Axis AKA/G Series User manual

USER
MANUAL
AKA/G SERIES
File: 2017-10-27 AKA_G AKAG122 GB

2USER MANUAL
____________________________________________________________________________________________________________________
Contents:
1. General description .................................................................................................................................................................. 3
2. Set .................................................................................................................................................................................... 3
3. Safety rules .............................................................................................................................................................................. 4
4. Technical data .......................................................................................................................................................................... 5
5. General balance description..................................................................................................................................................... 6
6. Keys and indicators .................................................................................................................................................................. 8
7. Preparing working environment................................................................................................................................................ 9
8. Preparing balance to work...................................................................................................................................................... 10
9. General operation principles .................................................................................................................................................. 11
10. Start-up................................................................................................................................................................................. 12
11. Internal calibration................................................................................................................................................................ 12
12. Checking the balance........................................................................................................................................................... 13
13. Connecting the balance to computer or printer .................................................................................................................... 14
14. Menu navigation ................................................................................................................................................................... 16
15. Setup .................................................................................................................................................................................. 17
15.1 Application selection –creating personalized menu................................................................................................. 18
15.2 Calibration with external weight / calibration options............................................................................................... 19
15.3 Auto-zeroing function................................................................................................................................................ 23
15.4 Unit selection ............................................................................................................................................................ 23
15.5 Interface parameters setting..................................................................................................................................... 25
15.6 Print setup................................................................................................................................................................. 26
15.7 LCD settings ............................................................................................................................................................. 28
15.8 Language selection................................................................................................................................................... 28
15.9 Setting date and time................................................................................................................................................ 29
15.10 Keyboard options...................................................................................................................................................... 29
15.11 Analog output............................................................................................................................................................ 30
15.12 Speed ....................................................................................................................................................................... 30
16. Applications .............................................................................................................................................................. 30
16.1 Product and user identification ................................................................................................................................. 32
16.2 Pieces counting ........................................................................................................................................................ 33
16.3 Unit ........................................................................................................................................................................... 34
16.4 Percentage ............................................................................................................................................................... 35
16.5 Animals weighing...................................................................................................................................................... 36
16.6 Tare setting............................................................................................................................................................... 37
16.7 Max or minimum value indication ............................................................................................................................. 38
16.8 Force indication (Newton)......................................................................................................................................... 39
16.9 Total.......................................................................................................................................................................... 40
16.10 Checkweighing function (thr) .................................................................................................................................... 41
16.11 Stats.......................................................................................................................................................................... 44
16.12 Density measurement............................................................................................................................................... 47
16.13 Recipe....................................................................................................................................................................... 51
16.14 Packing goods control (TP) ...................................................................................................................................... 52
17. Detailed information about balance communication ................................................................................................. 55
17.1 Long protocol description.......................................................................................................................................... 55
17.2 Protocol EPL description .................................................................................................................................................... 57
18. Troubleshooting and maintenance ..................................................................................................................................... 58

USER MANUAL 3
____________________________________________________________________________________________________________________
1. General description
AKA/G and AKZ/G series balances are destined for high accuracy weighing in laboratory practice.
Balances are equipped with graphical display and internal calibration system (only AKA/G) for accuracy
control during balance operations. Electronics system is based on new generation 32-bit microprocessor.
All balances are metrologically tested. According to an order balances can be calibrated or legally verified.
Balances with legal verification comply with certificate of type approval and are marked with the following
legal and securing items:
-green metrological mark placed on the balance name plate,
-notified body stamp (number of notified body) on the balance name plate,
-protective seals placed on: an edge of balance name plate, the casing mounting screw and in the place
of access to adjustment switch,
In order to renew legal verification please contact authorized service of AXIS.
Balance classification according to PKWiU: 33.20.31.
Certificates:
2. Set
A standard set consist of:
1. Balance,
2. Pan support and decorative pan,
3. Pan ring,
4. Tin floor of weighing chamber,
5. Feeder 12V / 1,2A,
6. User manual,
7. Guarantee card
Certificate of ISO quality system Certificate of balance type approval
DIN EN ISO 9001:2009

