INSTRUMENT OPERATION - Connect the probe(s)
via the miniature thermocouple socket(s) located
on the end of the instrument. Apply the tip of the
probe to the substance, medium or surface to be
measured. The instrument reading may take several
seconds to stabilise, depending on the nature of the
measurement and sensitivity of the probe.
BUTTON FUNCTIONS
Transfer/On-Off button
ON - Press to switch the instrument on.
OFF - Hold down for 3 seconds to switch off.
MEASURE/TRANSFER - Press to measure and
transfer result via Bluetooth (if connected).
CLEAR BONDING - See Bonding
Backlight button
Press to toggle the backlight on and off. Note
the backlight will switch off automatically
after ten seconds.
ALARMS - The default setting is off. The high/low
alarms can be switched on/off and adjusted through
the ThermaQ App. The current alarm status is
indicated by the instruments display and status LED.
SENSORS - Only use type K (Nickel Chromium/
Nickel Aluminium) thermocouple probes (to BS EN
60584-1:2013) fitted with suitable miniature type K
thermocouple plugs.
TEMPERATURE MEASUREMENTS - The instrument
measures its inputs at a set measurement interval
and displays the value in ºC or ºF. If Sensor 2 is
switched on, the display will alternate between the
two sensor readings every 5 seconds with 'Sen 1'&
'Sen 2'indicating which sensor reading is currently
being displayed.
CONNECTION - Use a Bluetooth®Low Energy 'Smart
or Smart Ready'iOS or AndroidTM host device with the
ThermaQ App installed to make connections to the
instrument. The ThermaQ app can be downloaded
free from the App Store (Apple devices) or on Google
PlayTM (Android devices). For full device compatibility
visit www.etiltd.com. Open the ThermaQ App and
connect to the device via the Devices screen, use the
instrument’s serial number to select the correct device
from the list. Once connected the ThermaQ App
will begin receiving and storing measurement data.
Please note: When not connected the instrument
continues to measure using its current settings but
no readings are stored in the instrument.
SETTINGS - Adjustable via the ThermaQ App include:
°C or °F, Measurement Interval, Auto-off Interval,
Sensor Name(s) and High/Low Alarm levels. Sensor
2 can be set on/off. All settings are stored in the
instrument and are downloaded to the ThermaQ
App on connection.
AUTO-OFF - If the instrument is not connected to
Bluetooth within the set Auto-off interval, it will shut
down. This function can be switched off via the
ThermaQ App.
ALARM STATUS LED - A red flash indicates a
temperature is in an alarm condition and a green
flash indicates the temperatures are within the set
alarm levels. The LED flashes every measurement
whether connected via Bluetooth or not.
OPEN CIRCUIT/FAULTY PROBE - If the probe has
developed an 'open circuit' fault or is not present the
instrument will display 'Err'.
BONDING - This instrument does not need to be
paired/bonded to your host device. Pairing/Bonding
may cause connection issues within Apps. If you
encounter connection problems please clear bonding
as described below.
Clear Bonding - start with the instrument switched
off, press and hold down the Transfer button for
approximately 5 seconds until the display shows
'clr bond'.
BATTERY STATUS - This is indicated by the ThermaQ
App and the instrument’s display using the battery
symbol. When the battery level falls below a safe level
the instrument displays 'Flat bat'before shutting down.
BATTERY REPLACEMENT - Unscrew the 4 screws
located on the back of the instrument and remove
rear case. This can be difficult especially after a
long period but if done gradually at each corner the
case will come apart. Replace the AA battery noting
polarity. Refit rear case and refit screws taking care
not trap/pinch the seal. Tighten the screws evenly but
do not overtighten as this will affect the instrument
water resistance.
WARNING: IPA and other solvents may cause
damage to the case and screen of this instrument.