” – negative value, digits,
dot – decimal separator), number of digits after dot should be the same as on
scale display,
Scale →omputer: without response,
Example:
⋅ in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order
should be sent:
S L 1 0 0 0 . 0 R LF (53h 4 h 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah),
⋅ in order to set low threshold 100kg in scale B150 (d=50g) the following order
should be sent:
S L 1 0 0 . 0 0 R LF (53h 4 h 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
Setting high threshold value (option):
omputer→ Scale: S H D1...DN R LF (53h 48h D1...DN 0Dh 0Ah),
D1...DN – threshold value (see )
Scale →omputer: without response.
Connecting cable WK-1 (scale – computer / 9-pin interface):

USER MANUAL
15
_____________________________________________________________________________________________________________________
Protocol ELTRON description
ommunication parameters: 8bits, 1stop, no parity, 9600bps,
After use key scale send:
Scale→Label printer : EPL-2 language instruction to initialize print of label:
US
-
control instruction
FR"0001"
-
number of labels
?
-
starting instruction
hh:mm
-
5 signs: hours:minutes
yyyy.mm.dd
-
10 signs: year.month.day
mass + unit
-
10 signs: scale indicate + mass unit
P1
-
ending instruction
Attention:
1. It is possible to place a constants sign (company name, product name).
2. The default number of label’s is 0001 (label number 1).Different numbers of
label are possible by using a LAbEL function.
3. Label formula must be saved in printer – label forma must be designed in
computer (by Zebra Designer program) and saved in printer.
4. Parameters and protocol of transmission must be compatible with printer label..

16
USER MANUAL
_____________________________________________________________________________________________________________________
13. Basic f nction
To make clear how to manage with each function, in further part of instruction
descriptions are replaced with pictures.
- put a load on the pan
- remove the load from the pan
-
press the key when indication is
displayed
- forced change
- automatic change
13.1 Normal weighing
Press
→0←
key (
→
T
←
key in non-
legalized scales), which zeros the
scale, operates only when the pan is
empty .
Weighing result should be read when
the indicator " " lights.
0.00
2.30
0
2.30
T

USER MANUAL
17
_____________________________________________________________________________________________________________________
13.2 Weighing with tare
The scale is equipped with tare equal
to its range.
Joint value tare and mass net can not
cross a maximum of scale.
To display gross weight press B/G
key.
Note:
Press B/G key to return to net weight
indication.
13.3 Increased readability
Press HR key to display the weighing
result (for 5s.) with the highest
readability possible. This function is
especially helpful in scales with legal
verification with d=e.
The weighing result with increased
readability can be used for
informational purposes only and
cannot be printed or sent to a
computer with key.
2.30
0.00
40.01
42.31
T
NET
NET
2.30
2.29
2.30

18
USER MANUAL
_____________________________________________________________________________________________________________________
14. Special f nction describes
List of available functions:
menu customization function (ACtIV),
removal of all function from menu (dE AUL).
autozeroing function (AutoZE),
pieces counting function (PCS),
function for setting serial port working (PrInt)
function for setting serial port (Port)
entering tare function (tArE)
recipe weighing function (rECIPE),
weighing large animals function (LOC)
force measuring function (nEWto)
maximum value indication function (UP)
anti-disturbance filter function ( ILtEr)
average calculating function (AVErA)
percentage weighing function (PErC),
extended calibration function (CALIb)
setting time of stabilisation function (Stb)
selecting label number function (LabEL)
automatic switching off scale function (AutoO )
entering reference zero (Zero)
determining solids and liquids density function (dEnSIt)
calculator for good packaging control (tP) - option*
statistical calculations (StAt)- option*
paperweight calculation
function (PAP) - option*
function with additional equipment require:
options with the clock:
- setting current date and time function (dAtE)
- total weight function (totAL)
options with the transoptors connectors:
- checkweighing function (thr)
* Functions offered with special version of scale software (with limited possibility
of using other special functions).
User create own menu by choosing function in ACtIV function (described in
chapter 14.1).

USER MANUAL
19
_____________________________________________________________________________________________________________________
14.1 Men c stomization f nction (ACtIV and dEFAUL)
Among available user functions it is
possible to select these, which should be
displayed after pressing MENU key. It
allows avoiding displaying whole list of
available functions, which makes
operation time longer.
Operation sequence shown in the
pictures on the left causes adding
function for setting serial interface
RS232 parameters (Port) to function
menu.
After switching on ACTIV function a
dot is displayed on the right side (to
distinguish from regular menu). hosen
functions are displayed with a dot on
the left side.
In every moment, it is possible to
restore primary (manufacture) settings
choosing dE AULt special function.
In order to remove function from menu
in the last operation in place of selecting
Port -1 choose Port -0.
0.00
ACtIV
Port -1
Port -0
PCS
.
.
.
Port
T
T
T

20
USER MANUAL
_____________________________________________________________________________________________________________________
14.2 A tozeroing f nction (A toZE)
When ..-Aut function is activated,
the scale automatically ensures stable
zero indication if the pan is empty or
if zero indication was acquired by
pressing
→
T
←
key.
To leave the function press MENU
key, then with
→
T
←
key chose
AutoZE and Aut-0.
Note:
Autozeroing function is activated
automatically for 10 min. after
switching-on.
0.01
AutoZE
Aut-1
Aut-0
0.00
T
T
This manual suits for next models
2
Table of contents
Other Axis Scale manuals

Axis
Axis ACG Series User manual

Axis
Axis BTA Series User manual

Axis
Axis 4B/NA Series User manual

Axis
Axis B/U Series User manual

Axis
Axis B0.6D User manual

Axis
Axis BD/TL Series User manual

Axis
Axis 4BA SERIES User manual

Axis
Axis ATA1200 User manual

Axis
Axis B/E Series User manual

Axis
Axis SE-12 Operator's manual

Axis
Axis BA/LG Series User manual

Axis
Axis ATN Series User manual

Axis
Axis BTA 2100 User manual

Axis
Axis 4BA/T Series User manual

Axis
Axis 4BA SERIES User manual

Axis
Axis BD2P User manual

Axis
Axis B Series User manual

Axis
Axis AG Series User manual

Axis
Axis LCD GB V4 User manual

Axis
Axis AD Series User manual