MSA PremAire System User manual

PremAire
®
OPERATION AND INSTRUCTIONS
REGULATOR
ESCAPE
CYLINDER
VORTEX
MAIN
INLET
DUAL
SUPPLY
This PremAire is certified by the National Institute of Occupational Safety and Health
(NIOSH).
The warranties made by MSA with respect to the product are voided if the product is
not installed, used and serviced in accordance with the instructions in this manual.
Please protect yourself and your employees by followin the instructions. Please read
and observe the WARNINGS and CAUTIONS inside. For any additional information
relative to use or repair, write or call 1-800-MSA-2222 durin re ular workin hours.
TAL 801 (L) Rev. 8 © MSA 2008 Prnt. Spe . 10000005389(A) Mat. 496958
Do . 496958
THIS MANUAL MUST BE READ CAREFULLY BY ALL PERSONS WHO HAVE OR WILL
HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any
complex piece of equipment, the PremAire from MSA will perform as desi ned
only if used and serviced accordin to the instructions. OTHERWISE, THE PROD-
UCT COULD FAIL TO PERFORM AS DESIGNED, AND PERSONS WHO RELY ON THE
PRODUCT COULD SUSTAIN SERIOUS PERSONAL INJURY OR DEATH.
See separate Inserts for NIOSH Approval information
(P/N 10032763 / 818364 / 10032761)

INTRODUCTION
NIOSH APPROVAL INFORMATION
CAUTIONS AND LIMITATIONS
A- Not for use in atmospheres ontaining less than 19.5
per ent oxygen.
B- Not for use in atmospheres immediately dangerous to
life or health.
C- Do not ex eed maximum use on entrations estab-
lished by regulatory standards.
D- Air line respirators an be used only when the respira-
tors are supplied with respirable air meeting the
requirements of CGA G-7.1 Grade D or higher quality.
E- Use only the pressure ranges and hose lengths spe i-
fied in the User’s Instru tions.
G- If airflow is ut off, swit h to filter/or artridge or an-
ister and immediately exit to lean air.
H- Follow established artridge and anister hange
s hedules or observe ESLI to ensure that artridges
and anisters are repla ed before break through
o urs.
I- Contains ele tri al parts whi h have not been evaluat-
ed as an ignition sour e in flammable or explosive
atmospheres by MSHA/NIOSH.
J- Failure to properly use and maintain this produ t ould
result in injury or death.
L- Follow the manufa turer’s User’s Instru tions for
hanging artridges, anister and/or filters.
M- All approved respirators shall be sele ted, fitted, used
and maintained in a ordan e with MSHA, OSHA and
other appli able regulations.
N- Never substitute, modify, add or omit parts. Use only
exa t repla ement parts in the onfiguration as spe i-
fied by the manufa turer.
O- Refer to Users Instru tions, and/or maintenan e man-
uals for information on use and maintenan e of these
respirators.
P- NIOSH does not evaluate respirators for use as surgi-
al masks.
S- Spe ial or riti al Users Instru tions and/or spe ifi
use limitations apply. Refer to user instru tions before
donning.
S - SPECIAL OR CRITICAL USER INSTRUCTIONS
1. An adequate respiratory prote tion program must
in lude knowledge of hazards, hazard assessment,
sele tion of proper respiratory prote tive equipment,
instru tion and training in the use of equipment,
inspe tion and maintenan e of equipment, and med-
i al surveillan e. [See OSHA regulations, Title 29 CFR
1910.134].
2. The program administrator and respirator users must
read and understand these instru tions before trying
to use or servi e this produ t.
3. The PremAire Supplied-Air Respirator System will per-
form as designed only if used and maintained a ord-
ing to these instru tions.
4. This respirator may be used only after proper instru -
tion and training in its use as spe ified in OSHA regu-
lations Title 29 CFR 1910.134.
5. Inspe t the respirator regularly and maintain it a ord-
ing to the instru tions. Repairs must only be made by
properly trained personnel.
6. Use only with an air sour e that meets ANSI
(Compressed Gas Asso iation) spe ifi ations. The air
delivered to the respirator's air-supply hose must be
respirable and of a purity equal to at least Quality
Verifi ation Level (Grade) D Air of the Compressed
Gas Asso iation Commodity Spe ifi ation for Air G-
7.1. Air pressure and flow rates must be within the
NIOSH approved ranges for the devi e.
7. Use stri tly in a ordan e with instru tions, labels and
limitations pertaining to this devi e.
8. This devi e may not provide a satisfa tory fa e seal
with ertain physi al hara teristi s (su h as beards
or sideburns) as outlined in ANSI Z88.2, resulting in
leakage in onne tion with the fa epie e, whi h voids
or limits the prote tion. If su h a ondition exists, the
user assumes all risks of death or serious bodily injury,
whi h may result.
9. Do not use the PremAire System for firefighting.
10. Do not use the PremAire System for underwater appli-
ations.
11. Thoroughly he k out the respirator on re eipt and
prior to use.
2
TAL 801 (L) Rev. 8 - 496958
TABLE OF CONTENTS
NIOSH Approval Information..............................................2
Spe ial or Criti al User Instru tions ...................................2
PremAire Respirator System Symbols ...............................3
PremAire System Options ..................................................3
Operating Prin iples ...........................................................5
General Instru tions/Air Sour e .........................................5
High-Pressure Air Cylinders ...............................................5
Air Compressors.................................................................5
Air-Supply Hose and Pressure Gauge................................6
Conne ting Inlet Pressure Gauge.......................................6
Inter onne ting Air-Supply Hoses......................................6
Preparing Respirator for Use..............................................7
Donning the Apparatus.......................................................7
Regulator Che k .................................................................7
Donning the Fa epie e.......................................................7
Using the Speed-On Harness ............................................8
Respirator Fit Test...............................................................8
Fa epie e Fit Che k...........................................................8
Donning and Using the Respirator .....................................8
Removing the Apparatus..................................................11
Air-Purifying Doffing (Removing).......................................11
Cleaning and Disinfe ting.................................................13
Inspe tion .........................................................................15
Storage .............................................................................16
Fun tional Che ks ............................................................17
Troubleshooting ................................................................18

