Axis BLM Series User manual

USER MANUAL
BLM Series
File: 2011-10-04-BLM-136 BL010 GB

2
OPERATION MANUAL
__________________________________________________________________________________________________________________
Index:
1.
General description .....................................................................................................3
2.
Completeness .............................................................................................................3
3.
Security rules...............................................................................................................4
4.
Technical data.............................................................................................................5
5.
General view of balance..............................................................................................
.
Keys and indicators ..................................................................................................... 7
7.
Preparation of work place............................................................................................9
8.
Preparation to work .....................................................................................................9
9.
General rules.............................................................................................................10
10.
Start of work.............................................................................................................. 11
11.
Pieces counting (PCS) .............................................................................................. 11
12.
Summation of measurements series (
Σ
.) .................................................................. 15
13.
Data library (
↔
LIB key ) ..........................................................................................1
14.
Connecting a computer or a printer ...........................................................................18
15.
Basic function description.......................................................................................... 20
15.1 Normal weighing ...................................................................................................... 20
15.2 Weighing with tare.................................................................................................... 20
1 .
Special function description....................................................................................... 21
1 .1 Autozeroing (Autot) ..................................................................................................22
1 .2 Calibration with external weight (CALIb)...................................................................23
1 .3 Serial port parameters setting (rS232) ..................................................................... 24
1 .4 Print-out configuration function (PrInt))..................................................................... 25
1 .5 Date and time setting function (dAtE) - option.......................................................... 2
1 .7 Customization function menu (ACtIV i dEFAULt) ..................................................... 27
17.
Maintenance and repairs of small defects ................................................................. 28
Declaration of Conformity........................................................................................... 29

OPERATION MANUAL
3
__________________________________________________________________________________________________________________
1. General description
BLM series scales are destined for identical pieces counting and can be used in
store houses control of production and in similar tasks.
Counting pieces function is based on calculation individual mass of detail from
sample weight or keyboard inscribed. cale avails a “learn algorithm” during
calculation and this algorithm is correcting inaccuracy conversion.
cale owns summing register. This register allows to addition or subtraction of
next result and observation current state of this register.
cales grant claims of regulations abort safeties in European Union, sign
inform about that on brand table CE.
This type of scale not requires legalization.
PKWiU classification: : 33.20.31.
2. Completeness
tandard set consist of:
1. cale
2. Carrying pan
3. Fish-plate pan
4. Feeder
5. User manual
6. Guarantee card

4
OPERATION MANUAL
__________________________________________________________________________________________________________________
3. Security rules
To avoid electrical shock or damage of the scale or
connected peripheral devices, it is necessary to follow
the security rules below.
• All repairs and necessary regulations can be made by authorised
personnel only.
• To avoid fire risk use a feeder of an appropriate type (supplied with the
scale). Pay attention that supply voltage is compatible with specified
technical data.
• Do not use the scale when its cover is opened.
• Do not use the scale in explosive conditions.
• Do not use the scale in high humidity.
• If the scale seems not to operate properly, unplug it from the mains and
do not use until checked by authorised service.
According to legal regulations it if forbidden to
dispose wasted electronic equipment in waste
containers.
• Please return wasted scale to the point of purchase or other company
specialised in recycling of wasted electronic components.

OPERATION MANUAL
5
__________________________________________________________________________________________________________________
4. Technical data
Scale type BLM1.5 BLM3 BLM6 BLM15 BLM30
Maximum loading (Max) 1.5kg 3kg 6kg 15kg 30kg
Reading unit (d) 0,1g 0,1g 0, g 1g 1g
Minimal sample weight 0,1g 0,1g 0, g 1g 1g
Pan dimension 50x180mm
Working temperature -10°C ÷ +40°C
Weighting time <3s
Scale base dimension 55x310x107mm
Scale weight ,6kg
Power supply ~ 30V 50Hz 6VA / =1 V 1, A

6
OPERATION MANUAL
__________________________________________________________________________________________________________________
5. General vie of balance
1 – pan welt (pan)
2 – carrying pan
(under welt)
3 – LED display
4 – keyboard
5 – level
6 – rotary legs
Connections views:
Zasilanie
Zasilacz
~230V
RxD (odbiór)
TxD (nadawanie)
masa
RS232C
1
2
3
5
6
7
8
9
6
5
1
2
4
3
Feeder
upply