4USER MANUAL
____________________________________________________________________________________________________________________
3. Safety rules
It is necessary to follow safety rules of work with the
balance shown below. Obeying those rules is the
condition to avoid electrical shock or damage of the
balance or connected peripheral devices.
Repairs and necessary regulations can be done by authorised personnel
only.
To avoid fire risk use a feeder of an appropriate type (supplied with the
balance) and supply voltage have to be compatible with specified
technical data.
Do not use the balance when its cover is opened.
Do not use the balance in explosive conditions.
Do not use the balance in high humidity environment.
If the balance seems not to operate properly, switch it off and do not use
until checked by authorised service.
According to current acts of low about protection of
natural environment, wasted balances should not be put
into waste containers together with ordinary waste.
Wasted balance after operation period can be delivered to units
authorized for gathering wasted electronic devices or to the place where it
was bought.

USER MANUAL 5
____________________________________________________________________________________________________________________
4. Technical data
Type
AKA120G
AKZ120G
AKA220G
AKZ220G
AKA320G
AKZ320G
AKA520G
AKZ520G
Capacity (Max)
120g
220g
320g
520g
Min load (Min)
0,02g
0,02g
0,02g
0,02g
Reading unit (d)
0,001g
0,001g
0,001g
0,001g
Verification unit (e)
0,01g
0,01g
0,01g
0,01g
Tare range
-100g
-200g
-300g
-500g
Accuracy class
II
Working temperature
+18 ÷+33°C
Weighing time
<3s
Pan dimension
115mm
Balance dimension (with
legs)
215(235 with legs)x345x90mm
Interface and equipment
RS232C, USB, PS2, clock (optional: LAN or Wi-Fi)
Balance weight
5kg
Power supply
~230V 50Hz 6VA / =12V 1,2A
Recommended external
calibration weight (OIML)
F2 100g
F2 200g
F2 200g
F1 500g
Type
AKA1200G
AKZ1200G
AKA2200G
AKZ2200G
AKA3200G
AKZ3200G
AKA4200G
AKZ4200G
AKZ10
Capacity (Max)
1200g
2200g
3200g
4200g
10kg
Min load (Min)
0,5g
0,5g
0,5g
0,5g
5g
Reading unit (d)
0,01g
0,01g
0,01g
0,01g
0,1g
Verification unit (e)
0,1g
0,1g
0,1g
0,1g
1g
Tare range
-1200g
-2200g
-3200g
-4200g
-10kg
Accuracy class
II
Working temperature
+18 ÷+33°C
Weighing time
<3s
Pan dimension
165x165mm
195x180mm
Balance dimension (with
legs)
215(235 with legs)x345x90mm
Interfaces and equipment
AKA/G or AKZ/G: RS232C, USB, PS2, clock (optional: LAN or Wi-Fi)
AKA or AKZ: RS232C (optional: USB, PS2, zegar, LAN or Wi-Fi)
Balance weight
5kg
Power
~230V 50Hz 6VA / =12V 800mA
Recommended external
calibration weight (OIML)
F2 1000g
F2 2000g
F2 5000g
Caution:
E2 is international symbol of calibration weight class according to O.I.M.L. Some requirements for weight
accuracy are connected with this class.