INTRODUCTION
12. Do not use ompressed oxygen with the PremAire
System.
13. Unless equipped with a pressure-demand fa epie e
and the Es ape-Cylinder option, the PremAire
Respirator MUST NOT be used and relied upon for
respiratory prote tion when the atmosphere ontains
on entrations of ontaminants whi h are unknown or
IDLH. Also, only the PremAire XV and the PremAire
Cadet XV with es ape ylinders are approved as an
entry and es ape respirator from IDLH atmospheres.
14. Users must wear suitable prote tive lothing and pre-
autions must be taken so that the respirator is not
worn in atmospheres that may be harmful to the
devi e.
15. Never alter or modify this devi e, ex ept as dire ted
by MSA during installation of NIOSH approved kits.
Use only MSA repla ement parts. If other than the
proper MSA parts are used, the NIOSH approval will
be voided.
16. Use only the listed hose lengths and air-pressure
range spe ified in these instru tions.
17. This approval applies only when the respirator is sup-
plied with respirable air through 8 to 300 feet of air-
supply hose within the pressure range of 60 to 100 psi
using Ultra Elite®and Ultravue®Fa epie es. When
using an Advantage 4000 Fa epie e ensure the pres-
sure range is between 70 to 100 psi. A maximum of
12 se tions of air-supply hose may be used in making
up the maximum working length of hose. Ea h se tion
of oiled hose is onsidered 50 feet in length (max. 6
se tions).
Particles and contaminants can enter an airline respi-
rator system when air-supply hoses are disconnected
and/or reconnected in a contaminated atmosphere.
This could result in serious injury or death depending
on the toxicity of the contaminant involved. It is the
responsibility of the user to determine the potential
risk and to take the necessary precautions which may
include a requirement that NO disconnection or recon-
nection of air-supply hoses be permitted in a contami-
nated atmosphere. If in doubt DO NOT disconnect
and/or reconnect. Failure to follow this warning can
result in serious personal injury or death.
18. When using Ultra Elite or Ultravue Fa epie es at tem-
peratures below 32˚F, nose up is required. A nose up
must always be used with the Advantage 4000
Fa epie e.
PREMAIRE RESPIRATOR SYSTEM SYMBOLS
Symbols are used to dire t you to other instru tions,
warnings and guidelines that apply to the type of
option(s). It is important that you familiarize yourself with
these symbols, along with the orresponding instru tions
before attempting to operate the respirator.
PREMAIRE SYSTEM OPTIONS
The PremAire Respirator is a pressure-demand, Type C
supplied-air respirator as defined by 42 CFR Part 84,
Subpart J. The respirator’s unique waist mounted mani-
fold, whi h serves as the air distribution enter for the
system.
These options an be ombined or used individually. A list
of all possible respirator onfigurations an be found on
the PremAire System Qui k Referen e Chart
(P/N 802999).
3TAL 801 (L) Rev. 8 - 496958
REGULATOR
ESCAPE
CYLINDER
VORTEX
MAIN
INLET
DUAL
SUPPLY
Pressure-Demand
Supplied-Air
Respirator
(P/N 49695 )
Escape Cylinder
Capability
(P/N 00 0 )
Duo-Twin
Capability
(P/N 1 370)
Vortex Tube
Capability
(P/N 497231)
Dual-Supply
Capability
(P/N 497230)