OPERATION MANUAL
7
__________________________________________________________________________________________________________________
6. Keys and indicators
Keys function description:
→
T
←
- tare (subtract package weight from weighed mass) / approve
options in menu,
→
0
←
- zero (option),
I/ - switch-on / switch-off (standby),
- special key,
LIB - database,
- print out,
MENU - special function menu,
UNITS - change of mass units on main display,
DATA - change data on displays DATA
HR - temporary increased indication resolution,
←
B/G - gross weight indication switch,
↔
LIB - get to record base,
→
PCS - start counting procedure with added sample
(long hold key: turn-on/turn-off function, MIN/OK./MAX
selection),
→
PCS- - start counting procedure with subtractive sample
(long hold key: turn-on/turn-off function, MIN/OK./MAX
selection),
→
SPW - set a sample detail mass by numeric keys and ENTER,
Σ
- added current indication to summing register,
Σ
- - subtracted current indication from summing register

8
OPERATION MANUAL
__________________________________________________________________________________________________________________
Main display indicators:
indicator - result stabilisation indicator,
indicator
→
0
←
- zero indicator,
indicator NET - net weight indicator (indication with subtracted tare),
indicator kg/g - change of unit displayed,
indicators:
lb,oz,ozt,ct
- units of masses current indications,
DATA display indicators:
indicator B/G - gross mass (key B/G),
indicator T - actual tare,
indicator PT - stored tare,
indicator APW - sample mass,
indicator Σ - contents of summing register,
COUNT display indicators:
indicator
∇
- not working,
indicator MIN - indication under minimum threshold,
indicator OK - indication between minimum and maximum thresholds,
indicator MAX - indication above maximum threshold,
indicator Σ - signalize indication of summing register contents.

OPERATION MANUAL
9
__________________________________________________________________________________________________________________
7. Preparation of ork place
Works place of scale should be chosen carefully. This place should be assured
a proper temperature and essential area for attendance. Weight should stand on
stable table.
Violent movements of air, dust, vibrations, violent changes of temperature or
humidity above 90% are inadmissible. cale should be disallow from strong
electromagnetically or magnetically fields.
8. Preparation to ork
1. Take the scale, pan and
feeder out of the package.
It is recommended to keep
the original scale package
in order to transport the
balance safely in future.
2. Place the scale on a
stable ground not affected
by mechanical vibrations
and airflows.
3. Put a carrying pan 2 in
wholes.
4. Put the fish plate pan
1.
5. Level the scale with the
rotating rear legs 6 so that
the air bubble in the
water-level 5 at the back
of the scale is in the
middle.
If scale has been
moved from places with low temperature to places with
upper temperature (for example in winter), please weight on time 4 hours
purpose of acclimatization.
1
2
5
6

10
OPERATION MANUAL
__________________________________________________________________________________________________________________
9. General rules
1. It is advised to check scale indication accuracy before and after series of
measurement using any load with known weight. To check the scale with legal
verification uses a calibration weight with valid calibration certificate. In case
permissible error is exceeded it is advised to contact the nearest service to
calibrate the scale.
2. Weighed sample should be placed in the centre of the pan.
3. The scale is equipped with a tare equal to its range. To tare the scale press
→T← key (left or right). toring a tare value does not extend measuring range,
but only subtracts it from a load placed on a pan. To make weight control
easier and to avoid range overdrawing, the scale is equipped with a load
indicator (graduated in percentages).
4. Weighing result should be read when the indicator " " lights, which
signalizes stabilization of a result.
5. When the scale is not used but it is necessary to be ready to work
immediately, it can be switched off by pressing I/ key. The scale reading
system is then switched off to "standby" mode. To switch the scale on press
I/ key. The scale is immediately ready to operate maximum accuracy (after
self tests).
6. Before each measurement make sure that zero indicator is displayed. If zero
indicator does not displayed or “----“ communicate appears, press →0← key
and wait until zero indication and zero indicator appears
7. The mechanism of scale is a precision devices and susceptible on congestions,
impacts and mechanical shakes.
Do not overload the scale more then 20% of maximum load (Max).
Do not press a pan by hand.
cale should be protected during the transport.