6USER MANUAL
____________________________________________________________________________________________________________________
5. General balance description
AKA100G-AKA520G, AKZ100G-AKZ520G view:
1 –pan
2 –pan support
3 –pan ring
4 –graphical display
5 –keys
6 –rotating legs
7 –level
8 –draft shield
AKA1200G-AKA4200G, AKZ1200G-AKZ10G:
1 –pan
2 –pan support
4 –graphical display
5 –keys
6 –rotating legs
7 –level
7
6
1
2
4
3
5
8
7
6
1
2
4
5

USER MANUAL 7
____________________________________________________________________________________________________________________
Feeder
~230V/50Hz
Printer
or computer
ON OFF
Adjustment
switch
RxD (receiver)
TxD (transmitter)
GND
Case
RS232C
Feeder
1
2
3
4
5
6
7
8
9
Data
+5V GND
Clock
PS2
12 3
4
5
6
Keyboard
USB
(option)

8USER MANUAL
____________________________________________________________________________________________________________________
6. Keys and indicators
T
-taring (enter mass subtracted from weighed mass)
0
-zeroing (option),
ENTER
-confirmation / choosing an option,
.
-decimal point,
1/F1 ... 5/F5
-numeric key / functional key,
6 /
0
-numeric key / zeroing (only for balances for direct sale),
7 /
-numeric key / result printout (transmission),
8 /
-numeric key / internal calibration,
9 / MENU
-numeric key / enter menu,
0 /
-numeric key / mode change,
∧
-navigation: go to option above,
∨
-navigation: go to option below,
>
-navigation: enter into option,
<
-navigation: exit from option,
CLR
-undo last operation,
I /
-switch on / switch off (standby),
indicator
-shows stabilization of weighing result,
linear indicator
-indicator of balance load (0-100%),
OFF indicator
-appears after the balance is switched off with I / key,
decrease of last digit
-informs that reading unit value is lower than acceptable indication error (balances with
legal verification, de)
Max, Min, d, e, II
-metrological parameters of the balance.
The use of keys during entering numeric values (special functions):
- increment current digit,
-insert comma,
T
- move to next position,
MENU - finish entering.
Max 200g Min 0.01g e=0.001g d=0.0001g
AUT
0% 100%

USER MANUAL 9
____________________________________________________________________________________________________________________
7. Preparing working environment
+18°C
+33°C
Location for the balance should be chosen with care in order to limit influence of the factors that can
interrupt working balance. This location has to maintain proper temperature for working balance and
necessary space for its operating. The balance should stay on stable table made of material that does not
influence magnetically on the balance.
Rapid air blasts, vibrations, dust, rapid temperature changes or air humidity over 75% are not allowed in
balance surrounding. The balance should be far from heat sources and devices emitting strong
electromagnetic or magnetic fields.

10 USER MANUAL
____________________________________________________________________________________________________________________
8. Preparing balance to work
1. Take the balance, the feeder and mechanical elements of the pan out. It is
recommended to keep the original scale package in order to transport the balance
safely in future.
2. Place the balance on a stable ground not affected by mechanical vibrations and
airflows.
3. Level the balance with
rotating legs 6so that the
air bubble in water level 7
at the back of the balance is
in the middle.
4. (for AG100-AG600) Gently
insert the mandrel of pan
support 2into balance
mechanism socket through
the pan ring 3 and the pan 1
on (AG600 balances have
not the pan ring).
5. (for AG1000-AG4000)
Place nuts 2on mandrels
that are visible in balance
cover holes, put the pan 1
on nuts.
6. If the balance was taken from a lower temperature surrounding to
a room with higher temperature, e.g. in winter, moisture can liquefy on
the balance casing. Do not connect power supply to the balance, because
this can cause damage or improper work of the balance. In this case leave
the balance for at least 4 hours unplugged for acclimatization.
7
6
123
7
6
1
2