NOTES
4
TAL 801 (L) Rev. 8 - 496958

OPERATING PRINCIPLES
OPERATING PRINCIPLES
The PremAire System Supplied-Air Respirator provides
respirable air to the user on demand — not onstantly, as
in a onstant-flow type devi e.
The PremAire Respirator is a pressure-demand regulator
mounted at the fa epie e. This mask mounted regulator
(MMR) maintains a slight positive pressure within the fa e-
pie e while regulating and redu ing the air supply to a
breathable pressure. This is a omplished by using a pilot
diaphragm that senses the breathing demands of the user
in a ontrolled feedba k state.
The PremAire Respirator is designed to maintain a slight
positive pressure of air inside the fa epie e, whether the
wearer is inhaling or exhaling, to prevent ontaminants
from entering the fa epie e, should there be slight fa e-
to-fa epie e seal leakage.
As su h, it is suitable for use with a large, ompressed air
ylinder of breathable air or a bank of ylinders set up in
as ade fashion. Alternatively, the PremAire Respirator
an be used with a ompressor system designed and
onfigured to supply respirable breathing air to the respi-
rator, within the NIOSH approved pressure range.
In normal use, the pressure-demand respirator is onne t-
ed to an air supply of a type and duration sele ted by the
user and will provide respiratory prote tion as long as the
user remains onne ted to the air supply.
DO NOT use the PremAire Respirator in an IDLH
atmosphere UNLESS equipped with the emergency
escape air cylinder and a pressure-demand only face-
piece. This is because if the air supply should fail, the
user has no adequate means to escape from an IDLH
atmosphere to fresh air. Duo-Twin versions of the
PremAire System are not NIOSH approved for use in
IDLH atmospheres. Failure to follow this warning can
result in serious personal injury or death.
IDLH means onditions that pose an immediate threat to
life or health or onditions that pose an immediate threat
of severe exposure to ontaminants, su h as radioa tive
materials, whi h are likely to have adverse umulative or
delayed effe ts on health [Title 42 CFR, Part 84].
Non-IDLH means any hazardous atmosphere whi h may
produ e physi al dis omfort immediately, hroni poison-
ing after repeated exposure, or a ute adverse physiologi-
al symptoms after prolonged exposure. [Title 42 CFR
Part 84].
GENERAL INSTRUCTIONS
AIR SOURCE
The purity of the air supply is the responsibility of the user.
The respirator is approved only when the air supplied to
the respirator meets the requirements of the Compressed
Gas Asso iation Spe ifi ation G7.1 for Quality Verifi ation
Level (Grade) D Breathable Air. This requires that the air
ontain no more than 10 parts per million (ppm) arbon
monoxide, not more than 1000 parts per million (ppm)
arbon dioxide, and not more than 5 milligrams per ubi
meter of oil vapor or oil parti ulates.
AIR-SUPPLY HOSE AND PRESSURE GAUGE
The PremAire System an be used with a wide range of
MSA air-supply hoses, whi h an be inter onne ted up to
a maximum length of 300 feet. A maximum of 12 se tions
of air-supply hose may be used in making up the maximum
300 feet working length of hose. Ea h se tion of oiled
hose is onsidered 50 feet in length (maximum 6 se tions).
MSA air-supply hoses have various temperature limita-
tions. DO NOT use the PremAire System whenever ambi-
ent or inlet-air temperatures exceed the limits specified
below for each hose material. Failure to follow this warn-
ing can result in serious personal injury or death.
INLET PRESSURE GAUGE
To ensure a urate readings, the inlet pressure-gauge
assembly must be installed at the air sour e.
INTERCONNECTING AIR-SUPPLY HOSES
MSA air-supply hoses an be inter onne ted up to a
maximum length of 300 feet without voiding the NIOSH
approval. MSA offers both threaded and lo king-type
qui k dis onne ts.
5TAL 801 (L) Rev. 8 - 496958
Air-Supply Hose
O-Ring
3/4" - 16 UNF
★Air-Supply Hose
This typi al threaded onne tor method an be used to inter onne t se -
tions of approved air-supply hose to the PremAire System.
HOSE RECOMMENDED
MATERIAL LIMITS
Polyvinyl Chloride 32˚F/120˚F
Neoprene -25˚F/212˚F
Nylon -20˚F/180˚F