OPERATION MANUAL
11
__________________________________________________________________________________________________________________
10. Start of ork
Leave the pan empty, plug the feeder to the mains (~230V/50Hz) and plug the
feeder connector to the 12V power socket at the bottom side of the scale.
cale will execute following step:
- test of electronic components,
- display type of scale and version of program,
- zeroing the indication.
After displayed an indication on the main (left) display scale is ready to work.
11. Pieces counting (PCS)
cale enables to count identical pieces, e.g. turnbuckles or buttons
Measurement with single pieces putting on pan:
A measurement is performed in two phases:
- first phase - single piece weight calculation on the basis of defined pieces amount
(default is 10 pieces.),
- second phase – pieces counting
1. Put on pan quantified details sample.
2. Turn a counting by press
→
→→
→
PCS+ key.
3. If the quantified sample consists 10 pieces press ENTER (solid line on the
drawing). If amount of detail is other write correct value by using numeric
keys and ENTER (dotted line on the drawing).
4. Number of details has been showed on right display with PCS indicator. A
sample mass has been showed in centered display with APW indicator.
5. Remove ample form pan and put o portion of pieces.
6. Result of calculation is read from right display.

12
OPERATION MANUAL
__________________________________________________________________________________________________________________
Measurement with single pieces removing from pan:
A measurement is performed in two phases:
- first phase - single piece weight calculation on the basic of removed pieces
(default is 10 pieces),
- second phase – recalculating a weighting portion.
1. Put on pan a box with details.
2. Turn a pieces counting by press
→
→→
→
PCS-key and confirm by ENTER key.
3. Remove one or more details from the portion.
4. If one detail is removed press ENTER key. If amount of removed is other write
correct value by using numeric keys and ENTER.
5. Number of details has been showed on right display with PCS indicator. A
sample mass has been showed in centered display with APW.
6. Result of calculation is read from right display.

OPERATION MANUAL
13
__________________________________________________________________________________________________________________
Change/write a sample mass by numeric keys:
During a pieces counting (indicator PCS) user can change actual sample mass.
Write a sample mass when a counting process is turn-off initiate a counting
process.
1. Press a
→
→→
→
SPW key.
2. Write sample mass by using a numeric keys and ENTER.
3. Number of details has been showed on right display with PCS indicator. A
sample mass has been showed in centered display with APW.
4. Result of calculation is read from right display.
ote:
1. It is advised that single piece weight is not less than one reading unit and
sample weight used in first phase is bigger than 100 reading units.
2. Err-PCS communicate signalizes that a sample was not put on the pan or
single piece weight is less than one reading unit (it is possible to count pieces but
measuring error is bigger).

14
OPERATION MANUAL
__________________________________________________________________________________________________________________
MI /OK./MAX selection, turn-on/turn-off counting detail
Hold a
→
PCS (
→
PCS-) key causes display following options:
- OFF – disable a counting details
- on – enable a counting details
- thr – setting a threshold selection MIN/OK./MAX
3sek.

OPERATION MANUAL
15
__________________________________________________________________________________________________________________
et a value of threshold causes a turn-on an indicators MIN, OK i MAX, state of
these indicators depends on value on indicator and thresholds.
If the result of details calculation is:
- lower than threshold undEr – scale signalizes MIN,
- between thresholds - scale signalizes OK, and during a pass a threshold scale
turn-on buzzer,
- bigger than threshold OVEr – and during a pass a threshold scale turn-on
buzzer.
ote:
On the first start scale value of thresholds are set to maximum.
During a setting a upper threshold remember that his value can’t be lower that
value of minimum threshold.
12. Summation of measurements series (Σ
ΣΣ
Σ.)
cale is equip In summation register. That registers enable addition a next result
of measurements.
1. Put element on the pan and press
→
→→
→Σ
ΣΣ
Σ
+ button. Contents of this register have
been showed temporary in centered display with indicator
Σ
.
Note: Using a
→
→→
→Σ
ΣΣ
Σ
- key causes a subtract indication from register.
2. Press several times a DATA key to belongs for continuous displaying of
register – in this case an indicator is turn-on.