USER MANUAL 11
____________________________________________________________________________________________________________________
9. General operation principles
1. After each location change user should level the balance and perform internal calibration.
2. Weighed sample should be placed in the centre of the pan.
3. Weighing result should be read when the indicator " " lights, which signalises stabilisation of a result.
4. The balance allows tarring in the whole measuring range. To tare the balance press
T
key (on the left
or on the right). Tarring does not extend measuring range, but only subtracts tare value from mass value
of a sample placed on the pan. To make the control of pan load easier and to avoid crossing
measurement range, the balance has a load indicator calibrated 0÷100% Max.
5. In direct sale use (d=e), make sure that
0
zero indicator is displayed before sample is placed on the
pan. If not, press
0
key and wait until the balance is zeroed and zero indicator appears. In other
balances the key does not operate.
6. When the balance is not used but should be ready to work immediately, it can be switched off by pressing
I/key. The backlight of balance reading system is then switched off and the balance enters into
"standby" mode, in which the balance maintains internal temperature and ability to start working with
maximum accuracy. Standby mode is signalled by the OFF indicator. To switch the balance on press I/
key.
7. The balance cannot be used to weigh ferromagnetic materials due to decrease of weighing accuracy.
8. Balance mechanism is a precise device sensitive to mechanical shocks and strokes.
Do not overload the balance more than 20% of maximum capacity.
Do not press the pan with a hand.
For transportation take off the pan (move it gently and lift it up) and pan support (lift it up)
and protect from any damages.

12 USER MANUAL
____________________________________________________________________________________________________________________
10. Start-up
Plug feeder into 230V power supply socket and feeder output connector into 12V socket at back of the
balance.
After switching-on, the balance displays AXIS logo
and performs automatic self-tests.
In case of test failure balance displays tests list. Lack
of mark means negative test result.
Afterwards the balance enters automatically into
internal calibration mode, which is described with
details in next chapter. Calibration can be terminated
using CLR key.
When internal calibration is finished, the balance
enters into normal weighing mode.
11. Internal calibration
The balance is equipped with internal calibration system, which general task is to maintain required
measurement accuracy on the balance.
Internal calibration is the process of putting internal weight on automatically by balance mechanism and
correcting accuracy in balance firmware. The correction is necessary because of differences between
values of gravitational acceleration in the place where the balance was manufactured and in the place
where it is operated, as well as due to changes of balance level and temperature.
Internal calibration is performed in the following situations:
-when key is pressed,
-after defined time interval (for legally verified balances - 2 hours),
-after temperature change (for legally verified balances –more than 2oC).
In legally verified balances time interval is set to 2 hours and defined temperature change is 2oC. In not
legally verified balances those values can be set as calibration options. The reason of starting internal
calibration is shown as an icon near weight picture.
In order to perform internal calibration proceed with the following:
Max ... Min ... e= ... d= ..
0.0000 g
0% 100%
Balance type INICIALIZATION firmware
- - - - - - - - -
AUTOCALIBRATION .
Please wait ...
ul.Kartuska 375B 80-125 Gdańsk
http//:www.axis.pl

13 USER MANUAL
____________________________________________________________________________________________________________________
Empty the pan.
Press key twice (double pressing the key helps to
avoid accidental starting calibration procedure).
During calibration internal weight is put three times on
and obtained results are compared.
Discrepancy of results is signalled with a message and
causes the balance being blocked.
Until calibration process is finished do not perform any
operation on the balance. Any vibrations and shocks
interfere calibration process and may delay it or
deteriorate accuracy of its result.
When internal calibration is performed successfully the
balance indicates zero on the display at empty pan.
Note:
In order to terminate calibration process press CLR key
and wait until balance mechanism is not settled in
initial position
AUTOCALIBRATION
Internal calibration:
( - confirmation)
AUTOCALIBRATION
Please wait ...
AUTOCALIBRATION
Please wait ...
Max ... Min ... e= ... d= ...
0.0000 g
0% 100%