OPERATING PRINCIPLES
Lo king qui k-dis onne ts easily onne t by pushing the
plug and so ket together. To separate, push the plug and
so ket together, then pull the sleeve from the plug.
Hoses must only be interconnected with either the
threaded connector (3/4 16 UNF) or the locking-type
quick-disconnects listed above. Do NOT use nonlock-
ing quick-disconnects to interconnect air-supply
hoses. Failure to follow this warning can result in seri-
ous personal injury or death.
6
TAL 801 (L) Rev. 8 - 496958
Quick Disconnect Socket with Check Valve 476956
Air Supply Hose (3/4 - 16 UNF)
Quick Disconnect Plug 476955 (No Check Valve)
(3/4 - 16 UNF)
Air-Supply Hose
O-Ring Pt. No. 455804
3/4" - 16 UNF
★
Air-Supply Hose
Quick Disconnect Plug 479015 (No Check Valve)
3/4" - 16 UNF
Quick Disconnect Socket with Check Valve 479032
CEJN - Chrome
Socket and Plug Assembly 479010
{
Socket and Plug Assembly 479009
{
SNAP-TITE — Aluminum
479026
479027

DONNING
PREPARING RESPIRATOR FOR USE
1. UPON RECEIPT, thoroughly he k the respirator
before use. Che k the fa epie e, regulator, hoses and
all fittings for shipping damage. Repla e any damaged
omponents.
2. BEFORE EACH USE, inspe t the fa epie e, regulator,
and hoses for wear. Repla e any damaged ompo-
nents. See inspe tion guidelines.
DONNING THE APPARATUS
1. Remove the fa epie e from the ase.
Do not use a cover lens in a high-temperature environ-
ment, such as firefighting. High temperatures may dis-
tort the cover lens. Or, moisture trapped between a
cover lens and the facepiece lens may condense and
distort vision. Always remove the cover lens before
donning the facepiece. Failure to follow this warning
can result in serious personal injury or death.
2. Fasten the waist-strap
and pull it tight for a
snug fit. The waist-strap
end must be tu ked in
and lay flat a ross the
body.
REGULATOR CHECK
1. Grasp the mask mounted regulator and push the shut-
off IN.
Note: The 1/4 Turn Se ond Stage MMR Regulator has a
shut-off button on one end of the regulator. The Firehawk
Se ond Stage MMR Regulator has a slide button (top
release button) on top of the regulator.
2. Conne t the air-supply hose to the “MAIN” inlet of the
PremAire manifold to initiate air flow.
3. No air should flow from the regulator. If it does, repeat
steps 1 and 2.
4. Dis onne t air-supply. Slowly bleed off pressure.
The apparatus must be checked and corrected for
proper operation by an MSA trained repairperson
before using. Failure to follow this warning can result
in serious personal injury or death.
Note: Before donning, he k that the regulator qui k- on-
ne t seal ring is seated properly in its groove, and that it
is not torn, gouged, or ni ked.
DONNING THE FACEPIECE
Do not wear eyeglasses under the facepiece. The tem-
ples or sidebars on eye glasses will prevent an airtight
seal. if you must wear glasses, install the spectacle
kit. Failure to follow this warning can cause inhalation
of contaminated air, resulting in serious respiratory
injury or death.
1. Extend the fa epie e
straps fully. Pla e ne k-
strap around your ne k
and don the fa epie e
by inserting your hin
first.
2. Pull the head harness
ompletely over your
head and tighten the
lower (ne k) straps.
3. Tighten the lower (ne k) harness straps first, by pulling
them straight ba k, not out. Tighten the temple straps
the same way.
4. Tu k in the ends of the straps so that they lay flat
a ross the head.
5. Push headband pad towards ne k and repeat step 2.
If ne essary, tighten the front strap for best visibility
and fit.
6. Tu k in the ends of the straps so they flat a ross the
head.
7TAL 801 (L) Rev. 8 - 496958

