NOTES ON ALARMS
- The acoustic signals that cause the error conditions can be stopped
by pressing the ON/STAND-BY key.
- There is the possibility of using one of the board relays to activate
an external alarm in the event that the error condition occurs (See
connection diagrams).
Settingthesetpoint(desiredtemperature)
‐PressSETfor1secthesetvaluewillstartflashingafterafewmoments.
‐IncreaseordecreasethevalueusingUPorDOWN.
‐PressSETtoconfirmthenewvalue.
SwitchingthedeviceON/OFF
PressUPformorethan3sec.Thecontrolanddefrostalgorithmsarenowdisabledandtheinstrument
displaysthemessage“OFF”alternatingwiththetemperaturereadbythesetprobe.
Manualdefrost(modelsS,X,YandConly)
PressforDOWNmorethan3sec(thedefroststartsonlythetemperatureconditionsarevalid).
Continuouscycle(modelsS,X,YandConly)
PressUPandDOWNtogetherformorethan3sec.
AccessandsettingtypeF(frequent)andtypeC(configuration)parameters
1. PressSETfor3sec(thedisplaywillshow”PS”)
2. ToaccessthetypeFandCparametermenu,enterthepassword”22”usingUP/DOWN;
ToaccesstheFparametermenuonly,pressSET(withoutenteringthepassword);
3. ScrollinsidetheparametermenuusingUP/DOWN;
4. Todisplay/setthevaluesoftheparameterdisplayed,pressSET,thenUP/DOWNandfinallySETto
confirmthechanged(returningtotheparametermenu).
Tosaveallthenewvaluesandexittheparametermenu,pressSETfor3sec.
Toexitthemenuwithoutsavingthechangedvalues(exitbytimeout)donotpressanybuttonforat
least60sec
Tableofalarms
Alam
code
buzzeran
alarmrelay
LEDDescription Parameters
involve
E
activeONprobe1error=contro
‐
E1inactiveONprobe2error=defrost [d0=0/1
E2inactiveONprobe3error=condenser [A4=10
IAactiveONexternalalarm [A4=10][+A7
dOractiveONopendooralarm [A4=7/8][+A7
LOactiveONlowtemperaturealarm [AL][Ad
HIactiveONhightemperaturealarm [Ah][Ad
EEinactiveONunitparametererror ‐
E
inactiveONoperatingparametererror ‐
E
inactiveONdefrostendedbytimeout [dP][dt][d4][A8]
d
inactiveONdefrostrunning [d6=0
chtinactiveONcondenserdirtypre‐alarm [A4=10
CHTactiveONcondenserdirtyalarm [A4=10
Et
inactiveONclockalarm
Ourproductshavebeenprecisselymodifiedbeforeleavingfactory.Toavoiddamagingcompressorunitor
othermalfunctions,usersmustnotmodifythemicrocomputerparametersprivately.