14 USER MANUAL
____________________________________________________________________________________________________________________
12. Checking the balance
In order to confirm correctness of the balance during its operation, before starting and after finishing every
measurement series it is advised to check weighing accuracy. It can be done by weighing external
calibration weight or other object with exactly known mass.
If exceeding of allowable measurement error is affirmed, the following things should be checked:
-if the balance stands stable and it is levelled,
-if the balance is exposed on rapid air blasts, vibrations, rapid temperature changes or air humidity,
-if the balance is not affected directly by heat source, electromagnetic radiation or magnetic field.
The cause of inaccuracy can be too low temperature of the balance as well, when it was unplugged from
power supply. In this situation leave the balance switched on for several minutes in order to adjust its
internal temperature.
If none of above causes of inaccuracy occurs, calibration with external weight should be performed to the
balance. Recommended external calibration weight (to buy for additional charge) is given in technical data
table. In order to calibrate the balance with external weight in legally verified balances verification seals
should be removed and another legal verification should be performed. In this case it is recommended to
contact authorized service centre.
Calibration with external weight is described in details in chapter 17.1.
13. Connecting the balance to computer or printer
The scale can be equipped with one or two serial interfaces RS232C, USB, LAN or Wi-Fi designed to
cooperate:
-with computer –the scale sends data after pressing key or after initiation signal from
computer,
-with printer - sending data after pressing key or automatically after putting on/off a sample
and measurement stabilization,
-with label printer –after pressing the scale sends set of instructions for label printer starting
from label number set in special funcion LabEL.
Set of send data is set using special function PrInt.
The following data can be send:
-Header (scale type, Max, d, e, serial number),
-Operator identification number,
-Successive printout number (measurement),
-Identification number or product bar code,
-Number of pcs (PCS function only),
-Single detail mass (PCS function only),
-Nett weight,
-Tare (package mass),
-Gross weight,
-Total mass (Total function only).
The way of sending data and transmission parameters is set using SErIAL special function.

USER MANUAL 15
____________________________________________________________________________________________________________________
If the scale is equipped with two serial joints (interfaces) Print and SErIAL function is set independently for
both interfaces.
If scale cooperates with a computer then the computer must have a special program. Dedicated programs
are also offered by AXIS.
Needed drivers and instructions can be found on the CD supplied with Axis scales.

16 USER MANUAL
____________________________________________________________________________________________________________________
14. Menu navigation
Balance’s display during weighing:
After pressing MENU key main menu shows up:
Main menu consists:
-Applications –user personalized applications
menu,
-Setup –creating personalized menu,
calibration, balance options,
-Info –information about the balance,
-Exit
To navigate following keys are used:
∧
-move cursor up,
∨
-move cursor down,
>
-enter option, choosing/scrolling suboption,
<
-exit actual option, choosing/scrolling suboption,
ENTER
-enter / option selection,
CLR
-exit actual option (undo last operation, auto-calibration stop),
F1, F2, ... , F5
-fast access to application (shortcut activated in application suboption),
MENU
-enter/exit from menu,
-turned on/off active application,
To use option or to choose application move cursor and press ENTER key.
Important tool to navigate are fast access keys, which are assigned by user. The keys enable direct
activation off chosen 5 applications by using F1, F2, ..., and F5 key.
Max ... Min ... e= ... d= .. I
NET AUT ANL
T10.1203 g
0% 100%
metrological data bar
net weight indicator
active autozeroing
active application mark
main indication
load bar
tip
active application icon
MENU
1. Applications
2. Setup
3. Info
4. Exit

USER MANUAL 17
____________________________________________________________________________________________________________________
15. Setup
Setup consists all options used for setting balance’s way of working:
Menu –applications selection to user’s
personalized menu,
Calibration –balance’s calibration,
Auto-zeroing –automatic zero indication hold
when pan is unloaded,
Unit –weight unit selection,
Interface –setting serial ports,
Print setup –data selection for transmission
(printout),
Time&date –inscribing actual date and time,
Keyboard –keys options,
Analog out –4-20mA (0-10V) out
configuration (option),
Speed –weighing speed selection,
Firmware update –firmware actualization
(only for service),
Defaults –back to factory settings,
Exit.
Attention:
Using Defaults option doesn’t change basic metrological balance parameters like: sensitivity, linearity (if the
calibration switch isn’t moved), but all other settings that have influence on balance’s work and
communication with other devices can be changed and need resetting by User.
MENU
1. Applications
2. Setup
3. Info
4. Exit
SETUP
1. Menu
2. Calibration
3. Auto-zeroing
4. Unit
5. Interface
6. Print setup
7. LCD settings
8. Language
9. Time&date
10. Keyboard
11. Analog output
12. Speed
13. Firmware update
14. Defaults
14. Exit