DONNING
USING THE SPEED-ON HEAD HARNESS
1. Loosen the ne k straps so the end-tabs are at the
bu kles.
2. Insert your hin into the fa epie e.
3. Pull the harness "net" over the rown of your head.
4. Tighten the ne k straps. If ne essary, tighten the tem-
ple or front strap adjustments.
5. Tu k in the straps so that they lay flat a ross the
head.
RESPIRATOR FIT TEST
A qualitative or quantitative fit test must be ondu ted for
ea h wearer of the respirator to determine the amount of
prote tion it will provide. Respirator fit tests are explained
fully in the Ameri an National Standard for Respiratory
Prote tion, ANSI Z88.2, whi h is published by the
Ameri an National Standards Institute.
Quantitative Test
If a quantitative fit test is used, a fit fa tor that is at least
1,000 shall be obtained before that respirator is assigned
to an individual.
Qualitative Test
If a qualitative fit test is used, only validated proto ols are
a eptable. The individual must pass a test designed to
assess a fit fa tor of at least 1,000.
Respirators must be qualitatively or quantitatively fit test-
ed in a negative-pressure mode.
The user must perform a respirator fit test and follow
all warnings and limitations specified. Failure to follow
this warning can result in serious personal injury or
death.
Respirator fit tests are explained fully in the Ameri an
National Standard “Pra ti es for Respiratory Prote tion,”
ANSI Z88.2 1980, published by the Ameri an National
Standards Institute.
Note: The user assumes all risks of death or serious bodi-
ly injury whi h an result if a fit test is not performed or
the respirator limitations are not followed.
FACEPIECE FIT CHECK
1. To he k the inhalation valve, inhale. If you do not
re eive suffi ient flow of air, remove and repla e the
fa epie e.
2. To he k for fa epie e fit, hold the palm of your hand
over the inlet onne tion and inhale. Hold your breath
at least 10 se onds. The fa epie e should ollapse
and stay ollapsed against your fa e. If it does not,
readjust the fa epie e and test again. If this does not
correct the leak, do not use the facepiece.
3. Test the exhalation
valve, take a deep
breath and hold it.
Blo k the inlet onne -
tion with the palm of
your hand and exhale.
If the exhalation valve
is stu k, you may feel a
heavy rush of air
around the fa epie e.
Note: You may need to exhale sharply to open the valve.
If this does not release the valve, do not use the fa e-
pie e. See PremAire Duo-Twin Instru tion Manual for Air
Tightness Test for Duo-Twin fa epie e (P/N 818370).
DONNING AND USING THE RESPIRATOR
Before donning and using the respirator, you must test
the apparatus for any leaks in the supplied-air system.
The respirator must be pressurized to conduct these
tests. See Instruction manual for the required leak test
procedures. Failure to follow this warning can result in
serious personal injury or death.
See PremAire Vortex Tube Instru tion Manual for
additional donning instru tions (P/N 497231).
This device may not seal properly with your face if you
have a beard, gross sideburns, or similar physical
characteristics (see ANSI Z .2). An improper facial
seal may allow contaminants to leak into the face-
piece, reducing or eliminating respiratory protection.
Do not use this device if such conditions exist. The
face-to-facepiece seal must be tested before each
use. Never remove the facepiece except in a safe,
non-hazardous, non-toxic atmosphere. Failure to fol-
low this warning can result in serious personal injury
or death.
Conne t the air-supply hose to the “Main” inlet of the
PremAire manifold to initiate air flow.
8
TAL 801 (L) Rev. 8 - 496958

DONNING
INSTALLING 1/4 REGULATOR
1. To onne t the Qui k-
Conne t regulator to the
fa epie e, push the reg-
ulator into the fa epie e
adapter.
a. Rotate the regulator
1/4 turn. The regula-
tor an be rotated in
either dire tion.
b. Listen for the release tab to “ li k” as the regulator
lo ks onto the fa epie e.
. Verify proper engagement by rotating the regulator
until it onta ts the release tab and stops. The reg-
ulator must only swivel 70 degrees.
d. Double he k proper engagement by pulling on the
regulator to ensure that the regulator is se urely
atta hed to the fa epie e.
Do not use the respirator unless the regulator is con-
nected properly. The regulator must swivel approxi-
mately 70 degrees, but must NOT rotate beyond the
tab stops. Do NOT use the respirator if the regulator
does not swivel approximately 70 degrees or rotates
beyond the tab stops. Return the respirator to an MSA
trained or certified repairperson to correct the condi-
tion. A regulator that is not installed correctly can sep-
arate from the facepiece unexpectedly. Failure to fol-
low this warning can result in serious personal injury
or death.
2. Inhale sharply to start the air flow. The shut-off button
should pop out automati ally.
a. Che k the bypass again. Turn the red knob ounter-
lo kwise until it lo ks in position.
INSTALLING FIREHAWK REGULATOR
1. Grasp regulator and orient regulator so that the slide
button is on top.
2. Slide regulator onto rail (fast tra k) of fa epie e over.
Slide regulator down the rail over until regulator
stops.
3. Insert regulator into fa epie e adapter by pushing
inward.
4. Listen for the button to “ li k” as the regulator lo ks
into the fa epie e.
5. Che k proper engagement by pulling on the regulator
to ensure regulator is se urely atta hed to fa epie e.
Do not use this respirator unless the regulator is con-
nected properly. A regulator that is not installed cor-
rectly can separate from the facepiece unexpectedly.
Return the respirator to an MSA trained and certified
repairperson to correct the condition. Failure to follow
this warning can result in serious personal injury or
death.
6. Inhale sharply to start the airflow.
If the apparatus passes all tests, the PremAire is ready
to use. Remember, you must make these tests every
time before you enter the hazardous atmosphere. If
the unit fails to meet any of these tests, the condi-
tion(s) must be corrected before using the apparatus.
Failure to follow this warning can result in serious per-
sonal injury or death.
INSTALLING FIREHAWK PUSH-TO-CONNECT
REGULATOR
1. Grasp regulator and insert regulator into fa epie e
adapter by pushing inward. Listen for the buttons to
“ li k” as regulator lo ks into the fa epie e. Che k
proper engagement by pulling on the regulator to
ensure regulator is se urely atta hed to fa epie e
adapter.
Do not use the respirator unless the regulator is con-
nected properly. A regulator that is not installed cor-
rectly can separate from the facepiece unexpectedly.
Return the respirator to an MSA trained or certified
repairperson to correct the condition. Failure to follow
this warning can result in serious personal injury or
death.
2. Inhale sharply to start the airflow.
9TAL 801 (L) Rev. 8 - 496958

