3
FUNCTIONS
OPERATION
• Portable four function instrument for calibrating process devices.
• Adjustable current source 0-24mA.
• Drives loads of up to 500Ω
• Large LCD display.
SPECIFICATIONS
Display LCD display, maximum display counts 1999.
13mm (0.5”) digit size
Function mA Current source, 2 ranges. 0 - 19.99mA × 0.01mA
and 0 - 24mA × 0.1 mA
Sampling Time Approx. 0.4 second
Operating Temperature 0°C to 50°C (32°C to 122°C)
Operating Humidity Less than 80% RH
Power Supply 006P DC 9V, MN1604/PP3 battery or equivalent Alkaline
type or heavy duty type
Dimension 185mm × 78mm × 38mm (7.3” × 3” × 1.5”)
Weight 265g/0.58lb (including battery)
Electrical Specication (at 23o±5oC)
Range Display Resolution Accuracy
0-19.99mA 0.01mA ±(0.25% FS+1d)
1-24mA 0.1mA ±(0.25% FS+1d)
Output 0-20 mA current for loads up to 500 ohms.
Output > 20 mA current for loads up to 400 ohms.
* FS = full scale
Note: The above specication are tested under the environment RF Field Strength less
than 3 V/M & frequency less than the 30 MHz only.
• Power on the instrument by sliding the power switch to the ON position.
• Connect the interface cable to the input socket and the clips to the device to be
calibrated.
• The red clip is the positive current output, the black clip is ground.
• Slide the range switch to either the 0-19.99mA display resolution or 1-24.0mA
position as required.
• Adjust the calibration control to adjust the current output to match the display value.