16
OPERATION MANUAL
__________________________________________________________________________________________________________________
13. Data library ( ↔
↔↔
↔LIB key )
The scale is equipped with data library where user can save detail characteristics
(number, single piece mass, default amount of samples and others) in chosen
record number and later when it is needed recall to it.
During scale work, scanner (configuration: prefix 04h, sufixx CR+LF)
readout of
detail number causes searching through scale record library and in case of finding
the right record recalling single piece mass and other data.
Record library edition
Longer pressing of
↔
↔↔
↔
LIB key enters edition menu.
On left display a Edit sign shows up and on the middle one a record identification
number ( from 000 to 999).
Next using
Σ
ΣΣ
Σ
+
and
Σ
ΣΣ
Σ
- keys choose record number and press Enter to edit it.
Chosen record id will show up on the left display. On the middle one following
options will successively display:
- PA no – detail number inserting,
- SPW – single piece mass,
- SPL – default sample quantity,
- PTARE – inserting constant tare value for the detail,
- undEr – lower threshold value,
- OVEr – upper threshold value,
- SAVE – saving settings,
- ErASE – erasing settings,
- out – exit the library (user can also use CLR key).

OPERATION MANUAL
17
__________________________________________________________________________________________________________________
To speed up options displaying user can press
Σ
ΣΣ
Σ
+
and
Σ
ΣΣ
Σ
- button.
We choose option by pressing E TER key (options that have inserted values
display together with a green diode in lower left corner of middle screen).
After choosing the proper option user can insert numeral value using keyboard
and confirm it with E TER key.
Entering record library
Shortly press
↔
↔↔
↔
LIB key to enter record library in order to readout inscribed
earlier record. User can differ saved records (green diode lights on middle screen)
from empty ones.
We choose record using
Σ
ΣΣ
Σ
+
and
Σ
ΣΣ
Σ
- keys and confirm our choice by pressing
E TER. It is also possible to inscribe directly the record number. In order to do
that press
and then insert numeral value using keyboard. Confirm pressing
E TER key.
election of empty record is signalized by „EMPtY” message. Currently used
record is marked by additional „in U E” message. If user wants to cancel
choosing of records press CLR key.

18
OPERATION MANUAL
__________________________________________________________________________________________________________________
14. Connecting a computer or a printer
The scale is equipped with R 232C which can be used to connect external
devices such as a computer or a printer
When cooperating with a computer data
is send after initiate signal from a
computer or after a press button.
When cooperating with the scale, a
computer should be equipped with a
program which enables processing data
from the scale
The Axis company offers programs to cooperating with scales. These programs
are available on site www.axis.pl :
- Test RS232C- program to tests serial connections (full version),
- ProCell – program enabling cooperation with Microsoft Excel or different
Windows applications (demo version).
Information for programmers (transmission protocol describe)
Signal description:
Computer
→
Scale: initialising signal S I CR LF (53h 49h 0Dh 0Ah),
Scale
→
Computer: weighing result according to the diagram below
(16Bytes, transmission parameters: 8bits, 1stop, no parity, 4800bps),
Byte description:
Byte 1 -sign „-” or space
" 2 - space
" 3
÷
4 - digit or space
" 5÷9 - digit, decimal dot or space
" 10 - digit
" 11 - space
" 12 - k, l, c, p or space
" 13 - g, b, t, c or %
" 14 - space
" 15 - CR
" 16 - LF

OPERATION MANUAL
19
__________________________________________________________________________________________________________________
Connecting cable WK-1 (connect scale with computer / interface 9-pin):
Connecting cable WD-1 (connect scale with printer KAFKA):
Settings og internal switch printer AXIS:
W-1 W-2 W-3 W-4
W-5
W-6
W-7
W-8
on off on off off on off off
SCALE
COMPUTER
SCALE
KAFKA PRINTER

20
OPERATION MANUAL
__________________________________________________________________________________________________________________
15. Basic function description
15.1 Normal eighing
Press the
→
T
←
key when indication
is different from zero at empty scale.
Weighing result should be read when
the indicator " " lights.
15.2 Weighing ith tare
The scale is equipped with tare equal
to its range.
Joint value tare and netto mass can not
surpass the range of weight (Max).
0.00 g
43.04 g
23.13 g
0.00 g
63.24 g
T
T
Table of contents
Other Axis Scale manuals

Axis
Axis LCD GB V4 User manual

Axis
Axis 4BA/T Series User manual

Axis
Axis SE-12 Operator's manual

Axis
Axis ATN Series User manual

Axis
Axis B Series User manual

Axis
Axis B/U Series User manual

Axis
Axis 4B/F Series User manual

Axis
Axis BD/TL Series User manual

Axis
Axis B/E Series User manual

Axis
Axis AKA/G Series User manual