DONNING
Note: If the respirator passes all tests, the unit is ready to
use. Remember, you must perform these tests every time
before you enter the hazardous atmosphere. If the unit
fails to meet any of the tests, the ondition(s) must be
orre ted before using the respirator.
Note: If using the PremAire Respirator with any of the fol-
lowing options, see the se tions indi ated for proper
instru tions for use.
See PremAire Es ape Cylinder Instru tion Manual
for use instru tions (P/N 800808).
See PremAire Duo-Twin Instru tion Manual for
use instru tions (P/N 818370).
See PremAire Dual-Supply Instru tion Manual for
use instru tions (P/N 497230).
See PremAire Vortex Tube Instru tion Manual for
use instru tions (P/N 497231).
Note: If a de ontamination pro edure, reated by a erti-
fied health and/or safety professional, has been estab-
lished for the appli ation in whi h this respirator is used,
that pro edure should take pre eden e.
10
TAL 801 (L) Rev. 8 - 496958

DOFFING
REMOVING THE APPARATUS
Note: When ready to leave the work area, pro eed to an
area not requiring respiratory prote tion*. Remain on-
ne ted to the air-supply hose until this "safe" area is
rea hed. Then remove the fa epie e. The air-supply hose
an then be dis onne ted.
REMOVING THE 1/4 TURN REGULATOR
1. To dis onne t the regulator, pull the release tab away
from the fa epie e.
a. Rotate the regulator
1/4 turn. Pull the reg-
ulator away from the
fa epie e as you turn
it so that it slides out
of the groove.
2. Press the shut-off to release system pressure.
3. Press the shut-off button IN. Stow the regulator in the
stand-by belt mount when it is not in use.
Note: Air will flow from the se ond stage regulator briefly
until system pressure is relieved.
REMOVING FIREHAWK REGULATOR
1. Grasp top of regulator.
2. Push the release buttons and pull regulator down and
out of fa epie e adapter.
Note: Regulator an hang on over rail in a stand-by
mode.
3. Slide regulator up over rail until regulator slide button
is free of over rail.
4. To remove the fa e-
pie e, fully loosen the
harness straps and pull
the fa epie e up and
away from your fa e.
DISCONNECTING THE FIREHAWK PUSH-TO-
CONNECT REGULATOR
1. Grasp top of regulator.
2. Push the release buttons and pull regulator out of
fa epie e adapter.
3. To remove the belt, press the belt bu kle release but-
ton IN.
Note: Inspe tion and Cleaning and Disinfe ting pro e-
dures are outlined in this manual. Ensure omplete appa-
ratus is lean and dry. Ensure that fa epie e head har-
ness straps and harness adjustment straps are fully
extended. Pla e the omplete apparatus in the storage
ase or suitable storage lo ation so it an be rea hed
easily for emergen y use.
STORAGE
Do not store the respirator near substan es that an
atta k respirator omponents, ausing them to fail to per-
form as designed.
The PremAire system should be stored in a ool, lean,
dry pla e, away from dire t sunlight. Heat and sunlight will
shorten the life of rubber parts.
AIR PURIFYING DOFFING (REMOVING)
If you are working in a contaminated atmosphere, or
you are exposed to contaminants while in the work
area, take the proper precautions to decontaminate
the facepiece and head harness before doffing the
facepiece. You must determine the potential risk and
take the necessary precautions. Failure to follow this
warning can result in serious personal injury or death.
Note: If a de ontamination pro edure ( reated by a erti-
fied health and/or safety professional) has been estab-
lished for the appli ation in whi h this respirator is used,
the pro edure should take pre eden e.
1. Return to an area that does not require respiratory
prote tion before removing the respirator.
2. Dis onne t the MMR from the fa epie e.
3. Press the MMR shut-off button in to stop the flow of
air.
4. Remove the fa epie e by fully loosening the harness
straps.
5. Pull the fa epie e up and away from your fa e.
6. See PremAire Duo-Twin Instru tion Manual for use
Instru tions (P/N 818370).
11 TAL 801 (L) Rev. 8 - 496958
*If a de ontamination pro edure, reated by a ertified health professional,
has been established for the appli ation in whi h the respirator is used, that
pro edure should take pre eden e.

