This meter complies with IEC61010 safety measurement requirement. Pollution
degree 2, CAT III 600V.
Please read the following safety information carefully before using or servicing
•
•
•
•
Do not apply more than 600V to the instrument
Do not use the instrument around explosive gas, vapour, or
dust Do not use the instrument in a wet environment
When using test leads, keep fingers away from lead contacts and behind
finger guards on the leads
When performing insulation resistance tests, disconnect and discharge all
power from the circuit to be measured
Do not short circuit testing leads during high voltage output or test
insulation resistance after high voltage output.
Do not measure over 10 seconds when:
Measuring resistance <2MΩ with use of 500V
Measuring resistance <5MΩ with use of 1000V
Measuring resistance <8MΩ with use of 1500V
Measuring resistance <10MΩ with use of 2500V
At any output voltage, when tested resistance is less than 10MΩ,
continuous testing time cannot exceed 10 seconds.
Operating the meter must be done very carefully as it outputs dangerous
voltages during measurement. Test object must be firmly clamped, and
hand must be away from measurement clips before pressing TEST button
to output high voltage.
Be careful when working above 33Vrms, 46.7V AC rms or 70V DC, such
voltages pose shock hazard
Discharge all loading of circuit under test after measuring high voltage
Place test leads in proper input terminals, make sure all test leads are
firmly connected to the meter’s input terminals
When performing resistance tests, remove all power from the circuit to be
measured and discharge all remaining power
Do not use the instrument with any parts or cover removed
When carrying out insulation measurement, do not contact the circuit under
test.
Do not use the instrument if it is damaged or if metal parts are
exposed. Check to make sure no cracks or missing plastic parts
before using the instrument
Do not change battery in wet environments
Make sure the meter is turn off when opening battery compartment.
When servicing the meter, use only test leads and power adaptor with the
same electrical specification as the original parts.
•
•
•
•
•
•
•
•
•
•
•
•