18 USER MANUAL
____________________________________________________________________________________________________________________
15.1 Application selection –creating personalized menu
All balances besides basic metrological functions: weighing and tare, have many applications (functions)
and configuration options.
In order to limit user applications quantity that
appear after pressing Menu key (and choosing
Applications option), user can choose several
applications and create his own menu.
Creating personalized menu:
During balance’s first start, after pressing MENU
key choose Menu option. List of applications will
show up. Precise description of all applications
in Applications chapter.
Adding applications to personalized menu is
done by pressing ENTER key when chosen
application is highlighted.
Added application is marked with „V” sign.
After choosing all necessary applications use
Exit option.
Defaults option deletes all applications from
personalized menu (return to factory settings).
SETUP
1. Menu
2. Calibration
3. Auto-zeroing
4. Unit
5. Interface
6. Print setup
7. LCD settings
8. Language
9. Time&date
10. Keyboard
11. Analog output
12. Firmware update
13. Defaults
14. Exit
MENU
Product ID
User ID
PCS
Unit
Percent
(Label)
Animal
Tare setting
MAX/MIN
Niuton
Total
Threshold
Stats
Paper
Recipe
Exit

USER MANUAL 19
____________________________________________________________________________________________________________________
15.2 Calibration with external weight / calibration options
Calibration with external weight should be performed if balance accuracy after internal calibration is not
satisfactory. Calibration weight stated in technical data table for the balance (or of better accuracy) with
valid verification certificate should be used then.
Calibration of legally verified balance requires violating a mark used to protect an access to
adjustment switch and results in loosing legal verification. To renew legal verification of the
balance, it is necessary to contact a service or notified body.
In balances comply with verification requirements performing calibration requires changing adjustment
switch position, which is placed behind protecting mark (sticker) of a notified body. An access to the switch
is possible only after removing the mark.
Before proceeding with calibration for balances comply with verification requirements, adjustment switch
should be set to ON position using thin screwdriver (the balance will display the message Pr ON).
When calibration process, described on next page, is finished, the balance will display the message Pr ON.
Adjustment switch should be set to OFF position using thin screwdriver (the balance will move to weighing).
ON OFF
Adjustment
switch

20 USER MANUAL
____________________________________________________________________________________________________________________
In order to calibrate the balance use MENU key, choose Setup option and then Calibration.
Load option enables inscribing standard
of mass value, which will be used to
calibrate (it is suggested to use standard
of mass value close to balance’s max).
SETUP
1. Menu
2. Calibration
3. Auto-zeroing
4. Unit
5. Interface
6. Print setup
7. LCD settings
8. Language
9. Time&date
10. Keyboard
11. Analog output
12. Firmware update
13. Defaults
14. Exit
CALIBRATION
1. Calibration:
2. Mode
3. Load: <...>
4. Report
5. Linearity
6. Exit
CALIBRATION
1. Calibration:
2. Mode
3. Load: 100g
4. Report
5. Linearity
6. Exit
This manual suits for next models
20
Table of contents
Other Axis Scale manuals

Axis
Axis BTA Series User manual

Axis
Axis AGN50 User manual

Axis
Axis AG Series User manual

Axis
Axis AGN/C Series User manual

Axis
Axis B0.6D User manual

Axis
Axis 4B/NA Series User manual

Axis
Axis SE-12 Operator's manual

Axis
Axis 4BA/T Series User manual

Axis
Axis B/E Series User manual

Axis
Axis BD/TL Series User manual