NOTES
12
TAL 801 (L) Rev. 8 - 496958

CLEANING AND DISINFECTING
CLEANING AND DISINFECTING
Respirators should be leaned and disinfe ted after ea h
use. The fa epie e should be leaned and disinfe ted
after every use. Remove the filter artridges (if used). MSA
re ommends using Confiden e Plus®Cleaning Solution
(P/N 10009971). Refer to the label for user instru tions.
ANSI suggests that users be trained in leaning pro e-
dure.
• Cleaning and disinfecting at or below 110˚F tem-
perature will avoid possible overheating and dis-
tortion of parts which would require replacement.
• DO NOT use any cleaning substances that can or
might attach any part of the apparatus.
• DO NOT use alcohol because it may deteriorate
rubber parts.
• If not rinsed thoroughly, cleaning agent residue
may irritate the wearer’s skin.
13 TAL 801 (L) Rev. 8 - 496958

NOTES
14
TAL 801 (L) Rev. 8 - 496958

INSPECTION
INSPECTION
DO NOT inspect the respirator before cleaning if there
is danger of contacting hazardous contaminants.
Clean and disinfect first, then inspect. Failure to follow
this warning can result in inhalation or skin absorption
of the contaminant and cause serious personal injury
or death.
Inspe t the respirator by sight and sound for normal oper-
ations after it has been leaned and disinfe ted. When
any part shows eviden e of damage, wear, or any other
adverse ondition explained in this se tion, it must be
repla ed and the ondition orre ted before the respirator
an be used again.
Note: Repla ement or repairs shall be done only by quali-
fied persons, using only MSA parts designed for the respi-
rator. No attempt shall be made to make adjustments or
repairs beyond the manufa turer's re ommendations.
Parts shall not be inter hanged among devi es of different
manufa turers. MSA authorizes levels of maintenan e and
repair for the PremAire System. (See new maintenan e
manual P/N 10017251.)
If there is no MSA Servi e Center in your area, return the
unit to MSA for servi e. Call 1-800-MSA-2222 for instru -
tions.
See PremAire Duo-Twin Instru tion Manual for other
general repair requirements.
Inspe t the entire apparatus after it is leaned and disin-
fe ted. ANSI Standards Z88.2 and Z88.5, des ribe three
levels of inspe tion pro edures whi h are to be per-
formed. Refer to these do uments or to an inspe tion
program prepared by a health professional in establishing
an inspe tion program. Detailed repair pro edures are
lo ated in PremAire User’s Maintenan e Instru tions.
If any of the following inspections do not function
properly, the apparatus must be removed from ser-
vice. Failure to follow this warning can result in seri-
ous personal injury or death.
Component Inspection
(After Each Use and Monthly)
1. Don the respirator following the instru tion pro e-
dures. These steps make up the Fun tional Test.
2. If all steps are performed su essfully, remove the res-
pirator and inspe t it following the steps below.
3. Fa epie e
a. Inspe t the fa epie e for rubber deterioration, dirt,
ra ks, tears, holes, or ta kiness.
b. Che k the harness headstraps for breaks, loss of
elasti ity, or missing bu kles and straps. Che k the
strap serrations for signs of wear.
. Inspe t the lens for ra ks, s rat hes, and a tight
seal with the fa epie e rubber.
d. The exhalation valve must be lean and operate
easily. The valve must move off the seat and return
when released.
e. Inspe t the fa epie e oupling for damage. Also
he k to be sure the spider gasket, O-ring, and
valve dis are present.
4. Harness
a. Inspe t all harness omponents for uts, tears,
abrasions, or signs of heat or hemi al-related
damage. Che k that the tee nuts, washers, and
s rews, if any, are se ure.
5. Re ord Keeping
a. Following inspe tion, the date and initials of the
designated person should be re orded on an
inspe tion tag. A more detailed re ord of the opera-
tions performed an be noted on an inspe tion and
maintenan e log. Inspe tion tags and inspe tion
and maintenan e logs are available from MSA.
When the inspe tion data has been re orded, the
apparatus is stored in a ready position.
Note: There are additional omponents to be inspe tion
for PremAire Cadet Respirators with Es ape Cylinder.
Refer to the Inspe tion se tion in P/N 497230.
15 TAL 801 (L) Rev. 8 - 496958

STORAGE
16
TAL 801 (L) Rev. 8 - 496958
PREPARATION FOR STORAGE
Note: Do not for e-dry the parts by pla ing them in a
heater or in dire t sunlight. The rubber will deteriorate.
When the fa epie e is thoroughly dry, store the fa epie e
in the plasti bag that it was shipped in.
1. In general, only the fa epie e requires leaning and
disinfe ting after ea h use. If the apparatus is soiled
or has dirt a umulation, use a damp sponge with
mild soap solution or use a soft/medium bristle brush
to remove deposits that may interfere with normal
operation.
a. Ensure se ond stage regulator is free of water, dirt
or debris.
2. Inspe t the entire apparatus as you reassemble it.
Follow the Inspe tion Instru tions.
3. Thoroughly dry the fa epie e and regulator after
leaning and disinfe ting. The fa epie e an trap
water, whi h ould enter the regulator.
STORAGE
DO NOT store the respirator near substances that can
attack respirator components, causing them to fail to
perform as designed.
The PremAire CADET Respirator must be stored in a ool,
dry pla e away from dire t sunlight. Heat and sunlight will
shorten the life of rubber parts.

FUNCTIONAL CHECKS
FUNCTIONAL CHECKS
(After Each Use and Monthly)
1. Che k that the regulator works properly.
a. Che k that the ylinder valve and se ond stage
shut-off are losed and the system is not pressur-
ized.
b. Hold the fa epie e against your fa e to reate an
effe tive seal.
. Atta h the regulator to the fa epie e and inhale
until the fa epie e begins to ollapse against your
fa e. Hold your breath for about 10 se onds. The
negative pressure should be maintained and the
fa epie e should remain ollapsed against your
fa e for the entire 10 se onds.
d. Do not use the apparatus if negative pressure an-
not be maintained in the fa epie e. Return the regu-
lator and fa epie e to a ertified repairperson.
2. Inspe t the Se ond Stage Shut-Off.
a. With the regulator pressurized, operate ea h valve
to be sure it operates. Venting of pressure relief
valves (or a ontinuing flow of air through the regu-
lator when the user is not inhaling) indi ates that the
regulator needs to be repaired.
b. Listen to the regulator. Any unusual sounds, su h as
whistling, hattering, li king, or rattling mean that
the regulator should be he ked further.
. If any of these symptoms o ur, the apparatus must
be removed from servi e. Return the regulator to a
ertified repairperson.
DO NOT disconnect when pressurized. Release all
pressure from the regulator by opening the bypass
valve. Removing the regulator when pressurized can
result in serious personal injury, death, or damage to
equipment.
17 TAL 801 (L) Rev. 8 - 496958

TROUBLE PROBABLE CAUSE REMEDY
Regulator has low flow performan e Air-supply not fully open or Be sure air sour e valve
pressure at wrong setting. is fully open and that inlet
pressure is between 60-100 psig.
Se ond-stage inlet filter may Remove the regulator from servi e and
be plugged. repla e the inlet filter. See repair
instru tions.
Se ond stage regulator may Return to MSA trained repair
require adjustments. te hni ian.
Air leaking at mask mounted Swivel O-rings leaking. Dis onne t the regulator hose assembly.
regulator. Inspe t and repla e the O-rings. See
repair instru tions.
Swivel not fully engaged. Return to MSA trained repair
te hni ian.
Fa epie e onne tion leaking. If using Ultra Elite or Ultravue Fa epie es,
remove the regulator and inspe t or
repla e seal ring and spider gasket within
the Qui k-Conne t Adapter. See repair
instru tions of inhalation he k valve
and spider gasket.
If using the Advantage 4000 Fa epie e,
inspe t the gray sili one valve for
perforations, uts, or other visible
irregularities that ould ompromise seal
integrity.
Air leaking from manifold. Manifold port(s) not plugged. Che k all manifold parts to make sure
they are plugged and sealed.
TROUBLESHOOTING
18
TAL 801 (L) Rev. 8 - 496958


Other manuals for PremAire System
5
Table of contents
Other MSA Safety Equipment manuals

MSA
MSA M1 User manual

MSA
MSA V-SHOCK User manual

MSA
MSA Gallet F2XR User manual

MSA
MSA Gravity Sure-Line Operation instructions

MSA
MSA S-Cap-Air User manual

MSA
MSA Latchways PRD 68220 User manual

MSA
MSA Latchways Freestanding Constant Force User manual

MSA
MSA AUER 3S-PS User manual

MSA
MSA HUB User manual

MSA
MSA GVC Series User manual
Popular Safety Equipment manuals by other brands

Dynatech
Dynatech PQ-4000 UD Instructions for use and maintenance

Dräger
Dräger Bodyguard 1500 Instructions for use

TM Technologie
TM Technologie TM.ONTEC G installation instructions

Guardian
Guardian CR3 SELF-RETRACTING LIFELINE instruction manual

ABS
ABS ZU Series manual

3M
3M DBI SALA ExoFit Nex 603XS1020 User instructions